Sa Zhangab,
Yan Li*a,
Liuqing Huangab,
Liuying Huangab and
Xuetao Luoab
aCollege of Materials, Department of Materials Science and Engineering, Xiamen University, Xiamen, Fujian Province 361005, China. E-mail: liyan@stu.xmu.edu.cn
bFujian Key Laboratory of Advanced Materials, Xiamen University, Xiamen, Fujian Province 361005, China
First published on 19th July 2022
Retraction of ‘In situ Raman investigation of the phase transition of NaVO2F2 under variable temperature conditions’ by Sa Zhang et al., RSC Adv., 2021, 11, 23550–23556, https://doi.org/10.1039/D1RA02827H.
A Raman spectrometer was used for the testing outlined in this research, utilising coupled variable temperatures to investigate the phase transition of NaVO2F2. The temperature change was controlled by a THMS600 high and low-temperature sample stand. After the publication of this research, the unstable temperature was found to be caused by the system of the THMS600 sample stage, which lead to the temperature error in the analysis of phase transformation during the heating process. This has led to errors in the analysis, and therefore the conclusions of this paper are not supported. The authors also regret that the funding information was incorrectly shown in the original article.
Signed: Sa Zhang, Yan Li, Liuqing Huang, Liuying Huang and Xuetao Luo, 23rd June 2022.
Retraction endorsed by Laura Fisher, Executive Editor, RSC Advances.
This journal is © The Royal Society of Chemistry 2022 |