Robust determination of low Ti contents in zircon using LA–ICP–MS/MS with NH3 reaction gas

Abstract

The Ti-in-zircon thermometer is widely used in geoscience for constraining the temperatures of geological settings. Precise and accurate determination of Ti contents (∼1–10 μg g−1) in zircon is a prerequisite for the application of this thermometer; however, traditional laser ablation–single quadrupole–inductively coupled plasma–mass spectrometry (LA–SQ–ICP–MS) analysis of Ti contents in zircon is challenging due to the low abundance of the interference-free isotope of 49Ti (isotope abundance: 5.51%), resulting in poor precision (>30%, 2RSD), in particular for the high spatial resolution (laser spot size <20 μm) analysis. We have developed a robust technique for determining low Ti contents in zircon using LA–ICP–MS/MS with NH3 as the reaction gas. Using an iCAP TQ ICP–MS/MS instrument (Thermo Scientific, USA), high-purity NH3 was a more effective reaction gas than the commonly used 1 : 9 NH3 : He mixture gas, and a three-fold improvement in sensitivity was achieved using an N2 flow rate of 5.0 mL min−1. The reaction products of Ti, Ca, and Zr with NH3 were identified over a mass range of 40 to 160 amu. The reaction product 48Ti(14N1H)(14N1H3)4 (a mass shift of +83, expressed as (48+83)Ti) was measured to separate 48Ti+ from 48Ca+ and 96Zr2+. Interfering ions 48Ca+ and 96Zr2+ have low reaction rates (∼0.0046% and ∼0.07%) at a mass shift of +83, and corrections are not required for glass reference materials and zircon samples. Five zircon reference materials with Ti contents of 2.5–5.0 μg g−1 were analysed, and the analytical precision was better than 10%. Compared with the single quadrupole (SQ) mode, our method is more robust for determining low Ti contents in zircon at high spatial resolutions (laser spot size <20 μm), making it useful for the analysis of complex zircon grains, e.g., metamorphic zircon.

Graphical abstract: Robust determination of low Ti contents in zircon using LA–ICP–MS/MS with NH3 reaction gas

Supplementary files

Article information

Article type
Paper
Submitted
23 Aug 2024
Accepted
24 Oct 2024
First published
24 Oct 2024

J. Anal. At. Spectrom., 2024,39, 3060-3068

Robust determination of low Ti contents in zircon using LA–ICP–MS/MS with NH3 reaction gas

J. Niu, S. Wu, Y. Yang, H. Wang, C. Huang, L. Xu and L. Xie, J. Anal. At. Spectrom., 2024, 39, 3060 DOI: 10.1039/D4JA00304G

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements