Issue 10, 2012

Modified high performance concentric nebulizer for inductively coupled plasma optical emission spectrometry

Abstract

The high performance triple-tube concentric nebulizer (HPCN) was modified and evaluated for sample introduction into inductively coupled plasma optical emission spectrometry (ICP-OES) at liquid flow rates of 0.25 to 0.8 mL min−1. The acceptable liquid flow rate for the use of HPCN was extended by replacing the tapered center capillary tube of the nebulizer (i.d./o.d.: 50 μm/150 μm) with a large bore tube (i.d./o.d.: 110 μm/170 μm). The nebulization efficiency was much improved by reducing the inner diameter of the nebulizer nozzle from 250 μm to 200 μm. At a liquid flow rate of 0.8 mL min−1 and a nebulizer gas flow rate of 1 L min−1, the Sauter mean diameter (D3,2) of the primary aerosol generated by the modified HPCN was 3.4 μm, and over 90% (v/v) of the aerosol droplets were below 10 μm in diameter. The D3,2 value was smaller than those generated by conventional nebulizers, Meinhard nebulizer type C (8.2 μm), Glass Expansion Conikal nebulizer (15.8 μm), SeaSpray (14.8 μm), and MiraMist (5.6 μm). The amount of the tertiary aerosol of the modified HPCN generated through a non-baffled cyclone chamber was approximately 1.8 to 3.1 times higher than those of the other nebulizers, with similar size distributions. The sensitivity in ICP-OES with the modified HPCN was 1.5- to 3.2-fold higher than those with the other nebulizers when the liquid flow rates were the same. The plasma robustness estimated from the commonly used ratio Mg(II)/Mg(I) was the same or slightly better than that of the conventional nebulizers. The HPCN also showed a good tolerance to high total dissolved solids (TDS) using 20% NaCl solution. Validation of the modified HPCN was performed by a recovery test of spiked seawater and analyzing the NMIJ CRM 7502-a white rice flour. The observed values for the ten elements Na, Mg, P, K, Ca, Mn, Fe, Cu, Zn and Cd were in good agreement with their certified values. We concluded the modified HPCN is a very useful nebulizer for ICP-OES with good performance at a large range of sample flow rates.

Graphical abstract: Modified high performance concentric nebulizer for inductively coupled plasma optical emission spectrometry

Article information

Article type
Technical Note
Submitted
17 Apr 2012
Accepted
27 Jun 2012
First published
28 Jun 2012

J. Anal. At. Spectrom., 2012,27, 1787-1793

Modified high performance concentric nebulizer for inductively coupled plasma optical emission spectrometry

A. S. Groombridge, K. Inagaki, S. Fujii, K. Nagasawa, T. Okahashi, A. Takatsu and K. Chiba, J. Anal. At. Spectrom., 2012, 27, 1787 DOI: 10.1039/C2JA30118K

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements