Internal standard optimization advances sensitivity and robustness of ratiometric detection method†
Abstract
We design a p-aminothiophenol (pATP) modified Au/ITO chip to determine nitrite ions in lake water by a ratiometric surface-enhanced Raman scattering (SERS) method based on nitrite ions triggering the transformation of pATP to p,p′-dimercaptoazobenzene (DMAB). Intriguingly, by using the SERS peak (at 1008 cm−1) from benzoic ring deforming as an internal standard instead of the traditional peak at 1080 cm−1, the detection sensitivity of the method was improved 10 times.