Issue 17, 2014

Rapid isolation and diagnosis of live bacteria from human joint fluids by using an integrated microfluidic system

Abstract

Arthroplasty is a general approach for improving the life quality for patients with degenerative or injured joints. However, post-surgery complications including periprosthetic joint infection (PJI) poses a serious drawback to the procedure. Several methods are available for diagnosing PJI, but they are time-consuming or have poor sensitivity and specificity. Alternatively, reverse-transcription PCR can detect live bacteria and reduce false-positive results but cannot avoid the cumbersome RNA handling and human contamination issues. In response, an integrated microfluidic system capable of detecting live bacteria from clinical PJI samples within 55 minutes is developed in this study. This system employs an ethidium monoazide (EMA)-based assay and a PCR with universal bacterial primers and probes to isolate and detect only the live bacteria that commonly cause PJI. The experimental results indicated that the developed system can detect bacteria in human joint fluids with a detection limit of 104 colony formation unit mL−1. Furthermore, nine clinical samples were analyzed using the microfluidic system. The results obtained from the microfluidic system were negative for all culture-negative cases, indicating that the proposed system can indeed reduce false-positive results. In addition, experimental results showed that the EMA sample pre-treatment process was crucial for successful detection of live bacteria. The culture-positive cases were diagnosed as positive by the proposed system only when the clinical samples were treated with EMA immediately after being sampled from patients. Based on these promising results, the developed microfluidic system can be a useful tool to detect PJI and potentially be applied in other clinical situations.

Graphical abstract: Rapid isolation and diagnosis of live bacteria from human joint fluids by using an integrated microfluidic system

Supplementary files

Article information

Article type
Paper
Submitted
22 Apr 2014
Accepted
11 Jun 2014
First published
12 Jun 2014

Lab Chip, 2014,14, 3376-3384

Author version available

Rapid isolation and diagnosis of live bacteria from human joint fluids by using an integrated microfluidic system

W. Chang, C. Wang, S. Yang, Y. Lin, J. Wu, M. S. Lee and G. Lee, Lab Chip, 2014, 14, 3376 DOI: 10.1039/C4LC00471J

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.