Hongling Shena,
Xu Jiaa,
Qingyan Mengab,
Wenjie Liu*ab and
Herbert H. Hill Jr.c
aCollege of Life Science, Tarim University, Alar, Xinjiang 843300, China. E-mail: lury@sina.com; Tel: +86 15569200251
bXinjiang Production & Construction Corps Key Laboratory of Protection and Utilization of Biological Resources in Tarim Basin, Alar, Xinjiang 843300, China
cDepartment of Chemistry, Washington State University, Pullman, WA 99163, USA
First published on 23rd January 2017
Fourier transform ion mobility spectrometry (FT-IMS) is a useful multiplexing method for improving the duty cycle (DC) of IMS from 1 to 25% when using an entrance and exit ion gate to modulate the ion current with a synchronized square wave chirp. To improve sensitivity further, this paper investigated the performance of FT-IMS using a frequency modulated single gate IMS system coupled to multi emitter array electrospray ionization. Signal to noise ratio, resolving power, linear range and limit of detection were evaluated using various tetra alkyl ammonium salts. The results demonstrated that a significant signal to noise ratio improvement of 4–6 times is achievable with the multi emitter array over that achieved with a single electrospray needle. The limits of detection for various tetraalkylammonium ions declined by 4 times on average and showed improved spectra quality compared to the signal averaging method. Multi emitter array electrospray ionization improved the ion intensities effectively, and Fourier transform IMS improved the duty cycle to 50% and hence improved the signal to noise ratio further.
Historically, Fourier transform ion mobility spectrometry (FT-IMS) was the first multiplexing ion mobility method used to improve the IMS duty cycle.14–16 By applying a synchronized frequency sweeping square wave to two Bradbury–Nielsen (BN) gates installed at the beginning and the end of the drift region, the duty cycle of FT-IMS was improved to 25%, producing an ion current interferogram. The interferogram could be transformed to an ion mobility spectrum by fast Fourier transform (FFT). Besides the improvement of duty cycle which resulted in a signal to noise ratio enhancement about 3 times, FT-IMS also reduced the peak tailing and baseline shift, thus improving confidence in the identification of the ions. However, when combine electrospray ionization to FT-IMS, only limited signal to noise (S/N) about 1.4 was achieved which is far lower than theoretic estimation.14 Another main concern about FT-IMS is the discriminative effect against bigger molecules because of BN gate depletion effect. For traditional FT-IMS uses two ion gate, this doubled the gate depletion effect and hence decreased the ion transmission for low mobility ion species. Another alternative multiplexing technique is Hadamard transform ion mobility spectrometry (HT-IMS). Instead of frequency modulated ion gate, HT-IMS utilizes a binary pseudo random sequence applied to the B–N gate, followed by Hadamard transform to retrieve the deconvoluted IMS spectrum.17 Though HT-IMS needs no modification to the IMS hardware and less computationally intensive, the false peaks caused by Hadamard transform are very hard to discriminate from real signal and hence increased the possibility of false alarm and misidentification.18
Multinozzle emitter array (MEA) has been applied to the mass spectrometry to improve the ionization efficiency and demonstrate many unique advantages such as the total electrospray current of the multinozzle emitter array is approximately proportional to the square root of the number of nozzles, thus increasing the sensitivity of mass spectrometry.19 Another attractive advantage is by reducing the flow rate of individual emitter, MEA showed high tolerance to water content, less ionization fluctuation, and equal mole ionization response to various compounds.20–23 The total flow rate of electrospray also improved to several micro liters and reduced the residence time in transfer tubing and improved the post-column performance when coupled to liquid chromatography separations.
For electrospray ionization and traditional drift tube ion mobility spectrometry are both atmosphere technique, thus MEA could improve the performance of ESI-IMS by generating and introducing more ions for separation.24 Also, the major challenge of MEA coupled to mass spectrometry is the limited transmission efficiency through the atmosphere-vacuum interface. This limitation does not exist for ambient pressure drift tube ion mobility spectrometry. In theory, all ions generated from an MEA could be introduced into an IMS drift tube for separation and detection.
The objective of this paper is to improve the performance of electrospray ionization ion mobility spectrometry by increasing the ion throughput via Fourier transform using the “external” ion gate, and by improving the ionization efficiency of electrospray, which generate more ions through multi emitter array electrospray ionization. The signal to noise ratio, resolving power, limit of detection, dynamic range and calibration curve of both traditional signal average mode and Fourier transform mode are compared and the sampling rate, chirp frequency range, chirp frequency rate and various apodization function are optimized to achieve higher S/N ratio.
The IMS tube, used in this study was 23.2 cm in total length, with a 6.4 cm desolvation region and a 16.8 cm drift region. The IMS drift tube was maintained at 120 °C temperature. A Bradbury–Nielsen type ion gate separated the desolvation region and drift region. The ion gate voltage was optimized to cut off the signal to zero while using corona discharge ionization source, which is ±32 V in this instrument. The first ring of the desolvation region was maintained at a constant electric potential of 10.00 kV and the gate reference voltage was measured 9.02 kV. The stacked stainless steel rings were separated from each other by insulating ceramic rings and were connected externally by resistors creating a potential gradient to guide the ions towards the Faraday plate. A counter current flow of dried nitrogen drift gas was introduced from the back of the Faraday plate to facilitate desolvation of ions and sweeping out any neutral molecules. A Keithley 427 amplifier (Keithley Instruments, Cleveland, OH) was utilized for signal amplification. A house made Labview (National Instruments, Austin, TX) based data acquisition (DAQ) software was used for chirp generation, gate driver control, and fast Fourier transform the interferogram to mobility spectra. The gating electronics is described in detail elsewhere. Two Bertan205B high voltage power supply boxes obtained from Bertan (Hicksville, New York) were used for constant voltage supply to the ESI needle and the first ring of drift tube (Table 1).
Experimental parameter | Conditions |
---|---|
ESI flow rate | 2 μL min−1; 10 μL min−1 (MEAs) |
ESI bias | 2.8 kV, 3.8 kV (MEAs) |
Drift tube voltage | 10.00 kV |
Voltage on the ion gate | 9.02 kV |
Drift gas | Nitrogen |
Drift gas flow rates | 900 mL min−1 |
Drift tube temperature | 120 °C |
Atmospheric pressure (Alar, Xinjiang) | 728 Torr |
Gate pulse | 0.2 ms |
Methanol was LC-MS grade and obtained from Fisher Scientific (Ottawa, Canada) without any further treatment. Individual 50 μM solutions of these salts were made in a 90:10:0.1% mixture of methanol, water, and acetic acid respectively. From these stock solutions a mixture containing all of the quaternary ammonium salts was constructed with concentrations ranging from ∼0.5 nM to 50 nM. This range was adequate to explore the limit of detection, linear range of the evaluated system on analytes in mixtures of various concentrations. All solutions were infused directly into the electrospray ionization source using a syringe pump (KD Scientific).
(1) |
(2) |
Reduced mobility constant K0 of T3A, T5A, T6A, T7A and T8A were calculated with eqn (2), with the results of 1.34, 1.00, 0.86, 0.78 and 0.71 cm2 V−1 s−1 respectively.
The resolving powers of each compound were calculated with the drift time (td) of the ion divided by the temporal peak width-at-half-height (w0.5) of the ion mobility peak. This relationship is given by eqn (3):
(3) |
For FT-IMS, the drift times of ions were not measured directly from arrive times of ions. Instead, the calculation of drift time was from time domain peaks transformed from frequency domain interferogram. Because FFT algorithm generates result from each time point of interferogram and expresses with a uniform unit of hertz, the drift time of each peak was calculated by eqn (4):
(4) |
(5) |
The generated sinusoid frequency sweeping signal was then converted to square wave signal by assign all positive values of eqn (5) to positive ones and all negative values to zeros using a sign function. For visual reason only the first 200 ms of a spectrum last for 1000 ms was shown in Fig. 1. This square wave was then sent to the gate controller to switch the ion gate and obtain a raw data which contain the frequency information via NI data acquisition board. The raw data is then multiplied with the square waveform to obtain the time domain interferogram. To avoid baseline distortion because of “frequency leakage”, a window function is used to apodize the beginning and ending part of interferogram to zero, then the truncated interferogram is transformed to frequency domain ion mobility spectra by fast Fourier transform (Fig. 2).
Fig. 1 Diagram of atmosphere pressure Fourier transform multi electrospray ionization ion mobility spectrometer. |
Fig. 3 shows ion mobility spectra from signal average mode and Fourier transform when using both single ESI and multi ESI. All four spectra are recorded with the same acquisition time. For signal average mode, the optimum flow rate of ESI solvent was screened using a 90:10 methanol to water with 0.1% glacial acetic acid to obtain a stable ion mobility spectrum. Though we use a relative high drift tube temperature, the limited desolvation capability decreased the performance of ion mobility resolving power for solvent peaks when ESI flow rate increased to 2 μL min−1. There is an obvious tailing for solvent peaks under signal averaging mode. Multi ESI using a flow rate up to 10 μL min−1 and the solvent peak tailing became more serious for insufficient desolvation. For Fourier transform IMS only responds to ions resonance between ion gate and detector with characteristic sweeping frequency, those tailing ions decompose in drift tube and exhibit variable drift time are discriminated to form a mobility peak and only contributed to noise. The solvent peaks under FT-IMS with both single ESI and multi ESI show improved resolving power compared to signal averaging mode respectively. For other peaks, the resolving power of FT-IMS are decreased about 10–15% compared to those of SA mode. One possible reason for this decreased resolving power is to suppress the “frequency leakage” for Fourier transform, both low frequency component and high frequency component of ion interferogram are curtailed using window function and thus decreased the high frequency end useful signal. Another reason is the drift time delay between the gating signal and detected signal, which means the high frequency modulated ion signal aren't detected by the IMS detector and decreased the resolving power of FT mode slightly. Table 2 lists a comprehensive comparison between SA mode and FT mode using single ESI and multi ESI. In all experiments, the acquisition time are set to 2 seconds. Signal averaging using a 0.2 ms gate pulse width. Correspondingly, FT mode using a maximum frequency of 5 kHz, producing a minimum gate pulse of 0.2 ms. Multi ESI also showed slightly decreased resolving power compared to single ESI, this could attribute to improved ion density results in higher ion–ion repulsion and extra peak broadening due to charge–charge repulsion.
Fig. 3 Typical ion mobility spectra from single ESI signal average (SA) mode; single ESI FT mode; multi ESI SA mode, multi ESI FT mode for T3A, T5A, T6A, T7A and T8A mixed standards. |
Ionization source | Mobility peaks | Fourier transform | Signal averaging |
---|---|---|---|
Multi ESI | Solvent peak | 24.39 ± 0.42 | 21.10 ± 0.68 |
T3A | 34.73 ± 0.42 | 44.26 ± 1.97 | |
T5A | 49.74 ± 0.39 | 63.44 ± 0.37 | |
T6A | 64.06 ± 1.60 | 71.97 ± 1.03 | |
T7A | 70.446 ± 1.60 | 73.78 ± 0.93 | |
T8A | 79.69 ± 0.29 | 82.31 ± 0.35 | |
Single ESI | Solvent peak | 25.22 ± 0.44 | 22.93 ± 0.69 |
T3A | 54.01 ± 0.95 | 56.23 ± 1.52 | |
T5A | 57.60 ± 1.26 | 65.67 ± 1.54 | |
T6A | 61.67 ± 0.73 | 79.35 ± 0.97 | |
T7A | 71.08 ± 1.26 | 86.41 ± 1.19 | |
T8A | 90.06 ± 0.32 | 97.05 ± 0.89 |
Fig. 4 The influence of sampling rate to Fourier transform ion mobility spectrometry using multi ESI and single ESI. |
The calibration curves of 5 tetraalkylammonium ions were obtained with six mixed standard solutions of T3A, T5A, T6A, T7A and T8A ranging from 0.5 to 50 nmol L−1, diluted in 90:10 ESI solvent with 0.1% acetic acid. For signal average mode, each spectrum was averaged from 10 iterations of work cycle, and each work cycle were the averaging of 20 single spectrum scan. For FT-IMS, the acquisition time was set to match the signal averaging. The calibration curves for all 5 TXAs showed good linear relationship between the analyte concentration and IMS response, over the concentration range studied with an instrumental detection limit of 0.0223, 0.0376, 0.0557, 0.0745, and 0.0875 μg mL−1 for TxAs studied for FT-IMS using multi ESI. For signal average mode using single ESI, the corresponding limit of detection are 0.1301, 0.1329, 0.1638, 0.2032 and 0.2267 respectively (Table 3). The limit of detection for each ion is calculated at three times the signal to chemical noise ratio. The calibration curves showed correlation coefficient of 0.9973, 0.9985, 0.9968, 0.9979 and 0.9984 for T3A, T5A, T6A, T7A and T8A using multi ESI FT-IMS respectively. Each point on the calibration curve is the average result of three replicate measurements. There are no significant differences in correlation coefficient between FT and SA mode (Table 4).
Ionization | Compound | LOD (FT) | LOD (SA) | ||
---|---|---|---|---|---|
nmol mL−1 | μg mL−1 | nmol mL−1 | μg mL−1 | ||
Multi-ESI | T3A | 0.0838 | 0.0223 | 0.2486 | 0.0662 |
T5A | 0.0996 | 0.0376 | 0.2657 | 0.1006 | |
T6A | 0.1282 | 0.0557 | 0.3742 | 0.1626 | |
T7A | 0.1509 | 0.0745 | 0.4115 | 0.2019 | |
T8A | 0.1602 | 0.0875 | 0.4774 | 0.2611 | |
Single ESI | T3A | 0.1301 | 0.0346 | 0.3253 | 0.0865 |
T5A | 0.1329 | 0.0503 | 0.3921 | 0.1484 | |
T6A | 0.1638 | 0.0842 | 0.4716 | 0.2049 | |
T7A | 0.2032 | 0.0997 | 0.5619 | 0.2757 | |
T8A | 0.2267 | 0.1243 | 0.6516 | 0.3563 |
ESI | Compounds | FT | SA | ||||
---|---|---|---|---|---|---|---|
Linear equations | R2 | Linear range (μmol mL−1) | Linear equations | R2 | Linear range (μmol mL−1) | ||
a y: peak high; x: sample concentration, μmol mL−1; R2: squared correlation coefficient.b FT: Fourier transform mode SA: signal-averaging mode. | |||||||
Multi-ESI | T3A | y = 71.04x + 107.79 | 0.9973 | 0.156–25 | y = 0.0064x + 0.0176 | 0.9981 | 0.312–25 |
T5A | y = 63.23x + 114.53 | 0.9985 | 0.156–25 | y = 0.0045x + 0.0178 | 0.9961 | 0.312–25 | |
T6A | y = 46.51x + 55.86 | 0.9968 | 0.312–25 | y = 0.0044x + 0.0118 | 0.9988 | 0.625–25 | |
T7A | y = 34.70x + 55.80 | 0.9979 | 0.312–25 | y = 0.0038x + 0.0081 | 0.9973 | 0.625–25 | |
T8A | y = 30.37x + 32.93 | 0.9984 | 0.312–25 | y = 0.0035x + 0.0057 | 0.9972 | 0.625–25 | |
Single ESI | T3A | y = 54.57x + 8.84 | 0.9957 | 0.312–50 | y = 0.0042x + 0.0148 | 0.9975 | 0.625–50 |
T5A | y = 51.57x + 51.33 | 0.9970 | 0.312–50 | y = 0.0034x + 0.0125 | 0.9958 | 0.625–50 | |
T6A | y = 32.43x + 27.59 | 0.9965 | 0.312–50 | y = 0.0032 + 0.0057 | 0.9960 | 0.625–50 | |
T7A | y = 21.32x + 24.09 | 0.9957 | 0.312–50 | y = 0.0026x + 0.0050 | 0.9963 | 0.625–50 | |
T8A | y = 17.10x + 15.73 | 0.9975 | 0.312–50 | y = 0.0024x + 0.0030 | 0.9952 | 0.625–50 |
This journal is © The Royal Society of Chemistry 2017 |