Kübra
Turan
a,
Ahmet
Üğe
b,
Bülent
Zeybek
b and
Gözde
Aydoğdu Tiğ
*a
aAnkara University, Faculty of Science, Department of Chemistry, Ankara, 06100, Turkey. E-mail: gaydogdu@science.ankara.edu.tr
bKütahya Dumlupınar University, Faculty of Science and Arts, Department of Chemistry, Kütahya, 43100, Turkey
First published on 1st December 2023
In this study, a poly(N-methyl aniline)-cerium oxide-functionalized MWCNTs (PNMA-CeO2-fMWCNTs) composite was synthesized in a one-step preparation technique. As a highly efficient modifier, the composite was used to modify the glassy carbon electrode surface for simultaneous detection of uric acid (UA) and 5-fluorouracil (5-FU). Morphological characterization of the GCE/PNMA-CeO2-fMWCNTs was studied using scanning electron microscopy. Structural characterization of the composite was performed using X-ray diffraction and Fourier-transformed infrared spectroscopy. Electron transfer properties of the prepared electrodes were carried out with electrochemical impedance spectroscopy and cyclic voltammetry. The linear working range for UA and 5-FU was found to be 0.25–50 μM and 0.5–750 μM, respectively. The limit of detection values for UA and 5-FU were 0.04 μM and 0.19 μM, respectively. The effects of various interfering substances on the electrochemical response of UA and 5-FU were investigated. The GCE/PNMA-CeO2-fMWCNTs sensor has excellent stability, reproducibility, anti-interference ability, and reproducibility. To demonstrate the practical application of the sensing platform, fetal bovine serum was selected and tested in the spiked samples, and satisfactory results were obtained. The prepared composite proved to be a promising platform for simple, rapid, and simultaneous analysis of UA and 5-FU.
Anticancer drugs are widely used in oncology patients for treatment.7 5-Fluorouracil (5-fluoro-1H pyrimidine-2,4-dione, 5-FU) is one of the most prominent chemotherapeutic medicines composed of a fluorine atom at the fifth position of uracil.8 It has been used worldwide since 1957 to treat different types of cancer, including metastatic colorectal and breast cancer colon, stomach, breast, pancreas, and cervix.7,9–12 The concentration of 5-FU in blood plasma or serum is regularly maintained in the range of 0.1–1.0 μM, despite its widespread use in chemotherapy.12 Overdosage of 5-FU can produce various side effects with an accumulation of harmful metabolites in cancer patients. Therefore, rapid, reliable, and accurate measurement of 5-FU concentration is necessary to maintain adequate levels of the drug in physiological fluids, regulate it by dosage, and ensure drug quality control and clinical diagnosis. In particular, trace-level quantification of 5-FU is essential in pharmaceutical quality control and clinical diagnosis in anticancer therapy.7,8 Thus, developing highly selective and sensitive analytical methods has gained considerable attention in detecting 5-FU.10 Among several techniques, electrochemical sensors can be considered efficient tools due to their high accuracy, fast response, and ease of application. However, the weak oxidation of 5-FU at the bare electrode poses a significant problem in the electrochemical method.8,12,13 Thus, there is a demand for establishing a high-performance sensor for the trace-level detection of 5-FU. Measuring UA and 5-FU concentrations in biological fluids is critical for diagnosing and treating various disorders. Since UA can also act as a potential inhibitor of the uridine mono-phosphate synthase enzyme, it can effectively reduce the sensitivity of cancerous cells to 5-FU.3 Consequently, developing a sensitive and selective electrochemical sensor platform is essential to detect two analytes simultaneously.
