Direct quantitative analysis of a cesium pencil in a simulated hot-cell using a remote-LIBS system equipped with a miniature collimator

Abstract

137Cs-containing borosilicate glass encapsulated in an SS tube, known as a Cs-pencil, is replacing the traditional 137CsCl powder used in blood irradiators for enhanced nuclear safety. Accurate quantification and investigation of the homogeneity of 137Cs distribution are integral aspects of its quality assessment. Herein, the first quantitative measurement of Cs-pencils using a remote-laser-induced breakdown spectroscopy (LIBS) system equipped with a miniature collimator is reported, which involves no sample preparation. A simulated hot-cell structure forcing the remote acquisition of the signal at >1 m was constructed for this analysis. The relationship between the signal intensity and solid angle of the collection created by the collimator with the remote distance was evaluated, and the result was verified both experimentally and theoretically. Moreover, studies were carried out to characterize the spectral noise embedded in the LIBS spectrum. The results reveal the inadequacy of the fitting algorithms in high noise spectra. With the use of a 266 nm laser for ablation and under ambient Ar conditions, the accuracy and precision for the determination of Cs was found to be 2.5–5% at the 1–1.5 m remote distances. The results obtained herein demonstrate that the miniature collimator-equipped remote-LIBS is suitable for analyzing the Cs-pencil produced by fissionogenic 137Cs and has significant potential to be employed as a routine analytical technique in the nuclear industry.

Graphical abstract: Direct quantitative analysis of a cesium pencil in a simulated hot-cell using a remote-LIBS system equipped with a miniature collimator

Article information

Article type
Paper
Submitted
17 Jun 2019
Accepted
23 Jul 2019
First published
23 Jul 2019

J. Anal. At. Spectrom., 2019,34, 1910-1918

Direct quantitative analysis of a cesium pencil in a simulated hot-cell using a remote-LIBS system equipped with a miniature collimator

M. Singh, R. K. Mishra, P. G. Jaison, C. P. Kaushik and A. Sarkar, J. Anal. At. Spectrom., 2019, 34, 1910 DOI: 10.1039/C9JA00207C

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements