Analytical Methods Committee AMCTB No. 97
First published on 1st July 2020
The past 30 years have seen increasing availability of methods and equipment using thermal desorption for the measurement of airborne pollutants. These methods: offer greater sensitivity than methods using solvent desorption; are more amenable to automation; and, are better suited to mass spectrometry (MS)-based detection. The greater sensitivity offered by thermal desorption means it is well suited to the analysis of samples collected through diffusive sampling with the additional benefits that this offers. This Technical Brief informs both analysts and less technically aware users of the capabilities and limitations of thermal desorption equipment and measurement methods.
Solvent desorption (SD) of air samples collected using sorbent tubes is a long established analytical procedure. However, in the 1980’s an alternative method, thermal desorption (TD), in which the sorbent is desorbed by heating and concentrated on a focusing trap (FT), began to become commercially available. Over the last 30 years the use of this technique, which allows measurement of airborne concentrations ranging from percent to parts per trillion has grown with advances in equipment and greater availability of validated methods. In addition to vapour phase organic compounds, TD can be used for the analysis of some inorganic gases such as nitrous oxide. TD is not suitable for the analysis of methane, compounds less volatile than C44 aliphatic hydrocarbons and most permanent inorganic gases (e.g. SO2, HCl, CO, CO2).
During analysis, the sample tube is initially leak tested then heated in a flow of carrier gas (typically helium) to desorb the collected analytes. The desorption temperature needs to be high enough (typically 200–300 °C) to extract all the analytes, but not so high as to generate artefacts from thermal degradation of the sorbent. The desorption volume should be as small as possible but sufficient enough to extract all the analytes from the sorbent. The desorbed analytes are concentrated on a low mass focusing trap (FT); usually maintained at sub-ambient temperature and often referred to as the ‘cold’ trap. If dilution is required, an inlet split allows some of the desorbed sample to be vented to waste or, in some systems, into a clean sorbent tube for subsequent reanalysis. At the end of the desorption period the FT is rapidly heated and the collected analytes are flushed onto the GC system via an outlet split where, if required, a second dilution may be made. In most cases the FT is back-flushed, as shown in Fig. 2, i.e. the analytes exit the trap at the same end they entered. This is particularly important where the FT contains more than one sorbent in order to reduce the likelihood of low volatility analytes coming into contact with the stronger sorbent and not being fully extracted.
(a) In a typical TD analysis around 2% of the desorbed analytes reach the GC detector compared with less than 0.1% for a 1 μl injection of a sample desorbed into 1 ml of solvent. Consequently, TD offers much greater sensitivity than SD.
(b) Chromatograms obtained using TD contain no solvent peak, the presence of which can mask the presence of some analytes and hinder the use of MS-based detection.
(c) Modern TD equipment sample tubes can be analysed with minimal sample handling thereby reducing labour costs and improving precision.
The primary drawback of TD is that analysis with some equipment is a “one-shot” process, i.e. should a problem occur during analysis there is no second opportunity. However, many TD systems now incorporate systems enabling partial recapture of the desorbed sample allowing repeat analysis, albeit of a slightly diluted sample.
Where the compounds of interest have a wide range of volatilities, multi-bed tubes containing two or more sorbents of differing strengths are sometimes used. When using multi-bed tubes it is important to orientate the sample tube so that the sample is collected through the weakest sorbent first (see Fig. 3). This is to avoid low volatility components coming into contact with the stronger sorbent and not being fully extracted during analysis. For the same reason, when desorbing a multi-bed tube the tube should be orientated with the flow passing through the strongest sorbent first, i.e. in the opposite direction to that during sampling.
The sorbent tubes used for TD are more expensive than those used with SD but can be reused after conditioning; typically more than 100 times. Conditioning involves heating the tube in a flow of gas to a temperature slightly above that used for desorption; typically 30–60 minutes with nitrogen.
The dimensions of both the diffusion head and the sample tube are engineered precisely so that components diffuse onto the sorbent bed following Fick’s laws of diffusion.3 Different substances have different diffusive uptake rates (UR). In addition, the same substance will have a different UR on different sorbents. As diffusive sampling does not use a pump, the amount of sample taken cannot be expressed in terms of a flow, i.e. ml min−1. Instead UR is typically expressed in terms of the mass of analyte collected, in ng, per ppm of analyte in the air being sampled, per minute of sampling. Hence, the unit of a UR value is ng ppm−1 min−1. UR values for a range of analytes and sorbents are available from sources such as HSE Method MDHS 104.4
The main benefit of diffusive sampling is that no pump or flow rate measurement is required, making the devices simple, cheap and reliable. These devices are widely used for personal monitoring, where they are less obtrusive to the wearer, and in locations such as operating theatres where a larger pumped device may be unacceptable. However, diffusive sampling is not suitable for short-term sampling of less than 15 minutes. Diffusive sampling is intended for use with tubes containing a single sorbent. The use of multi-bed tubes offers no advantage as airborne compounds collected onto the sample tube by this technique are likely to be retained in the first few mm of the front sorbent bed making the other sorbent bed(s) redundant.
As well as the tube sampler shown in Fig. 4, radial and badge-type diffusive samplers are also available. These devices collect the air sample onto a TD-compatible sorbent which, after sampling, is transferred to an empty TD tube and analysed in the same manner as the tube samples. These devices have higher uptake rates than tube samples, but require more sample preparation prior to analysis and are less amenable to reuse.
Ian Pengelly (Health and Safety Executive, HSE)
This Technical Brief was prepared for the Analytical Methods Committee (AMC), with contributions from members of the AMC Instrumental Analysis Expert Working Group, and approved by the AMC on 31 May 2020.
This journal is © The Royal Society of Chemistry 2020 |