Various analytical methods have been developed in the literature to detect UA and 5-FU analytes in biological samples.9,14–20 Compared to other analytical approaches, the electrochemical methods can allow the detection of the analytes in biological samples without a complex pretreatment method.12,21 Electrochemical techniques have remarkable features such as high sensitivity, fast response time, low cost, instrumental simplicity, and possible miniaturization and integration into portable devices.22,23 These methods are widely used to detect molecules that can easily participate in electrochemical reactions.23 Therefore, it becomes challenging to make simultaneous measurements with bare glass carbon electrodes (GCEs) with low electron transfer kinetics. Consequently, it is necessary to produce a new electrochemical sensor with novel surface modification techniques and advanced electrode materials. The electrochemical performance of the developed sensors depends on the materials used in their production, such as large specific surface area, redox properties, conductivity, good biocompatibility, stability, and good self-compatibility.24 Additionally, as the advanced nanomaterial on the electrode directly impacts the sensing performance, investigating a new functional composite is essential for the simultaneous, accurate, and sensitive determination of UA and 5-FU.18 The modification of electrode surfaces is usually performed using carbon-based nanostructures, metal-based nanomaterials, organic molecules, functionalized nanostructures, surfactants, and dyes with several methods such as electrodeposition, drop-casting, and electrochemical polymerization. Surface modification can provide a higher catalytic effect, higher sensing activity, lower ionization potential, higher electron affinity, higher electronic and ionic conductivity, and more active sites.25 Different conducting polymers, such as polypyrrole6,26 and poly(N-methyl aniline) (PNMA)27,28 were used to prepare modified electrodes targeting various substances. Among these, PNMA is an N-substituted polyaniline derivative polymer that can be synthesized by electrochemical and chemical polymerization. It has been determined to have different chemical properties from polyaniline due to the electronically inducing N-methyl group. However, the inherent disadvantages of PNMA-based sensors are their low electrical conductivity, stability, and selectivity.27,28 PNMA is often functionalized or incorporated with carbon-based materials and metal oxide nanoparticles (NPs) to solve these demerits.
Among the carbon-based nanomaterials, carbon nanotubes (CNTs) and multiwalled carbon nanotubes (MWCNTs) have been widely used as electrode materials in electroanalytical chemistry due to their high surface area, excellent electrical conductivity, and high electrocatalytic activity. Due to their fast electron transfer rates, high electrocatalytic effects, sensitivity, low detection limits, high chemical stability, and excellent biocompatibility, they have been extensively used in modifying sensors.29 Because of their elongated and tubular structure, CNTs may facilitate electron transport between the electrode surface and the electroactive species. The activity of the edge plane-like graphite regions at the CNT tips has been linked to the electrocatalytic effect of CNTs, which CNT functionalization can further improve.13 CNTs also reduce electrode fouling, which could considerably enhance the reusability of such sensors.30 Cerium oxide (CeO2) is a promising electrode material due to CeO2 electrochemical properties, environmental compatibility, and low cost. CeO2 has an exceptional capacity to store or release oxygen due to the oxidation state of the cerium ion varying from +3 to +4, depending on the reducing or oxidizing conditions.31
Recently, Ganesan et al. proposed a nitrogen-doped carbon quantum dots@Fe2O3/MWCNTs/GCE platform for detecting UA, xanthine (XA), and 5-FU. Differently from the literature, in this study, the developed sensor exhibited wide working ranges and lower LOD values about 3.5 and 8 times for UA and 5-FU, respectively.32 This study aimed to fabricate a highly efficient electrochemical sensor to detect UA and 5-FU simultaneously. For this purpose, one step prepared a poly(N-methyl aniline)-cerium oxide-functionalized MWCNTs (PNMA-CeO2-fMWCNTs) composite was fabricated. A GCE modified with PNMA-CeO2-fMWCNTs composite was used for electrochemical sensing of UA and 5-FU. The electroactive performance of the composite-modified GCE was evaluated by cyclic voltammetry (CV) and differential pulse voltammetry (DPV) techniques. The CV and DPV results showed that the PNMA-CeO2-fMWCNTs composite has high electroactivity and is suitable for the simultaneous determination of UA and 5-FU. The GCE/PNMA-CeO2-fMWCNTs sensor has excellent stability, anti-interference ability, and reproducibility. Moreover, the applicability of the sensor was tested with a synthetic human serum sample. The prepared composite was a promising platform for simple, rapid, and simultaneous analysis of UA and 5-FU.
Electrochemical studies were conducted using an AUTOLAB AUT302N equipped with NOVA 2.1.6 software. Electrochemical measurements were performed in a BASI C3 cell stand using a GCE (BASI MF-2012) as a working electrode, Ag/AgCl reference electrode (BASI, MF 2079), and counter electrode platinum wire (BASI, MW 1032). A Bandelin Sonorex ultrasonic bath was used to prepare the dispersion of composite materials and other solutions. In addition, a precision balance (KERN ABT 220-5DM), a magnetic stirrer (CHILTERN HS31), and a pH meter (Thermo ORION 720 A) were used for the preparation of solutions and materials during the study. ELGA Purelab Option-Q DV25 double-distilled water was used in all studies. CV and EIS techniques were used to obtain detailed information about the modification of electrode surfaces. For this purpose, CV between (−0.3 V) and (+0.6 V) and EIS measurements in the frequency range of 0.01 Hz to 100.0 kHz were recorded in 5 mM [Fe(CN)6]3−/4− redox probe containing 0.1 M KCl., DPV and CV analyses were conducted in PBS as a supporting electrolyte for determination of UA and 5-FU analytes simultaneously. DPV conditions were as follows: potential range from (+0.2 V) to (+1.4 V) at a modulation time of 0.02 s, modulation amplitude of 0.05 V, step potential of 0.003 V, and a scan rate of 0.015 V s−1.
An FEI Nova NanoSEM 650 microscope (The Netherlands) was utilized to record field emission scanning electron micrographs of powder samples. The sample surface was coated with Au–Pd alloy using a sputter coater under vacuum before imaging. The X-ray Diffraction (XRD) patterns of samples were obtained with an X-ray diffractometer (PANalytical-Empyrean, The Netherlands) with Cu Kα radiation. Fourier Transfer Infrared Resonance (FT-IR) spectroscopic analyses were conducted using a Bruker Optics Vertex 70 Fourier transform infrared spectrometer (Bruker Corporation Ltd., Germany). Baseline correction and smoothing were performed.
Fig. 1 FE-SEM images of (a) PNMA, (b) CeO2 NPs, (c) PNMA-CeO2, (d) PNMA-fMWCNT, and (e) PNMA-CeO2-fMWCNT. |
Fig. 2 illustrates the XRD pattern of (a) commercial CeO2 NPs, (b) PNMA, (c) PNMA-CeO2, (d) PNMA-fMWCNTs, and (e) PNMA-CeO2-fMWCNTs composites. The polycrystalline CeO2 with cubic fluorite structure shows the typic diffraction peaks in Fig. 2(a), where the 2θ values coincide with the standard XRD card (JCPDS file no 34-0394). PNMA pattern demonstrates four diffraction peaks at 2θ ∼ 17.2°, 24.4°, 34.9°, and 39.1° (Fig. 2(b)). These peaks can be attributed to the characteristic semi-crystalline structure of PNMA.35–37 As seen in Fig. 2(c), the PNMA-CeO2 pattern exhibits similar to the diffractogram of CeO2 NPs. In the event of PNMA-fMWCNTs (Fig. 2(d)), the peaks of fMWCNTs in the composite are clearly visible. This pattern shows mainly two peaks at 2θ ∼ 25.7° and 43°, which are compatible with the literature.38 The pattern of the PNMA-CeO2-fMWCNTs composite is very similar to that of CeO2 NPs. Also, a small peak of fMWCNTs in the ternary composite is observed at 2θ ∼ 25.7° (Fig. 2(e)). The detailed morphological characterization of the CeO2 NPs was examined by TEM (Fig S2†). The TEM micrograph showed the presence of particles smaller than 100 nm. The CeO2 nanoparticles had cube and truncated octahedral-like morphologies and were found together in small particles.34
Fig. 2 XRD pattern of (a) the CeO2 NPs, (b) PNMA, (c) PNMA-CeO2, (d) PNMA-fMWCNTs, and (e) PNMA-CeO2-fMWCNTs. |
Fig. 3 demonstrates the FT-IR spectra of the commercial CeO2 NPs, homopolymer, and composite materials. In the spectrum of the CeO2 NPs (Fig. 3(a)), the bands at 3424 and 1615 cm−1 can be attributed to the O–H vibration of the water adsorbed on the NPs. The bands at 1548, 1420, 1326, 1049, and 720 cm−1 are related to the Ce–O vibrations.39–41 The sharp band at about 472 cm−1 can be assigned to Ce–O39–42 stretching frequency. In the spectrum of PNMA (Fig. 3(b)), the broadband at 3266 cm−1 may be due to the O–H vibration of the water adsorbed onto the polymer and/or the N–H stretching motion of the aromatic units in the polymer chain.43 The bands located at 1598 and 1500 cm−1 are ascribed to the CC stretching vibrations of the quinoid and the benzenoid units, respectively.43,44 The band at about 1233 cm−1 pertains to the C–N stretching vibration in the benzenoid ring.45,46 The band at about 1111 cm−1 may correspond to the in-plane C–H bending vibrations of the aromatic units.44,46 The band at about 824 cm−1 demonstrates the out-of-plane C–H bending vibrations of the aromatic units, which confirm the presence of 1,4-disubstituted benzene rings, indicating polymer formation.33,43,47 The bands at 1042 and 580 cm−1 can be attributed to the presence of SO groups of SO42− ions as the dopant to be incorporated into the polymer structure.44
Fig. 3 FT-IR spectrum of (a) the CeO2 NPs, (b) PNMA, (c) PNMA-CeO2, (d) PNMA-fMWCNTs, and (e) PNMA-CeO2-fMWCNTs. |
As seen in Fig. 3(c) and (d), the FT-IR spectra of PNMA-CeO2 and PNMA-fMWCNTs are similar to the spectrum of the PNMA homopolymer. They exhibit characteristic bands of the polymer, although there are shifts in band locations. The FT-IR spectrum of the PNMA-CeO2-fMWCNTs ternary composite presents mainly bands of PNMA homopolymer and some bands of other components (Fig. 3(e)).
The electrochemical characterizations of the GCE and modified electrodes were carried out in 0.1 M KCl containing 5.0 mM [Fe(CN)6]3−/4. Fig. 4 shows (A) CVs and (B) Nyquist plots at (a) a GCE, (b) a GCE/PNMA, (c) a GCE/PNMA-CeO2, (d) a GCE/fMWCNTs and (e) a GCE/CeO2-PNMA-fMWCNTs. Compared to bare GCE (Fig. 4A(a)), the redox peak currents for GCE/PNMA were almost constant, and its peak-to-peak separation (ΔEp) was slightly reduced (Fig. 4A(b)). After adding CeO2 to the PNMA polymer (Fig. 4A(c)), the peak currents of the modified electrode decreased and showed the highest ΔEp (363.8 mV). These results indicate that the electron transfer kinetics for GCE modified with PNMA-CeO2 are slower than bare GCE (ΔEp = 224.6 mV). On the other hand, when the GCE surface was modified with PNMA-fMWCNTs (Fig. 4A(d)), the redox probe's anodic and cathodic peak currents considerably increased, and a significant decrease in ΔEp (112.3 mV) was observed. We have used these results to infer that the conductivity and electroactive surface area of the fMWCNTs-modified electrode has significantly increased. In the case of the GCE/PNMA-CeO2-fMWCNTs composite electrode, the lowest ΔEp value with 100.1 mV and the significant increase in redox peak currents were observed, as seen in Fig. 4A(e). The prepared ternary composite-modified electrode exhibited better electron transfer properties than bare GCE. EIS was then utilized to investigate the electron transfer features of these electrodes. The Nyquist plots of the bare GCE and modified electrodes are given in Fig. 4B.
Fig. 4 (A) CVs, (B) Nyquist plots of (a) GCE, (b) GCE/PNMA, (c) GCE/PNMA-CeO2, (d) GCE/PNMA-fMWCNTs and (e) GCE/PNMA-CeO2-fMWCNTs electrodes in 0.1 M KCl solution containing 5.0 mM [Fe(CN)6]3−/4−. |
The corresponding electron transfer resistance (Rct) of [Fe(CN)6]3−/4− at the bare GCE, GCE/PNMA, GCE/PNMA-CeO2, GCE/PNMA-fMWCNTs, and GCE/PNMA-CeO2-fMWCNTs electrodes were found to be about 659, 538, 1049, 9 and 224 ohms, respectively. As seen in Fig. 4B, the largest semicircular diameter and, therefore, the highest Rct value belong to GCE/PNMA-CeO2 (Fig. 4B(c)). The lowest Rct value was obtained at the GCE/PNMA-CeO2-fMWCNTs electrode. According to the results obtained from the CV and EIS measurements, the ternary hybrid composite modified GCE shows fast electron transfer kinetics.
The effect of scan rate on the performance of the GCE/PNMA-CeO2-fMWCNTs electrode was investigated using cyclic voltammetry. For this purpose, cyclic voltammograms of the prepared electrode versus Ag/AgCl were recorded in 5.0 mM [Fe(CN)6]3−/4− from (−0.3 V) to (+0.6 V) over a scan rate range of 10–250 mV s−1. The results obtained are displayed in (Fig. S3(a)†). As the sweep rate increased, the peak current signals increased, and the anodic and cathodic peak potentials shifted slightly to more positive and negative voltages, respectively. The anodic and cathodic peak currents calculated from the voltammograms against the square root of the scan rate were plotted, and a linear relationship was observed from the plot (Fig. S3(b)†). This result indicates that the electron transfer process at the solution–electrode interface is mainly diffusion-controlled.48 The electroactive surface area (EASA) of the platforms was determined using the Randles–Sevčik equation (eqn (1)) for semi-reversible electron transfer processes34,49,50. The EASA of the GCE/PNMA-CeO2-fMWCNTs electrode was calculated as 0.138 cm2, while that of GCE was 0.071 cm2. Similarly, there was a 2× increase in the surface area of the PNMA-CeO2-fMWCNT composite-modified electrode.
ip = (2.69 × 105)n3/2ACD1/2v1/2 | (1) |
Fig. 5 (A) CVs, (B) DPVs in PBS with a concentration of 0.3 mM UA and 1.2 mM 5-FU of (a) GCE, (b) GCE/PNMA, (c) GCE/PNMA-CeO2, (d) GCE/fMWCNTs and (e) GCE/PNMA-CeO2-fMWCNTs electrodes. |
The effect of scan rate on the analyte response of the GCE/PNMA-CeO2-fMWCNTs electrode was investigated separately for both analytes using the CV technique. For this purpose, the CVs of the prepared electrode were recorded at scan rates of 25–250 mV s−1 in PBS (pH 6.0). The CV curves for UA and 5-FU were illustrated in Fig. S4 and S5(a),† respectively. The CV measurements for UA and 5-FU were implemented in the potential range of (0.2 V)–(+1.0 V) and (0.9 V)–(+1.7 V), respectively. With a rising scanning rate, while the oxidation peak signals of UA and 5-FU enhance, the oxidation peak potentials move in an anodic direction. The graphs of ipvs. v1/2, ipvs. v, and logipvs. logv are demonstrated in Fig. S4 and S5(b)–(d).† As seen in Fig. S4 and S5(b) and (c),† the correlation coefficient of the Ipvs. v1/2 graph is higher than the Ipvs. v graph. Furthermore, the slope of the logIpvs. logv for UA and 5-FU analytes was found to be 0.568 and 0.567 (Fig. S4 and S5(d)†), respectively, which are close to the theoretical value of 0.5 for a diffusion-controlled process.51,52 These results indicate that the electrochemical oxidation reactions of UA and 5-FU analytes are the mainly diffusion-controlled process at GCE modified with PNMA-CeO2-fMWCNTs ternary composite.
The electron transfer coefficient (α) for the electrochemical oxidation reaction of UA and 5-FU at GCE/PNMA-CeO2-fMWCNTs can be calculated utilizing eqn (2), which is valid for a totally irreversible diffusion-controlled process:53,54
(2) |
The diffusion coefficient in a completely irreversible diffusion-controlled process may be determined using eqn (3), representing the linear relationship between the ip and v1/2:55
(3) |
k0 = 1.11 D01/2 (Ep − Ep/2)−1/2v1/2 | (4) |
The mean k0 value for UA and 5-FU at GCE/PNMA-CeO2-fMWCNTs electrode in the range of 0.025–0.200 V s−1 scan rates was obtained as (2.6 ± 0.7) × 10−3 and (4.8 ± 1.3) × 10−4 cm s−1, respectively.
Fig. 6 (A) DPVs, and (B) bar chart graph of electrodes in composite amounts (a) GCE, (b) 0.25 mg mL−1, (c) 0.50 mg mL−1, (d) 1.0 mg mL−1 (e) 2.0 mg mL−1 and (f) 3.0 mg mL−1. |
Different PBS buffers in the pH 2.0–8.0 were prepared to investigate the effect of pH on the electrochemical oxidation signals of UA and 5-FU at the GCE/PNMA-CeO2-fMWCNTs electrode. It was observed that the peak potential of both analytes shifted toward the negative direction as the pH of the buffer solution increased (Fig. 7A). This result indicates that the proton is involved in the electrochemical oxidation reactions of the analytes. The results show that the electrochemical response of UA and 5-FU increases up to pH 6.0 and decreases at higher pHs (Fig. 7A(f) and (g)). The highest current responses for each molecule were obtained at pH 6.0 (Fig. 7B). From this point of view, the optimum pH for the analysis medium was determined as 6.0. Subsequent studies have been carried out at this pH. The pH values were plotted against the peak potentials of UA and 5-FU, exhibiting linear relationships. The Epa (mV) = −66.33 pH + 1636.65 (R2 = 0.990) for UA and Epa (mV) = −62.48 pH + 777.05 (R2 = 0.994) for 5-FU are the linear regression equations (Fig. 7C).
Fig. 7 (A) DPV curves of 0.3 mM UA and 1.2 mM 5-FU at various pH values. Effect of pH on (B) DPV peak current and (C) DPV peak potential for the oxidation of 0.3 mM UA and 1.2 mM 5-FU in 0.1 M PBS. |
The slopes of the curves for UA and 5-FU are, respectively, 66.33 mV/pH and 62.48 mV/pH, which are almost near to the 59.0 mV/pH theoretical Nernst value (eqn (5)).58
Epa = −59.0 pH + constant | (5) |
According to this finding, a unity number of protons participating in the reaction equals a unity number of electrons transported.59,60
The LOD and LOQ values for UA and 5-FU were 0.04 μM, 0.15 μM, and 0.19 μM, 0.63 μM, respectively. The LOD and LOQ values were also calculated for each analyte at the GCE/PNMA-CeO2-fMWCNTs electrode. The LOD and LOQ values were 0.03 μM, 0.09 μM for UA (0.25–25 μM linear range) and 0.29 μM, 0.98 μM for 5-FU (0.5–1000 μM linear range) respectively. All analytical performance values are given in Table 1. The proposed electrode was proven to have a low LOD and a broad linear response for the simultaneous electrochemical detection of UA and 5-FU.
