Martin
Šala
*a,
Vid Simon
Šelih
*a,
Ciprian C.
Stremtan
b,
Tudor
Tămaş
c and
Johannes T.
van Elteren
a
aDepartment of Analytical Chemistry, National Institute of Chemistry, Hajdrihova 19, SI-1000 Ljubljana, Slovenia. E-mail: martin.sala@ki.si; vid.selih@ki.si
bTeledyne Photon Machines, 384 Gallatin Park Drive, Bozeman, MT 59715, USA
cDepartment of Geology, Babeş-Bolyai University, M. Kogălniceanu 1, 400084 Cluj-Napoca, Romania
First published on 20th November 2020
In the last few years elemental imaging by laser ablation inductively coupled plasma mass spectrometry (LA-ICPMS) has advanced rapidly, both due to hardware development associated with fast aerosol transport technologies and a deeper understanding of the influence of operational parameters on the image quality. Herein we describe the effect of dosage, i.e., the number of laser pulses per pixel, on the image quality attainable by LA-ICP-QMS as illustrated by mapping of a biological (murine brain tissue) and a mineralogical (asbestos fibers) sample. The usage of higher dosages results in better S/N ratios and is crucial when elements are present at lower concentration levels, or if mapping of more than one element is required. While this potentially increases the mapping time, elemental images with higher dosages will generally be of better quality.
Different approaches are being used for elemental imaging by LA-ICP-MS such as single pulse analysis (each pixel is generated by a single laser shot, i.e., dosage D = 1) and continuous scanning (each pixel is generated by multiple laser shots, i.e., dosage D > 1). Even though the latter approach samples material beyond the pixel size, potentially leading to extra blur, less noise is generated than in the former approach as more counts are accumulated per pixel. The dosage D is not only an important factor in the attainable image quality, but also defines the number of elements that can be measured per pixel with LA-ICP-QMS instruments.18,19 In combination with the latest generation of fast aerosol transfer devices, such instruments can only measure one element in single pulse mode (dosage D = 1); however, dosages of D ∼ 10 allow for measurement of more elements although the mapping time may be extended. Especially the development of new LA cells and transfer devices have led to much shorter particle washout times and therefore much faster responding systems. This has led to mapping times which are 1–2 orders of magnitude faster for the same beam size or increased spatial resolution by generating 10–100 times more pixels which are smaller in the same mapping time. However, smaller pixels generated by smaller beam sizes imply that the amount of material ablated is reduced, leading to a lower sensitivity. A higher dosage is therefore crucial for mapping of elements present at low concentration levels to keep S/N ratios as high as possible, even though higher dosages slightly increase the amount of blur. Image quality is thus a delicate balance between image noise and image blur, which are a direct function of the dosage.
The image quality obtainable by single pulse analysis (D = 1) and continuous scanning (D = 10) was compared by mapping of single or several elements in complex mineralogical (fibrous asbestos compounds) and biological (murine brain tissue sections) samples. Because of the heterogeneous nature of fibrous asbestos compounds, it is difficult to ablate them using single pulse analysis, especially if the laser fluence is not set above the ablation threshold of all the mineral components comprising the mixture. This is also valid for biological samples that contain tissues with different density. However, if the dosage is set correctly, in conjunction with the correct fluence, these types of samples can yield high quality elemental or isotopic images. This has been shown theoretically through modeling,18 but the underlying work shows the practical implications of different dosages used for mapping of a mineralogical and a biological sample.
While modelling allows us to fundamentally study LA-ICP-MS optimization of parameters such as beam size, repetition rate, scanning speed, dwell time and acquisition time in the quest for best image quality, practical application also requires a deeper understanding of instrumental intricacies such as differences in aerosol transport and communication problems between laser and ICP-MS (e.g., triggering, acquisition start, synchronization, etc.). For these reasons, this work focuses on practical assessment of dosage related to image quality for a mineralogical and a biological sample.
The murine brain tissue sample was prepared by freezing of tissue in LN2; sections of 16 μm thickness were then cut on a cryotome, placed on a glass slide and dried in air. No fixing or staining was performed (similar to the procedure described in Opačić et al.).20
Both types of samples were measured in single pulse (D = 1) and multiple pulse (D = 10) mode for imaging of one or several elements. Mapping was carried out on adjacent locations to clearly see the continuation of sample features and allow direct comparison of different mapping conditions. Table 1 summarizes the operational LA-ICP-MS settings; nuclides in bold are the ones related to images indicated in the text. The selection of elements was based on prior knowledge of the concentrations in the samples. In all experiments a beam size of 20 μm (square mask) was used, and other settings were based on model predictions for fastest possible mapping times, avoidance of aliasing, minimal blur and maximal S/N ratios.18 Data processing and image analysis were performed using the software packages HDIP (Teledyne Photon Machines Inc., Bozeman, MT) and ImageJ.21
Brain | Asbestos | |||||||
---|---|---|---|---|---|---|---|---|
LA (Analyte G2, ARIS) | ||||||||
Washout time, ms | ca. 20 | |||||||
Beam size (square), μm | 20 | |||||||
Fluence, J cm−2 | 0.5 | 4 | ||||||
Dosage | 1 | 10 | 1 | 10 | ||||
Repetition rate, Hz | 40 | 294 | 40 | 294 | ||||
Scanning speed, μm s−1 | 800 | 588 | 800 | 588 | ||||
He carrier flow rate, L min−1 cup|cell | 0.3|0.3 | |||||||
ICP-MS (Agilent 7900x) | ||||||||
R f power, W | 1500 | |||||||
Sampling depth, mm | 6.5 | |||||||
Ar makeup flow rate, L min−1 | 0.8 | |||||||
Isotopes measured, dwell time, ms | 56 Fe, 25 | 56 Fe, 10.5 | 55Mn, 7 | 56 Fe, 34 | 59 Co, 25 | 139 La, 25 | 52 Cr, 6 | 59 Co, 34 |
( Fig. 1a ) | 66Zn, 10.5 | 56 Fe, 7 | ( Fig. 1d ) | ( Fig. 3b ) | ( Fig. 3c ) | 59Co, 6 | ( Fig. 3a ) | |
( Fig. 1b ) | 63 Cu, 7 | 139 La, 6 | ||||||
66 Zn, 7 | 146 Nd, 6 | |||||||
( Fig. 1c ) | ( Fig. 3d ) | |||||||
( Fig. 2 ) | ( Fig. 4 ) | |||||||
Duty cycle time, ms | 25 | 34 | 25 | 34 | ||||
Mapping rate, kpx h−1 | 144 | 105.9 | 144 | 105.9 |
Commonly, biological samples are “thin” tissue slices with a thickness ranging from less than one to several tens of micrometres, which of course determines the useable dosage as too high a dosage will lead to complete consumption of the sample. Unless one chooses to do this intentionally to increase the spatial resolution in the scan direction, by removing only a small fraction of a pixel using a single laser shot with a high fluence, this may cause sensitivity issues. In geological samples usually this problem does not arise as samples are thick enough so that even at higher dosage, the sample is not fully ablated through.
Although quantification-related issues are not the topic of this paper, some discussion about this important issue is in place. The ability of high-speed, high-dosage LA-ICP-MS for quantitative mapping still has its restrictions as only several milliseconds are available for measurement of each and every pixel in the map. Consequently, only a handful of elements can be measured, and a calibration approach based on e.g. sum normalisation for elemental mapping (citations), requiring the measurement of tens of elements, is out of the question for sequential scanning QMS instruments but feasible for simultaneous TOFMS systems. However, samples with a homogeneously distributed element that can be used for internal standardisation can be easily quantified for a limited number of elements using LA-ICP-QMS mapping.
Fig. 1 LA-ICPMS images of murine brain tissue showing the distribution of 56Fei for measurement of (a) one nuclide (D = 1), (b) two nuclides (D = 1), (c) four nuclides (D = 10) and (d) one nuclide (D = 10). See Table 1 for details. |
Although Fig. 1a (measurement at D = 1 of one nuclide) shows a relatively high-detail image, it has to be compared to Fig. 1c and d (measurement at D = 10 of four and one nuclides, respectively). As expected, Fig. 1d has a ca. ten times higher count rate than Fig. 1a due to a dosage difference of ten, but still a (subjective) similar visual image quality. Fig. 1c has roughly the same count rate, and also a similar (subjective) visual image quality than Fig. 1a, in spite of the fact that in Fig. 1c three more nuclides are measured next to 56Fe. However, to accommodate the measurement of four elements the mapping rate decreases from 144 to 105.9 kpx h−1. This is due to the fact that we run into the repetition rate ceiling, i.e., 300 Hz, for multiple pulse analysis at a dosage D = 10 in the LA instrument used. A laser head with a repetition rate of 400 Hz would facilitate the measurement at 144 kpx h−1 as well, but then the four elements would need to be measured in 25 ms. According to Table 1 this would be possible as hopping and settling takes only 6 ms (hopping and settling times are instrument-specific), leaving 19 ms to be distributed among measurement of these elements. The use of high repetition rate laser heads for imaging applications have already been documented.17
A significant advantage of the multiple dosage analysis technique compared to the single dosage approach is its multielement capabilities as shown in Fig. 2 where an RGB overlay of the maps of three elements is presented. This map clearly shows different distributions and associations of elements in the sample that can (in general) help researchers understand physiological and metallomic implications connected to the samples analysed.
Fig. 2 Multielement RGB map of the murine brain tissue, overlaying the maps of three elements (56Fe, 63Cu and 66Zn). |
Fig. 3 LA-ICPMS images of asbestos fibres showing the distribution of 59Co upon measurement of (a) one nuclide (D = 10) and (b) one nuclide (D = 1), and 139La upon measurement of (c) one nuclide (D = 1) and (d) four nuclides (D = 10). See Table 1 for details. |
Fig. 4 Multielement RGB map of the asbestos fibres, overlaying the maps of three elements (52Cr, 139La and 146Nd). |
Although not visually obvious from the maps shown, higher dosages result in less image noise and higher S/N ratios as illustrated in Table 2. A NIST SRM612 glass reference standard was mapped with LA-ICP-QMS parameters similar to those used for mapping of the asbestos samples. We selected aluminium and lanthanum as elements representative for high (0.537 ± 0.011% m/m) and low (36 ± 7 μg g−1) concentrations, respectively. Al and La were mapped individually and simultaneously for dosages of 1, 2, 5 and 10. From Table 2 it is clear that higher dosages result in better S/N ratios, more pronounced for the lower concentration element La. Measurement of two elements simultaneously yields an overall lower sensitivity and thus an even higher influence of dosage on the S/N ratio.
Dosage D | Al | La | Al (+La) | La (+Al) |
---|---|---|---|---|
1 | 19.6 | 12.7 | 2.6 | 1.7 |
2 | 19.6 | 18.2 | 7.0 | 6.7 |
5 | 21.3 | 20.8 | 20.8 | 14.9 |
10 | 24.4 | 23.8 | 23.8 | 16.1 |
This journal is © The Royal Society of Chemistry 2021 |