C.
Holladay
,
M.
Keeney
,
B.
Newland
,
A.
Mathew
,
W.
Wang
and
A.
Pandit
*
Network of Excellence for Functional Biomaterials, National University of Ireland, NFB building, IDA business park, Dangan, Newcastle, Galway, Ireland. E-mail: abhay.pandit@nuigalway.ie; Tel: +353 91 49 2758
First published on 8th September 2010
Quantification of eluted nucleic acids is a critical parameter in characterizing biomaterial based gene-delivery systems. The most commonly used method is to assay samples with an intercalating fluorescent dye such as PicoGreen®. However, this technique was developed for unbound DNA and the current trend in gene delivery is to condense DNA with transfection reagents, which interfere with intercalation. Here, for the first time, the DNA was permanently labeled with the fluorescent dye Cy5 prior to complexation, an alternative technique hypothesized to allow quantification of both bound and unbound DNA. A comparison of the two methods was performed by quantifying the elution of six different varieties of DNA complexes from a model biomaterial (collagen) scaffold. After seven days of elution, the PicoGreen® assay only allowed detection of three types of complexes (those formed using Lipofectin™ and two synthesised copolymers). However, the Cy5 fluorescent labeling technique enabled detection of all six varieties including those formed via common transfection agents poly(ethylene imine), poly-L-lysine and SuperFect™. This allowed reliable quantification of the elution of all these complexes from the collagen scaffold. Thus, while intercalating dyes may be effective and reliable for detecting double-stranded, unbound DNA, the technique described in this work allowed reliable quantification of DNA independent of complexation state.
There are a variety of techniques used in the quantification of DNA. Ultraviolet spectroscopy,5,6 intercalating dyes,7–17 and detection of radiolabeled DNA18–20 are the most commonly employed methods. Ultraviolet spectroscopy is simple and straightforward, using absorption of ultraviolet light at 260 and 280 nm to calculate DNA content. However, this method is effective for detecting DNA in relatively limited sample types as proteins (among other biomolecules) have significant extinction coefficients at the wavelengths used to quantify DNA (260–280 nm).21 To minimize error, DNA release studies that quantify with UV spectroscopy use clear, aqueous buffers such as phosphate-buffered saline or Tris–EDTA.5,21,22
A more robust option is the use of intercalating dyes. These dyes can accommodate themselves between base pairs in double-stranded nucleic acids. Once bound in position, the fluorescence of the molecule is greatly increased. Examples of commonly cited intercalating dyes include PicoGreen® (PG), ethidium bromide and its derivatives, and Hoechst.23 PG is commonly used both for quantifying genomic DNA and in quantifying the release of plasmid DNA from biomaterial scaffolds. When DNA is released without any bound proteins or polymers, intercalating dyes should be effective. However, the recent trend in gene delivery has been towards complexation of plasmids with a variety of novel carriers in order to improve transfection efficiency.3,24–35 Theoretically, tightly bound DNA is not detectable by an intercalating dye as the binding sites are not accessible.23,36,37 If it is possible to fully dissociate the complexes, this is not problematic. However, one of the barriers that reduce the efficiency of non-viral gene delivery is the fact that some transfection reagents are difficult to separate from plasmids.4
In this case, the third technique, radiolabeling is by far the most reliable. Plasmids are modified to include radioactive bases which can later be quantified with radioactive detection techniques. This technique has been used successfully in applications such as plasmid release from a PEG-based hydrogel.20,23 However, both the consumables and equipment are expensive and considerable health and safety precautions must be in place, thereby limiting the popularity of radiolabeling.
The technique described here is conceptually similar to radiolabeling, but instead of radioactive bases a fluorescent dye is covalently attached to purified plasmids, allowing fluorescence-based quantification of the DNA content. Fig. 1 shows a pictorial representation of the concept. It is hypothesized that this technique will allow accurate quantification of DNA concentration independent of complexation. PG and the fluorescent labeling method will be used to detect complexes formed with a variety of transfection reagents. Ultimately, the two techniques will be used to assess the release of a variety of complexes from a standard cross-linked collagen scaffold. If found effective, the technique will simplify and improve the reliability of nucleic acid release studies which are crucial in the characterization of scaffolds intended for gene delivery.
Fig. 1 Schematic representation of plasmid DNA detection by PG and FL as DNA elutes from a biomaterial scaffold. The PG dye molecule (represented by the dim yellow circle attached to the green hexagon) is only activated (bright yellow) when associated with unbound DNA, as in the top right corner. The fluorescent label (represented by the glowing green triangle) is glowing in all cases. Thus, PG is only able to detect unbound DNA while FL can detect all DNA. |
Polymer | Mass of DNA prepared/µg | Polymer:DNA weight ratio | Polymer concentration |
---|---|---|---|
a Linear poly(DMEAMA)–hyperbranched(PEGMEMA/EGDMA) (44 kDa). b Hyperbranched poly(DMAEMA/EGDMA) (14 kDa) where DMAEMA is 2-dimethyl amino ethyl methacrylate, PEGMEMA is poly(ethylene glycol) methyl ether methacrylate and EGDMA is ethylene glycol dimethacrylate. | |||
Partially degraded polyamidoamine (SuperFect™) | 10 | 15:1 | 3 mg ml−1 |
Polyethylene imine (PEI) | 10 | 2:1 | 3 mg ml−1 |
Poly-L-lysine (PLL) | 10 | 2:1 | 3 mg ml−1 |
Lipofectin™ (Lipo) | 10 | 5:1 | 1 mg ml−1 |
PD-b-P/Ea | 10 | 10:1 | 3 mg ml−1 |
PD-Eb | 10 | 6:1 | 3 mg ml−1 |
PLL (40–60 kDa) and PEI (25 kDa) were used as purchased from Sigma, SuperFect™ was purchased from Qiagen and Lipofectin™ from Invitrogen. Deactivation enhanced atom transfer radical polymerization (ATRP) as described by Tai et al. was used to synthesize PD-b-P/E and PD-E.39 ATRP is a particular type of controlled radical polymerization developed in the lab of Dr K. Matyjaszewski at Carnegie Mellon University. It uses a special catalyst which adds one or more monomers at a time to a growing chain. The synthesis process is easily regulated by adjusting temperature or other reaction conditions, thus allowing precise control of polymer structure.40
Fig. 2 Comparison of PG fluorescence levels of 1 µg of plasmid after 7 days as a function of complexation agent. The blank-subtracted fluorescence intensity of the plasmid decreased approximately 20 fold when the plasmid was complexed with SF. The signal in the PEI and PLL samples was negligibly different to the blanks. Lipo, PD-E, and PD-b-P/E retained approximately 43%, 71%, and 69% of the fluorescent intensity obtained with uncomplexed plasmids. Data expressed as mean ± standard deviation, n = 3, p < 0.05. |
The standard curves in Fig. 3a have acceptably high Pearson's coefficients (>0.99), thus implying quantification of complexes prepared with Lipo, PD-E, and PD-b-P/E is feasible using the PG technique. However, complexes prepared with SF, PEI, and PLL were not reliably detected with PG after 7 days, as shown in both Fig. 2 and 3b, which shows the standard curves prepared with the three complexation agents. Regression analysis of the 4 curves in Fig. 3b indicated no significant relationship between the PG signal and the mass of plasmid, and thus the curves could not be used for quantification of samples.
Fig. 3 Sample standard curves measured with PG 7 days after preparation. The standard curves for Lipo, PD-E, and PD-b-P/E (a) were acceptable, with Pearson's coefficients of 0.99, 0.97, and 0.98 respectively. For SF, PEI, and PLL, however, regression analysis indicated no significant trend, and thus the standard curves could not be used for sample quantification (b). The standard curves prepared with free plasmid are included in both figures for comparison. The fluorescence intensity from the plasmid reached the detection maximum by 0.5 µg which is why the values reach a plateau. Thus, PG could not be used to quantify DNA content in containing SF, PEI, or PLL complexes. Data expressed as mean ± standard deviation, n = 3, p < 0.05. |
In contrast to the PG technique, the standard curves measured with FL after 7 days, as shown in Fig. 4, all have statistically significant slopes. The lowest Pearson's coefficient for these curves was 0.994. Thus, all of these standard curves were reliable and could be used for sample quantification. While there is quenching (up to 60% reduction in maximum fluorescence value), the standard curves for each type of complex should have the same amount of quenching as the samples, and thus can be reliably used for quantification.
Fig. 4 Sample standard curves measured via FL. While there is variability in the slopes of the curves, the lowest Pearson's coefficient was 0.994 and all trends were significant. Thus, the DNA quantification is reliable with all six varieties of complexes. A standard curve prepared with free plasmid DNA was included as well to show that there is a decrease in the slope due to complexation, but all slopes are linear. Data expressed as mean ± standard deviation, n = 3, p < 0.05. |
As a final test of the technique, elution from a common biomaterial scaffold was measured for each of the six complex varieties. The general shapes of the curves (as shown in Fig. 5) are similar to other elution curves described in the literature.10 Lipo was observed to have the lowest initial release, but had the highest rate of release over the first 24 hours (about 81% of plasmid was released from the scaffold), after which no significant elution was observed. PLL had relatively low encapsulation efficiency, with approximately 50% of the complexes detected in the wash solution at time 0. SF, PEI, PD-E, and PD-b-P/E were all observed to release continuously over 7 days. When the same wells were analyzed using PG, as shown in Fig. S1†, significantly lower plasmid release was observed (30–50% lower plasmid release was observed using the PG technique).
Fig. 5 Plasmid release measured with FL with a variety of complexation agents as a function of time. All scaffolds were loaded with 1 µg of complexes and treated identically afterwards. Data expressed as mean ± standard deviation, n = 3, p < 0.05. |
All of the FL standard curves shown in Fig. 4 had high determination coefficients (>0.988), while PG was only effective in detecting three out of six varieties of complexes (Fig. 3), and with slightly lower coefficients of determination (0.945 to 0.983). As discussed, the three varieties of complex that PG did not detect (SF, PEI, and PLL) are among the most popular transfection reagents described in gene delivery literature.28,30 These reagents bind DNA extremely effectively, thereby interfering with PG binding. However, the three of the six compounds were detectable. These were Lipo (Lipofectin™), a liposome, PD-E, a hyper-branched polymer, and PD-b-P/E, a linear polymer with a hyper-branched PEG-based modification. While this is not necessarily a reflection of their transfection efficiency, it may reflect differences in how the reagents bind DNA.
For example, in a recent release study, Hoechst, another intercalating dye, was seen to have similar fluorescence levels before and after DNA was combined with Lipo.46 This may be because Lipo allows the DNA to remain in an expanded form,47 and thus binding sites are still largely accessible. Both PD-E and PD-b-P/E use a tertiary amine (present in the DMAEMA monomer) in contrast to SF, PEI and PLL, which use primary amines (along with others) to bind DNA. Differences in DNA binding strength between these types of amines may arise due to the varying surrounding alkyl groups (both through steric and inductive effects). As PEI, PLL and SF are all high in primary amines, they may cause structural changes in the DNA which affects intercalation of PG. Lipo, PD-E and PD-b-P/E may complex DNA with less alteration of the structure of the molecule, thus allowing some PG binding to occur. It should be noted that while PG is able to detect the complexes, there is not necessarily a higher fraction of free DNA in the solutions.
A significant advantage of this technique is that the standard curves are prepared and detected under precisely the same conditions and at the same time as the samples and thus both user variability and batch-to-batch variability can be minimized. When standard curves are prepared at the time of analysis, a significant, systematic error is introduced because the signal intensity is up to 12% higher in freshly prepared samples. In this technique, the standard curves are prepared and detected at the same time as the samples, so while there is quenching in the FL signal after complexation (up to a maximum of 40% in the case of Lipo), all samples are affected by the same factors and no systematic errors should be introduced.
The elution curves shown in Fig. 5 are similar in shape to those described elsewhere in the literature,8,48 but with a higher maximum release than most curves measured using PG. Elution curves could not be calculated using the PG standard curves for SF, PEI, or PLL. Comparing the curves obtained by measuring the same samples with both techniques in Fig. S1†, it is evident that FL detects significantly more DNA than PG. The difference between the curves is difficult to explain, as the same wells are used and therefore each sample contains precisely the same amount of DNA. Potentially, breakdown products from the collagen scaffold could have bound some fraction of the DNA, interfering with PG detection, or otherwise altered the microenvironment so as to decrease the overall PG signal. The FL elution curves, however, should not be affected.
The only factor that appears to have affected the FL is that complexation results in some loss of signal intensity (0–60% reduction in total fluorescence). This may be related to quenching, but is irrelevant because the standard curves are prepared with the complexes and thus the standard curves have the same degree of signal reduction as the samples. The same is true of photo-bleaching, because all samples are prepared and detected at the same time, and thus all samples should be affected to the same extent by any external factors.
In a previous study, complex elution from a scaffold was calculated using PG.38 However, minimal elution was observed at all time points after 2 days while transfection was observed over almost 14 days. The authors suggested that cells migrating through the scaffold might be exposed to complexes bound to the proteins of the scaffold as a possible explanation for the extended high levels of transfection with essentially no elution.38 However, considering the finding that PG cannot readily detect SF complexes after 7 days, the most likely explanation is that the PG dye could not bind to the complexes and thus it was only able to detect the complexes prepared within the first 48 hours.
The comparison between the complex release in buffer TE and 10% serum media highlights the importance of the elution media. It has been observed in vivo that negatively charged proteins in serum interact with positively charged DNA complexes.44In vitro, Fig. 6 highlights the difference in elution in a typical serum-free buffer, and in 10% serum cell-culture media. In the presence of negatively charged serum proteins, the initial release of complexes was >15% higher and the average cumulative release after 7 days was approximately 10% higher. Thus, where possible, 10% serum media should be used as it is a far more accurate representation of in vivo conditions and yields significantly different elution curves. This decreases the applicability of UV spectroscopy for DNA quantification as the accuracy and sensitivity of UV-based DNA quantification are greatly decreased by the addition of proteins and carbohydrates into the solution.21
Fig. 6 Comparison between elution of SF complexes in serum-free buffer TE and in 10% serum media. The plasmid release is significantly higher in serum than in buffer, and the encapsulation is about 18% higher in buffer TE. Data expressed as mean ± standard deviation, n = 3, p < 0.05. |
A combination of the PG and FL techniques may provide a more complete understanding of the elution process. PG is highly effective at detecting free DNA, while FL indicates the total amount of DNA in the well. Thus, it is possible to calculate the total amount of DNA eluted with the FL technique and the fraction of the DNA that is unbound with PG. To illustrate, Fig. S2† compares the fractions of free DNA and total amount of DNA eluted after 7 days. The lowest fraction of free DNA was seen with PLL (∼1.1%) and the highest fraction with PD-b-P/E (15.4%). This could be valuable for understanding the mechanisms of elution-based transfection, as free DNA has very low transfection efficiency in vitro. It could also be valuable for the characterization of new DNA-binding polymers like PD-E and PD-b-P/E.
Quantification of complexed DNA via fluorescent labeling was found to be significantly more reliable than detection of those complexes with an intercalating dye such as PicoGreen®. The intercalating dye could not readily detect SF, PEI and PLL complexes and could only detect a fraction of Lipo, PD-E and PD-b-P/E. The fluorescently labeled plasmids yielded perfect standard curves 7 days after preparation, with minimal loss in sensitivity. Sample elution curves from a commonly used biomaterial were obtained. When compared to a previously published curve obtained with PG, these elution curves highlight the limitations of the PicoGreen® assay—limitations which the fluorescent labeling technique is not subject to. Thus, quantification of complexed DNA can be accurately preformed over time by covalently labeling plasmids with a fluorescent dye prior to complexation.
PG | PicoGreen® |
FL | fluorescent labeling technique |
PAMAM | polyamidoamine dendrimer |
GLuc | Gaussia princeps luciferase |
DMAEMA | 2-dimethyl amino ethyl methacrylate |
PEGMEMA | poly(ethylene glycol) methyl ether methacrylate |
EGDMA | ethylene glycol dimethacrylate |
PD-b-P/E | linear poly(DMEAMA)–hyperbranched(PEGMEMA/EGDMA) |
PD-E | hyperbranched poly(DMAEMA/EGDMA) |
SF | SuperFect™ |
PEI | poly(ethylene imine) |
PLL | poly-L-lysine |
Lipo | lipofectin™ |
TE | Tris–EDTA buffer |
Footnotes |
† Electronic supplementary information (ESI) available: Figures showing total and free plasmid release and a comparison of release profiles quantified by the two methods. See DOI: 10.1039/c0nr00456a |
‡ Research support: Funding for this project was provided by Science Foundation Ireland, Strategic Research Cluster under Grant No. 07/SRC/B1163. |
This journal is © The Royal Society of Chemistry 2010 |