Analyte | Linear equation | Linear range (μM) | LOD (μM) | LOQ (μM) |
---|---|---|---|---|
UA | I pa = 0.773x − 0.726 (R2 = 0.997) | 0.25–50.0 | 0.04 | 0.15 |
5-FU simultaneous | I pa = 0.0361x − 0.0583 (R2 = 0.998) | 0.5–750.0 | 0.19 | 0.63 |
UA | I pa = 1.334x − 0.642 (R2 = 0.995) | 0.25–25.0 | 0.03 | 0.09 |
5-FU | I pa = 0.0245 + 0.543 (R2 = 0.997) | 0.5–1000.0 | 0.29 | 0.98 |
In Table 2, the performance of the GCE/PNMA-CeO2-fMWCNTs is compared with other sensors that have been developed with similar materials to determine UA. The developed sensor demonstrates excellent analytical performance for UA. When the newly developed electrode was compared with previous studies, it was seen that the LOD value was three times lower than the electrode prepared with CeO2 (ref. 31) and about ten times lower than the electrode prepared only with MWCNTs62 in UA determination. The obtained sensor has lower LOD values compared to different sensors.32,48,59,66 The production of MOF-71 is generally tricky and complex. However, MOF-71 showed a worse analytical performance with a lower LOD value than the one-step prepared PNMA-CeO2-fMWCNTs composite.48
Sensor platform | Method | Linear range (μM) | LOD (μM) | Real sample | Ref. |
---|---|---|---|---|---|
a MOF-71: metal–organic frameworks, CCE: carbon ceramide electrode, P(Arg): poly(L-arginine), GO: graphene oxide. | |||||
Co–CeO2/GCE | SWV | 1.0–2200.0 | 0.12 | Human urine | 31 |
MGCE/MWCNT-COOH-AuNPs | DPV | 2.5–275.0 | 0.050 | Human serum and urine | 66 |
MOF-71/GCE | DPV | 50–1000.0 | 15.61 | — | 48 |
MWCNT/CCE | DPV | 0.55–90.00 | 0.42 | Human serum and some commercial pharmaceutical sample | 62 |
AgNPs/P(Arg)-GO | DPV | 0.5–150.0 | 0.142 | Human urine samples | 59 |
GO-AgNPs@MWCNTs | DPV | 0.5–6.5 | 0.0258 | Human serum | 67 |
N-CQD@Fe2O3/MWCNT/GCE | DPV | 0.5–265.0 | 0.106 | Human urine | 32 |
GCE/PNMA-CeO2-fMWCNTs | DPV | 0.25–25.0 | 0.040 | Fetal bovine serum | This work |
Table 3 compares the performance of the newly developed 5-FU sensor with previously developed sensors prepared with similar materials. The sensor presented an extensive linear range from 0.5 μM to 1000 μM with the LOD value of 0.19 μM for 5-FU which is lower than those reported in the literature63–65,68 The developed sensor presented a wider linear working range than numerous studies in the literature.32,64,65
Sensor platform | Method | Linear range (μM) | LOD (μM) | Real sample | Ref. |
---|---|---|---|---|---|
a BTB: bromothymol blue, N-CQD: nitrogen-doped carbon quantum dots, CPE: carbon paste electrodes modified, AuNps-PFR: porphyrin-capped gold nanoparticles. | |||||
MWNTs/BTB/GCE | CV | 0.8–5000.0 | 0.267 | Injection sample | 63 |
Poly(bromocresol purple); double-strand DNA | CV | 32.29–189.96 | 1.17 | Pharmaceutical samples | 65 |
CPE/AuNps-PFR | DPV | 29.9–234.4 | 0.67 | Pharmaceutical sample | 64 |
CPE/PFR | DPV | 29.9–234.4 | 5.36 | Pharmaceutical sample | |
N-CQD@Fe2O3/MWCNT/GCE | DPV | 0.5–120.0 | 0.019 | Human urine | 32 |
h-MoS2 | DPV | 200.0–7000.0 | 5.62 | Human serum | 68 |
GCE/PNMA-CeO2-fMWCNTs | DPV | 0.5–1000.0 | 0.19 | Fetal bovine serum | This work |
a N = 3 for the determination of mean recovery of three replications. | |||
---|---|---|---|
UA | 1/10 | 5 | 99.41 ± 1.72 |
10 | 98.44 ± 1.47 | ||
25 | 102.11 ± 2.04 | ||
5-FU | 1/10 | 25 | 98.16 ± 1.04 |
150 | 87.46 ± 4.25 | ||
1/100 | 50 | 99.70 ± 0.65 | |
100 | 98.82 ± 1.14 | ||
150 | 100.32 ± 1.67 |
The one-step prepared composites are perfect for utilization in electrochemical sensing because of their special qualities, which include high conductivity, a broad active surface, and high conductivity. Also, these electrode materials can be diversified with different polymeric or carbon materials. Sensors manufactured using composites prepared in a one-step process provide the practicality of sensing. Therefore, it is envisaged that the developed sensor will enable simultaneous, sensitive, and selective determination of different types of pharmaceuticals in future studies.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d3ay02099a |
This journal is © The Royal Society of Chemistry 2024 |