Karthikeyan K*a,
Senthil Kumar Kandasamyb,
Saravanan Pc and
Abdullah Alodhaybd
aDepartment of Electronics and Communication Engineering, M.Kumarasamy College of Engineering, Karur, Tamil Nadu, India. E-mail: karthimems@gmail.com
bDepartment of Electronics and Communication Engineering, Kongu Engineering College, Erode, Tamil Nadu, India. E-mail: senthilkumar6k@gmail.com
cDepartment of Self Development Skills, CFY Deanship, King Saud University, Riyadh, Saudi Arabia. E-mail: psaravanan.c@ksu.edu.sa
dDepartment of Physics and Astronomy, College of Science, King Saud University, Riyadh, Saudi Arabia. E-mail: aalodhayb@ksu.edu.sa
First published on 2nd February 2023
The objective of this study is the design, simulation, and performance optimization of a micromixer device using the three input parameters of device structure, flow rate and diffusion coefficient of gold nanoparticles while the output parameters are concentration, velocity, pressure and time domain analysis. Each input parameter in the microfluidic chip influences the system output. The data were gathered through extensive study in order to optimize the diffusion control. The fuzzy logic approach is used to optimize the performance of the device with respect to the input parameters. In this study, we have chosen three different flow rates of 1, 5, and 10 μL min−1, three different diffusion coefficient values of low, average and high diffusivity gold nanofluids (15.3 e−12, 15.3 e−11, 15.3 e−10 m2 s−1) which are used in three different shapes of micromixer device, Y-shaped straight channel micromixer, herringbone-shaped micromixer, and herringbone shape with obstacles micromixer, and we measured the output performance, such as mixing efficiency, pressure drop, concentration across the microchannel and time domain. The data were obtained by fuzzy logic analysis and it was found that the herringbone shape with obstacles micromixer shows 100% mixing efficiency within a short duration of 5000 μm, and complete mixing was achieved within 10 seconds with a low pressure drop of 128 Pa.
S. Camarri et al.20 studied the engulfment regime of a T-type and T-joint micromixer device and reported the mixing efficiency and pressure drop. They reported that the configuration of a CA1 or CA2 device shows better efficiency than an isolated T-micromixer. When comparing the pressure drop between two the different configurations of CA1 and CA2 devices, the configuration of the CA1 device shows a lower pressure drop compared to CA2. X. Zhan et al.21 designed a T-type micromixer with three different structural shapes: elliptical, rectangular and triangular shaped microchannels. Better mixing efficiency was found while using an elliptical cross-sectional microchannel. Z. Wu et al.22 reported the design and numerical simulation of a three-dimensional T-shaped passive micromixer with three different obstacles: square, triangular and cylindrical. They reported a mixing efficiency of 96% while using the triangular obstacles with Re = 100 and 18 kPa pressure drop. E. Tripathi et al.23 designed and reported a spiral shaped micromixer which was investigated for a wide range of Reynolds numbers between 0.1 and 100. Better mixing was observed within the range of Reynolds number from 0.1 to 50. E. Nady et al.24 studied the two inlets and two outlets of a Y-type passive micromixer with circular obstacles and tall walls. The total length of the device is 14 mm, the width of the channel is 200 μm, the thickness of the wall is 30 μm and the gaps between the walls are 170 μm and 70 μm. Better mixing was achieved in a short distance while using a circular channel with high number of tall wall structure and it act as an obstacles. Y. Liao et al.25 reported a passive micromixer device with staggered herring bone structure and split–recombination microchannel. The mixing efficiency was analysed using a wide range of flow rates, 1–12 μL min−1, and Reynolds numbers 3.3–40 and they achieved 98% mixing efficiency at 4.5–78 milliseconds. M. Ripoll et al.26 designed a Y-type ring shaped micromixer which was used to produce lipid nanoparticles through the mixing of lipids and biomolecules. They reported the mixing performance to be linked with the characteristics of the lipid nanoparticles. O. Ulkir et al.27 designed a T-shaped laminar diffusion-based micromixer with two inlets and two outlets. The mixing efficiency was studied using the diffusion coefficient 5 e−11 m2 s−1 and inlet flow rate of 15 e−15 m3 s−1. For the output value of the system, the velocity was 0.09 mm s−1, the pressure was 2 Pa and the concentration was 0.45 mol m−3. Karthikeyan et al.28 reported a Y-type herringbone shaped micromixer for mercury ion detection in water. They studied the pressure level and mixing efficiency of the device at different locations.
V. Vijayanandh et al.29 reported a T-type micromixer with different shapes of ridges, such as triangular, square and curved. The mixing efficiency of the device was optimised using different shapes of micromixers and they reported that the best mixing efficiency was achieved while using a micromixer with triangular ridges. Karthikeyan et al.16 studied the different shapes of micromixers such as a Y-type straight channel micromixer, and a serpentine shape micromixer with or without grooves. They reported the mixing efficiency and pressure drop. The best mixing efficiency was achieved with a short length while using a micromixer with grooves.
S. Hossain et al.30 reported a serpentine micromixer with the crossing of two layers. They studied a mixing efficiency of 96% at low Reynolds numbers from 0.2 to 10 and low pressure drop. X. Dong et al.31 designed a T-shaped micromixer for a non-Newtonian fluid. They studied a mixing efficiency of 93.84% at Re = 0.24 while using a non-Newtonian fluid and 93.90% mixing efficiency at Re = 8 while using a Newtonian fluid.
Karthikeyan et al.12 designed a Y-shaped micromixer with rectangular and triangular shaped obstacles to mix fluids with very low diffusivity. The mixing efficiency observed for the triangular shaped micromixer shows 100% mixing efficiency compared with other micromixers with rectangular shaped obstacles at a flow rate corresponding to the Reynolds number (Re) of 25. I. Ertugrul et al.32 reported a microfluidic device for platelet separation using the fuzzy logic technique.
M. Hejazian et al.33 reported a straight and serpentine shaped micromixer. The mixing efficiency of the device was studied using fluorescence intensity profiles with different flow rates of 20, 100 and 200 μL min−1. A. Usefian et al.34 designed a Y-shaped convergence and divergence-based micromixer for low flow rate applications. S. R. Bazaz et al.35 developed three different shapes of passive hybrid planar micromixer with repetitive obstacles, such as teardrop, nozzle, ellipse, pillar and tesla shaped obstructions inside the mixing zone. R. A. Taheri et al.36 reported a three-dimensional micromixer with split and recombine microchannel. They reported 96% mixing efficiency at a Reynolds number of 0.1, 90% mixing efficiency at a Reynolds number of 1 and 67% mixing efficiency at a Reynolds number of 10.
In this paper, we propose three dissimilar structures of micromixer: a Y-shaped straight channel micromixer (SCM), a herringbone serpentine channel micromixer (HSM), and a herringbone serpentine channel micromixer with obstacles (HSOM). The characteristic performance of the devices is discussed using the three input parameters of device structure, flow rate and diffusion coefficient, while the outputs are concentration, mixing efficiency, velocity, pressure and time domain analysis.
(1) |
∇·u = 0 | (2) |
The species transport in the systems can be described by the convection diffusion equation, as shown in eqn (3),
(3) |
Mathematically it can be represented as,
ΔP = Pinlet − Poutlet | (4) |
The mixing efficiency (M) of the micromixer can be calculated using the following formula,
(5) |
Fig. 6 Simulated result of Y-shaped herringbone serpentine channel micromixer with obstacles: (a) concentration field and (b) streamline distribution at 5000 μm. |
Fig. 4 shows the concentration level across the device of a Y-shaped straight channel micromixer. This device has two inlets in a Y-shape with a straight channel acting as a mixing zone followed by a sensing zone and outlet. Fig. 4 shows a simulation study of “Test case 1”, which is presented in Table 1. Fig. 5 shows the concentration level across the device of a Y-shaped herringbone serpentine channel micromixer. This device has two inlets in a Y-shape with a herringbone serpentine channel acting as a mixing zone followed by a sensing zone and outlet. Fig. 5 shows the simulation study of “Test case 10”, which is presented in Table 1. Fig. 6 shows the concentration level across the device of a Y-shaped herringbone serpentine channel micromixer with obstacles. This device has two inlets in a Y-shape along with a herringbone serpentine channel with obstacles acting as a mixing zone followed by a sensing zone and outlet. Fig. 6 shows the simulation study of “Test case 19”, which is presented in Table 1, Fig. 4, 5, and 6 show the mixing concentration profile across the device with different structures and the same input parameters. In Fig. 4, the fluid flow is laminar due to the microchannel and it requires a greater length of microchannel to achieve complete mixing: in this case complete mixing was achieved at 17500 μm. Therefore, Fig. 5 shows a Y-shaped herringbone serpentine channel with sharp edges for better mixing and mixing was achieved at 7500 μm. Then Fig. 6(a) and (b) present a Y-shaped herringbone serpentine channel with obstacles to achieve complete mixing within short length of under 5000 μm.
Test case | Inputs | Outputs | ||||
---|---|---|---|---|---|---|
Device structure | Flow rate | Diffusion coefficient | Velocity (mm s−1) | Pressure (Pa) | Concentration (mol m−3) | |
TC1 | Straight channel | 1 μL min−1 | 15.3 e−10 | Minimum | Minimum | Maximum |
TC2 | 1 μL min−1 | 15.3 e−11 | Average | Average | Average | |
TC3 | 1 μL min−1 | 15.3 e−12 | Maximum | Maximum | Minimum | |
TC4 | 5 μL min−1 | 15.3 e−10 | Minimum | Minimum | Maximum | |
TC5 | 5 μL min−1 | 15.3 e−11 | Average | Average | Average | |
TC6 | 5 μL min−1 | 15.3 e−12 | Maximum | Maximum | Minimum | |
TC7 | 10 μL min−1 | 15.3 e−10 | Minimum | Minimum | Maximum | |
TC8 | 10 μL min−1 | 15.3 e−11 | Average | Average | Average | |
TC9 | 10 μL min−1 | 15.3 e−12 | Maximum | Maximum | Minimum | |
TC10 | Herring bone serpentine channel | 1 μL min−1 | 15.3 e−10 | Minimum | Minimum | Maximum |
TC11 | 1 μL min−1 | 15.3 e−11 | Average | Average | Average | |
TC12 | 1 μL min−1 | 15.3 e−12 | Maximum | Maximum | Minimum | |
TC13 | 5 μL min−1 | 15.3 e−10 | Minimum | Minimum | Maximum | |
TC14 | 5 μL min−1 | 15.3 e−11 | Average | Average | Average | |
TC15 | 5 μL min−1 | 15.3 e−12 | Maximum | Maximum | Minimum | |
TC16 | 10 μL min−1 | 15.3 e−10 | Minimum | Minimum | Maximum | |
TC17 | 10 μL min−1 | 15.3 e−11 | Average | Average | Average | |
TC18 | 10 μL min−1 | 15.3 e−12 | Maximum | Maximum | Minimum | |
TC19 | Herring bone serpentine channel with obstacles | 1 μL min−1 | 15.3 e−10 | Minimum | Minimum | Maximum |
TC20 | 1 μL min−1 | 15.3 e−11 | Average | Average | Average | |
TC21 | 1 μL min−1 | 15.3 e−12 | Maximum | Maximum | Minimum | |
TC22 | 5 μL min−1 | 15.3 e−10 | Minimum | Minimum | Maximum | |
TC23 | 5 μL min−1 | 15.3 e−11 | Average | Average | Average | |
TC24 | 5 μL min−1 | 15.3 e−12 | Maximum | Maximum | Minimum | |
TC25 | 10 μL min−1 | 15.3 e−10 | Minimum | Minimum | Maximum | |
TC26 | 10 μL min−1 | 15.3 e−11 | Average | Average | Average | |
TC27 | 10 μL min−1 | 15.3 e−12 | Maximum | Maximum | Minimum |
The diffusion coefficient and inlet flow rate of the A and B fluids entering the micromixer are the parameters that make up the system input. The system output parameters are the velocity, pressure, and concentration of the liquids. The diffusion of liquids A and B is made possible by the values of the diffusion coefficient. The pressure and velocity of fluids in the micromixer channel are also influenced by the inlet flow rate ratio. Fig. 7 displays the inputs and outputs of the fuzzy logic system.
According to the values of the upper and lower limits of the input and output parameters, the membership function values written for each input and output value are updated in the fuzzy logic approach. The COMSOL Multiphysics application has been used for dozens of different analytical procedures. The results of the analysis are used to develop rules and parameter values. Nine criteria were developed to specify the connection between the parameters after the upper and lower bounds for modelling the necessary parameters using the membership function were chosen. The following table is the fuzzy logic test case (Table 1). We chose three different structures of micromixer devices: Y-shaped straight channel micromixer (SCM), Y-shaped herringbone serpentine shape micromixer (HBM) and Y-shaped herringbone serpentine shape micromixer with obstacles (HBM-OB). Each device structure has two input parameters of flow rate (1, 5 and 10 μL min−1), diffusion coefficient (15.3 e−10, 15.3 e−11 and 15.3 e−12 m2 s−1), and three output parameters of velocity, pressure and concentration. We chose 27 test cases using the fuzzy logic test case table below.
Fig. 8 shows the velocity across the Y-shaped straight channel micromixer with different flow rates of 1 μL min−1, 5 μL min−1 and 10 μL min−1. The peak velocity was achieved in the middle of the microchannel and the velocity was reduced at the wall of the microfluidic channel due to fluid sticking onto the channel wall.
Fig. 9 shows the velocity across the Y-shaped herringbone serpentine channel micromixer with different flow rates of 1 μL min−1, 5 μL min−1 and 10 μL min−1.
The peak velocity was achieved in the middle of the microchannel and the velocity was reduced at the wall of the microfluidic channel due to fluid sticking onto the channel wall. Fig. 10 shows the velocity across the Y-shaped herringbone serpentine channel micromixer with obstacles, with different flow rates of 1 μL min−1, 5 μL min−1 and 10 μL min−1. The peak velocity was achieved in the middle of the microchannel and the velocity was reduced at the wall of the microfluidic channel due to fluid sticking onto the channel wall.
Fig. 11 Concentration across the Y-shaped straight channel micromixer at 1 μL min−1 (A), 5 μL min−1 (B), and 10 μL min−1 (C) with a diffusion co-efficient of 15.3 × 10−10 m2 s−1. |
Fig. 11(A), (B) and (C) show the concentration across the fluidic channel at different locations of 1000 μm, 5000 μm, 7500 μm, 10000 μm, 15000 μm and 20000 μm. When fluids enter into the straight channel from the inlets, the fluid flow is laminar and the fluid–fluid interaction time is greater when the fluid flow is at a low flow rate of 1 μL min−1 (11(A)), so better mixing concentration is observed. Similarly, when the fluid flow is increased to 5 μL min−1 (11(B)) and 10 μL min−1 (11(C)), the fluid–fluid interaction is reduced, so the mixing concentration level is reduced at different locations.
Fig. 12 shows the concentration across the Y-shaped straight channel micromixer at different flow rates of 1 μL min−1 (12(A)), 5 μL min−1 (12(B)), and 10 μL min−1 (12(C)) with a diffusion co-efficient of 15.3 × 10−11 m2 s−1. At different locations along the fluidic channel (1000 μm, 5000 μm, 7500 μm, 10000 μm, 15000 μm, and 20000 μm), Fig. 12(A), (B) and (C) show the concentrations along the fluidic channel at different locations. In a straight channel, fluid enters from the inlets in a parallel flow. The observed mixing concentration is higher when the fluid flow rate is 1 μL min−1 (12(A)), which results from greater fluid–fluid interaction time.
Fig. 12 Concentration across the Y-shaped straight channel micromixer at 1 μL min−1 (A), 5 μL min−1 (B), and 10 μL min−1 (C) with a diffusion co-efficient of 15.3 × 10−11 m2 s−1. |
Similarly, the mixing concentration level is reduced at different locations when the fluid flow is increased to 5 μL min−1 (12(B)) and 10 μL min−1 (12(C)). The mixing concentrations of test cases TC4, TC5 and TC6 are comparatively lower than the previous test cases TC1, TC2 and TC3.
Fig. 13 shows the concentration across the Y-shaped straight channel micromixer at different flow rates of 1 μL min−1 (13(A)), 5 μL min−1 (13(B)), and 10 μL min−1 (13(C)) with a diffusion co-efficient of 15.3 × 10−12 m2 s−1. In Fig. 13(A), (B) and (C), the concentration across the fluidic channel is depicted at various points of 1000 μm, 5000 μm, 7500 μm, 10000 μm, 15000 μm, and 20000 μm.
Fig. 13 Concentration across the Y-shaped straight channel micromixer at 1 μL min−1 (A), 5 μL min−1 (B), and 10 μL min−1 (C) with a diffusion co-efficient of 15.3 × 10−12 m2 s−1. |
Fluid flow is laminar when it enters the straight channel from the inlets. At low flow rates, such as 1 μL min−1 (13(A)), the fluid–fluid interaction time is greater, resulting in a better mixing concentration. Additionally, as the fluid flow increases to 5 μL min−1 (13(B)) and 10 μL min−1 (13(C)), the fluid–fluid interaction reduces, thereby decreasing the mixing concentration at different locations. Fig. 14 shows the concentration across the Y-shaped herringbone serpentine channel micromixer at different flow rates of 1 μL min−1 (14(A)), 5 μL min−1 (14(B)), and 10 μL min−1 (14(C)) with a diffusion co-efficient of 15.3 × 10−10 m2 s−1. This figure illustrates the concentration of fluids across the fluidic channel across a number of locations of 1000 μm, 5000 μm, 7500 μm, 10000 μm, 15000 μm, and 20000 μm. The fluid flow in a herringbone serpentine channel is laminar when it enters from the inlets. When fluid flow is maintained at a low flow rate of 1 μL min−1 (14(A)), the amount of fluid–fluid interaction is greater, resulting in a better mixing concentration. Furthermore, the mixing concentration level is reduced at different locations as the fluid flow increases to 5 μL min−1 (14(B)) and 10 μL min−1 (14(C)). Fig. 15 shows the concentration across the Y-shaped herringbone serpentine channel micromixer at different flow rates of 1 μL min−1 (15(A)), 5 μL min−1 (15(B)), and 10 μL min−1 (15(C)) with a diffusion co-efficient of 15.3 × 10−11 m2 s−1. At different locations within the fluidic channel, of 1000 μm, 5000 μm, 7500 μm, 10000 μm, 15000 μm and 20000 μm, Fig. 15(A), (B) and (C) illustrate the concentrations.
Fig. 14 Concentration across the Y-shaped herringbone serpentine channel micromixer at 1 μL min−1 (A), 5 μL min−1 (B), and 10 μL min−1 (C) with a diffusion co-efficient of 15.3 × 10−10 m2 s−1. |
Fig. 15 Concentration across the Y-shaped herringbone serpentine channel micromixer at 1 μL min−1 (A), 5 μL min−1 (B), and 10 μL min−1 (C) with a diffusion co-efficient of 15.3 × 10−11 m2 s−1. |
Laminar flow occurs when fluid is introduced into the herringbone serpentine channel from the inlets. At a low flow rate of 1 μL min−1 (15(A)), there is more fluid–fluid interaction time, resulting in a higher mixing concentration. In a similar manner, as the fluid flow increases to 5 μL min−1 (15(B)) and 10 μL min−1 (15(C)), the fluid–fluid interaction is reduced, which results in a reduction in mixing concentration levels at various locations.
Fig. 16 shows the concentration across the Y-shaped herringbone serpentine channel micromixer at different flow rates of 1 μL min−1 (16(A)), 5 μL min−1 (16(B)), and 10 μL min−1 (16(C)) with a diffusion co-efficient of 15.3 × 10−12 m2 s−1. In Fig. 16, the concentrations are depicted at different points in the fluidic channel for different distances of 1000 μm, 5000 μm, 7500 μm, 10000 μm, 15000 μm, and 20000 μm. From the inlets, fluid flows in a laminar fashion through the herringbone serpentine channel. With a low fluid flow rate of 1 μL min−1 (16(A)), the fluid–fluid interaction time is greater, resulting in a better mixing concentration. In the same way, as the fluid flow increases to 5 μL min−1 (16(B)) and 10 μL min−1 (16(C)), the fluid–fluid interaction is reduced, resulting in a reduction in mixing concentration levels at differing locations.
Fig. 16 Concentration across the Y-shaped herringbone serpentine channel micromixer at 1 μL min−1 (A), 5 μL min−1 (B), and 10 μL min−1 (C) with a diffusion co-efficient of 15.3 × 10−12 m2 s−1. |
Fig. 17 shows the concentration across the fluidic channel at different locations of 1000 μm, 5000 μm, 7500 μm, 10000 μm, 15000 μm and 20000 μm with different flow rates of 1 μL min−1 (17(A)), 5 μL min−1 (17(B)), and 10 μL min−1 (17(C)) with a diffusion co-efficient of 15.3 × 10−10 m2 s−1. In the present study, we observe that a laminar fluid flow is observed when fluid enters the micromixer devices from the inlets. A higher mixing concentration is observed when the flow rate is low at 1 μL min−1 (17(A)), so the fluid–fluid interaction time is greater. Furthermore, an increase in fluid flow rate decreases the time required for fluid–fluid interaction, resulting in a decrease in mixing concentration. For the proposed device, quadrant-shaped obstacles are introduced inside the microchannel for improved mixing within a short period of time, which allows for a reduction in the length of the device. It is considered that the mixing process has been completed once the fluid concentration reaches the average concentration of the fluid inflow (5.5 mol m−3). According to the results, at a flow rate of 1 μL min−1, the concentration level is saturated throughout a distance of 5000 μm.
This study shows that a better mixing concentration is achieved compared with the other two micromixer devices: Y-shaped straight channel micromixer and Y-shaped herringbone serpentine shape micromixer without obstacles. When the flow rate is increased to 5 and 10 μL min−1 (17(B) and 17(C)), the mixing concentration level is reduced slightly compared with the other two micromixer devices.
Fig. 18 shows the concentration across the fluidic channel at different locations of 1000 μm, 5000 μm, 7500 μm, 10000 μm, 15000 μm and 20000 μm with different flow rates of 1 μL min−1 (18(A)), 5 μL min−1 (18(B)), and 10 μL min−1 (18(C)) with a diffusion co-efficient of 15.3 × 10−11 m2 s−1. In this study, we can observe that when fluids enter into micromixer device from the inlets, the fluid flow is laminar and fluid–fluid interaction time is greater when the fluid flow is at low flow rate of 1 μL min−1 (18(A)) so a better mixing concentration is observed. Similarly, when the fluid flow rate is increased, the fluid–fluid interaction time is reduced because the mixing concentration is reduced. The proposed device has quadrant shaped obstacles introduced inside the microchannel for better mixing within a short duration, reducing the length of the device. It is discovered that the concentration level is saturated throughout a 5000 μm length at a flow rate of 1 μL min−1.
This study shows that better mixing concentration is achieved compared with the other two micromixer devices: Y-shaped straight channel micromixer and Y-shaped herringbone serpentine shape micromixer without obstacles. When the flow rate is increased to 5 and 10 μL min−1 (18(B) and 18(C)), the mixing concentration level is reduced slightly compared with the other two micromixer devices. Fig. 19 shows the concentration across the fluidic channel at different locations of 1000 μm, 5000 μm, 7500 μm, 10000 μm, 15000 μm and 20000 μm with different flow rates of 1 μL min−1 (19(A)), 5 μL min−1 (19(B)), and 10 μL min−1 (19(C)) with a diffusion co-efficient of 15.3 × 10−12 m2 s−1. It was observed that the fluid flow into the micromixer device is laminar when it enters the mixing zone. When the fluid flow is low, 1 μL min−1 (19(A)), the fluid–fluid interaction time is greater, so the mixing concentration is better. When the flow rate is low (19(A)), the fluid–fluid interaction time is longer, resulting in a better mixing concentration. Similarly, an increasing flow rate reduces the fluid–fluid interaction time, which reduces the mixing concentration. An obstacle of quadrant shape has been introduced into the microchannel for better mixing within a short period of time and to reduce the size of the device. The concentration level is saturated over a length of 5000 μm by mixing at a rate of 1 μL min−1. In this study, it was shown that the mixing concentration was higher compared with the other two micromixers: Y-shaped straight channel micromixers and Y-shaped herringbone serpentine micromixers without obstructions. The mixing concentration level is slightly reduced when the flow rate increases to 5 μL min−1 and 10 μL min−1 (19(B) and 19(C)), in comparison with the other two micromixers.
Fig. 20 Mixing efficiency across the micromixer devices at 1 μL min−1 with a diffusion co-efficient of 15.3 × 10−10 m2 s−1. |
From Fig. 20, we are able to see three different micromixer devices: Y-shaped straight channel micromixer (SCM), Y-shaped herringbone serpentine shape micromixer (HBM) and Y-shaped herringbone serpentine shape micromixer with obstacles (HBM-OB).
In this study, we can observe that the mixing efficiencies of the SCM device at the above-mentioned locations are 50.68%, 88.33%, 95.17%, 97.98%, 99.66% and 99.96%. The mixing efficiencies of the HBM device are 49.55%, 87.93%, 99.89%, 99.97%, 99.97%, 99.97%, 99.97%, 99.97% and 99.97%. Similarly, the mixing efficiencies of HBM-OB are 55.62%, 97.02%, 99.93%, 99.93%, 99.93%, 99.93%, 99.93%, 99.93% and 99.93%. When comparing the mixing efficiency of all three types of micromixer device and mixing length, the best mixing efficiency was achieved in the HBM-OB device due to the structural dimensions of the device and the obstacles. The obstacle-induced fluid–fluid interaction caused better mixing, which was achieved in a short duration compared with the other two types of micromixer device.
Fig. 21 shows the mixing efficiency across the fluidic channel at different locations of 1000 μm, 2500 μm, 5000 μm, 7500 μm, 10000 μm, 12500 μm, 15000 μm, 17500 μm and 20000 μm with a constant flow rate and a diffusion co-efficient of 15.3 × 10−11 m2 s−1. In Fig. 21, we can see three different Y-shaped micromixer devices: a Y-shaped straight channel micromixer (SCM), a Y-shaped herringbone serpentine shape micromixer (HBM), and a Y-shaped herringbone serpentine shape micromixer with obstacles (HBM-OB). During this study, we could observe that the mixing efficiencies of the SCM device at the above-mentioned locations were 42.58%, 61.33%, 71.30%, 78.74%, 88.78% and 95.96%, respectively. In terms of mixing efficiency, 41.12%, 70.12%, 97.14%, 99.89%, 99.95%, 99.95%, 99.95%, 99.95%, 99.95% and 99.95% were achieved for the HBM device. As for HBM-OB, the mixing efficiencies were 48.33%, 91.37%, 99.92%, 99.93%, 99.93%, 99.93%, 99.93%, 99.93% and 99.93%. A comparison of the mixing efficiency of all three types of micromixer device and mixing length revealed that the HBM-OB device had the best mixing efficiency, owing to the structure and the obstacles of the device. In comparison with the other two types of micromixer device, the obstacles induce fluid–fluid interaction, which results in improved mixing in a short period of time.
Fig. 21 Mixing efficiency across the micromixer devices at 1 μL min−1 with a diffusion co-efficient of 15.3 × 10−11 m2 s−1. |
Fig. 22 shows the mixing efficiency across the fluidic channel at different locations of 1000 μm, 2500 μm, 5000 μm, 7500 μm, 10000 μm, 12500 μm, 15000 μm, 17500 μm and 20000 μm with a constant flow rate and a diffusion co-efficient of 15.3 × 10−12 m2 s−1. From Fig. 22, we are able to see three different micromixer devices: Y-shaped straight channel micromixer (SCM), Y-shaped herringbone serpentine shape micromixer (HBM) and Y-shaped herringbone serpentine shape micromixer with obstacles (HBM-OB). In this study, we can observe that the mixing efficiencies of the SCM device at the above-mentioned locations are 41.15%, 57.02%, 65.59%, 71.84%, 83.01% and 93.07%. The mixing efficiencies of the HBM device are 39.46%, 66.06%, 94.94%, 99.68%, 99.94%, 99.95%, 99.95%, 99.95%, and 99.95%. Similarly, the mixing efficiencies of HBM-OB are 46.77%, 89.01%, 99.91%, 99.94%, 99.94%, 99.94%, 99.94%, 99.94% and 99.94%. When comparing the mixing efficiency of all three types of micromixer device and mixing length, the best mixing efficiency was achieved in the HBM-OB device due to the structural dimensions of the device and the obstacles. The obstacles induce fluid–fluid interaction, which causes better mixing, which was achieved over a short duration compared with the other two types of micromixer device.
