Abstract

This study presents our new observation on mass-independent fractionation (MIF) during Mo isotope measurement by MC-ICP-MS and investigates its implication for the application of the double spike (DS) technique. A mixture of NIST SRM 3134 Mo solution with a calibrated 97Mo–100Mo DS (1 : 1) was prepared for measurements over ∼15 h. The Mo isotopic data in this study show a good linear relationship in the scatter plots of ln(97/95Mo) vs. ln(98/95Mo), ln(97/95Mo) vs. ln(100/95Mo), and ln(98/95Mo) vs. ln(100/95Mo). Considering measurement errors on both vertical and horizontal axes, York linear regression was applied to linear fit. The slopes derived from the York linear regression are 1.33 ± 0.12 (2SE), 2.30 ± 0.20 (2SE), and 1.73 ± 0.14 (2SE), respectively. Given the error, the slopes are not exactly consistent with their theoretically expected values (1.49, 2.46, and 1.64) based on the assumption of mass-dependent fractionation (MDF). This study, for the first time, identifies MIF during Mo isotope measurement by MC-ICP-MS. DS calculation using theoretical mass fractionation factors shows that the analyzed NIST SRM 3134 Mo solution yielded δ98/95Mo values of 0.00 ± 0.04‰ (2SD, n = 17) relative to the originally calibrated NIST SRM 3134 Mo. However, using experimental mass fractionation factors calculated from the slopes of line relationships, there is a deviation of about 0.46‰ for δ98/95Mo. The results indicate that the calibrated isotope ratios of the standard and DS could not be the true values but approximate values for better matching with the MDF-based equations. After DS calculation using experimental and theoretical mass fractionation factors, respectively, an additional standard-sample bracketing (SSB) method was conducted. Both calculations gave identical and accurate δ98/95Mo values (0.00 ± 0.06‰), indicating that the SSB method is able to correct deviation from MIF during DS calculation. Accordingly, we proposed that an additional SSB method is recommended to acquire accurate isotope ratios during the application of the DS technique. This study highlights that researchers should pay attention to the possible existence of instrument-related MIF during isotope measurement.

Graphical abstract: Mass-independent fractionation during Mo isotope measurement by MC-ICP-MS: implications for application of the double spike technique

Supplementary files

Article information

Article type
Technical Note
Submitted
23 Dec 2022
Accepted
02 Mar 2023
First published
03 Mar 2023

J. Anal. At. Spectrom., 2023,38, 785-791

Mass-independent fractionation during Mo isotope measurement by MC-ICP-MS: implications for application of the double spike technique

Y. Zhao, W. Yang and C. Leng, J. Anal. At. Spectrom., 2023, 38, 785 DOI: 10.1039/D2JA00427E

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements