To detect drug concentration in tacrolimus solution, an anchor planar millifluidic microwave (APMM) biosensor is proposed. The millifluidic system integrated with the sensor enables accurate and efficient detection while eliminating interference caused by the fluidity of the tacrolimus sample. Different concentrations (10–500 ng mL−1) of the tacrolimus analyte were introduced into the millifluidic channel, where it completely interacts with the radio frequency patch electromagnetic field, thereby effectively and sensitively modifying the resonant frequency and amplitude of the transmission coefficient. Experimental results indicate that the sensor has an extremely low limit of detection (LoD) of 0.12 pg mL−1 and a frequency detection resolution (FDR) of 1.59 (MHz (ng mL−1)). The greater the FDR and the lower the LoD, the more the feasibility of a label-free biosensing method. Regression analysis revealed a strong linear correlation (R2 = 0.992) between the concentration of tacrolimus and the frequency difference of the two resonant peaks of APMM. In addition, the difference in the reflection coefficient between the two formants was measured and calculated, and a strong linear correlation (R2 = 0.998) was found between the difference and tacrolimus concentration. Five measurements were performed on each individual sample of tacrolimus to validate the biosensor's high repeatability. Consequently, the proposed biosensor is a potential candidate for the early detection of tacrolimus drug concentration levels in organ transplant recipients. This study presents a simple method for constructing microwave biosensors with high sensitivity and rapid response.
During the measurement process, microfluidic technology and RF biosensor are combined to achieve quantitative and stable testing to solve the problem of sample wastage and biological contamination. Microfluidic technology is a rapidly developing technology, which has the characteristics of miniaturization and integration.18,19 It can be used to detect very small amounts of liquid samples, such as 10−9–10−18 L.20 Microfluidics technology is characterized by its small volume, rapid analysis speed, automatic completion of the entire process of sample analysis, and increasing integration scale, allowing for the development of high-cost, compact, and one-time detection instruments.21–23 Currently, many effective detection technologies have been validated in microfluidic devices, including electrochemical methods,24 fluorescence detection methods,25 and RF sensing methods.26 For example, Xu et al. detected tumor-derived exosomes using a magnetic microfluidic chip;27 Peng et al. used online fluorescence derivatization flow injection in a microfluidic device for the detection of Cr(III) and Cr(VI) in water samples after solid-phase extraction;28 Liu et al. used a microfluidic RF biosensor to monitor cell growth.29 However, they have inherent limitations in terms of detection, including low sensitivity, high destructiveness, and high detection limits.
Organ transplantation is the most effective way to treat various end-stage organ diseases. The rejection of a transplanted organ is the main cause of failure of transplantation, and immunosuppressive therapy is the key to the success of organ transplantation. Tacrolimus (FK506) is one of the most commonly used calcineurin inhibitors, mainly used to combat organ transplantation rejection.30 After oral absorption, tacrolimus mainly binds to tacrolimus-binding protein in a prototype, acts on T cells, and plays an immunosuppressive role. However, the binding mode of tacrolimus to the tacrolimus-binding protein in red blood cells is non-linear; therefore, the whole blood concentration of tacrolimus should be measured in clinical treatment. At present, drug concentration monitoring methods mainly include chromatographic analysis31 and immunoassay. Immunoassays to monitor tacrolimus concentration can lead to unpredictable overestimation of true tacrolimus concentration due to cross-reactivity of metabolites accounting for up to 30%, and overestimation of true tacrolimus concentration is bound to affect clinical efficacy.32 In addition, the operation procedure of this method is complex and each link has time requirements, which leads to a long time limit for the final concentration measurement. Thus, there is an urgent need for a technique and equipment that can detect tacrolimus concentration quickly, simply, cheaply, and stably to improve the current detection instruments.
This paper proposes an anchor planar millifluidic microwave (APMM) biosensor for the accurate and quantitative detection of tacrolimus samples. The overall size of the APMM structure is 16.87 mm × 15.94 mm, and a sample with a volume of 1 μL is dropped into the millifluidic channel for detection to minimize the consumption of the sample. Since electromagnetic waves can easily penetrate the material, changes in the resonance frequency and transmission coefficient are effectively characterized by tacrolimus concentration when the APMM biosensor detects tacrolimus samples. The integration of a microwave biosensor and a millifluidic channel has the advantages of a very LoD and high sensitivity, as demonstrated by experimental data. This integration enables the detection of biological samples quantitatively and has the benefit of a very low LoD. Moreover, based on the regression analysis of the dual-frequency characteristics of the APMM biosensor, the results demonstrate a very strong linear correlation between the tacrolimus concentration and the two resonance peaks. The good repeatability and reusability of the proposed biosensor are validated by performing five measurements on each sample. Therefore, the proposed RF biosensor is a new method for the detection of tacrolimus concentration, with more economical and simple advantages, and would further promote the development of organ transplant operation so that more organ transplant patients can benefit.
The resonator is modeled as a parallel RLC circuit, as shown in Fig. 1(f). Its capacitance is affected by the dielectric constant of the material under test (MUT) and the substrate itself, resulting in changes in the resonator's resonant frequency. The formula is as follows.
(1) |
Equivalent admittance of capacitance is expressed as follows.33
(2) |
CAPMM = CUc(N − 1)L | (3) |
Unit capacitance is expressed as
(4) |
(5) |
(6) |
(7) |
(8) |
When the electric field is applied to the port of the microstrip line, a closed magnetic field will be generated around the microstrip line, and an electric field will be generated between the microstrip line and the substrate. In fact, electromagnetic energy exists as a time-varying electromagnetic field perpendicular to the surface of the flat open-loop resonator (Fig. 1(c)). Therefore, the solution sample dripping along the microstrip line will have an effect on the time-varying electromagnetic field surrounding the resonator, and this change will be displayed by the network analyzer.
The reflection coefficient S21 of the APMM sensor can be observed in Fig. 2(c) and (d). Whether a PDMS millifluidic channel is present or not, the results of the ADS simulation are in good agreement with the measurements. After loading the PDMS millifluidic channel on the APMM sensor, although its resonant frequency and notch amplitude were reduced, this result had no effect on the MUT detection. For the transmission coefficients of various resonator harmonics, formula (11) can be used to express feven and fodd as37
(9) |
(10) |
(11) |
Considering the relative permittivity of air, if the relative permittivity of the sample is constant at the operating frequency, the nth harmonic frequency shift will be n times the fundamental frequency. To accomplish the measurement, the first and second harmonics are selected during data acquisition, and the effect of the change in dielectric constant on the harmonic frequency is considered. Fig. 2(c) and (d) depict the simulated and measured results of S21 for biosensors without and with millifluidic channels, respectively.
The effects of the length (d) and width (s) of the anchor plane microwave sensor gap on the frequency and S21 were investigated and are shown in Fig. 3(a) and (b). It has been found that these parameter changes have little effect on the odd mode but significantly alter the frequency shift of the even mode and the response of S21. To validate the sensing capability of the proposed APMM sensor, the MUT was placed within the measurement area and simulated using ADS software. When the dielectric constant is 1, the response of the resonator to the change of loss factor (tanδ) is studied, as shown in Fig. 3(c). At the sensor's resonant frequency, when the loss factor is increased from 0.05 to 0.25 with a step size of 0.05, the absolute amplitude of S21 decreases significantly while the resonant frequency remains nearly unchanged. In addition, the dielectric constant (ε) of the variable MUT was also studied, whose ε varies from 1 to 5 with a step of 1, and the loss factor is considered to be 0.0019. As the dielectric constant rises, the simulation results reveal that the sensor's resonant frequency shifts to a low frequency in an obvious range. In addition, as shown in Fig. 3(d), the absolute amplitude of S21 decreases as the dielectric constant increases.
Fig. 3 The simulated S21 of the APMM at different parameters of (a) s and (b) d. ADS simulation resonant profile of the sensor loaded MUT with varying (c) loss tangent and (d) permittivity. |
For this experiment, 0.5 mg Astellas (Ireland) Procofotacrolimus capsules and 10 mL phosphate-buffered saline (PBS, pH = 7.2–7.4) produced by DaTech (Quanzhou) Co., Ltd. were prepared. The two substances were then combined and dissolved in a beaker that had been ultrasonically cleaned. To completely dissolve the powder into PBS, the beaker was placed on a magnetic stirrer and stirred for eight hours at 460 rpm. Next, 50 μg mL−1 tacrolimus stock solution was diluted with PBS solution. Finally, seven groups of samples were prepared and packaged separately, with the following concentrations: 10, 50, 100, 200, 300, 400, and 500 ng mL−1. The molecular formula of FK506 is shown in Fig. 1(d).
The morphology of the biosensors along with or without FK506 samples were characterized by SEM images, as shown in Fig. 4. The SEM images of the bare APMM biosensor are shown in Fig. 4(a). The enlarged SEM imaging (Fig. 4(b)) showed clean copper particles on the surface of the electrode before the FK506 sample was deposited. The SEM image of the biosensor with tacrolimus solution is shown in Fig. 4(c). It is demonstrated that the boundaries of the copper particles become blurred due to the deposition of FK506 solution. Moreover, smaller particles of FK506 can be found in the enlarged SEM image (Fig. 4(d)). Compared with the bare copper electrode, the electrode surface with samples presented a fuzzy and dusky state.
(12) |
At present, the concentration of tacrolimus in clinical monitoring is different in different time periods after organ transplantation. Tacrolimus concentrations in early transplant patients approach homeostasis at 10–15 ng mL−1. In addition, for patients with post-transplantation or autoimmune diseases, the concentration requirements of tacrolimus in the maintenance period and stable period are relatively low; generally, 5–10 ng mL−1 can be basically maintained for a long time. Low detection of limit refers to the lowest amount of the tested substance in the sample that can be detected, which is used to indicate the lowest concentration detected by the determination method on the tested substance in the sample under the said conditions.
According to formula (13),40 the LoD of the proposed biosensor is 0.12 pg mL−1, demonstrating the sensor's extremely low detection limit. In addition, Table 1 provides a summary of the comparison between the proposed APMM biosensor and other published microwave sensors.
