Abstract

The International Atomic Energy Agency (IAEA) uses information from characterization of uranium-bearing material samples to support the verification of states' declarations and to look for indications of possible undeclared materials and activities. Parameters of interest include, among others, the concentrations of elemental impurities, the patterns of rare earth elements (REE), and the isotope ratios of uranium and some impurity elements. These data along with other physical and chemical properties are being used to assess material quality, monitor chemical processes, and verify material origin. This manuscript describes a Sector Field-Inductively Coupled Plasma-Mass Spectrometry (SF-ICP-MS) procedure for the determination of 67 minor and trace elements, including REE, in uranium ore concentrates (UOC) and related compounds without matrix separation. All elements are measured within one instrumental sequence, which improves the timeliness of the analyses. A particular focus is set on developing a reliable quality control mechanism in the absence of uranium-bearing reference materials with certified concentrations of all analysed elements. In order to provide objective experimental evidence that the particular requirements for each specific parameter are fulfilled the validation process included (i) assessment of measurement uncertainties, (ii) analysis of available certified reference materials with uranium matrix (validated for 31 elements), and (iii) measuring test samples in the frame of four inter-laboratory comparison exercises (validated for 54 elements). The majority of the analysed elements could be measured with a satisfactory accuracy but determination of difficult-to-analyse elements, such as silicon and tungsten, requires a modified sample preparation and measurement procedure. An alternative ‘non-destructive’ analytical method, such as XRF or laser ablation ICP-MS, can be applied for direct analysis of non-dissolved residues that remain in the digests.

Graphical abstract: Development, validation and verification of an ICP-MS procedure for a multi-element analysis of uranium ore concentrates

Article information

Article type
Paper
Submitted
08 May 2017
Accepted
13 Sep 2017
First published
19 Sep 2017

J. Anal. At. Spectrom., 2017,32, 2226-2237

Development, validation and verification of an ICP-MS procedure for a multi-element analysis of uranium ore concentrates

S. F. Boulyga, J. A. Cunningham, Z. Macsik, J. Hiess, Maxim V. Peńkin and S. J. Walsh, J. Anal. At. Spectrom., 2017, 32, 2226 DOI: 10.1039/C7JA00180K

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements