Adel Ehab Ibrahim*a,
Hisham Hashemb,
Magda Elhenaweeb and
Hanaa Salehb
aPharmaceutical Analytical Chemistry Department, Faculty of Pharmacy, Port-Said University, Egypt. E-mail: pharmacist_adel_2005@yahoo.com; Tel: +20 1112332345
bPharmaceutical Analytical Chemistry Department, Faculty of Pharmacy, Zagazig University, Egypt
First published on 8th January 2020
By 2025, it's estimated that 322 million males worldwide will suffer from sexual disorders. This can give an estimation for the size of the pharmaceutical and counterfeit products industry for the next few years. Meanwhile, green analytical chemistry forced itself to decrease the massive environmental pollution and hence new analytical methodologies are needed to replace the old ones that consume large amounts of hazardous solvents. In this research, two new methods were validated for determination of seven recognized drugs used in treatment of male impotence, premature ejaculation as well as enhancing sexual libido by HPLC on RP-C18 core–shell particulate and monolithic columns. The study was extended to compare the capabilities of those stationary phases to accommodate greener chromatography concepts without loss of efficiency. Both morphologies shortened the analysis time relative to the previously reported conventional HPLC methods by different approaches. Core–shell particles had higher efficiency in terms of theoretical plates' number and enhanced resolution power which enabled lower detection limits. However, the monolithic column had lower column backpressure which enabled the use of ethanol as a greener alternative solvent at even higher flow rates. The methods were finally applied successfully for the determination of drugs under study in pharmaceutical dosage forms, counterfeit products and in human plasma.
The seven drugs under study are used to treat male sexual disorders. Avanafil (AVN), sildenafil (SDN), apomorphine (APM) and yohimbine (YHB) are recognized drugs in treatment of male impotence. Tramadol (TMD) and trazodone (TZD) are drugs abused in treatment of premature ejaculation. Dapoxetine (DPX) is the first drug to be approved for on-demand treatment of premature ejaculation. TZD and YHB were reported also to enhance sexual libido. SDN and AVN are phosphodiesterase-5 (PDE-5) inhibitors used in treatment of ED. SDN was the first drug approved as PDE-5 inhibitor by the United States Food and Drug Act (US FDA), while AVN is a new ultra short acting PDE-5 inhibitor.5 APM is a dopaminergic agonist (D1 and D2) that can be used as sublingual tablets to treat ED.6 YHB is a natural alkaloid that was reproduced synthetically because of its α2-adrenoceptor antagonist effect which was found to enhance ED with psychological origin.6 TMD is a strong opioid analgesic that has much lower dependence than morphine. It has side effect of delaying ejaculation so it was abused in treatment of PE,7 beside its effect in mode elevation. DPX is the first short acting drug approved mainly for on-demand treatment of PE. DPX is a selective serotonin reuptake inhibitor (SSRI). TZD is a unique drug that is abused in treatment of ED and PE, simultaneously. TZD is a serotonin reuptake inhibitor so it delays ejaculation. It also causes a condition known as priapism which is a condition in which a penis remains erect for hours in the absence of stimulation or after stimulation has ended so it was also tried for treatment of ED.8 Another added value for TZD was reported to increase sexual libido in both men and women.9,10
The stationary phase is the main core where separation of different analytes takes place. The choice of stationary phase became one of the main and most critical steps in the development of LC methods because it requires compromise between separation efficiency, analysis time and environmental safety. Lots of analyses are done daily by research and industrial quality control laboratories which must serve high through output, so the time required for analysis is a critical parameter. Environmental safety and greening the analytical methodologies is another important consideration. Conventional LC equipment can generate around half liter of organic solvents daily,11 that's why LC is targeted by green chemistry concepts.
