Krzysztof
Swiderski
*,
Krzysztof
Greda
,
Pawel
Pohl
and
Piotr
Jamroz
*
Department of Analytical Chemistry and Chemical Metallurgy, Faculty of Chemistry, Wroclaw University of Science and Technology, Wyb. Wyspianskiego 27, 50-370 Wroclaw, Poland. E-mail: piotr.jamroz@pwr.edu.pl; krzysztof.swiderski@pwr.edu.pl; Fax: +48-71-320-3807; Tel: +48-71-320-3807
First published on 31st January 2022
The application of hanging drop cathode (HDC) atmospheric pressure glow discharge (APGD) coupled with a Dove prism (DP) as an efficient excitation microsource for optical emission spectrometry (OES) was evaluated in details. The design of experiment (DoE) approach was applied to optimize the operating conditions, i.e., the discharge current, the graphite tube diameter, and the HNO3 concentration, of this excitation source. The effect of the addition of low molecular weight organic compounds (LMWOCs) on the intensity of the analytical lines of Ag, Ca, Cd, Hg, Mg, Mn, Pb and Tl was evaluated, and the most beneficial results were obtained for 8% (m/m) formic acid. Using a Dove prism, the 2-D spatial distribution profiles of the emission lines of the selected elements (Ag, Ca, Cd, Hg, Mg, Mn, Pb and Tl) and the background species (H, O I, O II, N2, OH, and NO) were obtained. In the developed system, the enhancement of the Ag, Pb and Tl atomic emission lines was the most spectacular and resulted in the improved limits of detection (LODs) of these elements, being 0.54, 14 and 2.2 μg L−1, respectively. For this reason, this method was used for the sensitive determination of these analytes. The trueness and the precision of the developed method were verified by the analysis of a certified reference material (CRM) of tea leaves (INCT-TL-1).
The newest APGD systems can obtain the limits of detection (LODs) of various elements comparable to or even better than those offered by commercially available ICP spectrometers.1,2 In addition, the past decades of research devoted to APGD brought a much better understanding of the processes occurring at the plasma–liquid interface and being responsible for the transport of analytes from the delivered solutions into the discharge column, in addition to the atomization and excitation processes occurring therein. To date, numerous parameters affecting the analytical performance of APGD-OES methods were investigated, including the type of acid used for acidification of the delivered solutions,1,4 pH of these solutions,1,5,6 the concentration of the added ionic and non-ionic surfactants7 or low molecular weight organic compounds,5,8,9 and the discharge current,6,10 as well as the polarization of the liquid electrodes.11 All these parameters were related to the discharge itself or to the analyzed sample solutions.
Unfortunately, there was very little interest in the optimization of the imaging of the discharge on the entrance slit of the spectrometer because the interelectrode gap was overwhelmingly 1:1 imaged using appropriate focusing lenses.1 Since APGD operated in contact with flowing solutions is rather non-uniform plasma, the spatially resolved acquisition may influence the analytical response of this excitation source, as reported.12,13 Therefore, an interesting modification of the optical system for APGD can be the application of a Dove prism (DP), which allows the rotation of the discharge image and influences the measured signals.
In this work, the application of a newly developed hanging drop cathode atmospheric pressure glow discharge (HDC-APGD) system12 along with the OES detection, realized by a CCD imaging spectrometer and a DP, is presented. A rotation of the discharge image by 90° was established to reduce the intensity of the background species and increase the analytical signals of the studied elements. This effect led to a significant improvement of the LODs of the elements, particularly of those for which the analytical lines are observed in the vicinity of the OH, NO and N2 molecular bands. Using the design of experiments (DoE) approach, which involves the change of several parameters at the same time due to the Box–Behken response surface design, the discharge current, the graphite tube (to deliver the solutions) diameter, and the HNO3 concentration (to acidify these solutions) were optimized. Additionally, the influence of low molecular weight organic compounds (LMWOCs) added to the sample solutions was evaluated. Using the DP, 2-D spatial distribution maps of the analytical lines of the selected elements and the background species were obtained. Finally, the application of the developed method in the trace analysis (Ag, Pb and Tl) of a certified reference material (CRM) of tea is presented.
The radiation emitted by HDC-APGD was imaged 1:1 using an achromatic quartz lens (f = 78.5 mm) on the entrance slit (10 μm) of a Shamrock SR-500i (Andor, UK) spectrometer, equipped with a Newton DU-920P-OE (Andor, UK) CCD camera and two holographic gratings (1800 and 1200 groves per mm). The interelectrode gap was imaged vertically and horizontally on the entrance slit of the spectrometer. A quartz DP (Lambda Research Optics, USA), placed in a kinematic rotation mount (Thorlabs, Sweden), was applied in order to rotate the obtained image (90°) on the spectrometer entrance slit.
To handle the spectrometer and control its configurations, Solis S (Andor, UK) software was used. The intensities of the atomic emission lines and the molecular band heads were background corrected using the two-point correction method on either sides of their profiles and the baseline signal subtraction, respectively. The CCD camera was operated in a full vertical binning (FVB) mode as well as in an image mode (1024 × 255 px). In the FVB mode, the integration time was 1 s with 10 repetitions. In the case of the image mode, in order to obtain more reliable data, the integration time was set to 10 s, which was caused by the lower sensitivity of the CCD camera in this mode.
A CRM of tea leaves (INCT-TL-1), obtained from the Institute of Nuclear Chemistry and Technology, Poland, was used to assess the trueness of the developed method. Its samples (0.5 g, n = 3) were wet digested in 10 mL of a 65% HNO3 solution (Sigma-Aldrich, Germany), using a programmable DigiPrep Jr digestion block (SCP, Canada) and appropriate quartz digestion tubes (SCP, Canada). The digestion temperature was set to 100 °C for 2 h and then increased to 120 °C for further 2 h. The resulting sample solutions of the INCT-TL-1 CRM were cooled down to room temperature and a portion of 5 mL of a 30% H2O2 solution (Avantor Performance Materials, Poland) was added to them. The heating of these sample solutions at 110 °C was continued until they were evaporated to near dryness. The residues left were dissolved in water (5 mL) and filtered through 0.45 μm nylon syringe filters. Subsequently, the prepared sample solutions were acidified with HNO3 to a concentration of 0.6 mol L−1 and formic acid was added (a final concentration of 8%, m/m). For the quantification of Ag, Pb and Tl in the INCT-TL-1 CRM samples solutions, the triplicate standard additions method was applied.
Coefficients of regression equation | R-statistics of models | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Const | A | B | C | A 2 | B 2 | C 2 | AB | AC | BC | R 2 | R (adj) 2 | R (pred) 2 | |
Ag | −1.25 × 105 | 3.24 × 103 | −1.67 × 105 | 1.58 × 104 | — | 1.07 × 105 | — | 4.24 × 103 | −3.44 × 102 | −7.24 × 103 | 99.3 | 98.7 | 96.7 |
Ca | −3.78 × 104 | 1.30 × 103 | −4.94 × 104 | 4.81 × 103 | −7.63 | — | −177.1 | 5.81 × 102 | −5.41 × 101 | 2.09 × 103 | 93.9 | 85.8 | 38.5 |
Cd | −6.10 × 104 | 1.28 × 103 | −4.41 × 104 | 9.18 × 103 | — | — | — | 3.00 × 103 | −1.69 × 102 | −6.26 × 103 | 94.5 | 90.3 | 83.3 |
Hg | −2.08 × 103 | 1.81 × 101 | −6.56 × 103 | 5.17 × 102 | 1.52 | 5.53 × 103 | 20.94 | 2.55 × 102 | −1.87 × 101 | −6.82 × 102 | 99.9 | 99.7 | 98.6 |
Mg | −6.01 × 105 | 1.78 × 104 | −9.00 × 105 | 8.14 × 104 | — | 5.96 × 103 | — | 1.83 × 104 | −1.85 × 103 | −3.06 × 104 | 99.1 | 98.1 | 95.3 |
Mn | −3.66 × 104 | 1.07 × 103 | −8.41 × 104 | 6.29 × 103 | — | 4.96 × 104 | — | 1.78 × 103 | −1.33 × 102 | −3.02 × 103 | 98.8 | 97.5 | 91.5 |
Pb | −2.24 × 104 | 5.78 × 102 | −2.97 × 104 | 3.30 × 103 | — | — | — | 6.76 × 102 | −6.90 × 101 | — | 91.6 | 87.0 | 69.7 |
Tl | −1.25 × 105 | 6.02 × 103 | −1.70 × 105 | 9.32 × 103 | −47.80 | — | — | 2.81 × 103 | −1.88 × 102 | — | 83.8 | 71.7 | 31.9 |
Considering the coefficient of determination (R2), the closer to 100% its value, the more accurate the obtained model. As can be seen from Table 1, the models describing the effect of the studied experimental factors on the net intensities of the analytical lines of Ag, Ca, Cd, Hg, Mg, Mn, and Pb explained 92% to 100% of the measured responses, which is a very satisfying result. Only in the case of Tl, the R2 value was 84%. Nevertheless, small differences between the R2 values and the adjusted R2 (R(adj)2) values for all studied elements indicated that the obtained models were not over-fitted. Considering the predicted R2 (R(pred)2) values, the models could well predict the new responses, particularly in the case of Ag, Cd, Hg, Mg and Mn.
As could be expected, an increase in the discharge current resulted in an enhancement of the intensity of all investigated analytical lines of the studied elements (Fig. 2a). The increased discharge current certainly resulted in a more powerful discharge and a higher transport of the solution components into its phase because the sample consumption changed. In a certain period of time, when the discharge was sustained, it was possible to increase the sample uptake rate necessary to form the drop and operate the discharge. It was established that the sample uptake rate increased from 0.40 mL min−1 to 1.10 mL min−1 as the discharge current increased from 35 to 65 mA. This higher discharge power influenced not only the evaporation/sputtering rate of the solution (and the analyte transport) but also the atomization/excitation efficiency in APGD, which directly influenced the analytical signal of the elements. The enhancement of the analytical signals, caused by the increased discharge current, was also observed in other APGD systems but sustained in contact with flowing solutions.1,6
The increase of the HNO3 concentration, used for the acidification of the solutions, also enhanced the analytical signals of the studied elements, but this effect was not that strong as in the case of the discharge current (Fig. 2b). The intensity of the analytical lines of Ca and Tl was almost unchanged by the change of the solution acidity. Considering the lowest intensity of the analytical line of the studied elements acquired in the series acquired for the solutions acidified with HNO3 at concentrations changed in the range of 0.1 to 0.6 mol L−1, the most significant, almost 3-fold enhancement was observed for Cd and Hg. The change of the sample consumption was also not that significant as in the case of the discharge current; it increased from 0.70 mL min−1 (for 0.1 mol L−1 HNO3) to 1.10 mL min−1 (for 0.6 mol L−1 HNO3). For other APGD systems, an increase in the sample acidity usually led to the enhancement of the analytical signals of the studied elements and the deterioration of the discharge stability.1,6 However, it should be pointed out that in this case, when the discharge was operated in contact with flowing solutions, the concentration of HNO3 or HCl in solutions was not higher than 0.1 mol L−1. In the present system, where HDC-APGD was used and operated in contact with a solution drop, its impaired stability was observed when the HNO3 concentration exceeded 0.6 mol L−1, which indicates an increased tolerance to this acid.
The effect of the outer diameter of the graphite tube was not investigated so far for any APGD system. We hypothesized that the increase in the tube size would stabilize HDC-APGD due to the enlargement of the solution drop, causing the enhancement of the analytical performance of this system. However, as can be seen in Fig. 2c, the effect of the tube diameter was opposite to the expected one. For the system operated with large-sized tubes, a longer stabilization time was needed, the stability of HDC-APGD was decreased, and the intensity of both the atomic lines and the molecular bands was reduced.
In summary, the optimal operating conditions of the developed discharge system are as follows: the discharge current – 65 mA, the concentration of HNO3 – 0.6 mol L−1, and the graphite tube diameter – 6 mm. Under these conditions, the net intensities of the analyte emission lines were measured and compared to the values predicted by the appropriate models (see Table 2). In the case of Cd and Tl, there was an excellent agreement between the predicted and the experimentally measured values. In the case of Ag, Ca, Hg, Mg, Mn, and Pb, the measured intensities of their analytical lines were not strictly in line with the predicted values; the differences were found to be changed from −21% (Pb) to +21% (Ca). Therefore, the conditions proposed by the models were considered as fairly optimal and chosen in all further experiments.
Element | Intensity in optimal conditions | Difference (%) | |
---|---|---|---|
Predicted | Measured | ||
Ag | 1.24 × 105 | 1.00 × 105 | −19.4 |
Ca | 1.60 ×104 | 1.93 ×104 | +20.6 |
Cd | 5.62 × 104 | 5.70 × 104 | +1.4 |
Hg | 7.61 × 103 | 6.49 × 103 | −14.7 |
Mg | 6.00 × 105 | 6.68 × 105 | +11.3 |
Mn | 4.48 × 104 | 3.73 × 104 | −16.7 |
Pb | 1.66 × 104 | 1.31 × 104 | −21.1 |
Tl | 5.42 × 104 | 5.72 × 104 | +5.5 |
In this study, a series of solutions containing methanol, ethanol, formic acid, acetic acid or formaldehyde in the concentration range of 0–10% (m/m) were prepared, and the net intensities of the analytical lines of Ag, Ca, Cd, Hg, Mg, Mn, Pb and Tl were measured. The results of this experiment, expressed as enhancement factors of the acquired analytical signals of the studied elements in the solution with the added LMWOCs in reference to the analytical signals obtained for the solution without any organic additive, are given in Fig. 3. It was established that in most cases, the evaluated enhancement factors are in good agreement with those reported in the literature.1,5,9 The greatest improvement of the analyte responses was obtained in the cases of formic acid, i.e., ∼2.4–2.6-fold for Ag, Cd, Hg and Pb and formaldehyde, i.e., ∼1.2–1.4-fold for Mg, Mn and Tl. Due to the better stability of the HDC-APGD system and considering the high enhancement factors, formic acid (8%, m/m) was used in the analysis of the CRM.
Fig. 3 The influence of addition of the examined LMWOCs on the response of Ag, Ca, Cd, Hg, Mg, Mn, Pb and Tl. |
Usually, the entrance slit of a spectrometer has a vertical orientation; hence, only a small part of the emitted radiation from the negative glow zone can be acquired by this spectrometer. According to the results obtained in our previous work,12 it was hypothesized that the spectrum acquisition from the whole negative glow zone length would significantly enhance the detectability of the studied elements. For this reason the interelectrode gap discharge image was rotated using a quartz DP.16,17
To efficiently use the DP, it was necessary to identify the exact position of the negative glow zone. For this purpose, 2-D maps of the intensity distribution of the analytical lines of Ag I (338.3 nm), Ca I (422.7 nm), Cd I (228.8 nm), Hg I (253.7 nm), Mg I (285.2 nm), Mn I (279.8 nm), Pb I (368.3 nm), and Tl I (535.0 nm), as well as Hβ (486.13 nm), O I (777.4 nm) and O II (441.5 nm) and the band heads of the N2 (337.4 nm), NO (247.2 nm), and OH (306.5 nm) molecules were acquired and background corrected. To move precisely the HDC-APGD image against the entrance slit of the spectrometer, the microdischarge system was placed on an optical platform equipped with two micrometer screws, allowing to position it in the horizontal and vertical directions. The abovementioned 2-D maps were obtained by vertically measuring the discharge image and the radiation emitted by it with a step of 10 μm from the central axis to the outside of the discharge volume. This step length of 10 μm was chosen as the entrance slit width of the spectrometer and covered all discharge cross-sections. The measurements were carried out using the optimal conditions selected at the previous step: the discharge current – 65 mA, the concentration of HNO3 – 0.6 mol L−1, and the graphite tube diameter – 6 mm. The results of this experiment are presented in Fig. 4 and 5. In both figures, the x-axis represents the distance from the discharge center (0), while the y-axis represents the discharge height, where 0.0 mm is the anode tip, while the solution surface is at 2.5 mm. It can be seen that HDC-APGD does not create an ideal cone with the top at the end of the anode tip but the plasma is formed partially around the end of the sharped solid electrode.
Fig. 4 2-D emission profiles of the analytical lines of the elements in HDC-APGD. The position of the solid anode (pin electrode) is 0 mm. |
Only for few elements (Ag, Pb, and Tl), the majority of the emission from their atoms occurred in the vicinity of the liquid cathode (see Fig. 4). In the case of Ca, Mg, and Mn, the maximum of the emission of their atoms was moved toward the positive column zone. It should be noted that this phenomenon was the strongest in the case of Ca. A similar effect was observed for Cd, but in this case, the strong emission from the atoms of this element started in the negative glow zone and reached its maximum in the vicinity of the anode. In general, there was a very weak emission from the Hg atoms observed in the vicinity of the liquid cathode and in other zones of the discharge. For this element, the maximal emission from its atoms occurred (similar to the case of Cd) near the anode but was not that significantly extended to the other discharge zones, and rather concentrated in the anode glow zone. Except for O II and H, all other background species demonstrated their maximum emission in the vicinity of the anode (see Fig. 5). There was almost no emission from the N2 molecules near the cathode zone, possibly due to high evaporation and expansion rates of the water vapor, which did not allow to diffuse N2 from the air into that discharge region (Fig. 5). Similarly, as in the case of N2, the emission from O I was observed mainly in the vicinity of the anode (see Fig. 5). The emission from the NO radicals, whose presence in the APGD spectra is caused by a direct reaction between N and O from air, was also mostly observed at the end of the anode, similar to the case of N2 and O I.
To the best of our knowledge, there are no similar 2-D emission maps of different plasma (atomic and molecular) compounds available in the literature for APGD systems sustained in contact with liquids. There were few studies where the vertical or horizontal emission profiles of some plasma compounds were collected.6,12,18–20 Therefore, the results obtained in the present work are difficult to compare with them, although there are some similarities, particularly in the cases of Cd, Mg and Pb,6 Ag, Cd and Zn,12 or Pb.20
A series of single-element standard solutions of Ag, Pb and Tl, acidified with 0.6 mol L−1 HNO3, were prepared, covering the concentration range from 0.010 mg L−1 to 100 mg L−1. As can be seen from Table 3, a good linearity (R2 > 0.998) was obtained up to 20, 60 and 10 mg L−1 for Ag, Pb and Tl, respectively. Based on the measurements of the respective blank solution (0.6 mol L−1 HNO3), the LODs of Ag, Pb and Tl were determined for the HDC-APGD-DP-OES system (3σ criterion, where σ was the standard deviation of the background level for a repeatedly measured n = 10), and they were 0.54 μg L−1 (Ag), 14 μg L−1 (Pb) and 2.2 μg L−1 (Tl). It is worth noting that the LODs achieved for the optical system with the DP were much better than those offered by the HDC-APGD system without the DP (see Table 4). The average precision of measurements of the intensities of the analytical lines of Ag, Pb and Tl, expressed as relative standard deviations (RSD), were 2.0%, 1.2%, and 1.1%. It should be noted that these LODs as well as RSDs were comparable to or better than those reported for other APGD systems (Table 4).
Element | λ (nm) | Slope/103 (a.u.) | LOD (μg L−1) | URL (mg L−1) | R 2 |
---|---|---|---|---|---|
Ag | 338.3 | 156 | 0.54 | 20 | 0.9989 |
Pb | 368.3 | 4.21 | 14 | 60 | 0.9998 |
Tl | 535.0 | 22.7 | 2.2 | 10 | 0.9998 |
Method | LOD (μg L−1) | Reference | ||
---|---|---|---|---|
Ag | Pb | Tl | ||
a With the Dove prism. b Without the Dove prism. c Chemically enhanced. d SCGD – solution cathode glow discharge. e HDE – hanging drop electrode. f ac-EALD – alternating current electrolyte atmospheric liquid discharge. | ||||
HDC-APGDa | 0.54 | 14 | 2.2 | This work |
HDC-APGDb | 1.9 | 135 | 14 | This work |
SCGDc,d | 0.1 | 1 | — | 9 |
HDE-APGDe | 2.1 | — | — | 12 |
FLC-APGD | 2 | 180 | 8 | 21 |
ac-EALDf | 1 | 45 | — | 22 |
SCGDd | 1 | 10 | — | 23 |
To evaluate the applicability of the developed HDC-APGD-DP-OES method in the analysis of real samples, the CRM of tea leaves (INCT-TL-1) was chosen due to the presence of certified concentrations of Pb and Tl. Since it was found that there were relatively high interferences coming from K and Na at 100 mg L−1 and higher concentrations in the case of the studied elements, the method of three standard additions (at 100, 200 and 300 μg L−1 of Ag, Pb and Tl) was applied for the calibration. Notably, the presence of Ca and Mg in the range of 10–200 mg L−1 did not result in such interferences.
The results of this analysis for Pb and Tl are given in Table 5, and in the case of Ag present in the analyzed CRM at a very low level (below 0.011 mg kg−1), the appropriate recovery is given. In the case of Pb and Tl, the developed method revealed the very good trueness, since the recoveries of the certified values were in the range 96–103%. The recovery of Ag was also quantitative (102%). All these measurements confirmed the applicability of the method in the analysis of real samples.
Element | Found value (mg kg−1) | Certified value (mg kg−1) | Recovery (%) | |t-value|a |
---|---|---|---|---|
a A t-test for comparison of found and certified values was applied (n = 3, α = 0.05, and tcritical = 4.303). b The CRM was enriched with Ag. | ||||
Ag | <0.011 | — | — | — |
1.02 ± 0.08 | 1.00b | 102 ± 8.0 | — | |
Pb | 1.71 ± 0.16 | 1.78 ± 0.24 | 96.1 ± 8.9 | 0.76 |
Tl | 0.065 ± 0.005 | 0.063 ± 0.005 | 103.2 ± 7.9 | 0.69 |