Anri Mochizukia,
Takuya Maeyama*a,
Yusuke Watanabeb and
Shinya Mizukamib
aDepartment of Chemistry, School of Science, Kitasato University, 1-15-1 Kitasato, Minami, Sagamihara, Kanagawa 252-0373, Japan. E-mail: maeyama@kitasato-u.ac.jp
bSchool of Allied Health Sciences, Kitasato University, 1-15-1 Kitasato, Minami, Sagamihara, Kanagawa 252-0373, Japan
First published on 4th August 2020
Dosimetry of spatial dose distribution of ionizing radiation in tissue equivalent materials is particularly important for cancer radiotherapy. Here, we describe a radio-fluorogenic gel-based dosimeter that has achieved 16 times higher sensitivity by incorporating surfactants and halogenides. The gel dosimeters were prepared from dihydrorhodamine 123 (DHR123) and small amounts of nano-sized clay and a radiosensitizer. By comprehensively changing the type of additives for the sensitizer (three surfactants: Triton X-100, sodium dodecyl sulfate (SDS) and cetyltrimethylammonium bromide, and three halogenides: trichloroacetic acid, tribromoacetic acid and 2,2,2-trichloroethanol), the increase in sensitivity can be explained by an increase in relative fluorescence quantum yield and an increase in radiation chemical yield. These highly sensitive gel dosimeters also show dose rate independent sensitivity under irradiation at 0.64 and 0.77 Gy min−1 using a 6 MV X-ray therapeutic beam from the medical linac.
Dihydrorhodamine 123 (DHR123) has been used as one of the sensitive probes in the production of reactive oxidizing species (ROS) in cells.4 Non-fluorescent DHR123s are converted to highly fluorescent forms, rhodamine 123 (RD123) in the presence of ROS, such as OH radical; hence, the fluorescence intensity is proportional to the amount of reacted ROS. Recently, the DHR123 probe was applied to ionizing radiation dosimeter,5 since this DHR123 probe also work for ionizing radiation induced ROS.6 The fluorescence intensity increases with the increase in the absorbed dose. While the fluorescence distribution corresponding to dose distribution was maintained in the hydro nano-clay gel matrix, DHR123 radio-fluorogenic nano-clay gel (DHR123RFG) dosimeter produced two-dimensional (2D) dosimetry with a 2D fluorescent scanner. This hydrogel dosimeter prepared mainly with water and a small amount of gelling materials that is tissue equivalent material is expected to be used as a verification method of dose distribution in radiotherapy.
As a typical gel dosimeter,7 there are Fricke gel8–11 and polymer gel12–15 dosimeters, and many radiosensitizers2,16–20 have been investigated because improving sensitivity is an issue, up to now. On the other hand, there are not many reports in the gel dosimeters utilizing fluorometry using radiosensitizers, although there are several type of radio-fluorogenic gels.21–27 Thus, the purpose of this study is investigation of sensitizer effects on DHR123RFG in detail.
The reaction mechanism of the DHR123RFG6 is thought to be the same as the radio-chromic hydrogels using leuco dye such as leuco crystal violet and leuco malachite green,28,29 although the DHR123RFG is a specific gel dosimeter which uses the fluorescence characteristics. In the radio-chromic hydrogels, the sensitivity enhancement is reported by the addition of a surfactant and a halide. The surfactant in these phantoms helps to solubilize the leuco-dye molecules, which are only sparingly soluble in water.28 The halide increases the efficiency of product yield.30 The dose rate dependence of these addition effects was also discussed.31 In this study, we apply sensitizers to the DHR123RFG based on the abovementioned research. Its sensitivity characteristics and dose rate dependence are evaluated, and the mechanism of the sensitization effect also is discussed.
No. | Surfactant | Conc. [mM] | DHR123 [μM] | Nanoclay [wt%] |
---|---|---|---|---|
1 | 100 | 2.5 | ||
2.1 | Tx100 | 0.25 | 100 | 2.5 |
2.2 | Tx100 | 0.5 | 100 | 2.5 |
2.3 | Tx100 | 1 | 100 | 2.5 |
2.4 | Tx100 | 4 | 100 | 2.5 |
2.5 | Tx100 | 38 | 100 | 2.5 |
2.6 | SDS | 0.25 | 100 | 2.5 |
2.7 | SDS | 0.5 | 100 | 2.5 |
2.8 | SDS | 1 | 100 | 2.5 |
2.9 | SDS | 2.5 | 100 | 2.5 |
2.10 | SDS | 17 | 100 | 2.5 |
2.11 | CTAB | 0.5 | 100 | 2.5 |
No. | Halogenide | Conc. [mM] | DHR123 [μM] | Nanoclay [wt%] |
---|---|---|---|---|
1 | 100 | 2.5 | ||
3.1 | TCAA | 0.25 | 100 | 2.5 |
3.2 | TCAA | 0.5 | 100 | 2.5 |
3.3 | TCAA | 1 | 100 | 2.5 |
3.4 | TCAA | 5 | 100 | 2.5 |
3.5 | TBAA | 0.25 | 100 | 2.5 |
3.6 | TBAA | 0.5 | 100 | 2.5 |
3.7 | TBAA | 1 | 100 | 2.5 |
3.8 | TBAA | 5 | 100 | 2.5 |
3.9 | TCE | 0.25 | 100 | 2.5 |
3.10 | TCE | 0.5 | 100 | 2.5 |
3.11 | TCE | 1 | 100 | 2.5 |
3.12 | TCE | 5 | 100 | 2.5 |
3.13 | TCE | 50 | 100 | 2.5 |
No. | Surfactant | Conc. [mM] | Halogenide | Conc. [mM] | Dose rate [Gy min−1] |
---|---|---|---|---|---|
1 | 4.64 or 0.77 | ||||
4.1 | Tx100 | 38 | |||
4.2 | Tx100 | 38 | TBAA | 0.5 | |
4.3 | Tx100 | 38 | TBAA | 1 | |
4.4 | SDS | 17 | |||
4.5 | SDS | 17 | TBAA | 0.5 | |
4.6 | SDS | 17 | TBAA | 1 |
Gel dosimeter irradiation was performed using a 6 MV X-ray beam generated with a Varian Linear Accelerator (TrueBeam, Varian Medical Systems, Palo Alto, USA). The radiation dose ranges for all the prepared samples are 0–4 Gy. A dose rate of 4.66 Gy min−1 was used for samples shown in Tables 1 and 2, while two dose rates (4.66 and 0.77 Gy min−1) were used for samples shown in Table 3. Within one week of irradiation, fluorescence measurements of the irradiated samples were conducted using an F-4500 spectrofluorometer (Hitachi, Japan). The exciting wavelength is 480 nm and the emission and excitation slit widths were set at 1 nm and 20 nm, respectively, and these parameters were kept constant for the measurements of all samples.
The yield of RD123 [G(RD123)] was calculated by using the formula (1) wherein ΔI is the increasing rate of fluorescence intensity per 1 Gy and Φf is the relative fluorescence quantum yield of RD123 calculated from the slope of the calibration curve (No. 5.1 in Table 4). The effects of the sensitizer on relative fluorescence quantum yield of RD123 also evaluate under several condition as shown in Table 4 (No. 5.2 to 5.9).
(1) |
No. | Additive | Conc. [mM] | RD123 [μM] | DHR123 [μM] |
---|---|---|---|---|
5.1 | 0.026–0.21 | 100 | ||
0.25–2.1 | ||||
5.2 | TBAA | 1 | 0.21 | 100 |
5.3 | Tx100 | 0.95 | 0.21 | 100 |
5.4 | Tx100 | 3.8 | 0.21 | 100 |
5.5 | Tx100 | 38 | 0.21 | 100 |
5.6 | SDS | 1 | 0.21 | 100 |
5.7 | SDS | 2.6 | 0.21 | 100 |
5.8 | SDS | 17 | 0.21 | 100 |
5.9 | 0.026–0.21 | 0 | ||
0.25–2.1 |
Fig. 2(a) and (b) show the effect of surfactant concentrations of Tx100 and SDS on dose sensitivity, respectively. The values of dose sensitivity with standard error, its ratio, and initial fluorescence intensity are also listed in Tables S1–S3 in the ESI.† Fig. 2(c) and (d) shows the effect of surfactant concentrations on relative fluorescence quantum yield of RD123 calculated from the calibration curve which added surfactants. The right-hand axis shows the ratio which is normalized by the slope of standard calibration curve (No. 5.1 in Table 4).
From these results, it was found that the sensitivity was increased by up to about 2.5 times by Tx100 (0.25–4 mM), and by further addition of Tx100 (38 mM), the sensitivity showed 7 times higher than the standard sample. It was also found that the relative fluorescence quantum yield of the RD123 increased with the increase in Tx100 concentration. This is a reasonable explanation for sensitivity enhancement. In other words, the RD123 which is produced by the radiation-induced reactions of the DHR123RFG shows stronger fluorescence intensity by the presence of Tx100, and the sensitivity also increased. When 0.25 to 2.5 mM of SDS was added, there was no increase in the sensitivity (gel 2.6 to 2.9 in Table 1); however, on the addition of 17 mM SDS, the sensitivity was found to increase by 8 times (gel 2.10 in Table 1). Unlike in Tx100, from Fig. 2(d), the relative fluorescence quantum yield of RD123 only got higher by adding 17 mM SDS, while other samples (gel 5.6, 5.7 in Table 4) one did not change, and it agrees with above dose sensitivity results. In generally, it is considered that the relative fluorescence quantum yield changes due to two factors, the aggregation effects of the fluorescence dyes and the solvent effect. In the former, the surfactant prevents the aggregation of dyes, and increase the quantum yield of RD123 as reported in the literature using another fluorescence dye.32,33 In the latter case, the surfactant increases the relative fluorescence quantum yield by giving the different environment than simple aqueous solution such as micelles.34 These phenomena show the different degrees of effect depending on the dye, but an increase in quantum yield of fluorescent dyes using surfactants and clay has been reported.35
To confirm the sensitizing effects of surfactants in detail, we compared the surfactant concentration dependence of (black square) and relative fluorescence quantum yield (red square) which normalized the value of standard sample (Fig. 3). Relative fluorescence quantum yield is higher than relative dose sensitivity in both surfactants. Thus, it was suggested that radiation chemical yields decreased by addition of surfactants.
Fig. 3 Surfactant concentration dependence on the relative fluorescence quantum yield and the relative dose sensitivity ((a) Tx100, (b) SDS), and on the yield, G(RD123) ((c) Tx100, (d) SDS). |
To discuss in more detail, the effects of surfactant concentrations on G(RD123) [μmol J−1] were shown in Fig. 3. G(RD123) calculated using formula (1). On both surfactants, G(RD123) was reduced to one-sixth by the addition of 0.25 mM; therefore, it is considered that the surfactants that inhabit reaction of DHR123 with hydroxyl radicals, gets the final production efficiency of RD123 lower. Although, for the Tx100, G(RD123) did not depend on surfactant concentration in a range 1–38 mM, for the SDS, G(RD123) made with 17 mM increases to 0.125 μmol J−1. This observation showed that a high concentration of SDS did not inhibit the reaction of DHR123 with water decomposition radicals and it effectively increased the relative fluorescence quantum yield of RD123. On the addition of 17 mM SDS, the relative fluorescence quantum yield increased rapidly; hence, the higher dose sensitivity is attributed to the process that involves the interaction between DHR123RFG and surfactant. It was an interesting result because, in LCV gel dosimeters, the effect of the presence or the absence of surfactant on dose sensitivity was not discussed due to the low solubilities of LCV.31
The measurement results of the calibration curve with and without non-fluorescent DHR123 are shown in Fig. 4 (comparison between No. 5.1 and No. 5.9 in Table 4). The fluorescence intensity of RD123 varied greatly depending on the presence or absence of DHR123, which revealed that DHR123 acts as a quencher of RD123. Since the DHR123 gel dosimeter after irradiation also include an excess of DHR123 (100 μM order) with respect to RD123 (several μM order), it is considered that the quenching effect of DHR123 on the sensitivity of the gel dosimeter is large. In other words, as shown in the Fig. 5, a nonfluorescent uncharged compound, DHR123, has a low solubility in water, so the two surfactants become efficient DHR123 dispersants, thereby reducing their role as quenchers. And it is speculated that an increase in sensitivity has occurred. Clay-dispersed hydrogel and water cannot be considered as the same solvent, but SDS and Tx-100 do not contribute to the improvement of the quantum yield of RD123 alone in the studies of investigating the influence of the surfactant of DHR123 in the aqueous solution.36 Therefore, it is also possibly caused that the dye aggregates due to the presence of clay.
Next, comparing Tx100 and SDS, SDS contains Na+ counter ions. However, this Na+ ion has low reactivity with water decomposed radicals,37 and is considered not to participate in the reaction of DHR123. Nonionic Tx100 and anionic SDS have different micellar structures. It has also been pointed out that SDS, which produces highly polar micelles, has a higher H3O+ concentration.38 It is presumed that there is a good influence on the reaction yield or fluorescence quantum yield due to the change in pH. It is necessary to control each component of RD123 and DHR123, and to investigate the effect of adding a surfactant for further consideration. Moreover, when these two surfactants have different dispersion effects, it is supposed that adding both surfactants at the same time is also effective.38,39
Fig. 6 (a)–(c) Effect of halogenide concentrations on dose sensitivity ((a) TCAA, (b) TBAA and (c) TCE) (d) effect of TBAA concentrations on the fluorescence quantum yield. |
(1) The hydroxyl radical scavenging reaction by DHR123 (k1), (2) the hydroxyl radical scavenging reaction by halogenide (k2), (3) the oxidation reaction of DHR123 by halogen radical (k3) produced by the second reaction. It is considered that the higher the concentration of halogenide, the faster the rate of the second reaction (k2) than that of the first reaction (k1) and the production quantity of halogen radical gets larger. Therefore, the oxidation reaction of DHR123 occurs more easily and the production quantity of RD123 gets larger. However, in the case of adding excess halogenide, it is presumed that other minor reactions, such as the reaction of halogenides occur against the reaction of DHR123 with a halogen radical (k3). As a result, the production quantity of RD123 decreased a little. Considering each rate constant, for example, in terms of TCE, the reactivity with hydroxyl radical is reported. The following showed the rate constants of DHR123 (ref. 6) and TCE37 with hydroxyl radical in eqn (2) and (3) respectively.
DHR123 + ˙OH → DHR–OH˙, k1 = (3.2 ± 0.1) × 1010 M−1 s−1 | (2) |
CCl3CH2OH + ˙OH → product, k2 = 4.2 × 108 M−1 s−1 | (3) |
Comparing these two rate constants, k1 is 100 times larger than k2, the ratio of scavenging hydroxyl radical by TCE (0.5 mM) is 0.06%. It showed a low value and the reaction of TCE with hydroxyl radical rarely occurs. On the other hand, experimental results show the sensitivity enhancement by addition of TCE. Thus, we considered that the rate constant of DHR123 with hydroxyl radical is about 108 M−1 s−1, since RD123 which had a similar structure with DHR123 with hydroxyl radical reported 1.6 × 108 M−1 s−1.6 In this case, the ratio of the scavenging hydroxyl radical calculated with TCE (0.5 mM) using the rate constant k1 = 108 M−1 s−1 is 95%, therefore DHR123 can react with only 5% of the hydroxyl radical. This rate constant is thought to be appropriate because the dose sensitivity did not change despite adding more than 0.5 mM TCE.
Fig. 8 Effect of surfactant and halogenide, TBAA concentrations on dose sensitivity ((a) Tx100, (b) SDS). |
The fluorescence intensity, If, can be expressed by the formula (4) wherein ε is a molar extinction coefficient, c is a concentration of the fluorescent substance, d is an optical path-length (1 cm), I0 is an intensity of excitation light, and Φf is a relative fluorescence quantum yield.
If ∝ ε × c × d × I0 × Φf | (4) |
It is considered that the anticipated increasing rate of sensitivity, ΔIf, can be expressed by formula (5) wherein Δc is the increasing rate of the production concentration of RD123 by adding halogenide and ΔΦf is the increasing rate of the relative fluorescence quantum yield of RD123 (considering inhabitation of reaction by surfactant) by adding surfactant.
ΔIf ∝ Δc × ΔΦf | (5) |
The increasing ratio of dose sensitivity by adding halogenide (TBAA) is 1.72 (±0.05) in 0.5 mM and 1.95 (±0.05) in 1 mM (Table S2†); the increasing ratio of dose sensitivity by adding surfactant is 6.36 (±0.45) in 38 mM Tx100 (average of relative dose sensitivity of No. 2.5 and No. 4.1 in Tables S1 and S3†) and 9.96 (±0.29) in 17 mM SDS (average of relative dose sensitivity of No. 2.10 and No. 4.4). From the calculated anticipated ΔIf using these values and the formula (5), the increasing ratio is 1.72 × 6.36 = 10.9 in 0.5 mM TBAA and 12.4 in 1 mM TBAA. Comparing the measured value, 7.93 ± 0.55 in 0.5 mM TBAA (No. 4.2), 11.2 ± 0.8 in 1 mM TBAA (No. 4.3), and the calculated value, the calculated value was close to the measured value.
In the case of SDS, the anticipated ΔIf which was calculated using the formula (5) like in Tx100, the increasing ratio is 17.1 in 0.5 mM TBAA and 19.4 in 1 mM TBAA. Comparing the measured value, 14.7 ± 1.0 in 0.5 mM TBAA (No. 4.5), 16.2 ± 0.8 in 1 mM TBAA (No. 4.6), and the calculated value, the calculated value was close to the measured value, as well as in Tx100. Therefore, the effect of surfactants and halogenides on the dose sensitivity can be expressed by the formula (5), and it showed that the contributions of the surfactant and the halogenide did not interfere with each other.
No. | Surfactant and additive | Dose rate [Gy min−1] | Dose sensitivity [a.u.] | Ratio | |
---|---|---|---|---|---|
1 | 0 | 0 | 4.64 | 38.23 ± 1.78 | 1.02 |
0.77 | 38.94 ± 2.27 | ||||
4.1 | 38 mM Tx100 | 0 | 4.64 | 217.1 ± 7.15 | 1.15 |
0.77 | 250.5 ± 11.65 | ||||
4.2 | 38 mM Tx100 | 0.5 mM TBAA | 4.64 | 303.2 ± 21.06 | 1.20 |
0.77 | 364.6 ± 19.72 | ||||
4.3 | 38 mM Tx100 | 1 mM TBAA | 4.64 | 426.4 ± 31.67 | 1.02 |
0.77 | 434.3 ± 14.29 | ||||
4.4 | 17 mM SDS | 0 | 4.64 | 381.0 ± 16.17 | 1.04 |
0.77 | 394.9 ± 10.82 | ||||
4.5 | 17 mM SDS | 0.5 mM TBAA | 4.64 | 561.2 ± 36.40 | 1.06 |
0.77 | 597.0 ± 33.46 | ||||
4.6 | 17 mM SDS | 1 mM TBAA | 4.64 | 617.7 ± 28.63 | 1.04 |
0.77 | 643.4 ± 2.25 |
Footnote |
† Electronic supplementary information (ESI) available. See DOI: 10.1039/d0ra02717k |
This journal is © The Royal Society of Chemistry 2020 |