Beta Z.
Poliquit
,
Paul L.
Burn
* and
Paul E.
Shaw
*
Centre for Organic Photonics & Electronics, School of Chemistry and Molecular Biosciences, The University of Queensland, Brisbane, Queensland 4072, Australia. E-mail: p.burn2@uq.edu.au; p.shaw3@uq.edu.au
First published on 22nd June 2022
Pre-concentration is an important tool to improve the detection of low vapour pressure analytes. Quantification of the amount of analyte sorbed and released and how it relates to the pre-concentrator material and analyte of interest is important for analyte detection and pre-concentrator material development. In particular, establishing the diffusion mechanism enables correlations between the type of sorbing phase and the sorption mode to be identified, which may then provide strategies to improve pre-concentrator materials. We describe the development of a UV-vis absorption spectroscopy method to monitor the real time uptake of analyte vapours by pre-concentrator materials and apply it to characterising the uptake of the nitroaromatic 4-nitrotoluene (pNT) by three pre-concentrator materials under continuous flow conditions: poly(dimethylsiloxane) (PDMS), PDMS:1,4-divinylbenzene (DVB), and PDMS-co-DVB. The results show that PDMS and PDMS:DVB films up to ∼100 μm thick initially exhibit pNT uptake consistent with Case II diffusion as opposed to a Fickian process. In contrast, the PDMS-co-DVB films exhibited more complex behaviour, with mass uptake kinetics consistent with anomalous diffusion that varies with film thickness, which we attribute to the presence of different phases in the films. The pNT uptake for the thicker DVB-containing films under continuous flow was lower than that of PDMS although the rate of diffusion was faster, which was attributed to the presence of large pores in these materials. The results demonstrate that the mechanism of diffusion of analyte vapours into PDMS-based pre-concentrators is composition dependent and impacts the balance between the rate of analyte uptake and the total analyte uptake during continuous flow measurements.
Pre-concentrators generally have poor selectivity and sorb a wide range of analytes with similar properties (for example, polarity), relying on the analytic instrument to provide differentiation of the chemicals in the sample mixture.3 Furthermore, the quantity of analyte collected is typically inferred by measuring how much is desorbed from the pre-concentrator using the analytical instrument, which requires numerous time-based sequential sampling-measurement iterations.4,5 This approach for determining how much analyte is sorbed is not ideal as only the analyte released from the SPME is detected, which may be less than the amount sorbed. Indeed, the rate and amount of analyte released can have a strong temperature dependence.4 It is therefore of interest to be able to study the analyte sorption process over time both in terms of the rate of up-take and quantitatively assessing the amount of analyte sorbed. It has been reported that the sorption of analytes into SPMEs is governed by Fick's law of diffusion, whereby diffusion is controlled by a concentration gradient.5,6 However, polymeric materials can swell upon absorption of the analyte, which is a characteristic of Case II and Super Case II diffusion.7,8 Case II and Super Case II diffusion differ in that while the analyte diffuses into the film as a distinct front, in the latter case the front accelerates during the diffusion process. These subtleties in diffusion cannot be observed by the standard sorption-release measurements. In this regard, establishing the diffusion mechanism enables correlations between the type of sorbing phase and the sorption mode to be identified, which may then provide strategies to improve SPMEs, including the choice of materials and their selectivity.
In this study, we describe the development of a novel approach, based on in situ UV-vis spectroscopic measurements, to continuously monitor the analyte sorption with respect to time in PDMS-based pre-concentrator films under a constant supply of analyte vapour. The advantage of using absorbance measurements is that it provides an unambiguous quantitative measurement of the analyte present in the film. Analyte vapour uptake measurements were performed on three different types of pre-concentrator materials – poly(dimethylsiloxane) (PDMS), PDMS:1,4-divinylbenzene (DVB) blends, and PDMS-co-DVB (copolymerised material). We demonstrate the method using p-nitrotoluene (pNT) as an exemplar for explosives detection. The observed time-dependence of the uptake is consistent with Case II diffusion, i.e., the swelling driven process that results in a defined front propagating through the film at constant velocity. From the data, we calculated the front velocity for each PDMS-based material. The analyte diffusion mechanism as well as the calculated values of the front velocity confirm the spontaneous nature of the analyte uptake process. They also provide concrete evidence of the relationship between the physical and chemical attributes of the different film compositions, such as, the presence of pores, and the crystallinity and polarity of the polymer, to the analyte uptake. This new approach clarifies how nitroaromatic vapours diffuse into PDMS-based pre-concentrator materials and provide a quantitative approach for evaluating future pre-concentrator materials.
