Jianhua Wang‡
a,
Lishen Zhang‡b,
Xiaochun Li*a,
Xiaoliang Zhanga and
Hua-Zhong Yu*ab
aCollege of Biomedical Engineering, Taiyuan University of Technology, Taiyuan, Shanxi 030024, China. E-mail: lixiaochun@tyut.edu.cn
bDepartment of Chemistry, Simon Fraser University, Burnaby, British Columbia V5A 1S6, Canada. E-mail: hogan_yu@sfu.ca
First published on 26th July 2019
Nowadays quantitative chemical analysis is usually costly, instrument-dependent, and time-consuming, which limits its implementation for remote locations and resource-limited regions. Inspired by the ancient papercutting art (kirigami), we herein introduce a novel cut-and-paste protocol to fabricate 3D microfluidic paper-based analytical devices (μPADs) that are suitable for on-site quantitative assay applications. The preparation of the device is fast, simple, and independent of any lithographic devices or masks. Particularly designed reaction “channels” were pre-cut from a piece of filter paper, then assembled back to the silanized, superhydrophobic paper pads. The different layers of the device were assembled using a chemically-inert adhesive spray. The fabricated device has high efficiency of liquid handling (up to 60 times faster than conventional methods) and it is particularly inexpensive. Beyond the benchtop fabrication advantage, in conjunction with a custom mobile app developed for colorimetric analysis, we were able to quantify representative environmental contaminants (i.e., the amount of Cr(VI) and nitrite ions) in various water samples with the cut-and-paste μPADs (namely kPADs). Their detection limits (0.7 μg mL−1 for Cr(VI) and 0.4 μg mL−1 for nitrite ions, respectively) are comparable with conventional spectrophotometric methods, which confirm the potential of kPADs for on-site environmental/sanitary monitoring and food toxin pre-screening.
For the assembly of 3D μPADs, the challenge is to ensure uninhibited transfer of solutions from one layer to another.14–16,22,23 Martinez et al. assembled patterned paper substrates with punched double-side tapes (punched and refilled with cellulose powders), to create a prototype 3D device that tests 4 different samples for up to 4 different analytes.14,21 Liu and Crooks prepared a device by folding a single piece of paper patterned lithographically for colorimetric and fluorescence assays.15 Such a paper-folding (origami) approach to fabricate 3D μPADs has been lately expanded to other novel designs, stacking methods, and chemiluminescence immunoassays for real-world samples.23 Most recently, Maria et al. reported a unique photolithographic method to create perpendicular channels by patterning both sides of the same sheet of photoresist-treated filter paper.24 The other method of rapid prototyping of multiplex microfluidic devices (not particularly applicable to μPAD), namely xurography, uses a digital cutting plotter for making microstructures in various polymeric films (with limited exploration of paper substrates).25 which has recently shown great advantages in creating PVC micromixers and micro-mixing arrays.26,27
Inspired by the ancient papercutting art (kirigami),28 we introduce herein a cut-and-paste protocol to fabricate 3D μPADs (namely kPADs) that are suitable for on-site quantitative assay applications. The novelty of our approach is beyond the cutting aspect that has been reported previously,29,30 rather, on the modification of the pre-cut paper substrate to be superhydrophobic and the reassembly with the untreated, hydrophilic reaction “discs” or “channels” with a chemically inert adhesive spray to prepare the kPADs. While the pioneering origami approach to prepare 3D μPADs indeed provided a unique and simplified device assembly protocol,15,23 the kirigami (particularly cut-and-paste) protocol reported herein aims to solve both assembly and patterning challenges (vide supra). In the following sections, we will describe the benchtop fabrication of kPADs with moderate complexity that is entirely instrumentation-free, i.e., they can be created with a pair of scissors and a pack of laboratory filter paper in a standard wet laboratory setting. More importantly, in conjunction with smartphone-reading colorimetric protocol such devices are demonstrated for quantitative detection of environmental contaminants (i.e., Cr(VI) and nitrite) in water samples.
Binary silane solutions were prepared by mixing two organosilanes (OTS/MTS = 7:3) in n-hexane with a total concentration of 0.2% (v/v) to prepare the hydrophobic hollow pads in the preparation of the kPADs. Stock solutions (1000 μg mL−1) of Cr(VI) and NO2− were prepared with K2CrO4 and NaNO2, respectively. Standard solutions of NO2− (at pH 7.0) and Cr(VI) (at pH 4.0) were freshly diluted from respective stock solutions. Solutions of NaNO3, Na2CO3, NaHCO3, NaOH, NaHPO3, Na2PO3, Na2B4O7, PbCl2, CrCl3, CdCl2, CaCl2, MgCl2, MnSO4, NaCl and KCl were prepared at 100 μg mL−1, respectively. “Real-world” samples were collected from the Fen River (37.858055992N, 112.5312312824E), a local waste water pipeline (37.8595855992N, 112.5165812824E), a tap water pipe in the lab (37.861355992N, 112.5181812824E) and a home fish tank. Their pH values were adjusted to either 7.0 (for NO2− detection) or pH 4.0 (for Cr(VI) testing) with NaOH (1 M) and HCl (1 M). The chromogenic reagent for nitrite (Griess reagent) contains 50 mM sulfanilamide, 330 mM citric acid and 10 mM N-(1-naphthyl)ethylenediamine in methanol. The chromogenic reagent for Cr(VI) was prepared with 0.2% 1,5-DPC and 1.0% methyltrioctylammonium chloride in acetone. Unless otherwise noted, all solutions were prepared with deionized water (>18.2 MΩ cm), produced from a Barnstead EasyPure Water System (Barnstead/Thermolyne, Dubuque, IA), both standard solutions and water samples were kept at 4 °C in the dark prior to the measurements.
Fig. 1 (a) Schematic view of the cut-and-paste fabrication of a kPAD. Color-coding: superhydrophobic hollow pads – dark gray; chromogenic reagent treated discs – blue, yellow, and pink; pre-cut hydrophilic discs and channels – light gray. (b) Cross section of the kPAD showing the “pathway” from one inlet to 9 outlets. The detailed fabrication procedure was presented as Fig. S1 in ESI.† |
Fig. 2 Screen snapshots of the “kPAD_DET” app running on an Android smartphone. (a) Launching menu; (b) measuring/testing menu; (c) reporting (result output) menu. |
Differing from the method reported by Martinez et al.,14 in which the “holes” were filled with cellulose powder, we utilized the original paper discs cut out from the same piece of filter paper; this was to ensure the best fit between the hydrophilic reaction zone and the hydrophobic “barrier”. With this design, the natural delivery of liquid samples across different layers remained and the second-step patterning is no longer needed. Compared with those devices created by directly folding or sticking two layers together,14,15 our design, inserting a “channel” layer between two pads ensures reliable transport of solutions within the 3D device. Very recently, Cardoso et al. have explored the use of readily available scholar glue to create hydrophobic flow barriers, and the glue-coated paper was then exposed to UV/Vis light for cross-linking to maximize chemical resistance and stability.7 Conceptually different from their innovation, herein we applied a minimal amount of adhesive spray, which is chemically stable and insoluble in water,32 to assemble different layers of μPADs. A detailed comparison between the kPADs with other existing 3D μPADs in terms of equipment requirements, patterning method, and assembly approach has been presented as Table S1 in the ESI.†
In comparison with the digital cutting-based xurography approach,25 which has been demonstrated to be powerful in creating 3D microfluidic devices with various polymeric materials (e.g., micromixers and micromixing arrays),26,27 the cut-and-paste fabrication of kPADs promises a bench-top, instrumentation-free method to make disposable, paper-based devices for on-site or in-field applications (vide infra). It should be noted that the size, shape, and number of hydrophobic discs (reaction zones) can be flexibly altered to meet any practical demands, i.e., the multiplicity and throughput can be adjusted as needed. The reproducibility of fabrication procedure and the stability of thus prepared kPADs have been confirmed (as shown in Fig. S3 and S4,† respectively). Particularly, the sizes of the holes on the hollow paper substrates and the cut-out discs show excellent correlations with the preset dimensions (i.e., sizes of hole punchers), and can be prepared with high reproducibility (Fig. S3†).
We have demonstrated the function of thus fabricated kPAD by adding 20 μL of 1:5 diluted inks (magenta, yellow, and black inkjet printer inks) to each of the inlets (Fig. 3a), and examining the color uniformity at corresponding outlets. As shown in Fig. 3b, all 27 outlets were wetted with respective ink in about 20 s, which is 20 to 50 times faster than previously created 3D μPAD (for which 5 to 15 min was needed for the solution to wet through, Fig. S5†).14,22 The facile transfer of solutions between different layers in our devices is due to the fact that we have retained the original porosity of filter paper. This is certainly an improved alternative from the compressed cellulose powder filling or UV patterning adapted in previous studies.14,22 The bottom right inset of Fig. 3b shows a zoom-in view of the outlet, which reveals the uniform color distribution across the entire reaction zone. The bottom left picture shows a water droplet on the modified paper base, whose contact angle was determined to be 155 ± 2° (which is superhydrophobic).31 In the case of wax or photoresist was used to pattern the paper base (to build hydrophobic barriers),14–16 typically water contact angles of slightly above 90° were achieved. Clearly, a device with patterned superhydrophobic substrate and unaltered paper disc shows optimized refinement in the reaction zone yet maintains the ideal liquid flow properties in the 3D configuration. The stability and solvent-resistance of these kPADs are guaranteed by the silanization treatment, which has been demonstrated to produce a structurally robust, superhydrophobic surface coating.31
The next step is to test the analytical performance of the kPADs fabricated via the cut-and-paste protocol for running quantitative assays that have potential applications. As the trial experiments, two environmental contaminants Cr(VI) and NO2−, representative heavy metal and inorganic salt, were chosen as they are typical sanitary contaminants in water. Long term intake of them both have persistence toxicity to human being, and likely results in esophageal cancer and gastric cancer.33,34 The detection of Cr(VI) was traditionally based on the colorimetric reaction between 1,5-diphenylcarbazide (DPC) and Cr(VI), which generates a magenta product [Cr(III)DPCO](3−n)+.35–37 While the NO2− concentration in water samples was quantitated with the Griess reaction, which also generates a magenta product (the principle of colorimetric reaction were shown in Fig. S6†).38–40 One of the questions that should be addressed is how to correlate the concentration of analytes with the color information in the assay image, i.e., which color should be analyzed for quantitation. The M (magenta) value from the CMYK (Cyan-Magenta-Yellow-Black) color space is naturally a good option to analyze the magenta products in both assays of Cr(VI) and NO2−. However, the picture acquired from a phone camera is based on the RGB (Red-Green-Blue) color space; besides, transferring an image format from RGB to CMYK may cause color loss.41 Herein, we directly use the original image (in RGB color space) for the analysis,41–44 which is based on the linear relationship between RGB to CMYK color spaces (eqn (1)).42 In practical applications, CMYK values usually use normalized expression (100%), i.e., the value obtained from eqn (1) divided by 255.
C = 255 − R |
M = 255 − G | (1) |
Y = 255 − B |
K = min (C, M, Y) |
The feasibility of using G (green) value to represent the magenta (M) intensity was verified by comparing the G and M values of the same color. In this case, different magenta dots were printed on a piece of filter paper (Whatman Grade 4), and then analyzed with the phone app. A linear relationship between M intensities and corresponding G values is present in the ESI (Fig. S7 and S8†), which is in good agreement with the fact that magenta and green are complementary colors.
Due to the high-quality camera optics and open-source operation system, smartphones with mobile apps have been recently explored for signal reading and data analysis of colorimetric and scanometric assays.45–47 Hence, both assays of Cr(VI) and nitrite, conducted with kPAD, have been analyzed with our customized app. Nonetheless, comparative studies with standard spectrophotometric method have been carried out to optimize the reaction conditions and to validate the reliability of kPADs methods (see Fig. S9 in the ESI†). For the detection of Cr(VI) and nitrite carried out with kPADs, five standard solutions were used in each case to construct the calibration curve and four target samples were tested, respectively. The detection zones (outlets) on a kPAD were pretreated with the chromogenic reagent for either nitrite or Cr(VI) prior to the final assembly step (selectivity tests for nitrite and Cr(VI) were shown in Fig. S10†). As shown in Fig. 4, the calibration curves established based on the correlation between the G value and the concentration for both NO2− and Cr(VI) standard solutions showed perfect linear relations with R2 values of 0.995 and 0.998. The detection limits for Cr(VI) and NO2− were calculated to be 0.7 μg mL−1 and 0.4 μg mL−1, respectively, according to the “3σ/b” rule (where σ is the standard deviation of the y-intercept of the regression line and b is the slope). These values are comparable with the detection limits obtained with other colorimetric assays reported in literature.34,37 Nevertheless, these calibration curves and the deduced equations for the signal–concentration correlations were incorporated to the kPAD_DET app that can be directly applied for the quantitation of unknown samples (vide infra).
Fig. 4 Calibration curves constructed with the mobile app (kPAD_DET) for the detection of (a) nitrite and (b) Cr(VI) on kPADs, respectively. |
Upon sample testing, we capture a picture of the outlet-side of the kPAD (Fig. 2b) and analyze the results with the mobile app. For proof-of-concept, the concentration of nitrite and Cr(VI) of tap water was detected by spiking standard nitrite (0, 2, 4, 6 μg mL−1) and Cr(VI) (0, 2, 5, 10 μg mL−1) solutions, respectively; meanwhile comparative detections have been carried out with a standard spectrophotometer. For the detection of NO2− and Cr(VI) with the spectrophotometry method, the amount of chromogenic reagent used were respectively 400 μL and 60 μL for every 2.0 mL standard or sample solution. In contrast, as little as 20 μL-standard or sample solutions were needed for the kPAD detection. As shown in Fig. 5, the concentrations determined with the “kPAD_DET” system correlated well with those obtained with a spectrophotometer within the experimental uncertainties, and were both close to the actual concentration (spiked value). The statistical consistency between the two methods (Fig. S11†) further confirms the reliability of the kPADs detection over a wide range of analyte concentrations.
To further validate the application potential of the kPADs fabricated with the simple cut-and-paste method, the concentration of NO2− in several “real-world” samples were tested and compared with the standard spectrophotometry method (Fig. 6a). The concentrations determined with the kPAD_DET system are again in good agreement with those obtained with spectrophotometry measurements. As a “gold standard” method,48 we have further shown the satisfactory comparison between the measurement techniques with the Bland–Altman plot (Fig. 6b). While the detection limit determined above (0.4 μg mL−1, Fig. 4a) meets the primary drinking water standard (1.0 μg mL−1) stated by the US Environmental Protection Agency (EPA),49 we in fact can quantitate much lower concentration of NO2− (from 0.03 to 0.4 μg mL−1, as shown in Fig. 6a); this is in fact much lower or at the same order of magnitude of the nitrite concentration in typical drinking water (0.1 μg mL−1) as reported by the World Health Organization (WHO).50 It is not debatable that the traditional spectrophotometric method provides greater sensitivity and accuracy (with rather smaller experimental uncertainties as shown in Fig. 5a and 6a), which requires advanced instrumentation and professionally trained operators. These pre-assembled kPADs are portable and disposable; they are suited for applications in field tests or other remote setting, as the only requirements would be samples of interests and a smartphone with the installed app. That is to say, the low cost, sufficient sensitivity and ease of use highlight the potential of the cut-and-paste protocol to prepare kPADs for rapid, on-site colorimetric assay applications, where traditional analytical instrumentation is not readily accessible. While we have shown the capability of kPADs for the quantitation of two representative environmental contaminants in water samples, the application can be readily expanded to a broad array of analytes, by simply transforming established colorimetric reactions to paper substrates.
Footnotes |
† Electronic supplementary information (ESI) available. See DOI: 10.1039/c9ra04014e |
‡ These authors contributed equally to this work. |