Fig. 22 Mixing efficiency across the micromixer devices at 1 μL min−1 with a diffusion co-efficient of 15.3 × 10−12 m2 s−1. |
Fig. 23 shows the mixing efficiency across the fluidic channel at different locations of 1000 μm, 2500 μm, 5000 μm, 7500 μm, 10000 μm, 12500 μm, 15000 μm, 17500 μm and 20000 μm with a constant flow rate and a diffusion co-efficient of 15.3 × 10−10 m2 s−1. From Fig. 23, we are able to see three different micromixer devices: Y-shaped straight channel micromixer (SCM), Y-shaped herringbone serpentine shape micromixer (HBM) and Y-shaped herringbone serpentine shape micromixer with obstacles (HBM-OB).
Fig. 23 Mixing efficiency across the micromixer devices at 5 μL min−1 with a diffusion co-efficient of 15.3 × 10−10 m2 s−1. |
In this study, we can observe that the mixing efficiencies of the SCM device at the above-mentioned locations are 41.21%, 56.69%, 65.23%, 71.50%, 82.72% and 92.88%. The mixing efficiencies of the HBM device are 40.58%, 66.80%, 95.18%, 99.67%, 99.91%, 99.92%, 99.92%, 99.92% and 99.92%. Similarly, the mixing efficiencies of HBM-OB are 50.60%, 89.97%, 99.88%, 99.90%, 99.90%, 99.90%, 99.90%, 99.90% and 99.90%. When the mixing efficiency of all three types of micromixer device and mixing length are compared, the best mixing efficiency was achieved in HBM-OB.
Fig. 24 shows the mixing efficiency across the fluidic channel at different locations of 1000 μm, 2500 μm, 5000 μm, 7500 μm, 10000 μm, 12500 μm, 15000 μm, 17500 μm and 20000 μm with a constant flow rate and a diffusion co-efficient of 15.3 × 10−11 m2 s−1. This figure displays three different types of micromixers: a Y-shaped straight channel micromixer (SCM), a Y-shaped herringbone serpentine shape micromixer (HBM) and a Y-shaped herringbone serpentine shape micromixer with obstacles (HBM-OB). As a result of this study, we observed 39.94%, 53.05%, 58.60%, 62.59%, 72.95% and 86.81% mixing efficiencies for the SCM device at each of the above-mentioned locations. This graph shows the mixing efficiencies of the HBM device as 38.98%, 62.83%, 92.28%, 99.24%, 99.88%, 99.93%, 99.93%, 99.93% and 99.93%. It was found that the mixing efficiencies of HBM-OB were 43.99%, 85.86%, 99.84%, 99.91%, 99.91%, 99.91%, 99.91%, 99.91%, 99.91%, 99.91% and 99.91%. A comparison of the mixing efficiency of all three types of micromixers and mixing length shows that the HBM-OB device has the best mixing efficiency due to its structural dimensions and the obstacles. Compared with the other two types of micromixer device, the obstacles induced fluid–fluid interaction, resulting in better mixing in a short period of time.
Fig. 24 Mixing efficiency across the micromixer devices at 5 μL min−1 with a diffusion co-efficient of 15.3 × 10−11 m2 s−1. |
Fig. 25 shows the mixing efficiency across the fluidic channel at different locations of 1000 μm, 2500 μm, 5000 μm, 7500 μm, 10000 μm, 12500 μm, 15000 μm, 17500 μm and 20000 μm with a constant flow rate and a diffusion co-efficient of 15.3 × 10−12 m2 s−1. From Fig. 25, we are able to see three different micromixer devices: Y-shaped straight channel micromixer (SCM), Y-shaped herringbone serpentine shape micromixer (HBM) and Y-shaped herringbone serpentine shape micromixer with obstacles (HBM-OB). In this study, we can observe that the mixing efficiencies of the SCM device at the above-mentioned locations are 39.72%, 52.23%, 52.23%, 60.42%, 69.42% and 85.54%. The mixing efficiencies of the HBM device are 38.79%, 62.14%, 92.10%, 99.29%, 99.91%, 99.96%, 99.96%, 99.96% and 99.96%. Similarly, the mixing efficiencies of HBM-OB are 43.71%, 85.07%, 99.84%, 99.94%, 99.94%, 99.94%, 99.94%, 99.94% and 99.94%. When the mixing efficiency of all three types of micromixer device and mixing length are compared, the best mixing efficiency was achieved in the HBM-OB device due to the structural dimensions of the device and the obstacles. The obstacles induce fluid–fluid interaction, causing the better mixing to be achieved over a short duration compared with the other two types of micromixer device.
Fig. 25 Mixing efficiency across the micromixer devices at 5 μL min−1 with a diffusion co-efficient of 15.3 × 10−12 m2 s−1. |
Fig. 26 shows the mixing efficiency across the fluidic channel at different locations of 1000 μm, 2500 μm, 5000 μm, 7500 μm, 10000 μm, 12500 μm, 15000 μm, 17500 μm and 20000 μm with a constant flow rate and a diffusion co-efficient of 15.3 × 10−10 m2 s−1. From Fig. 26, we are able to see three different micromixer devices: Y-shaped straight channel micromixer (SCM), Y-shaped herringbone serpentine shape micromixer (HBM) and Y-shaped herringbone serpentine shape micromixer with obstacles (HBM-OB). In this study, we can observe that the mixing efficiencies of the SCM device at the above-mentioned locations are 41.16%, 57.02%, 65.59%, 71.84%, 83.01% and 93.07%. The mixing efficiencies of the HBM device are 38.83%, 62.38%, 92.19%, 99.31%, 99.89%, 99.93%, 99.93%, 99.93% and 99.93%. Similarly, the mixing efficiencies of HBM-OB are 43.60%, 85.37%, 99.81%, 99.88%, 99.88%, 99.88%, 99.88%, 99.88% and 99.88%. When the mixing efficiency of all three types of micromixer device and mixing length are compared, the best mixing efficiency was achieved in the HBM-OB device due to the structural dimensions of the device and the obstacles.
Fig. 26 Mixing efficiency across the micromixer devices at 10 μL min−1 with a diffusion co-efficient of 15.3 × 10−10 m2 s−1. |
Fig. 27 shows the mixing efficiency across the fluidic channel at different locations of 1000 μm, 2500 μm, 5000 μm, 7500 μm, 10000 μm, 12500 μm, 15000 μm, 17500 μm and 20000 μm with a constant flow rate and a diffusion co-efficient of 15.3 × 10−11 m2 s−1. Fig. 27 illustrates three different types of micromixers: a Y-shaped straight channel micromixer (SCM), a Y-shaped herringbone serpentine shape micromixer (HBM) and a Y-shaped herringbone serpentine shape micromixer with obstacles (HBM-OB). The mixing efficiencies of the SCM device at the above-mentioned locations are 39.62%, 50.93%, 54.95%, 57.37%, 64.53% and 83.67%, respectively. The mixing efficiencies achieved by the HBM device are 38.67%, 62.39%, 92.24%, 99.25%, 99.90%, 99.95%, 99.95%, 99.95% and 99.95%, respectively. As for the mixing efficiencies of HBM-OB, 44,28%, 84.61%, 99.81%, 99.91%, 99.91%, 99.91%, 99.91%, 99.91%, 99.91%, 99.91%, 99.91%, 99.91% and 99.91% have been recorded. Due to the structural dimensions of the device and the obstacles present, the HBM-OB device showed the best mixing efficiency when comparing the mixing efficiency of the three types of micromixer. In comparison with the other two types of micromixer device, the obstacles induce fluid–fluid interaction, which results in improved mixing in a short period of time.
Fig. 27 Mixing efficiency across the micromixer devices at 10 μL min−1 with a diffusion co-efficient of 15.3 × 10−11 m2 s−1. |
Fig. 28 shows the mixing efficiency across the fluidic channel at different locations of 1000 μm, 2500 μm, 5000 μm, 7500 μm, 10000 μm, 12500 μm, 15000 μm, 17500 μm and 20000 μm with a constant flow rate and a diffusion co-efficient of 15.3 × 10−11 m2 s−1.
Fig. 28 Mixing efficiency across the micromixer devices at 10 μL min−1 with a diffusion co-efficient of 15.3 × 10−12 m2 s−1. |
From Fig. 28, we are able to see three different micromixer devices: Y-shaped straight channel micromixer (SCM), Y-shaped herringbone serpentine shape micromixer (HBM) and Y-shaped herringbone serpentine shape micromixer with obstacles (HBM-OB). In this study, we can observe that the mixing efficiencies of the SCM device at the above-mentioned locations are 39.58%, 50.81%, 54.56%, 56.81%, 63.70% and 83.89%. The mixing efficiencies of the HBM device are 38.63%, 62.38%, 92.27%, 99.25%, 99.91%, 99.97%, 99.97%, 99.97% and 99.97%. Similarly the mixing efficiencies of HBM-OB are 43.22%, 84.29%, 99.83%, 99.94%, 99.94%, 99.94%, 99.94%, 99.94% and 99.94%. When the mixing efficiency of all three types of micromixer device and mixing length are compared, the best mixing efficiency was achieved in the HBM-OB device due to the structural dimensions of the device and the obstacles.
Generally, a conventional mixer device requires a greater volume of samples and reagents and other existing micromixer devices also work in high flow rates to achieve complete mixing. This high flow rate will create a greater pressure drop (more than 10 kPa). This proposed and optimized micromixer device provides complete mixing in a shorter length with shorter timing and with a lower pressure drop.
Fig. 29 shows the pressure drop of micromixer devices with respect to three different flow rates of 1, 5 and 10 μL min−1. This figure shows the lowest pressure drop was observed when using an SCM device, such as 10.67, 53.302 and 106.74 Pa. The observed pressure drop levels in the HBM device are 49.37, 246.56 and 494.05 Pa and the pressure drop levels of HBM-OB are 128.01, 640.55 and 1289.7 Pa. We can arrange the pressure drop level of the device in the following order: HBM-OB > HBM > SCM device.
Average concentration at 5000 μm | Number of elements |
---|---|
5.51616 mol m−3 | 40948 |
5.49954 mol m−3 | 55750 |
5.49793 mol m−3 | 65516 |
5.49609 mol m−3 | 137008 |
Table 3 shows a comparison of different types of passive micromixer device with different specifications.
Type | Channel width | Channel height | Typical velocity | Mixing efficiency | Pressure drop | Reference |
---|---|---|---|---|---|---|
Y-shaped with obstacles | 0.100 mm | 0.100 mm | 1.06 m s−1 | 0.05 MI | 3200 Pa | 12 |
Y-shaped with grooves | 0.200 mm | 0.100 mm | 1–100 μL min−1 | 99.24% | 0.01–1.1 × 105 Pa | 16 |
T-shaped with three different cross sectionals | 0.3 mm | 0.6 mm | 1 × 10−4 m s−1 | 95% | 0.5 × 105 Pa | 21 |
T-shaped with obstacles | 0.150 mm | 0.150 mm | 0.1 ≤ Re ≤ 100 | 99.1% | 18135.8 Pa | 22 |
Y-shaped with circular channel | 0.200 mm | 0.030 mm | 1–6 μL min−1 | 97.08% | — | 24 |
Y-shaped with ring | 0.150 mm | 0.155 mm | 4–20 ml min−1 | 96–98% | — | 26 |
T-shaped with 2 inlets and 2 outlets | 0.150 mm | 0.100 mm | 0.09–0.5 mm s−1 | 0.45 mol m−3 | 2–6 Pa | 27 |
T-shaped with obstacles | 0.100 mm | 0.100 mm | 0.006 m s−1 | 0.500 mol m−3 | — | 29 |
Y-shaped serpentine channel | 0.100 mm | 0.100 mm | 9–75 Re | 100% | 23 Pa | 34 |
T-shaped with obstacles | 0.300 mm | 0.100 mm | 0.001–0.1 and 40–45 Re | 90–100% | — | 35 |
Y-shaped serpentine shape | 0.300 mm | 0.300 mm | 0.28 to 30 Re | 0.9 to1.0 MI | 7500 Pa | 39 |
T-shaped with different obstacles | 0.2 mm | 0.2 mm | 0.04 m s−1 | 81.2% | 2600 Pa | 40 |
3D T-shaped | 0.10 mm | 0.050 mm | 25 to 250 Re | 54% | 29 kPa | 41 |
Y-shaped with square and circle | 0.3 mm | 0.2 mm | 100 Re | 99.9% | 65 MPa | 42 |
Tesla micromixer | 0.200 mm | 0.200 mm | 1–100 μL min−1 | 0.45 mol m−3 | 52.868 Pa | 43 |
Y-shaped herringbone serpentine channel micromixer with obstacles | 0.200 mm | 0.100 mm | 1, 5, 10 μL min−1 | 100% at 5000 μm | 128 Pa | Present work* |
Fig. 31 Time domain study of Y-shaped straight channel micromixer at 1 μL min−1 with a diffusion co-efficient of 15.3 × 10−10 m2 s−1. |
Fig. 32 shows a time domain study of the HBM device at a low flow rate of 1 μL min−1 with a diffusion co-efficient of 15.3 × 10−10 m2 s−1 (TC10). At different locations of the device, of 500 μm, 2500 μm, 5000 μm, 10000 μm, and 20000 μm, the x-axis represents time in seconds and the y-axis represents concentration in mol m−3.
Fig. 32 Time domain study of Y-shaped herringbone serpentine channel micromixer at 1 μL min−1 with a diffusion co-efficient of 15.3 × 10−10 m2 s−1. |
An analysis of concentration levels over time at different locations and timings is presented in this study. During this experiment, the inlet and outlet were monitored for 120 seconds. Initially, the concentration level was wider at 500 μm, ranging from 1–10 mol m−3, while at 2500 μm, the concentration range was between 3.93 and 7.10 mol m−3. A similar concentration level was observed at 5000 μm when it was measured at 21 seconds, and it was 5.5 moles m−3 when it was measured at 10000 μm and 20000 μm at 41 seconds and 91 seconds, respectively.
Fig. 33 shows a time domain study of the HBM-OB device at a low flow rate of 1 μL min−1 with a diffusion co-efficient of 15.3 × 10−10 m2 s−1 (TC19). The x-axis denotes the time (seconds) and the y-axis denotes the concentration (mol m−3) at different locations of the device of 500 μm, 2500 μm, 5000 μm, 10000 μm and 20000 μm. This time domain study shows the change in concentration level at different locations and timings. This study was carried out for 120 seconds from inlet to outlet. At the initial stage of 500 μm, there is a wide concentration level between 1 and 10 mol m−3 and there was not complete mixing; when it was measured at 2500 μm, the concentration range was between 4.8 and 6 mol m−3. Similarly, at 5000 μm, the concentration level was almost narrow at 5.5 mol m−3 at 16 seconds, and at this stage the fluid was completely and well mixed; when it was measured at 10000 μm and 20000 μm, the concentration level was almost saturated at 5.5 mol m−3 at 40 and 48 seconds. When comparing Test cases 1, 10 and 19, Test case 19 shows the best mixing efficiency within a short duration compared with the other test cases.
Fig. 33 Time domain study of Y-shaped herringbone serpentine channel micromixer with obstacles at 1 μL min−1 with a diffusion co-efficient of 15.3 × 10−10 m2 s−1. |
A fuzzy logic program was used to classify the data obtained from the analyses, and optimization procedures were performed on the data. During the optimization process, the parameters were changed to obtain the data. Changes in input parameters were applied to the same design in order to obtain output data.
As a result of the analysis and optimization processes, the optimum input parameters that should be applied to the HBM-OB micromixer device in order to achieve complete mixing from the inlet fluids with flow rates of 1, 5, and 10 μL min−1 and a wide range of diffusivity 15.3 e−10, 15.3 e−11, and 15.3 e−12 m2 s−1 were determined. If the input parameters are applied to the microfluidic device in these value ranges, it is understood that the pressure in the output channel is in the range of 128–1289 Pa and complete mixing was achieved within a short length of less than 5000 μm and a short time of 10 seconds due to the structural dimensions of the device as well as the input parameters (TC19). The proposed HBM-OB micromixer device is most suitable for low-diffusivity fluids and its applications such as biosensing, blood plasma analysis, blood cell analysis and heavy metal ion sensing.
This journal is © The Royal Society of Chemistry 2023 |