![]() | ||
Fig. 1 Schematic diagram of (a) the custom-built sample chamber and (b) the pNT vapour delivery system based on 2 mass flow controllers (MFC). |
Films of all three PDMS-based systems showed similar behaviour with an immediate increase in their optical densities when the flow of pNT into the chamber was initiated followed by a decrease of that rate with equilibrium reached at later times. Control measurements without a pre-concentrator film showed a much smaller change in absorbance, on the order of 0.06 after 60 s (ESI† Fig. S3), which is attributed to pNT in the atmosphere along the optical path. This confirms that sorption of the analyte into the pre-concentrator film is the primary source of the measured absorbance change. While all three systems present the same initial sorption trend with a rapid and steady uptake of analyte, they differ in terms of the time taken to achieve equilibrium and the equilibrium concentration. For example, for the PDMS films (Fig. 2a), the time taken to reach equilibrium and the equilibrium concentration both increase with film thickness. This is consistent with the analyte being sorbed into the bulk of the film rather than adsorbed onto a surface. In the latter case analyte uptake would be near-identical as the samples have the same surface area. In contrast, the films of PDMS:
DVB (Fig. 2b) equilibrate at similar absorbance values (∼1.5) independent of film thickness with the thicker films equilibrating faster. This behaviour suggests that the pNT uptake might be limited to the surface of the PDMS
:
DVB films rather than in the bulk. Finally, for the films made of PDMS-co-DVB (Fig. 2c) the behaviour was intermediary, initially showing an increase in the amount of analyte sorbed at equilibrium with increased film thickness and then a decrease for the thickest film.
The sorption kinetics can therefore be divided into two key stages noting that as the chamber is supplied with a continuous supply of analyte vapour, the concentration of analyte in the atmosphere is constant and therefore any changes to the rate of uptake can be attributed to the sample: the first stage (stage 1) is the initial sorption phase where the film is not saturated with analyte and the analyte uptake is limited by either the rate of deposition onto the surface of the film or its diffusion into the film. Where the analyte diffuses into the bulk of the film, this process will continue until the diffusion front has reached the substrate and can no longer proceed. In this work, this process is characterised by a linear increase of the optical density over time and therefore a constant rate of mass uptake. The second key stage is the attainment of an equilibrium, which appears as a plateau in the kinetics data. That is, the rate of analyte sorption equals the rate of desorption and the amount of pNT within the film remains constant under the conditions of the measurement. There is often not a sharp transition between these two states, which can arise from, for example, the reorganisation of the polymer chains to enable the analyte to diffuse into all available sites.
The data from stage 1 in the sorption kinetics can be used to extract details of the mechanism by which the analyte diffuses into the film. Specifically, each diffusion mechanism will exhibit a distinct time-dependence of the mass uptake and can be identified from8,12
mt = ktn, | (1) |
At = ktn, | (2) |
Fig. 3 shows the change in the optical density of the PDMS-based pre-concentrator films over time plotted on a log–log scale. The curves show similar trends implying that the diffusion mechanism for pNT was similar regardless of the film composition or thickness, with the exception of thicker PDMS-co-DVB films. The value of n for each pre-concentrator film was taken from fitting to the data with eqn (2), where each curve was fitted and then the results averaged (see ESI† Table S1). For the PDMS and PDMS:
DVB blend films the average value of n was 0.99 ± 0.01 and 0.99 ± 0.08, respectively. In other words, the initial mass uptake increased linearly with time, which is consistent with Case II diffusion. That is, a front of the analyte diffused into the films at a constant rate despite the PDMS
:
DVB films having voids in the surface. The value of n for the diffusion of pNT into PDMS-co-DVB was 0.92 ± 0.07, which indicates that the diffusion has “anomalous” characteristics. Anomalous diffusion of analyte is often attributed to non-uniform film morphology. For example, a rigid material with strained swelling is likely to experience non-linear analyte diffusion compared to a highly flexible material.15–17 We have previously reported that the PDMS-co-DVB films contain two phases in addition to surface voids. The spheroidal phase was assigned to poly(1,4-divinylbenzene) rich regions of the film, which would be expected to be more rigid than the PDMS containing components.9 Thus, the non-uniform film morphology is consistent with the films having anomalous diffusion characteristics.9 Furthermore, the fact that n for the PDMS
:
DVB film was around 1 suggests that the components within the blend were more uniformly distributed.
