Ion mobility spectrometry experiments should be carried out at high temperatures to reduce uncertainties in the measurement of reduced mobilities

Abstract

In ion mobility spectrometry (IMS), reduced mobility (K0) is an identification parameter of gas-phase ions but, frequently, these values are different whether there is contamination with moisture and other volatile compounds or not. We studied the effect of 2-butanol as a contaminant in IMS using electrospray ionization-IMS-mass spectrometry. The ion mobilities of valinol, phenylalanine, and tryptophan were measured with 0.0, 1.7, 3.4, 5.1, and 6.8 mmol m−3 of 2-butanol (0.065, 0.131, 0.196, and 0.261 ppmv, respectively) in the buffer gas at 100, 150, 200, and 250 °C. At 100 °C, K0 values of valinol, phenylalanine, and tryptophan ions decreased by −14.2%, −11.9%, and −10.3%, respectively, with 6.8 mmol m−3 of 2-butanol in the buffer gas. These different changes in mobilities were due to the formation of large 2-butanol : analyte ion clusters. The largest mobility reductions were obtained at 100 °C due to increased 2-butanol : analyte ion interactions, calculated using Gaussian. At 250 °C, the formation of clusters was hindered, and the injection of 2-butanol did not affect mobilities. That the largest mobility changes when contamination was present were obtained at the lowest temperature, as is almost always the case for field IMS instruments, indicates that drift-tube ion mobility measurements must be carried out at high temperatures to obtain reliable reduced mobilities. Contaminants of mass or polarity larger than that of 2-butanol will produce larger mobility changes. These mobility changes will be much larger at the relatively low temperatures used by most field instruments due to a balance between heating power consumption and the benefits of high temperatures. IMS at high temperatures could be the choice for the detection of illegal drugs, explosives, and biological and warfare agents in airports, ports, and customs to avoid false positives, ease the transit of passengers and cargo, increase the seizure of illegal substances, and save money and discomfort due to unnecessary delays produced by false positives.

Graphical abstract: Ion mobility spectrometry experiments should be carried out at high temperatures to reduce uncertainties in the measurement of reduced mobilities

Supplementary files

Article information

Article type
Paper
Submitted
19 Apr 2021
Accepted
31 May 2021
First published
31 May 2021

Anal. Methods, 2021,13, 2878-2887

Ion mobility spectrometry experiments should be carried out at high temperatures to reduce uncertainties in the measurement of reduced mobilities

R. Fernandez-Maestre and M. C. Daza, Anal. Methods, 2021, 13, 2878 DOI: 10.1039/D1AY00674F

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements