Various types of passive and active micromixers have been successfully developed to address the problem of mixing in microfluidic devices. However, many applications do not need fluids to be mixed at all times, or indeed require mixing to be turned on and off at will. Achieving such on-demand mixing is not feasible for passive mixers, particularly when the flow rate cannot be used as a control parameter. On the other hand, active mixers are usually not designed to be able to turn mixing off completely, and they often have complicated fabrication processes and special operation requirements, limiting the range of applications. In this work, we demonstrate an on-demand micromixer based on the actuation of magnetic microwalls. These are made by replica micromoulding and can be easily integrated within commercial microfluidic devices, such as the ibidi® 3-in-1 μ-Slide. Using a simple magnet, the microwalls can be actuated between a fully upright ‘on’ state, which turns on mixing by creating a meandering path in the main channel, and a fully collapsed ‘off’ state, which completely turns off mixing by opening up the channel leaving it unobstructed. Besides the increase in path length when the microwalls are activated, inertia effects also play a significant role for mixing due to the tight bends in the meandering flow path. We quantify the mixing effect using coloured fluids of different viscosities and at different flow rates, and we show that the microwalls can effectively enhance mixing across a wide range of operational conditions.
Despite these developments, the adoption of many of the solutions into commercial products has been slow. One reason is that the fabrication and/or actuation of the active micromixers are usually complicated,5,6 and the processes involved may not be easily translated into mass production and/or established workflows.7,8 On the other hand, static micromixers such as the herringbone9 and inertial effect mixers10 are now commercially available. However, the mixing effect of those types of devices cannot be turned off. Thus, the possibility of having the mixing and/or inertial effects activated on-demand can broaden the range of applications for these devices (see more in section 4).
In this paper, we describe the fabrication and actuation of magnetic microwalls for fluid mixing in microfluidic channels. The microwalls are made by replica micromolding in a hot-embossing setup, using a mixture of magnetic particles and a thermoplastic elastomer (TPE), and with a glass mould fabricated beforehand in a femtosecond-laser-assisted-etching (FLAE) process. Using a simple magnet, the microwalls can be actuated between a fully upright state, which creates a meandering path in the existing channel, and a fully collapsed state, which opens up the channel unobstructed. We have shown that the microwalls can be easily integrated in existing microfluidic devices, such as the ibidi® 3in1 μ-Slide commercial chip. The mixing effect of the microwalls has been quantified using fluids of different viscosities and at different flow rates, and we have shown that the microwalls can effectively enhance mixing across a wide range of these operational conditions. We also found that, besides the increase in path length when the microwalls are activated, inertia effects also play a significant role for mixing due to the tight bends in the meandering flow path.
Fig. 1 Fabrication and actuation of the magnetic microwall mixer. (A) The fabrication process: i. femtosecond laser patterning; ii. etching in 45% potassium hydroxide solution; iii. rinsed and dried mould; iv. surface treatment with trichloro(1H,1H,2H,2H-perfluorooctyl)silane; v and vi. first hot-embossing with magnetic SIBS – poly(styrene-block-isobutylene-block-styrene) pre-mixed with magnetic particles; vii. second hot-embossing step with polystyrene; viii and ix. demoulded microwalls and bounded with a microfluidic channel. (B) The microwall mixer integrated in an ibidi® 3in1 μ-Slide. The assembled channel is 3 mm wide, 0.6 mm high and 25 mm long. (C) Exposed microwalls at their initial resting position without magnetic actuation, which is 45 degrees to the surface; pictured without the μ-Slide. The microwalls are 600 μm-high, 2.65 mm-wide and 30 μm-thick and fabricated with a 350 μm alternating shift in the width direction, with 450 μm distance between each microwall in the direction of the channel. (D) The experimental setup showing one coloured and one clear stream pumped into the side inlets, and a 15 mm cubic neodymium magnet connected to a servo motor that is placed 4 mm beneath the μ-Slide for actuation. The insert shows the meandering path created by the microwalls. (E and F) The microwalls in the ‘on’ and ‘off’ state, respectively; the inserts on the bottom left show the corresponding magnet orientation. (G and H) Snapshots from a movie showing top-view of 2 mL min−1 water flow when the microwalls are in the ‘on’ and ‘off’ state, respectively. The pictures are partially enhanced to show the microwalls. The movie showing the actuation of the microwalls can be found in the ESI.† All scale bars in this figure are 2 mm. |
When not actuated, the microwalls are oriented at an inclined angle determined by the mould (Fig. 1(C)). When a magnetic field is applied perpendicular to the substrate, a magnetic torque is generated and the microwalls move to an upright orientation, see Fig. 1(E), and they force fluids to follow a serpentine path through the channel as shown in the insert in Fig. 1(D), thereby inducing fluid mixing (Fig. 1(G), more details in section 3.1). The actuation mechanism is similar to the actuation of magnetic artificial cilia,12 which has been described in detail both experimentally13–15 and numerically.16 Here we used a 15 mm cubic permanent magnet with a surface magnetic flux density of 390 mT and placed 4 mm below the microfluidic chip to actuate the microwalls, as shown in Fig. 1(D). Such actuation can also be achieved using electromagnets (see ESI† II for a home-made electromagnetic setup for turning ‘on’ the microwalls). When the poles of the magnet are aligned horizontally, the microwalls collapse onto the bottom substrate, and the channel opens unobstructed, as shown in Fig. 1(F) and (H).
The degree of mixing was quantified based on the evaluation of the intensity of segregation (IoS).17 A detailed description of the implementation of this quantification method can be found in the ESI III.† In short, for every set of experiment (water or glycerol as base fluid), a calibration measurement in a channel without microwalls was first performed, and the image of a clearly separated double stream was used to determine the intensity values of the dyed and undyed streams. The IoS method we adopted in this paper maps the grey value of each point to a concentration value, and the mixing index (MI) was then calculated by:
Note that this method of quantifying mixing has its limitations. First, it uses 2D images, so the information is based on the combined intensity throughout the depth of view, and it cannot be used to examine flow structures in 3D and their effect on mixing. Secondly, the mapping from the grey value to the concentration is calibrated by 0%, 50% (fully mixed) and 100% fluids (mapping curves can be seen in the ESI† III), and there are unavoidably inaccuracies in the mapping. However, since the width of the channel (3 mm) is much larger than the height (0.6 mm), the 2D images can reasonably represent the quality of mixing in the double flow (side-by-side) configuration of our experiments. And since the difference in MI when the microwall mixer is turned ‘on’ and ‘off’ is quite large (see results section 3.1), the errors from the concentration estimation becomes less significant due to the high signal-to-error ratio.
The mixing effect is also reversible and reproducible. For 1 mL min−1 and 10 mL min−1 of water, Fig. 2(B) shows that the microwalls can be reliably turn on and off, and the mixing indices can be switched between two distinctive levels. The mixing effect is stronger at 10 mL min−1 than at 1 mL min−1, judged by the higher mixing index achieved. The increase in the mixing index at higher flow rate is rather counterintuitive because there is less time for the fluid to mix in the meandering channels. It is possible, however, that the increase in mixing index is a result of inertial effects, which will be discussed in more detail in section 3.2.
Fig. 2(C) shows the mixing results in 87 ± 1 wt% glycerol in water (with viscosity 220 ± 40 mPa s). It can be seen that the mixer can induce mixing when it is turned on at time 0, but the resulting mixing indices are lower than those with water at the same flow rates, by comparison with Fig. 2(A). The lower indices are the result of the higher Péclet number due to the higher viscosity, meaning that the two fluids will take longer to mix with the same flow condition. At relatively low flow rates, the increase in mixing index is temporary, as shown in the insert of Fig. 2(C) for 0.1 mL min−1, but also seen at higher flow rates to a lower degree (see ESI† movies). This temporary increase in mixing index is caused by the agitation of the fluids in the volume occupied by the mixer when it changed its configuration from off to on. In principle, one can also make use of this effect and continuously agitate the fluid to mix high viscosity fluids, for example by using a continuously rotating magnet.13,14 However, for some applications, a static on/off configuration might be desirable in order to simplify the mechanical design and reduce vibration. In the scope of this work, we aim at quantifying the mixing effect only in static configurations for a better comparison between different viscosities.
It can be also inferred from Fig. 2(C) that at relatively high flow rates of glycerol, e.g. 4 mL min−1, higher mixing indices are achieved when the mixer is turned on, and the effect does not diminish afterwards, unlike for lower flow rates. This effect is caused by the uncontrolled flow pattern as a result of the deformation of the microwalls, as shown in the enhanced bottom left part of Fig. 2(E). The deformation results from the high fluid pressures at high flow rates when such a highly viscous medium is used. For comparison, Fig. 2(D) shows that the microwalls are not deformed at the same flow rate (4 mL min−1) in water. The deformation of the microwalls creates openings between the ceiling of the channel and the top of the microwalls, where random streams of high and low concentration fluids can pass through without restrictions. The resulting flow pattern is composed of a large number of horizontal streamlines as shown on the right side of Fig. 2(E), most likely close to the ceiling of the channel. This is different from the streamline pattern that results from of the two fluids passing through a regular meandering channel, where all streamlines start from the opening of the last microwall, as shown in Fig. 2(D). The mixing indices at higher flow rates in glycerol are therefore less representative of the level of mixing, which is also a limitation of our chosen method for quantifying mixing, as discussed in section 2.2. However, for completeness, we keep these results in Fig. 2(C).
Note that having the microwalls in the ‘on’ state will increase the hydrodynamic resistance and hence the pressure drop across the channel. By making a simplifying assumption of a Poiseuille flow in the meandering channel, we can estimate a pressure drop of about 240 times larger for the ‘on’ state compared to the ‘off’ state. For example, for water flow at 1 mL min−1, the pressure drop in the ‘on’ state across the microwalls is about 850 Pa, compared to about 3.5 Pa when they are in the ‘off’ state. The detailed calculation can be found in the ESI.†
Fig. 3 shows simulation results of the flow patterns in water, obtained using COMSOL Multiphysics®. The detailed simulation setup can be found in ESI† IV. The microwalls are implemented as static divisions in the channel in the ‘on’ position. Because the coloured and clear streams in the experiments have the same viscosity, and we are only investigating the flow pattern, the fluid is considered homogenous in the simulation. Fig. 3(A–D) show the laminar flow pattern with water at 0.2 mL min−1 in the meandering channel as well as around the bends. In contrast, Fig. 3(E–H) shows that at 2 mL min−1, inertial effects become significant, which can be seen from the vortices and dean flow patterns in Fig. 3(F–H). Such effects are the key working mechanism for inertial micromixers, such as the spiral channel mixers.10,19,20 The effect can be characterized by a Dean number κ = δ0.5Re, with δ being the ratio of the channel hydrodynamic radius to the radius of curvature, and Re is the Reynolds number. With our channel geometry and the applied flow rate, δ0.5 is close to 1, and Re is approximately 10 for 0.2 mL min−1 and 100 for 2 mL min−1 water flows, resulting in a κ value of about 10 and 100 for 0.2 mL min−1 and 2 mL min−1, respectively. The relation between Re, κ and the observed enhancement of mixing is consistent with the literature, where similar relations were discovered using static channels with various types of geometries.21
The possibility to introduce inertial effects on-demand could also benefit other applications besides mixing,20 such as cell sorting22 and separation of particles.23 Specific dynamic structures tailored to those applications can be designed, and their manufacturability and actuation need to be taken into account early on. Certain aspects, such as mould geometry, filling, ease of demoulding and magnetic field strength and orientation can pose practical limitations to the configurations that can be realized, which might require further development of fabrication techniques as well as magnetic control methods.
We have experimentally quantified the effects of the microwalls on mixing in different fluids (water and 87 wt% glycerol) by pumping coloured and clear streams of the same base fluids side by side into the channel and calculating the mixing indices of the unmixed and mixed fluids from the videos taken during the experiments. The results clearly show a substantial enhancement of mixing when the microwalls are turned on, and the effect is both reversible and reproducible. The experiments also show that higher flow rates result in better mixing for water, due to inertial effects around the tight bends created by the microwalls. Results of numerical simulations performed using COMSOL® Multiphysics indeed show a significant change in flow patterns due to inertial effects when increasing the flow rate from 0.2 mL min−1 to 2 mL min−1.
The ability to induce mixing easily and reversibly in a microfluidic channel expand the opportunities for interesting applications. Multi-stream co-flowing channels, such as the μ-Slide used in this work, are often used in studies of chemotaxis, cancer metastasis and general cell migration.24,25 In such setups, the spatial–temporal gradient is controlled by tuning the flow rates of the fluids with different concentration of the substances in question, so the gradient cannot be changed independently from the flow rate. With our approach, one can change the gradient without changing the flow rate in the same device, simply by turning on and off the microwalls, with the possibility of creating a wide range of combinations of both static and dynamic conditions for the mentioned biochemical studies. Many biological analysis and processes are significantly influenced by the level and duration of mixing of different components, such as PCR, antibody tests and flow cytometry, and our approach can also be used to switch between configurations with different levels of mixing in the same setup.
The possibility to activate inertial effects on-demand with the microwalls in a channel is another interesting aspect of our approach. Inertial effects can be used for flow focusing, cell sorting, and cell enrichment.20,21 Being able to activate these applications on-demand can be useful, but it could be even more advantageous to be able to tune the inertial effects by changing the configuration of the microwalls, so the effect can be tuned to tailor-fit to the change of the application requirements and conditions, such as cell size and shape, flow rate and fluid viscosity. Due to the flexibility of the FLAE/molding process, other shapes and curvatures can also be made (see ESI†), which could bring a wider range of control and flexibility to the applications of inertial microfluidics. Of course, further detailed investigations of the microwall geometry and arrangement, as well as the resulting flow structures need to be carried out for specific applications.
We have shown that the microwall mixers can be made by replica moulding and can be easily integrated with commercially available chips, and they can be actuated simply by a magnet. This means that there is a relatively low threshold for mass production and adoption of such devices in various laboratory settings. We also hope that our findings can inspire new developments and new applications in areas yet unforeseen.