![]() | ||
Fig. 3 The change in the optical density over time in log–log of (a) PDMS, (b) PDMS![]() ![]() |
The differences in the morphologies would be expected to impact the velocity of the diffusion front, which is a useful parameter for assessing the capabilities of the sorbing material and making comparisons. Furthermore, it also provides guidance on how long sampling needs to be undertaken before an appropriate concentration of analyte is reached for a specific analytical instrument. One simple approach to estimate the velocity of the front is based on the inflexion point of the optical density (analyte sorption) versus time plot, which for Case II and Super Case II diffusion will approximately correspond to the travel time required for the analyte front to reach the substrate. Hence, the time-averaged velocity of the diffusion front can be estimated by dividing the film thickness by the travel time. Fig. 4 shows the velocity of the diffusion front for films of each material for different thicknesses. For thin films (<20 μm), the velocity of the front is comparable across all the PDMS-based materials. However, as the film thickness was increased (>20 μm), the front velocity in the PDMS:DVB and PDMS-co-DVB films exceeded that of the PDMS films. This observation is consistent with the voids on the surface of the PDMS:
DVB and PDMS-co-DVB films. That is, the voids provide pathways for the analyte to diffuse into the film structures, thus increasing the surface area for the analyte to sorb into the pre-concentrator material. It might be thought that given none of the films sorbed the analyte through a Super Case II mechanism that the front velocity would be constant independent of the film thickness. However, this was not the case with the thicker pre-concentrator films having higher front velocities. It is not entirely clear what the origin of this effect is, but it may arise from the thicker films requiring several dip-coating steps that could give rise to subtly different film morphologies. The different morphologies could arise from differences in solvent evaporation from or drying of the film with thickness, or the ageing of the sol–gel solution during the fabrication process. To probe the potential effect of ageing of the sol solution on the film morphology and velocity of the diffusional front, a series of PDMS films were made in succession from 2 dip-coating cycles using the same sol solution. The results are shown in ESI† Fig. S4 and indicate that film thickness rather than the order of film fabrication had the greatest influence on the front velocity. While the link between dipping cycles and morphology is unclear, the dip-coating cycles promote strong film surface features (e.g., surface bumps) which clearly contribute to the overall film morphology. This point is further supported by the increased light scattering with film thickness, that is, increased optical density in the transparent region of thick films compared to thin films.
In the final part of the study, we quantitatively compared the amount of sorbed analyte at equilibrium by pre-concentrator films of each material of similar thickness (Fig. 5). Note the concentration of the pNT was 19 ± 5 ppm for these measurements. For films with a thickness of 13 μm, the PDMS:
DVB films were found to sorb the greatest amount of analyte, followed by the pristine PDMS film, with the PDMS-co-DVB pre-concentrator film the least. These results are opposite to what was reported from static sorption measurements where the PDMS-co-DVB pre-concentrator films had higher analyte uptake.9 However, it is important to note at this stage that independent of the measurement type – static or dynamic – pre-concentrator films that contained a DVB or a DVB-derived component always performed better than PDMS alone. The reason why the films with DVB or a DVB-derived component perform better than PDMS arises from the fact that the π–π interactions between the analyte and pre-concentrator film plays a role in the sorption process. In the static measurements the films were placed in the headspace vapour of the analyte for defined time periods whereas in the dynamic measurements there is a continuous air flow over the films. In both cases there are both sorption and desorption processes that can occur. Given that the first step in the sorption into the film is the adsorption of the analyte onto the film surface then there might be expected to be a difference in the desorption rate from the films under the two conditions. Intuitively, one would expect that the rate of surface desorption would be higher for dynamic conditions due to the nitrogen flow. Indeed, it has been previously shown that the speed of airflow across a PDMS/DVB SPME fiber strongly affected the sorption of toluene.18 In that study, the extraction profiles for toluene were investigated using a 65 μm PDMS/DVB coating and airflow speeds ranging from 0.8 to 83.2 cm s−1 for sampling times ranging from 5 s to 1 h. It was found that the maximum adsorbed toluene was lower and occurred faster for higher airflow, which was attributed to a thinner pre-concentrator/analyte boundary layer and a more competitive adsorption process at higher flow speeds. Thus, these reported results would suggest that for the PDMS-co-DVB films the analyte is less able to sorb efficiently thoughout the film, which is consistent with the anomalous diffusion process. The results also suggest that for the PDMS-co-DVB pre-concentrator films that under dynamic conditions the analyte is preferentially sorbed in the PDMS rich regions of the film with less diffusing into the more rigid DVB-derived spheroids. In contrast, the enhanced up-take of the PDMS
:
DVB blend over that of the PDMS pre-concentrator film arises primarily from the π–π interactions between the analyte and DVB in the pre-concentrator film.19
We find that the relative amount of pNT sorbed by each of the PDMS-based pre-concentrator materials under a continuous flow is different from what was previously observed under static conditions. Specifically, we find that the amount of pNT vapour sorbed by PDMS was generally greater than that sorbed by PDMS-co-DVB and PDMS:
DVB for the thicker films. For the thinner films the order was changed with the PDMS
:
DVB blend having the largest uptake at equilibrium and the PDMS-co-DVB having the least. The results show that under dynamic conditions there is a subtle trade-off between film surface area and intermolecular interactions of the analyte and the pre-concentrator material. Nevertheless, independent of the film thickness the speed of the diffusion front was greater for the films that had greater porosity, that is, the PDMS
:
DVB blend and PDMS-co-DVB.
The results confirm that the diffusion of analyte vapours into PDMS-based concentrators cannot be assumed to be a concentration-driven process, and that the composition of the pre-concentrator can affect the diffusion process and the balance between the rate of analyte uptake and the total analyte uptake under dynamic conditions. The development of a method for characterising analyte uptake over time therefore provides a new perspective with which to evaluate pre-concentrator materials and optimise their performance.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d2sd00045h |