Issue 9, 2015

A fluorescein labeled aptamer switch for thrombin with fluorescence decrease response

Abstract

We developed a simple fluorescent switch for rapid thrombin detection by using a high-binding affinity 27-mer aptamer with fluorescein (FAM) labeled at the 3′ end. This FAM-labeled aptamer showed remarkable fluorescence quenching upon thrombin binding. This assay enabled the detection of thrombin ranging from 0.062 nM to 2 nM. The assay was selective, and other tested proteins and diluted serum did not interfere with thrombin analysis.

Graphical abstract: A fluorescein labeled aptamer switch for thrombin with fluorescence decrease response

Supplementary files

Article information

Article type
Paper
Submitted
20 Feb 2015
Accepted
01 Apr 2015
First published
01 Apr 2015

Anal. Methods, 2015,7, 3888-3892

A fluorescein labeled aptamer switch for thrombin with fluorescence decrease response

L. Hao and Q. Zhao, Anal. Methods, 2015, 7, 3888 DOI: 10.1039/C5AY00464K

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.