An improved in situ technique for the analysis of the Os isotope ratio in sulfides using laser ablation-multiple ion counter inductively coupled plasma mass spectrometry

Abstract

A new technique is developed for in situ measurement of 187Os/188Os in sulfides using LA-MC-ICP-MS, with the improvement of the spatial resolution and quantification limits. In this study, an all-ion counter (all-IC) configuration is adopted for static concurrent measurement of 187Os + 187Re and 188Os along with 185Re. To cross-calibrate the detection efficiencies of the different ion counters, an Au standard solution is introduced into the instrument and its signal is measured in each ion counter in a dynamic peak jumping mode. Then, the signal intensities of IC1, IC2 and IC3 are all normalized to that of IC0. The sample-standard bracketing (SSB) method is adopted not only for the Re interference and Os mass fractionation correction but also for the monitoring of detection efficiency during continuous measurement. In detail, an in-house sulfide standard CR-1P (only spiking Re) is specifically employed for correcting interference 187Re on 187Os and monitoring detection efficiency variation in IC0 and IC1. The mass fractionation of Os and detection efficiency variation of IC1 and IC2 are calibrated against another in-house sulfide standard CO-1P (only spiking Os). Even though the 188Os intensity decreases approximately 20 times, our improved method still provides accurate and precise results of the 187Os/188Os ratio, whereas published methods that employ Faraday detectors for the signal collection do not work. Furthermore, the threshold of reliable interference correction of 187Re on 187Os is estimated by analysing synthetic sulfides with different Re/Os ratios. The analytical results demonstrate that the relative deviations (RD) of the calculated 187Os/188Os ratio will be limited to within 0.8%, 1%, 3% and 5%, when 187Re/188Os is lower than 0.055, 0.09, 0.42 and 0.75 (i.e., Re/Os: 0.012, 0.019, 0.089 and 0.159), respectively. In general, this new method significantly improves the spatial resolution and quantification limits of in situ measurements, which is beneficial for analysing sulfides in the depleted subcontinental lithospheric mantle (SCLM).

Graphical abstract: An improved in situ technique for the analysis of the Os isotope ratio in sulfides using laser ablation-multiple ion counter inductively coupled plasma mass spectrometry

Supplementary files

Article information

Article type
Technical Note
Submitted
20 Feb 2019
Accepted
10 Jun 2019
First published
10 Jun 2019

J. Anal. At. Spectrom., 2019,34, 1546-1552

An improved in situ technique for the analysis of the Os isotope ratio in sulfides using laser ablation-multiple ion counter inductively coupled plasma mass spectrometry

L. Zhu, Y. Liu, S. Jiang and J. Lin, J. Anal. At. Spectrom., 2019, 34, 1546 DOI: 10.1039/C9JA00066F

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements