Issue 15, 2015

Metal–organic framework MIL-101(Cr) as a sorbent of porous membrane-protected micro-solid-phase extraction for the analysis of six phthalate esters from drinking water: a combination of experimental and computational study

Abstract

An attractive metal–organic framework (MOF) MIL-101(Cr) material was synthesized at the nanoscale and applied as a sorbent in the porous membrane-protected micro-solid-phase extraction (μ-SPE) device for the pre-concentration of phthalate esters (PAEs) in drinking water samples for the first time. Parameters influencing the extraction efficiency, such as the selection of sorbent materials, pH adjustment, the effect of salt, magnetic-stirring extraction time, the desorption solvent and the desorption time, were investigated. Under the optimum conditions, the limits of detection from gas chromatography-mass spectrometric analysis for PAEs varied from 0.004 to 0.02 μg L−1. The linear ranges were from 0.1 to 50 μg L−1 or from 0.2 to 50 μg L−1 for the analytes with the relative standard deviations fluctuating from 0.8 to 10.9% (n = 5). The enrichment factors (EFs) for the target PAEs were varied from 143 to 187. MIL-101(Cr) exhibited remarkable advantages compared to activated carbon and MIL-100(Fe). On the other hand, the computational method was first used to predict the adsorption of MIL-101(Cr) towards PAEs. The molecular interactions and the free binding energies between MIL-101(Cr) and PAEs were observed and calculated in terms of the molecular modeling method. MIL-101(Cr) showed high potential in the analysis of PAEs at trace levels in drinking water. The computational result was consistent with the detected enrichment factors. The computational modeling accurately predicted the extraction efficiency of MOF-based material towards the target analytes. Therefore, the combination of experimental and computational study provided a new strategy on the trace contaminant analysis.

Graphical abstract: Metal–organic framework MIL-101(Cr) as a sorbent of porous membrane-protected micro-solid-phase extraction for the analysis of six phthalate esters from drinking water: a combination of experimental and computational study

Supplementary files

Article information

Article type
Paper
Submitted
20 Mar 2015
Accepted
24 May 2015
First published
27 May 2015

Analyst, 2015,140, 5308-5316

Metal–organic framework MIL-101(Cr) as a sorbent of porous membrane-protected micro-solid-phase extraction for the analysis of six phthalate esters from drinking water: a combination of experimental and computational study

T. Wang, J. Wang, C. Zhang, Z. Yang, X. Dai, M. Cheng and X. Hou, Analyst, 2015, 140, 5308 DOI: 10.1039/C5AN00553A

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements