Jonathan G.
Shackman
a,
Gabriella M.
Dahlgren
a,
Jennifer L.
Peters
a and
Robert T.
Kennedy
*ab
aDepartment of Chemistry, University of Michigan, Ann Arbor, MI, USA
bDepartment of Pharmacology, University of Michigan, Ann Arbor, MI, USA
First published on 22nd July 2004
A microfluidic device that incorporates continuous perfusion and an on-line electrophoresis immunoassay was developed, characterized, and applied to monitoring insulin secretion from single islets of Langerhans. In the device, a cell chamber was perfused with cell culture media or a balanced salt solution at 0.6 to 1.5 µL min−1. The flow was driven by gas pressure applied off-chip. Perfusate was continuously sampled at 2 nL min−1 by electroosmosis through a separate channel on the chip. The perfusate was mixed on-line with fluorescein isothiocyanate-labeled insulin (FITC-insulin) and monoclonal anti-insulin antibody and allowed to react for 60 s as the mixture traveled down a 4 cm long reaction channel. The cell chamber and reaction channel were maintained at 37 °C. The reaction mixture was injected onto a 1.5 cm separation channel as rapidly as every 6 s, and the free FITC-insulin and the FITC-insulin-antibody complex were separated under an electric field of 500 to 600 V cm−1. The immunoassay had a detection limit of 0.8 nM and a relative standard deviation of 6% during 2 h of continuous operation with standard solutions. Individual islets were monitored for up to 1 h while perfusing with different concentrations of glucose. The immunoassay allowed quantitative monitoring of classical biphasic and oscillatory insulin secretion with 6 s sampling frequency following step changes in glucose from 3 to 11 mM. The 2.5 cm × 7.6 cm microfluidic system allowed for monitoring islets in a highly automated fashion. The technique should be amenable to studies involving other tissues or cells that release chemicals.
Another important physiological function is secretion of chemical products from cells. A variety of chemicals are released from cells, including signaling molecules such as hormones or neurotransmitters, trophic factors, and metabolic products. Temporally resolved measurements of cell releasates are important in studying the regulation of the secretory process as well as technological pursuits such as drug development. Such measurements are typically performed by perfusing cells, collecting fractions, and then performing off-line analysis by immunoassays or other appropriate methods. The development of a microfabricated device that could miniaturize and automate such measurements would be a valuable tool in physiological studies. Recently, initial efforts toward such systems have been reported including a system to measure dopamine release that used microfluidics to trap a cell while an external microelectrode was manually positioned over the cell for measurement.18
In this work we report on a microfluidic device that perfuses a cluster of cells and monitors secretion at 6–10 s intervals using a rapid electrophoresis-based immunoassay. The system is applied to monitoring insulin secretion from single islets of Langerhans. Islets are 75–200 µm diameter spheroid microorgans located in the pancreas that contain 2000–4000 endocrine cells each. 70–80% of islet cells are β-cells that secrete insulin as part of the glucose homeostasis mechanism (for a review see ref. 19). Insulin secretion, stimulated primarily by glucose, has complex dynamics upon exposure to step increases in glucose concentration that include an initial burst of insulin release (first phase) followed by a lower rate of release (second phase) that is frequently oscillatory (for reviews see ref. 20 and 21). Insulin secretion regulation is of interest because impaired secretion is associated with type 2 diabetes, and drugs that treat diabetes target this process.
In previous work we demonstrated a microfabricated device that could detect insulin secreted from single islets using an electrophoresis-based immunoassay.22 The islet was housed in a small chamber on the chip that was continuously sampled by electroosmotic flow into a narrow channel. The resulting sample stream was mixed online with fluorescein isothiocyanate-labeled insulin (FITC-insulin) and anti-insulin antibody (Ab). The mixture was allowed to react as it was electroosmotically pumped along a heated channel prior to injection onto an electrophoresis channel where the bound (B) and free (F) FITC-insulin were separated. The ratio of the bound to free FITC-insulin (B/F) was used to quantify insulin. This device allowed electropherograms to be acquired at 15 s intervals and had a detection limit of 3 nM. Although this system demonstrated the feasibility of using microfabricated devices for chemical monitoring of live cells, it was limited in that cells were maintained in a quiescent solution and not perfused. Lack of perfusion resulted in serious limitations including inability to: (1) obtain truly dynamic measurements of secretion because insulin secreted from the islet continually increased in concentration within the chamber, (2) rapidly raise and lower concentration of drugs or secretagogues on the cell as required for many experiments, and (3) continually provide fresh nutrients to the cells, which limited the time that cells could be viable on the chip. In this work we describe modification of the chip that enables perfusion of cells while maintaining sampling. It is demonstrated that this modification enables vastly improved temporal resolution for monitoring, facile control of the cellular environment, and compatibility with longer-term measurements. The system is also demonstrated to be compatible with different types of physiological media.
Fig. 1 Microfluidic device to continuously monitor insulin release from islets of Langerhans with perfusion. (a) Layout of the device. Channels (lines) and access holes (circles) are drawn to scale. All channels were 6 µm deep. Electrical connections are shown as dashed lines. (b) Side-on, cutaway view of chip at the islet chamber. A single islet was contained in a 300 µm diameter chamber with an 80 µL fluid reservoir above. Pressure-driven fluid bathed the islet via the perfusion channel and flowed into the reservoir above. Solution around the islet was sampled by EOF through the sample channel. (c) Photograph of a completed 2.5 cm × 7.5 cm device. The larger ports contained 80 µL of solution and were capped during experimentation. The high pressure perfusion inlet (lower center port) was connected via PEEK™ tubing to a stainless steel pressure bomb containing a 2 mL vial of the perfusion solution (d). |
Microfluidic reservoirs applied after bonding were purchased from Upchurch Scientific (Oak Harbor, WA). 80 µL reservoirs (Upchurch N-131) were used for all ports except the high-pressure perfusion inlet, which used a 1/32″ tubing interconnect port (Upchurch N-124H). A photograph of the completed device is shown in Fig. 1c. A thin-film resistive heater (Minco Products, Inc., Fridley, MN) was taped to the chip underside to maintain the islet, immunoassay reagents, and the reaction channel at 37 °C, as monitored by digital thermometer (Fisher). During calibration and islet monitoring, 80 µL of the appropriate solution were placed in all reservoirs (excluding the perfusion inlet) and covered with plastic caps drilled with Pt electrode access holes. The cell chamber was perfused using a gas-pressure system off-chip illustrated in Fig. 1d. The perfusion inlet was connected to the high-pressure bomb via 80 cm PEEK™ tubing (0.20″ id, 1/32″ od, Upchurch); the PEEK™ tubing was flushed with water followed by the perfusion solution prior to chip connection. To change the perfusion media, the pressure was released from the pressure bomb via the purge valve, the fluid changed, and the PEEK™ tubing flushed with perfusion solution prior to reapplying pressure. Any air within the perfusion lines would cause the islets to be blown off the glass and disperse; hence, utmost care was taken when making the perfusion connections to maintain the outlet lower than the pressure bomb to sustain siphon flow.
Chips were conditioned daily by electroosmotically flowing 1 M NaOH through all channels, followed by 150 mM HEPES, pH 7.4, followed by the experimental solutions. After calibration or islet experiments, the chip was conditioned with 150 mM HEPES, pH 7.4, followed by the experimental solutions. This procedure allowed the same chip to be repeatedly used on successive days. Experiments on islets were performed using two different perfusion media. One medium was RPMI 1640 (a cell culture media) supplemented with 25 mM HEPES and either 3 or 11 mM glucose, adjusted to pH 7.4. When this media was used, 150 nM FITC-insulin was placed in the FITC-insulin reservoir and 75 nM antibody placed in the Ab reservoir (Fig. 1a). Both of these reagents were dissolved in PETA buffer (see composition above). Electrophoresis buffer in the gate and waste reservoirs consisted of 150 mM HEPES adjusted to pH 7.4 with NaOH. The applied voltage was 4 kV, with an electric field of 500 V cm−1 in the separation channel and an immunoassay reaction time of 60 s. Sample injection time was 1 s applied at 9 s intervals. Cells were perfused at 1.5 µL min−1 by applying 100 psi (He) to the pressure bomb.
Experiments were also performed with cells perfused with BSS (see composition above). For these experiments the immunoassay reagent concentrations were 100 nM for FITC-insulin and 50 nM for antibody. These reagents were dissolved in HEAT40 buffer (see composition above). Electrophoresis buffer was 150 mM HEPES adjusted to pH 7.4 with NaOH. The applied voltage was 5 kV, with an electric field of 600 V cm−1 in the separation channel and an immunoassay reaction time of 60 s. Sample injection time was 0.5 s applied at 5.5 s intervals. Cells were perfused at 0.6 µL min−1 by applying 50 psi (He) to the pressure bomb.
Concentrations of insulin were determined from each electropherogram by comparing B/F peak height ratios to a calibration curve. The calibration curve was obtained by pumping different concentrations of standard insulin in the perfusion media via the high-pressure bomb to the islet reservoir. Perfusion media flow rates were determined by weighing the chip (n = 3) before and after 30 min of perfusion and converting to volume using the density of the perfusion buffer.
Several methods were considered for pumping media into the chip. EOF was inadequate to produce the necessary flow rates, especially when using high ionic strength islet media, and would place a damaging electric field across the islet. A second method was use of a mechanical syringe pump (Model 55–3206, Harvard Apparatus, Inc., Holliston, MA) to drive the perfusion buffer. This technique failed due to high irreproducibility of flow. We then investigated the use of gas-pressure driven flow. Helium was used to minimize degassing and avoid oxidation of buffer constituents or insulin. As some physiological buffers, including the RPMI 1640 cell culture media, are based on bicarbonate and intended to be used in a 5% CO2 atmosphere, it was necessary to add a different buffering agent to these solutions (25 mM HEPES). The use of gas pressure proved to be simple and reliable route to pumping into the chip.
The perfusion inlet to the 300 µm diameter cell chamber was a comparatively small 6 µm deep channel, leading to the possibility of heterogeneous distribution of molecules that flow into the chamber. This situation was expected to be exacerbated by the presence of a 100–200 µm diameter islet within the cell chamber. To test this possibility, the cell chamber was imaged with and without an islet as 100 nM fluorescein was pumped into it from the perfusion channel at 1.5 µL min−1 (Fig. 2). Fig. 2a and b illustrates that the fluorescence intensity was homogenous across the cell chamber within the limits of the measurement technique. (Imaging was performed at 10× magnification resulting in a roughly 250 µm laser spot illuminating the 300 µm cell chamber. The Gaussian laser beam profile results in a peaked fluorescent intensity in the center of the chamber.) With an islet present, the fluorescence distribution was still relatively even, although it was apparent that the side of the islet opposite the flow input received a slightly lower concentration of fluorescein than the inlet side (Fig. 2d and e).
Fig. 2 Images and analysis of solution replacement within the islet chamber. (a) Fluorescence intensity across a cell chamber perfused with 100 nM fluorescein at 1.5 µL min−1. Numbers correspond to the pixel row scanned. Lines correspond to different locations in chamber as depicted in the fluorescent image in (b). (c) Brightfield image of chamber containing an islet. Perfusion inlet channel, sample outlet channel, and fluid flow directions are marked. (d) Same as (a) except with an islet present in the chamber. Scan line positions are shown in fluorescent image in (e). (f) Fluorescence intensity detected in reaction channel as perfusion media switched from 0 nM to 250 nM to 0 nM fluorescein in BSS. Solution changes are marked with arrows. 10% and 90% of maximal intensity are marked with vertical bars. |
In order to determine the efficiency of sampling a rapid change in the cell chamber, a 250 nM fluorescein solution in BSS was pumped into the chamber followed by a 0 nM fluorescein solution at 0.6 µL min−1 (Fig. 2f) while monitoring fluorescence in the reaction channel adjacent to the intersection of the immunoassay reagents and islet sampling channels. A delay time of 80 s was observed from commencing flow to detecting a fluorescence signal that was 10% of the maximum intensity. The rise time of the fluorescence increase from 10% to 90% maximum intensity was 30 s. Upon switching to blank solution the 80 s delay was again observed and the fluorescence dropped from 90% to 10% intensity over 70 s. These times represent the upper limit of temporal resolution because at least part of the spread of signal was due to the time required to rinse out the perfusion channel and cell chamber. As illustrated with actual islet measurements (see below), faster responses can be measured when the analyte molecule is produced directly within the chamber. These results also confirm that the perfusion system was adequate for rapidly changing concentration of drugs or nutrients (e.g. glucose) that were applied to the islet chamber.
Fig. 3 Calibration curves plotting B/F peak height ratios versus insulin concentrations obtained on the device while perfusing RPMI 1640 at 1.5 µL min−1. (a) Calibration of 5 nM to 1 µM insulin standards. Points were fitted by a simple logarithmic function with R2 = 0.995. (b) Calibrations obtained on consecutive days (Day 0 and Day 1) and two weeks later (Day 14) on the same device. |
We also determined the stability of immunoassay during continuous operation in a single day as this will limit the time that a single islet can be monitored. Fig. 4a illustrates the B/F calculated from 720 electropherograms collected at 10 s intervals for 2 h while perfusing the chip with 50 nM insulin. After an initial increase in B/F associated with the device reaching a stable temperature, the relative standard deviation (RSD) of B/F was 6%. Example electropherograms acquired at 20 min and 2 h are compared in Fig. 4b. At the end of 2 h the baseline in the electropherograms had begun to rise, apparently due to degradation of HEPES in the electrophoresis buffer; the electropherograms in Fig. 4b and c have been normalized to zero, resulting in the apparent lower absolute intensities in Fig. 4c relative to 4b due to the elevated baseline after 2 h. In addition, evaporation of all solutions except the perfusion media became a factor after this length of time. (The reservoirs storing the solutions were covered; however, they were only loosely capped to prevent negative pressure build-up in the reservoir as reagents were consumed.) This performance was a significant improvement over the non-perfusion system which could only be maintained for 30 min without intervention.22 The non-perfusion device may have suffered from degradation of insulin standards over time at atmospheric conditions, whereas the new device constantly replenished the standards that were maintained in a pressurized helium environment. Continual or periodic replenishment of the immunoassay reagent and electrophoresis solutions would likely allow operation of the perfusion chip even longer. Such replenishment could be achieved automatically by adding perfusion lines to the appropriate reservoirs or manually by pipetting fresh solutions into the chip.
Fig. 4 Stability for continuous operation. (a) Bound to free ratios (B/F) during a continuous 2 h perfusion of 50 nM insulin. Dashed box shows region used to calculate relative standard deviation (RSD). (b) Electropherograms collected during the experiments depicted in (a) at 20 min and 2 h. Peaks corresponding to free FITC-insulin (F) and bound to antibody (B) are marked. |
In addition to long term operation over a single day, we found that it was possible to re-use individual chips for multiple experiments. A single chip was utilized for over 6 months and several islets (discussed below) as part of this study. Such long-term operation was attributed to the use of the cleaning procedures outlined in the experimental section and using purified solutions within the chip.
Fig. 5 Monitoring insulin release during glucose step changes during perfusion of various buffers. (a) Electropherogram collected during on-line monitoring of a single islet during RPMI 1640 perfusion. (b) Insulin release of an islet as glucose was raised from 3 to 11 mM, followed by a return to 3 mM. Islet was perfused RPMI 1640 cell culture media at 1.5 µL min−1. (c) Electropherogram collected during on-line monitoring of a single islet during balanced salt solution (BSS) perfusion. (d) Insulin release of an islet as glucose was raised from 3 to 11 mM, followed by a return to 3 mM. Islet was perfused at BSS at 0.6 µL min−1. |
The results of monitoring single islets in RPMI 1640 demonstrate that the system has both the sensitivity and temporal resolution to detect oscillations in insulin secretion. After first phase and during oscillations, insulin levels returned to nearly basal levels in as little as 30 s, demonstrating that secreted insulin was effectively removed by the perfusion. Additionally, when islets were returned to 3 mM glucose from 11 mM, the levels returned to basal amounts within 60 s without any further pulses of secretion demonstrating that the glucose was rapidly washed out of the cell chamber as expected from the results in Fig. 2. Detection of such rapid changes in secretion also illustrate that sampling occurs with high temporal resolution.
Islets were monitored for up to an hour without any morphological signs of damage, such as cell lysis or dispersion of the islet membrane. Additionally, unregulated insulin secretion was not observed; islets maintained either regular oscillatory or stable, raised second phase secretion upon glucose stimulation, and steady basal levels of insulin release following glucose stimulation were observed, providing further evidence of the lack of cellular damage during the course of the experiments. The stability of islets indicates that the constant flow of buffer and electrical effects within the cell chamber are not detrimental to the cells over the time scale of the measurements used here. The possibility of electrical effects within the islet chamber arises from the use of electroosmotic flow, achieved by grounding the islet chamber and applying voltage at a downstream reservoir, for sampling. We estimate that the voltage dropped within the islet chamber is inconsequential (3 × 10−18 V) and the current density is just 38 µA mm−2. Given these low values, it is not surprising that the islets retained structural integrity and normal physiological responses within the chamber.
It is reasonable to expect that islet monitoring would be limited only by the reagent stability (2 h as mentioned above), because the islets are maintained on the device in conditions similar to those in an incubator, i.e. at 37 °C with a constant supply of fresh cell culture media. Longer-term measurements may be possible with chips that incorporate continual immunoassay reagent and separation solution replacement.
Ksp = [Ca2+]γCa2+[HPO42−]γHPO42− = 2.6 × 10−7 | (1) |
We then explored phosphate-free balanced salt solution (BSS), which is also commonly used as a defined media for physiological experiments.26,27 Utilization of BSS in combination with HEAT40 as the reagent solvent eliminated precipitation in the channels even during extended operation. Under these conditions it was necessary to thoroughly rinse out NaOH, used to clean the chips, during conditioning to prevent Ca(OH)2 from forming when BSS was introduced to the chip.
Upon switching to the BSS solution for perfusion and HEAT40 as reagent solvent, several smaller changes in the operation of the device were made including: lowering the perfusion flow rate from 1.5 µL min−1 to 0.6 µL min−1, reducing the immunoassay reagent concentrations from 150 nM to 100 nM FITC-insulin and 75 nM to 50 nM Ab, and increasing the separation field strength from 500 V cm−1 to 600 V cm−1. With these modifications, resolution of B and F was still adequate (see Fig. 5c); however, electropherograms could be stably acquired at 6 s intervals instead of 10 s.
Using the new operation parameters, the chip was used to monitor insulin secretion from individual islets perfused with BSS while the glucose concentration was stepped from 3 to 11 to 3 mM. Similar to RPMI 1640, biphasic responses were observed in all islets studied. Average basal levels prior to increasing the glucose concentration were 47 ± 34 pg min−1. The maximal response during first phase was 2360 ± 400 pg min−1, which was much higher than the first phase response observed with RPMI 1640. This greater initial burst of insulin release was likely due to the higher concentration of Ca2+ in BSS (2.4 mM) relative to RPMI 1640 ([Ca2+] = 0.424 mM),30 because entry of extracellular Ca2+ is a primary trigger of insulin secretion. The second phase of secretion was similar in the two buffers, both in terms of absolute level and oscillations. With BSS, 2 of 3 islets displayed oscillations with a period of 2.3 ± 0.2 min. (It is not uncommon for only a fraction of islets to display oscillations.)
These results demonstrate that the microfluidic device can perfuse and sample from both complex cell culture media and a minimal salt solution, while still performing an on-line immunoassay. The primary problem in changing buffers was clogging of microfluidic channels associated with precipitation of phosphate salts. These problems can be circumvented with appropriate tailoring of solutions.
This journal is © The Royal Society of Chemistry 2005 |