(13) |
Structure | Specimen | FDR (MHz (ng−1 ml)) | LoD (pg ml−1) | Millifluidic | Ref. |
---|---|---|---|---|---|
Patch biosensor | Glucose solution | 1.13 × 10−4 1.97 × 10−4 | 26.54 × 107 15.22 × 107 | No | 41 |
Radio-frequency (RF) biosensor | Streptavidin and complementary DNA (cDNA) | — | 1000 | No | 42 |
Microstrip patch sensor | NaCl | 4.8 × 10−12 | \ | No | 43 |
SRR | Prostate specific antigen (PSA) | \ | 100 | No | 44 |
SRR | Glucose solution | 8.89 × 10−7 | — | Yes | 45 |
APMM | Tacrolimus in aqueous solution | 1.59 | 0.12 | Yes | This work |
Fig. 5(a) illustrates the capacitance of the biosensor for samples with varying concentrations of tacrolimus. The inset picture demonstrates the capacitance peaks in the frequency range of 0.55–0.80 GHz. The capacitance of the formant is positively correlated with the concentration of the tacrolimus solution. When the tacrolimus concentration changes as 10, 50, 100, 200, 300, 400, and 500 ng mL−1, the net capacitance values at the cavity's center frequency were calculated to be 0.59, 0.37, 0.24, 0.19, 0.14, and 0.14 pF, respectively. As shown in Fig. 5(b), the inductance of the biosensor designed for various samples of tacrolimus exhibits formant peaks in the frequency range of approximately 1.1 to 1.5 GHz. The frequency and amplitude of the inductance peak are positively correlated with tacrolimus concentration. The inductance values at the center frequency of tacrolimus samples were 32.56, 46.04, 67.88, 75.79, 99.98, 105.66, and 115.57 nH. Therefore, the C and L at the resonant center frequency are proportional to the tacrolimus solution concentration. Fig. 5(c) and (d) depict the real and imaginary components of the sensor's admittance to different concentrations of tacrolimus samples. The inset pictures demonstrate the real and imaginary part of Y(1,1) peaks in the frequency range of 0.55–0.8 GHz. The admittance of the formant of the planar millifluidic microwave biosensor correlates positively with the concentration of tacrolimus solution.
The linear relationship between the resonant frequency parameters F and S at the first resonant peak of the device and the concentration of the tested sample is depicted in Fig. 6(a). Among them, the coefficient (R2) of frequency F and concentration C is 0.914, and the linear correlation between the concentration and frequency of the device is relatively general. The linear relationship between the S parameter and the concentration of the tested sample was observed. The coefficient of determination (R2) of S21 and concentration C was 0.995. The device demonstrated an excellent linear correlation between concentration and S21. Next, the linear relationship between the resonant frequency F and S parameters of the device at the second resonant peak and the sample concentration was analyzed using Fig. 6(b). The determination coefficient (R2) of frequency F and concentration C is 0.919. Nonetheless, the linearity R2 between S21 and sample concentration C reached 0.951, and the device demonstrated a linear correlation between concentration and S21. The above analysis demonstrates that the linear correlation between the two formant parameters of the device is not ideal. Therefore, this paper proposes to calculate the linear correlation between the frequency changes of the two formant peaks of the RF device. Similarly, it also calculates the linear correlation between the ΔS21 of the two formant peaks, as shown in Fig. 6(c) and (d). Firstly, as shown in Fig. 6(c), the linear relationship between the difference of resonant frequency ΔF of the two peak values of the device and the concentration of the tested sample is observed. The determination coefficient (R2) of ΔF and concentration C is 0.992, indicating a good linear correlation. At the same time, the linearity (R2) between ΔS parameters and the concentration of the sample under test is increased to 0.998, indicating a very good linear correlation. In summary, by measuring the correlation between S21 and the resonant frequency offset, the optimal frequency response of the resonator is obtained.
To test the difference and reliability of sample concentration, the P-value test is presented in this paper, and the P-values of the seven sample concentrations used in the experiment are calculated from the perspectives of frequency change and loss variable, respectively. The results of the calculations are depicted in Fig. 6(c) and (d). In general, a P-value greater than 0.05 is regarded as insignificant, whereas a P-value less than 0.01 is regarded as highly significant. The results of the experiment indicate that the P-values for both frequency and loss changes are less than 0.01, or even less than 0.001. The structure indicates that there is a substantial difference between the concentrations of the samples that were tested.
The real-time response and stability results of the biosensor are shown in Fig. 6, with the change in the tacrolimus solution concentration over time and the real-time peak of the transmission coefficient (S21). Fig. 6(e) depicts the biosensor's response stability to different concentrations (0–500 ng mL−1) of tacrolimus in blood samples. Fig. 6(f) illustrates the stable variation over time of the difference between the two resonance peaks (ΔS21) measured each time during the same test period, demonstrating a stronger linear correlation. In the meantime, the response time for the proposed RF biosensor is about 109 ms, as shown in Fig. S5(b).† The experimental results demonstrate that the sensor exhibits a highly stability and fast response trend over a specific measurement period and has the potential for real-time monitoring in the field.