Older trends for enhancing separation efficiency depended on decreasing stationary phase particles size; however this was limited by the high column backpressure developed. To shorten analysis time, shorter columns were used, but this affected separation efficiency. Another means was to increase flow rates, but it was counteracted also by the developed column backpressure. Greener chromatography focuses on decreasing the consumption of organic solvents and/or replacing environmentally toxic solvents by safer ones.
In the years of 2003 and 2004, sub-2 μm particles and ultra performance liquid chromatography (UPLC) were introduced successfully on commercial level. UPLC reduced analysis time, lowered solvent consumption and enhanced sensitivity and separation efficiency. However, it's not a widespread technique in economical laboratories due to the short chromatographic column lifetime, low reproducibility and high maintenance costs required which arise from the high backpressure developed on the UPLC system.12 So, the new stationary phase geometries, core–shell particles and silica monolith, became a promising tool for the purpose of economic, simple, green and efficient methodologies in short analysis time. The ideas of these stationary phase morphologies are old. Several papers described and compared the conventional totally porous reversed phase particles to these new morphologies.13–15 However, new trends in developing efficient, economical and faster, together with greener analysis methods utilized these new stationary phase morphologies on the conventional HPLC systems.
Literature survey revealed that individual determination of the drugs under study was reported extensively. Some papers were reported for simultaneous determination of drugs used in treatment of ED16–18 and some others were reported for simultaneous determination of some antidepressant drugs that can be abused in treatment of PE.19,20 However, as discussed before, the co-administration of drugs for treatment of ED and PE is possible. Few methods were reported for co-determination of two drugs used for ED and PE such as SDN/DPX,21,22 AVN/DPX,23,24 SDN/TMD25 and SDN/YHB.26 To the best of our knowledge, only one method was reported for simultaneous determination of three PDE-5 inhibitors and four SSRIs, and was applied on pharmaceutical dosage forms.27
This study provides a new comparison tool between monolithic rods and core–shell 2.7 μm particles columns and their capabilities to be used for greener chromatography. The study covers the areas of column's efficiency, selectivity, analysis time, and solvent consumption. Also, the proposed study offers two new highly efficient methods that were successfully validated for the identification, separation and quantification of seven drugs used in treatment of male sexual disorders for the first time. The methods were applied successfully for identification and determination of these drugs in pharmaceutical dosage forms, counterfeit products and human plasma.
AVN and DPX as hydrochloride salts were kindly supplied by Al-Andalus pharmaceutical company, Egypt. SDN as citrate, APM, TMD, TZD and YHB, all as hydrochloride salts, were kindly supplied by EIPICo., Egypt. All drugs were pharmaceutical grade (chemical structures Fig. 1).
Pharmaceutical dosage forms; Vigor forte® chewable tablets (batch number: 61212), and Trittico® tablets (batch number: 1401046) were kindly supplied by EIPICo, Egypt. Yohimbex® tablets (batch number: 11051), Joybox® tablets (batch number: 171436) and Tramundin® tablets (batch number: 153385) were purchased from Egyptian market. Erovanafil® tablets (batch number: 160367) was kindly supplied by Al-Andalus pharmaceutical company, Egypt (for composition refer to Section 3.1.4). Counterfeit products were purchased from the illegal Egyptian market under names; Tiger-King®, Fox®, Big-P®, and Love-zone®.
Blank human plasma samples were purchased from the Egyptian national blood bank.
pH meter (model-713) from Metrohm industries, Switzerland was used to adjust pH during the study.
Two chromatographic columns were used in the proposed study. Monolithic Chromolith® Performance RP-18e (100 × 4.6 mm) column was purchased from Merck, Germany. Core–shell, Poroshell® EC-C18 2.7 μm (150 × 4.6 mm) column was purchased from Agilent technologies, Germany.
a For determination of drugs in plasma, run time 12–18 min. | ||
---|---|---|
Column I | Monolithic Chromolith® | |
Flow rate | 2 mL min−1 | |
Mobile phase gradient programming | Mobile phase A (buffer) | Mobile phase B (ethanol) |
0.0–6.0 min | 65.0% | 35.0% |
6.0–8.0 min | 65.0–40.0% | 35.0–60.0% |
8.0–10.0 min | 40.0% | 60.0% |
Column II | Core–shell Poroshell® | |
Flow rate | 1 mL min−1 | |
Mobile phase gradient programming | Mobile phase A (buffer) | Mobile phase B (aetonitrile) |
0.0–11.0 min | 65.0% | 35.0% |
11.0–12.0 min | 65.0–35.0% | 35.0–65.0% |
12.0–16.0 mina | 35.0% | 65.0% |
All analyses were done by gradient elution technique at the flow rates as listed in Table 1. Auto-sampler injection volume was 20.0 μL and column compartment was kept at 35 °C. Detection was done at 210 nm using UV-detector.
The working solutions were prepared by dilution of the stock solution using the solvent mixture to prepare calibration standards at concentrations 2.0, 10.0, 25.0, 50.0, 75.0 and 100.0 μg mL−1. The quality control (QC) samples were prepared at three different concentrations, low (QCL), medium (QCM) and high (QCH) which were 5.0, 50.0 and 80.0 μg mL−1 by spiking a freshly prepared placebo solution. Placebo solution was prepared by dispersing 10.0 g of commonly used excipients per liter water (e.g. starch, magnesium stearate, titanium dioxide, sodium starch glycolate and carboxymethyl cellulose sodium). All stock solutions, calibration standards and QC samples were stored in amber containers in a refrigerator at 2–8 °C.
For determination of drugs in plasma, linearity calibration standards were prepared at concentrations 1.0, 10.0, 20.0, 30.0, 40.0 and 50.0 μg mL−1 by diluting the stock solution in blank plasma samples then vortex mixed for 10 minutes. According to FDA guidelines, the total organic solvent added should not exceed 2% of biological sample.28 QC samples were prepared in four concentrations at 2.5, 5.0, 25.0 and 40.0 μg mL−1 by dilution in blank plasma samples and vortex mixed in the same way. All calibration standards and QC samples were stored in freezer (−20 °C).
For Trittico®, Tramundin®, Joybox® and Erovanafil® tablets, the working solutions were prepared by dilution of 5 mL of the filtrate to 100.0 mL using the solvent mixture. For Vigor forte® and Yohimbex® tablets, the working solutions were prepared by dilution of 20.0 mL of the filtrate to 100.0 mL using the solvent mixture. Working solutions of counterfeit products' tablets were prepared directly by dissolving the average weight of one tablet in 200.0 mL solvent mixture, sonicated for 10.0 min then filtered.
Frozen plasma samples were thawed before analysis at room temperature. An aliquot of 400 μL of calibration standards and QC samples were added in 1.5 mL Eppendorf centrifuge tube then 600 μL of ACN was added. The mixture was shaken on vortex mixer for 5 min then centrifuged at 10000 rpm for another 10.0 min. 20 μL of the clear supernatant was injected on the HPLC-UV system.
Fig. 2 Chromatograms showing separation of the seven drugs under study on core–shell and monolithic columns. Chromatographic conditions: refer to Table 1, sample QCM at concentration 50.0 μg mL−1. (1) APM, (2) TMD, (3) YHB, (4) AVN, (5) TZD, (6) SDN, (7) DPX. |
Three independent blank human plasma samples from different human sources were treated and injected on the HPLC-UV system and were compared to spiked plasma samples to test interferences from any endogenous substances. Fig. 3 shows separation of drugs using QCL sample (5.0 μg mL−1) spiked in blank human plasma. Independent blank human plasma samples were injected and no interference from any endogenous material that may be found in plasma was detected.
Fig. 3 Chromatograms showing separation of drugs under study in spiked human plasma. Chromatographic conditions: refer to Table 1, sample QCL at concentration 5.0 μg mL−1. Column: core–shell Poroshell® EC-C18, 2.7 μm. (1) APM, (2) TMD, (3) YHB, (4) TZD, (5) AVN, (6) SDN, (7) DPX. |
Limits of detection (LOD) and quantification (LOQ) were calculated according to signal-to-noise ratios; (10:1) for LOQ and (3:1) for LOD. LODs and LOQs as shown in (Table 2) indicated the sensitivity of the proposed methods.
Parameter | Chromolith® Performance RP-18e | ||||||
---|---|---|---|---|---|---|---|
APM | TMD | YHB | AVN | TZD | SDN | DPX | |
a Number of theoretical plates per 100 mm long stationary phase, for QC-sample at concentration 5 μg mL−1 of each drug.b Column backpressure at zero time.c Volume consumed per chromatographic run.d Results = average ± relative standard deviation.e The comparison is irrelevant in the gradient elution region of the chromatogram. | |||||||
Average tR (minutes)d | 1.7 ± 0.2 | 2.2 ± 0.2 | 3.1 ± 0.0 | 3.8 ± 0.4 | 4.4 ± 0.2 | 5.6 ± 0.0 | 8.9 ± 0.2 |
Selectivity, K | — | 1.6 | 1.6 | 1.2 | 1.2 | 1.31 | 1.71 |
Resolution | — | 3.8 | 4.9 | 2.9 | 2.1 | 3.3 | 13.9 |
Symmetry factor | 0.6 ± 0.4 | 0.7 ± 0.8 | 0.7 ± 0.6 | 0.8 ± 0.4 | 0.7 ± 0.6 | 0.7 ± 0.8 | 0.6 ± 0.8 |
Na,d | 2800.0 ± 1.2 | 3300.0 ± 2.0 | 3300.0 ± 0.6 | 3200.0 ± 1.8 | 3200.0 ± 0.6 | 3400.0 ± 1.2 | —e |
Pb | 120 bar | ||||||
LOD ng mL−1 | 100.0 | 500.0 | 200.0 | 200.0 | 100.0 | 200.0 | 100.0 |
LOQ ng mL−1 | 400.0 | 1000.0 | 600.0 | 600.0 | 400.0 | 500.0 | 400.0 |
Total run time | 9.5 minutes | ||||||
Volume of organic solvent consumedc | 8.8 mL per run + 1.2 mL per re-equilibration |
Parameter | Poroshell 120® EC-C18 2.7 μm | ||||||
---|---|---|---|---|---|---|---|
APM | TMD | YHB | TZD | AVN | SDN | DPX | |
Average tR (minutes)d | 3.2 ± 0.0 | 4.7 ± 0.2 | 5.3 ± 0.0 | 7.2 ± 0.4 | 9.8 ± 0.2 | 10.5 ± 0.2 | 15.0 ± 0.2 |
Selectivity, K | — | 1.8 | 1.2 | 1.6 | 1.4 | 1.2 | 1.4 |
Resolution | — | 12.9 | 4.6 | 12.7 | 12.0 | 2.9 | 23.5 |
Symmetry factor | 0.8 ± 0.2 | 0.8 ± 0.6 | 0.8 ± 0.4 | 0.8 ± 0.6 | 0.8 ± 0.2 | 0.8 ± 0.4 | 0.8 ± 0.8 |
Na,d | 11200.0 ± 1.2 | 15200 ± 1.6 | 15100.0 ± 0.8 | 14700.0 ± 0.8 | 15500.0 ± 2.2 | 16900.0 ± 1.8 | —e |
Pb | 200 bar | ||||||
LOD ng mL−1 | 50.0 | 200.0 | 100.0 | 50.0 | 100.0 | 50.0 | 50.0 |
LOQ ng mL−1 | 200.0 | 500.0 | 300.0 | 200.0 | 300.0 | 200.0 | 200.0 |
Total run time | 16.0 minutes | ||||||
Volume of organic solvent consumedc | 7.0 mL per run + 1.8 mL per re-equilibration |
Dosage-form/counterfeit product | Labeled amount | Chromolith® found amounta | Poroshell® found amounta |
---|---|---|---|
a Results = recovery% ± RSD.b Counterfeit products have no labeled data for drugs under study; they are supposed to be of natural herbal origin. | |||
Vigor-forte® | APM 3 mg | 100.8 ± 3.8 | 100.8 ± 5.4 |
SDN 50 mg | 96.8 ± 2.2 | 94.0 ± 1.6 | |
Trittico® | TZD 100 mg | 99.2 ± 3.2 | 96.4 ± 3.0 |
Joybox® | DPX 60 mg | 96.0 ± 1.6 | 100.8 ± 1.6 |
Tramundin® | TMD 100 mg | 92.6 ± 1.4 | 99.0 ± 0.6 |
Yohimbex® | YHB 5 mg | 95.2 ± 3.0 | 96.4 ± 3.0 |
Erovanafil® | AVN 100 mg | 92.0 ± 1.2 | 91.4 ± 1.0 |
Tiger-King®b | TMD | 10.2 mg per tablet ± 3.2 | — |
SDN | 20.8 mg per tablet ± 2.0 | ||
DPX | 20.6 mg per tablet ± 1.4 | ||
Fox®b | SDN | 20.4 mg per tablet ± 1.4 | — |
DPX | 30.4 mg per tablet ± 2.8 | ||
Big-P®b | SDN | 25.2 mg per tablet ± 1.2 | — |
Love-zone®b | SDN | 55.8 mg per tablet ± 1.6 | — |
DPX | 30.2 mg per tablet ± 2.8 |
The use of EtOH instead of ACN slightly affected the order of elution of drugs under study for AVN which has switched its elution order with TZD (Table 2). This may be attributed to the different polarities of EtOH and ACN. Since elution of analytes in RPLC depends on difference of partitioning between the mobile and the stationary phases, it seems like AVN has higher partitioning into EtOH/buffer mobile phase than into ACN/buffer mobile phase.
On the other hand, core–shell particles exhibited higher separation efficiency in terms of resolution and selectivity. Number of theoretical plates was higher and peaks symmetry was better in case of core–shell particles. Moreover, the LOD and LOQ were lower on core–shell than on monolithic column which enhanced sensitivity of the method and this may be attributed to the better peak symmetry (Table 2). It should be noted that N was compared for peaks eluted within the time of isocratic elution only, because it's irrelevant to compare N for peaks within the gradient elution period (DPX peak only).
Analysis was shorter on monolithic column (less than 9.0 min), while core–shell particles required double that time (16.0 min) due to its shorter length and ability to withstand higher flow rates. To re-equilibrate the columns between chromatographic runs, a mobile phase volume equivalent to 5 times the void volume (VM) is required for washing. Despite having shorter length, monolithic column has larger VM than core–shell column, so monolithic column requires larger re-equilibration volume. But since the flow rate on silica monolith is double that on core–shell particles, the final re-equilibration time between runs was found to be nearly the same on both columns due to difference of applied flow rates (4.0 & 5.0 min on monolithic and core–shell columns, respectively).
Both methods were found to be economical since the total volume of organic modifiers consumed per chromatographic run was minimal, taking in consideration the higher flow rate used on monolithic column and the better environmental safety of EtOH than ACN (Table 2). The shorter analysis time means lower solvent consumption which is safer for environment and also more economical.
Two simple fast LC methods were validated for the simultaneous determination of apomorphine, sildenafil, avanafil, yohimbine, trazodone, tramadol and dapoxetine and applied successfully for their determination in the corresponding dosage forms, in plasma and in checking the adulteration of counterfeit products. The overall analysis time didn't exceed 9.0 and 16.0 min. on monolithic and core–shell columns, respectively.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: 10.1039/c9ra08717f |
This journal is © The Royal Society of Chemistry 2020 |