Issue 31, 2021

Simultaneous determination of four fertility-related hormones in saliva using disposable multiplexed immunoplatforms coupled to a custom-designed and field-portable potentiostat

Abstract

This work reports the first amperometric immunosensor for the simultaneous determination of four fertility-related hormones in saliva: progesterone (P4), luteinizing hormone (LH), estradiol (E2), and prolactin (PRL). The immune platform involves direct competitive (P4 and E2), and sandwich (LH and PRL) assays implemented onto functionalized magnetic microbeads (MBs). The amperometric transduction was performed upon placing the MBs-immunoconjugates onto each of the four working electrodes of the SPCE array (SP4CEs) and applying a detection potential of −0.20 V (vs. Ag pseudo-reference electrode) using the H2O2/hydroquinone (HQ) system. The achieved analytical and operational characteristics of the developed multiplexed immunoplatform showed a sensitivity that allows the determination of these hormones in saliva, and an adequate selectivity to analyse complex clinical samples. The bioplatform was employed for the determination of the set of four hormones in human saliva samples collected from individuals with different hormonal profiles. The results obtained using a conventional potentiostat were compared with those provided employing a novel low-cost custom-designed and field-portable quadruple potentiostat. Similar results were found which also agreed with those obtained by applying ELISA methods for the determination of single hormones.

Graphical abstract: Simultaneous determination of four fertility-related hormones in saliva using disposable multiplexed immunoplatforms coupled to a custom-designed and field-portable potentiostat

Article information

Article type
Paper
Submitted
22 Jun 2021
Accepted
16 Jul 2021
First published
21 Jul 2021

Anal. Methods, 2021,13, 3471-3478

Simultaneous determination of four fertility-related hormones in saliva using disposable multiplexed immunoplatforms coupled to a custom-designed and field-portable potentiostat

B. Arévalo, V. Serafín, J. F. Beltrán-Sánchez, J. Aznar-Poveda, J. A. López-Pastor, A. J. García-Sánchez, J. García-Haro, S. Campuzano, P. Yañez-Sedeño and J. M. Pingarrón, Anal. Methods, 2021, 13, 3471 DOI: 10.1039/D1AY01074C

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.