Precise measurement of selenium isotopes by HG-MC-ICPMS using a 76–78 double-spike

Abstract

Selenium (Se) stable isotopes are a new geochemical tool with great potential as a tracer of redox processes and chemical cycling of chalcophiles and volatile elements. However, Se isotope measurements in low-Se samples present a formidable analytical challenge. In this study, we report a new method to measure Se stable isotopes (δ82/78Se; per mil deviation relative to Se NIST SRM 3149) with extremely high precision. Selenium has six stable isotopes and therefore is a good candidate for isotope analysis using a double spike approach, which has the advantage that it can correct for any stable isotope fractionation that may occur during sample processing. We have calibrated a novel 76Se–78Se double spike and have developed a rapid and precise analytical protocol on a multi-collector inductively coupled plasma mass spectrometer using an ESI hydride generation introduction system. Sensitivity is over 1000 V per ppm for the total Se signal; a measurement typically requires 25 ng of natural Se. Argon dimer interferences on masses 76, 78 and 80 were corrected in situ by measuring mass 80. Germanium interferences on masses 74 and 76 were corrected by measuring mass 73 and mass 75 was monitored to correct for arsenic hydride on mass 76. Wash-out times were in the order of 180 s, greatly reduced compared to previous studies that rely on on peak zero argon dimer corrections (wash-out times of up to one hour). The 2 s.e. error for a single analysis typically ranges from 0.01 to 0.025‰ (n = 80) for δ82/78Se. Long-term reproducibility and accuracy were estimated by multiple analyses of the Se Merck standard over numerous different analytical sessions, resulting in a mean δ82/78Se value of −0.989 ± 0.034‰ (n = 93; 2 s.d.), which is in excellent agreement with previous studies.

Graphical abstract: Precise measurement of selenium isotopes by HG-MC-ICPMS using a 76–78 double-spike

Associated articles

Supplementary files

Article information

Article type
Paper
Submitted
01 Oct 2019
Accepted
02 Dec 2019
First published
09 Dec 2019

J. Anal. At. Spectrom., 2020,35, 320-330

Precise measurement of selenium isotopes by HG-MC-ICPMS using a 76–78 double-spike

M. Pons, M. Millet, G. N. Nowell, S. Misra and H. M. Williams, J. Anal. At. Spectrom., 2020, 35, 320 DOI: 10.1039/C9JA00331B

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements