Tetyana
Sergeyeva
*a,
Daria
Yarynka
a,
Vitaly
Lytvyn
b,
Petro
Demydov
b,
Andriy
Lopatynskyi
bc,
Yevgeny
Stepanenko
a,
Oleksandr
Brovko
d,
Anatoly
Pinchuk
e and
Volodymyr
Chegel
*bc
aInstitute of Molecular Biology and Genetics, National Academy of Sciences of Ukraine, 150, Zabolotnogo Str., 03143 Kyiv, Ukraine. E-mail: t_sergeyeva@yahoo.co.uk
bV.E. Lashkaryov Institute of Semiconductor Physics, National Academy of Sciences of Ukraine, 41, prospect Nauky, 03680 Kyiv, Ukraine. E-mail: vche111@yahoo.com
cInstitute of High Technologies, Taras Shevchenko National University of Kyiv, 64/13, Volodymyrska Street, 01601 Kyiv, Ukraine
dInstitute of Macromolecular Chemistry, National Academy of Sciences of Ukraine, 48, Kharkivske shosse, 02160 Kyiv, Ukraine
eUniversity of Colorado Colorado Springs, 420 Austin Bluffs, Parkway, CO, USA
First published on 10th February 2022
We demonstrate a novel sensor platform with enhanced sensitivity and selectivity for detecting aflatoxin B1 – a common food toxin in cereals. The approach is based on a molecularly imprinted polymer film that provides selective binding of the aflatoxin B1 and fluorescence signal from the analyte molecule enhanced by the local electric field induced in close proximity to the surface of a silver nanoparticle excited at the localized surface plasmon resonance (LSPR) wavelength. Molecularly imprinted polymers (MIPs) with supramolecular aflatoxin-selective receptor sites and embedded spherical silver nanoparticles (with diameters 30–70 nm, the LSPR band 407 nm) were prepared in the form of a thin polymer film on the surface of a glass slide using in situ polymerization. The detection limit of the sensor for aflatoxin B1 is 0.3 ng mL−1, which is significantly lower than for a fluorescent sensor without silver nanoparticles. The plasmon-enhanced fluorescence factor is 33, and the linear dynamic range of the sensor is 0.3–25 ng mL−1.
Metal nanoparticles can increase the quantum yield and improve the photostability of the analyte fluorophore molecules at optimized distances around 5–90 nm.10–12 This phenomenon is commonly referred to as metal-enhanced fluorescence (MEF) or plasmon-enhanced fluorescence (PEF).13,14 A few physical–chemical mechanisms are suggested to be responsible for the metal-enhanced fluorescence. First, the nanoparticle's highly enhanced local electrical field at the wavelength of localized surface plasmon resonance (LSPR) results in modified optical characteristics of the nearby fluorophore when overlapping LSPR and fluorescence spectra is provided. Specifically, metal nanoparticles act as antennas for the fluorophore, which typically results in higher emission intensity.11,12 The shape and size of the metal nanostructures are critical for metal-enhanced fluorescence. Second, the non-radiative energy transfer between the metal nanoparticle and the fluorophore that depends on spectral overlap between the metal surface and the fluorophore plays a crucial role in fluorescence enhancement. This process is explained by Förster resonance energy transfer (FRET).15
The plasmon-enhanced fluorescence is widely used to increase the sensitivity of fluorescent sensors.16,17 Typically, PEF provides ten to hundreds of enhancement factors and enables the fluorimetric registration of various analytes with an attomolar limit of detection and single-molecule detection by exploiting the localized surface plasmon excitation in metal nanoparticles.18–20
PEF-based immunosensors were reported to detect different mycotoxins such as ochratoxin A, fumonisin B1, deoxynivalenol with the limits of detection in the ng kg−1 – μg kg−1 range.21,22 For example, gold nanoparticles (AuNPs) were used as signal enhancers for fluorescence-based detection of aflatoxin B1, resulting in developing highly-sensitive immunosensor assay.23 AuNPs were obtained using green synthesis and further conjugated with anti-AFB1 antibodies. Thus enhancement of the sensitivity of the immunosensors for aflatoxin B1 detection was demonstrated.23 PEF technique in different implementations for enhanced sensitivity of fluorescent immunosensor detection of aflatoxins was reported for reaching the limits of detection down to 50 fg mL−1.23–26
Moreover, emerging applications of molecularly imprinted polymers (MIPs) as artificial receptors in sensors based on plasmon- and surface-enhanced spectroscopies were reported.27,28 PEF-based sensor exploiting core–shell Ag@SiO2 hybrid nanostructures and boronate MIP was demonstrated for sensitive detection and specific recognition of riboflavin in human urine with the limit of detection of 3.3 ng mL−1.28 Silica shell of the thickness 15 nm has been used a spacer on the surface of Ag nanoparticles with an average diameter of 60 nm to produce the fluorescence enhancement factor of 27.6. The sensor exhibited a linear relationship between the fluorescence intensity and riboflavin concentration in the range 10–700 ng mL−1, imprinting efficiency of 51%, and fast reaching binding - equilibrium within 20 min. Several MIP-plasmonic sensors based on surface-enhanced Raman scattering (SERS) readout were developed recently,29–31 confirming the possibility of combining the plasmonic enhancement technique with MIP technology for the selective detection of various analytes. The difference of signal enhancement mechanisms in MIP-plasmonic sensors using PEF and SERS phenomena was reported, implying more complex nature of enhancement factor in PEF as a compromise between the electric field enhancement and fluorescence quenching.28
To achieve the highest fluorescence enhancement factor, - the PEF approach - usually requires precise control and complex lithographic techniques to tune the plasmonic nanoparticles into a spectral-spatial configuration with the fluorophore and provide the largest plasmonic “hot spots” intensity while keeping the fluorescence quenching low.18–20 At the same time, relatively simple and tunable in situ syntheses of plasmonic nanoparticles inside MIPs were demonstrated by chemical or photoreduction methods.32–35 Importantly, these fabrication techniques allow optimization of the plasmonic enhancement of the intramolecular optical processes, e.g., in SERS spectroscopy34 by changing the concentration of the precursor, which defines the size and concentration of the synthesized nanoparticles as well as molecule-nanoparticle distance.
MIP films or thin films with incorporated metal nanoparticles can serve as an alternative to natural antibodies and receptors for producing sensing chips for the fluorescent affinity sensor. MIPs are generally as sensitive and selective as natural receptors, while they normally demonstrate superior stability and inexpensive, user-friendly production procedures. On the other hand, AgNPs can be used as an enhancer for AFB1 fluorescence detection and a sensor signal amplifier. Gold and silver nanoparticles can be formed during the in situ synthesis in the structure of a number of polymers, e.g., poly(ethyleneimine), poly(hydroxyethyl methacrylate), poly(vinylpyrrolidone), novolak, poly(4-vinylphenol), poly(4-vinylphenol)-co-(methyl methacrylate), poly(styrene-co-allyl alcohol),36,37 and MIPs.38,39 The size and shape of the described nanoparticles as well as the distance between them, which are the important parameters determining their ability to enhance fluorescence intensity, can be controlled.36,37
This paper introduces the novel nanostructured composite polymeric films with embedded in situ-synthesized Ag nanoparticles and a portable sensor device prototype for the PEF-based MIP-assisted detection of aflatoxins. We describe the fabrication protocol of MIP films and their sensing performance in terms of concentration-dependent response, selectivity, and optimization of Ag nanoparticles content. To the best of our knowledge, the proposed in situ synthesis route of the sensor films by simultaneous UV photoreduction of Ag nanoparticles and UV polymerization of the monomer- mixture was used here for the first time for the fluorescent detection of aflatoxins. We hypothesize that the metal-enhanced fluorescence significantly increases AFB1 quantum yield and brightness and, consequently, improves the AFB1 detection limit. The proposed prototype of portable sensor system can be used for rapid and reliable aflatoxin B1 monitoring at extremely low concentration levels.
Monomer/oligomer | MIP | NIP |
---|---|---|
Ethyl-2-oxocyclopentanecarboxylate | 10 mg | — |
Acrylamide | 9 mg | 9 mg |
TGDMA | 162 mg | 162 mg |
OUA | 29 mg | 29 mg |
DMF | 100 μL | 100 μL |
Polyethyleneglycol (Mw 20![]() |
30 mg | 30 mg |
2,2′-Dimethoxy-2-phenylacetone | 0.2 mg | 0.2 mg |
The monomer mixtures were heated at 80 °C for 2 min to dissolve PEG 20000. This was followed by the addition of the initiator and a final 30 min-polymerization step initiated by UV light (λ = 365 nm, intensity 3.4 W m−2).
To obtain AFB1-selective MIP films with AgNPs embedded in their structure, 0.6–291 mM AgNO3 were added to the initial monomer mixtures used for the synthesis of AFB1-selective films. AgNPs were formed in situ during the pre-heating step and further UV-initiated polymerization procedure.
The MIP films were synthesized using a dummy template-based approach with ethyl-2-oxocyclopentanecarboxylate as a template molecule and acrylamide as a functional monomer.41 The corresponding non-imprinted polymer (NIP) films were obtained without ethyl-2-oxocyclopentanecarboxylate so that no aflatoxin B1-selective sites were formed in their structure. The polymerization was followed by an 8 h Soxhlet extraction procedure in ethanol to remove the dummy template and the other monomers, which were not included in the structure of the MIP and NIP films. That was followed by the other 8 h extraction procedure in distilled water at 80 °C to remove PEG 20000 responsible for the pore formation in the polymers. Finally, the fully-formed MIP and NIP films immobilized on the glass slides were dried and stored at room temperature for further investigations. The thickness of the MIP films immobilized on glass surfaces was estimated using Electronic Digital Caliper 0–300 micrometer (“Adoric”, Ningbo, China).
After incubation, the MIP and NIP films were rinsed in 20 mM Na-phosphate buffer pH 6.0, containing 10% acetonitrile, and dried at room temperature. The sensor responses were registered after UV-irradiation of the sensor chips (MIP films covalently attached to the surface of glass slides) fixed in a solid samples’ holder of the spectrofluorimeter (Ex bandwidth 5 nm, Em bandwidth 2.5 nm, excitation wavelength 365 nm, measurement range 395–550 nm, and emission wavelength 420 nm) directly at the films’ surface (Fig. S4†).
![]() | ||
Fig. 1 Optical setup of portable fluorimeter: 1, 6 – collimators, 2 – bandpass filter (365 nm), 3 – sample holder, 4 – nanochip with highly-conductive nanoparticles, 5 – longpass filter (400 nm). |
The portable fluorimeter uses ultraviolet (365 nm) LED as an excitation light focused by a collimator (Ocean Insight 74-UV) and sent through an optical cable to the sample holder with a built-in bandpass filter (365 nm). The sensor chips were fixed by a specially designed holder for solid samples. The fluorescent signal amplified by the plasmonic nanochip is collected by a collimator and is registered with a photodiode through a longpass filter (400 nm). The recorded signal is transmitted via Bluetooth to the receiving gadget or, after focusing, with an optical cable to the spectrometer and computer.
The registration principle of the proposed MIP-AgNPs sensor film is based on the selective binding of aflatoxin B1 in analyzed samples by nanoreceptor sites and registration of the plasmon-enhanced fluorescence signal of aflatoxin molecules excited by UV irradiation. A home-built portable fluorimeter registered the fluorescent sensor signals. Silver nanoparticles in the MIP films’ structure amplify the fluorescence signal due to the enhanced local electric field resulting from localized surface plasmon resonance excitation in nanoparticles.
The developed sensor chips based on immobilized acrylamide-containing MIP thin films were produced and tested for their ability to generate fluorescent sensor responses due to AFB1 binding. NIP films were obtained using the same compositions of monomers except for the dummy template (ethyl-2-oxoceclopentanecarboxylate)41 that was added to the MIP compositions only. The thickness of the MIP films immobilized on glass surfaces was estimated as 80 μm. The aflatoxin B1 detection is possible due to its natural ability to fluorescence. The sensor signals were excited with UV-irradiation (λ = 365 nm) after incubation of the developed sensor chips in 15–500 ng mL−1 solutions of AFB1, while blue fluorescence of the toxin of interest selectively adsorbed on the surface of the chips was registered at λ = 420 nm directly at the sensor chip surface with both the standard laboratory spectrofluorimeter and the home-built portable fluorimeter for AFB1 in-field detection (section 2.4).
Fig. 2 shows a typical calibration plot for the fluorescent sensor systems based on acrylamide-containing MIP films immobilized on the surface of a glass chip. Fig. 2a graph demonstrated a typical form indicating the presence of affinity receptor sites for AFB1 binding in MIP film's structure in contradistinction to the NIP film, where the calibration plot is linear. As seen for MIP at high concentrations, receptor binding sites are saturated.
The limits of detection were determined based on European standard ISO 11843-2:2000.42 The lowest detection limit (LOD) for AFB1 detection was estimated to be 10 ng mL−1, while the linear dynamic range was 10–250 ng mL−1 (Fig. 2b). The storage stability for the sensing chips for AFB1 fluorescent detection was assessed for 18 months at room temperature.
The sensor described in this paper has a better LOD (10 ng mL−1) as compared to the previously-reported sensor based on the application of 60 μm-thick free-standing MIP films41 whose LOD was 15 ng mL−1, as well as better reproducibility (R = 107% vs. R = 87%). Moreover, the proposed sensor chips are also characterized by superior mechanical stability.
The ability of the developed fluorescent sensors based on MIP-based sensor chips to discriminate between the toxin of interest and its potential interferents that are often present in the extracts of the analyzed cereal samples was investigated using the other structurally similar fluorescent mycotoxins, i.e., aflatoxins B2 and G2 (AFB2, AFG2) as well as ochratoxin A (OhA).
The differential fluorescent sensor signals, generated with the immobilized MIP-based sensor chips in response to the addition 100 ng mL−1 AFB1, AFB2, AFG2, and OhA, were evaluated (Fig. 3).
Fig. 3 exemplifies that the MIP films immobilized on the glass surface exhibit preferential binding of the toxin of interest as compared to the potentially interfering fluorescent substances. However, the achieved LOD for AFB1 detection (10 ng mL−1) required further improvement to ensure sensitive detection of the target toxin in food and feeding stuffs. Therefore, the initial MIP composition was further modified by incorporating Ag nanoparticles in their structure.
Here we demonstrate the in situ formation of Ag nanoparticles directly in the MIP thin film during polymer formation. Silver nanoparticles (AgNPs) were synthesized using a reduction of silver nitrate (AgNO3) added to the initial composition for the imprinted polymer synthesis in specific concentrations. The proposed synthetic route is based on silver nanoparticles synthesis in MIP thin film structure due to AgNO3 reduction during the pre-heating step of the initial monomer mixture used for the MIP formation followed by UV-initiated photopolymerization. To the best of our knowledge, this is the first demonstration of Ag nanoparticles synthesis in the MIP structure for the effective fluorescent detection of aflatoxins. AgNO3 was added to the initial monomer mixture to the final concentrations varying from 0.6 mM to 291 mM.
The morphology of the AgNPs formed in the structure of the AFB1-selective MIP films was investigated using TEM microscopy. Fig. 4 shows TEM images of the MIP thin films obtained by in situ polymerization with the AgNPs embedded in the polymer structure.
![]() | ||
Fig. 4 TEM images of AgNPs formed in the structure of MIP thin polymer films synthesized with AA as a functional monomer taken at different magnifications (scale 100 nm (a) and 200 nm (b)). |
As one can see, the shape of AgNPs formed in the structure of the MIP films at the optimized conditions (from the monomer compositions containing 1.5 mM AgNO3 in the initial mixture of monomers for the MIP synthesis) was close to spherical shape with the size of 30–70 nm. The nanoparticles were evenly distributed in the structure of the polymer. According to the numerical calculations of the system with spherical silver nanoparticles (Fig. S1 in ESI†), this size of AgNPs provided some fluorescence enhancement, and, at the same time, there is a possibility to improve the already reached enhancement results by increasing the nanoparticle size (Fig. S2 in ESI†). Some deviations from the spherical shape (i.e., rectangular, hexagonal) were also observed, as one can see in Fig. 4. This limits the results of the used model and the possibility of further size growth for spherical nanoparticles under the used polymerization conditions. We note that the presence of copper grid during polymerization procedure, which could influence the reduction of silver ions and, consequently, the morphology of synthesized nanostructures, should be considered.
The rule of plasmonic enhancement of fluorescence stating that the surface plasmon resonance bands are supposed to overlap with the excitation or emission bands of the fluorophore molecules43–45 was confirmed in this work. Taking into account that the measured LSPR wavelength position of the prepared Ag nanoparticles was about 407 nm, the LSPR band of spherical AgNPs overlaps significantly with the excitation and emission bands of aflatoxin B1 exhibiting peaks at 365 nm (ref. 41) and 420 nm, respectively (Fig. S3†), which satisfies a condition of plasmon resonance energy transfer between the nanoparticle and the fluorophore. Moreover, due to the proposed AgNO3 reduction by the initial monomer mixture, the prepared Ag nanoparticles are mainly surrounded by the polymer that creates an advantageous condition for the mechanism of fluorescence enhancement in comparison with possible fluorescence quenching.
All the MIP and NIP Ag-containing polymers immobilized on the surface of glass slides were used as sensor chips for aflatoxin B1 detection. The ability of AgNPs formed in the polymer structure to improve the detection limit of the target mycotoxin due to the plasmon-enhanced fluorescence phenomenon was investigated. All synthesized films were tested for their ability to bind AFB1 from aqueous solutions and generate the fluorescent sensor response, as shown in Fig. 5.
For the MIP films synthesized from the mixtures containing 1.5 mM AgNO3, a 14-fold increase in the value of the fluorescent sensor responses compared to the AgNPs-free MIP films due to AFB1 addition was observed Fig. 6. The higher and lower AgNO3 concentrations added to the MIP composition resulted in significantly lower (2.1–4.5-fold) enhancement effects. Higher concentrations of silver nitrate had the opposite effect on AFB1 fluorescence. For MIP films synthesized from the mixtures containing above 15 mM AgNO3, quenching of aflatoxin B1 fluorescence was observed.
Fig. 6 shows the enhancement of the fluorescence intensity of aflatoxin B1 due to the surface plasmon resonance in AgNPs. All measurements were carried out using Fluoromax_PLUS_PR928P spectrofluorimeter, as well as portable fluorimeter (excitation wavelength 365 nm, measurement range 395–550 nm). According to these results, the fluorescence spectra obtained using standard laboratory spectrofluorimeter Fluoromax_PLUS_PR928P (Horiba, Japan) and portable fluorimeter were identical.
The addition of AgNO3 to the initial composition of monomers leading to the formation of Ag nanoparticles in the polymer structure results in a significant increase in values of the sensor responses and a significant increase of the aflatoxin B1-selective MIP films’ imprinting factor. We observed a 4.7-fold increase in the imprinting factor at 100 ng mL−1 AFB1 concentration.
The AgNP-containing MIP-based sensor chips were tested as selective elements of the fluorescent sensor, and the ability of AgNPs to improve the minimal detection limit of aflatoxin B1 detection was studied. Typical calibration plots for the fluorescent sensor systems based on acrylamide-containing MIP-films synthesized with addition 1.5 mM AgNO3 are presented in Fig. 7.
The minimal detection limit of the aflatoxin B1-selective sensor system decreased by a factor of 33 because of the plasmon-enhanced fluorescence: the LOD was decreased down to 0.3 ng mL−1 for AgNPs-based MIP films as compared to 10 ng mL−1 for the unmodified MIP-based sensor systems.
The linear dynamic range of the fluorescent sensor is 0.3–25 ng mL−1, as can be seen in Fig. 7 (see Fig. S5† for low AFB1 concentrations).
The fluorescent sensor intensity signals of the sensors based on AgNPs-containing MIP films were 40–80% higher than those of AgNPs-free MIPs, as shown in Fig. 8.
AgNPs-containing MIP thin films demonstrate superior properties as sensitive elements of a portable fluorescent sensor system as compared to unmodified MIPs. They can generate significantly higher sensor responses, which resulted in the 33-fold decrease of the minimal detection limit for AFB1.
The selectivity of the newly synthesized AgNPs-containing sensor chips was also evaluated (Fig. 9).
The differential fluorescent sensor signals, generated with the AgNPs-containing MIP sensor chips in response to the addition of 15 ng mL−1 AFB1, AFB2, AFG2, and OTA, were evaluated.
The developed sensor system based on AgNPs-containing MIP sensor chips exhibits better selectivity. The sensor responses initiated by the addition of the potentially interfering AFB1 structural analogues (AFB2, AFG2, and OhA) were negligible compared to those generated by the target mycotoxin (AFB1). Moreover, the selectivity of the AgNPs-containing MIP-based sensors was much better than the selectivity of the unmodified MIP films shown in Fig. 3.
In order to evaluate the performance of the developed portable fluorescent sensor system based on AgNPs-containing MIP sensor chips for the food quality monitoring, the real cereal samples were tested for AFB1 contamination using the developed method. Five different maize flour samples were tested. The first three samples (sample no. 1, 2, and 3) were naturally contaminated with AFB1 and provided by Romer Labs (Kyiv, Ukraine): sample no. 1 – “Romer Labs CheckSample Survey Aflatoxins in corn (CSSMY013 – M17411A)”; sample no. 2 – “Quality control material Aflatoxins in corn, low level”, sample no. 3 – “Quality control material Aflatoxins in corn, mid-level”. The manufacturer characterized all the naturally contaminated samples for the AFB1 content by standard analytical methods (HPLC and ELISA). Also, AFB1-free maize flour samples produced by two different manufacturers (“Lavka Tradytsiy”, Ukraine (sample no. 4), “Dobrodiya Foods”, Kyiv, Ukraine (sample no. 5)) and purchased in the local supermarket were used in the investigation. According to the standard procedure described earlier, the procedure of AFB1 extraction from maize flour was made using 80:
20 v/v acetonitrile
:
H2O solution.41,46 The AFB1-free flour extracts were spiked with 1 (sample no. 4) and 5 (sample no. 5) ng mL−1 AFB1. Results of aflatoxin B1 detection in real maize flour samples using the fluorescent sensor system based on AgNPs-containing AFB1-selective MIP thin films are presented in Table 2. Fluorescent sensor responses were estimated by the home-built portable fluorimeter.
Sample No. | Amount of AFB1 in the sample (according to the traditional method) | Amount of AFB1 in the sample (according to portable fluorescent MIP-based sensor system) | Recovery (R, %) |
---|---|---|---|
1 | 7 μg kg−1 | 6.4 ± 3.2 μg kg−1 | 109 |
2 | 4.2 ± 1.7 μg kg−1 | 4.5 ± 2.0 μg kg−1 | 93 |
3 | 7.3 ± 2.9 μg kg−1 | 7.7 ± 0.9 μg kg−1 | 94 |
4 | 1 ng mL−1 | 1.9 ± 0.9 ng mL−1 (0.8 μg kg−1) | 90 |
5 | 5 ng mL−1 | 5.4 ± 0.8 ng mL−1 (2 μg kg−1) | 93 |
The results presented in Table 2 clearly demonstrate the efficiency of the proposed AgNPs-containing MIP-based sensor system for monitoring AFB1 contamination of the cereal samples. The obtained results of the analytical identification of aflatoxins corresponded to the traditional instrumental and immunochemical methods.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: 10.1039/d1an02173g |
This journal is © The Royal Society of Chemistry 2022 |