Optimization of a methodology for simultaneous determination of twelve chlorophenols in environmental water samples using in situ derivatization and continuous sample drop flow microextraction combined with gas chromatography-electron-capture detection

Abstract

Continuous sample drop flow microextraction (CSDFME) combined with gas chromatography-electron-capture detection (GC-ECD) has been developed as a high preconcentration technique for the determination of chlorophenols (CPs) in environmental water samples. In this method, a few microliters of organic solvent (11.0 μL chlorobenzene) is transferred to the bottom of a conical test tube. Then 10.0 mL of aqueous solution transforms into fine droplets while passing through the organic solvent. At this stage, CPs are extracted into the organic solvent. Under the optimum conditions, enrichment factors and extraction recoveries are in the range of 630–1770 and 31.5–88.5, respectively. The calibration graphs are linear in the range of 0.01–300 μg L−1 and limits of detection (LODs) are in the range of 0.005–0.50 μg L−1. The relative standard deviations (RSDs, for 100 μg L−1 of MCPs, 50 μg L−1 of DCPs, 2.00 μg L−1 of TCPs, 1.00 μg L−1 of TeCPs and PCP in water) with and without using an internal standard are in the range of 0.9–5.5% (n = 7) and 1.2–6.4% (n = 7), respectively. The relative recoveries of well, tap and river water samples which have been spiked with different levels of CPs are in the range of 90.6–104.2%.

Graphical abstract: Optimization of a methodology for simultaneous determination of twelve chlorophenols in environmental water samples using in situ derivatization and continuous sample drop flow microextraction combined with gas chromatography-electron-capture detection

Article information

Article type
Paper
Submitted
27 Feb 2017
Accepted
11 Apr 2017
First published
12 Apr 2017

Anal. Methods, 2017,9, 2865-2872

Optimization of a methodology for simultaneous determination of twelve chlorophenols in environmental water samples using in situ derivatization and continuous sample drop flow microextraction combined with gas chromatography-electron-capture detection

M. Karimaei, K. Sharafi, M. Moradi, H. R. Ghaffari, H. Biglari, H. Arfaeinia and N. Fattahi, Anal. Methods, 2017, 9, 2865 DOI: 10.1039/C7AY00530J

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements