Issue 2, 2001

Fully integrated PCR-capillary electrophoresis microsystem for DNA analysis

Abstract

A fully integrated genomic analysis microsystem including microfabricated heaters, temperature sensors, and PCR chambers directly connected to capillary electrophoretic separation channels has been constructed. Valves and hydrophobic vents provide controlled and sensorless sample positioning and immobilization into 200 nL PCR chambers. The use of microfabricated heating and temperature sensing elements improves the heating and cooling rates for the PCR reaction to 20 °C s−1. The amplified PCR product, labeled on-column with an intercalating fluorescent dye, is injected into the gel-filled capillary for electrophoretic analysis. Successful sex determination using a multiplex PCR reaction from human genomic DNA is demonstrated in less than 15 min. This device is an important step toward a microfabricated genomic microprocessor for use in forensics and point-of-care molecular medical diagnostics.

Article information

Article type
Paper
Submitted
04 Oct 2001
Accepted
19 Oct 2001
First published
21 Nov 2001

Lab Chip, 2001,1, 102-107

Fully integrated PCR-capillary electrophoresis microsystem for DNA analysis

E. T. Lagally, C. A. Emrich and R. A. Mathies, Lab Chip, 2001, 1, 102 DOI: 10.1039/B109031N

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.