Andrii Tupysa,
Klaudia Tetfejera,
Ludwik Haliczab,
Ewa Bulskaa and
Jakub Karasiński*a
aFaculty of Chemistry, Biological and Chemical Research Centre, University of Warsaw, 101 Żwirki i Wigury St., 02-089 Warsaw, Poland. E-mail: jkarasinski@cnbc.uw.edu.pl
bGeological Survey of Israel, 32 Yeshayahu Leibowitz St., Jerusalem 9692100, Israel
First published on 30th October 2024
In this research work, we investigated the possibility of using mixtures of monoisotopic elements (93Nb/89Y, 165Ho/159Tb and 209Bi/197Au) as calibrators for accurate determination of isotope ratios by multicollector mass spectrometry (MC-ICP-MS). The potential advantage of using monoisotopic elements for calibration lies in their significantly lower uncertainty in atomic masses. This would certainly affect the uncertainty of a measured isotope ratio after data processing, particularly if absolute isotope ratios are being evaluated. Model isotopic systems (87Sr/86Sr–93Nb/89Y, 92Zr/90Zr–93Nb/89Y, 208Pb/207Pb–209Bi/197Au and 167Er/166Er–165Ho/159Tb) were selected and the Internal Standard method (Russell's law) and optimized regression model (ORM) for isotope measurements, using the mentioned mixtures of monoisotopic elements as calibrators, were applied. The precision and trueness of the obtained results were assessed. For example, the 167Er/166Er isotope ratio in a certified reference material of erbium was measured using a gravimetric mixture of Ho and Tb as well as a 160Gd/158Gd pair (NRC GADS-1) as instrumental isotopic fractionation (IIF) calibrators in ORM. Although the average result for both approaches was in the range of instrumental error (0.68223 ± 0.00351 with 165Ho/159Tb and 0.68236 ± 0.00034 with 160Gd/158Gd), the precision was ten times lower when a mixture of monoisotopic elements was used for correction. Based on the results obtained, it was assessed that Russell's law does not provide a correct description for systems with a mixture of two monoisotopic elements, probably due to the assumption of equal IIF for the analyte and internal calibrator. In our opinion, such an assumption is not true in a system when the internal calibrator is a pair of two different elements. A new mathematical description of such a system could enable the introduction of a suitable correction. The ORM method is free from assumptions about the same IIF for individual chemical entities and for this reason it looks more encouraging and shows the potential to be useable even with a mixture of monoisotopic elements as calibrators. In this paper we present experimental results confirming this assumption, although the current challenge is to ensure a sufficiently high precision of calibrator ratio measurement and, consequently, better regression linearity and higher precision of the obtained results.
Natural isotopic fractionation occurs with low intensity for non-traditional stable isotopes. For this reason, natural differences in isotopic composition are very slight, and thus it is necessary to maintain high precision of results to be able to distinguish the isotopic composition of the given element in the analyzed objects. The rapid development of technology has relatively recently enabled the construction of analytical instruments that allow measurement of natural isotope fractionation with satisfactory precision. That is why the analytical chemistry of isotope ratios is a relatively new area, but not a sufficiently explored branch of science. Fortunately, it has been developing very fast recently since the invention of multiple collector mass spectrometry with ionization in inductively coupled plasma (MC-ICP-MS) which is now accepted as a gold standard in providing reliable isotope fractionation data.1,2 This technique offers efficient ionization, high mass resolution and high-precision measurements with relatively low sample consumption and high throughput. On the other hand, MC-ICP-MS requires the application of a very complex instrument, which is sensitive to memory effects, sample matrix composition and suffers from spectral interferences.2 In fact, the most important limitation of the measurements performed with MC-ICP-MS is the necessity to correct the disturbance in instrumental fractionation which causes the measured isotope ratio to never equal the actual value.
For monoisotopic elements, the atomic mass is simply the mass of the single naturally occurring isotope. There is no need to account for the relative abundances and atomic masses of multiple isotopes, which simplifies the determination process. Moreover, in polyisotopic elements, variations in natural isotopic abundances due to isotopic fractionation can introduce uncertainty. Monoisotopic elements do not have this issue, reducing potential sources of error. IIF corrections used in accurate isotope ratio measurements often rely on reference standards whose isotopic compositions and atomic masses are well-characterized. Any uncertainty in the atomic mass of these standards propagates through to the measurements of samples calibrated against them. In addition correction for this bias requires knowledge of the accurate atomic masses. Uncertainty in atomic masses thus propagates into uncertainty in the corrected isotope ratios. It is also worth mentioning that statistical methods used to process and interpret isotopic data often depend on the precision of the input values, including atomic masses. Uncertainty in atomic masses can introduce additional variability in statistical analyses, thereby increasing the overall uncertainty in isotope ratio measurements.
Over the last few decades, various IIF correction methodologies have been used for the determination of isotope amount ratios.3–5 However, it should be noted that for some elements none of those methods can be applied, due to them not possessing at least 4 isotopes (double spike model) or having no neighboring element with a close atomic mass that could be used as a calibrator (Internal Standard or regression model). That is why for such elements, e.g., magnesium, there is no other possibility but to use the standard-sample bracketing (SSB) isotopic fractionation correction model.6 However, this model also has its advantages and limitations. Correctly performed calibration of this type allows results with high trueness and precision to be obtained. Additionally, since each isotope ratio is corrected separately, mass independent fractionation (MIF) has no effect on the accuracy of this model.5 The main limitation of SSB is the need to perfectly match the matrix between the standard and the sample, which is often difficult (due to the difficulty in isolating a sufficiently pure fraction of the analyzed element, e.g., Mg). Failure to meet this condition usually gives results subject to systematic error.
The second calibration strategy that is most frequently used is the inter-elemental internal calibration, also known as the Internal Standard (IS) model. It requires the addition of a foreign element to the sample.7–9 In some cases intra-elemental internal calibration (e.g., Sr) can be carried out, but the number of isotope pairs with a known and stable ratio is limited, so there is no other possibility but to check for neighboring elements as calibrators. This model is based on an assumption that both the analyte and calibrator undergo the same mass discrimination in the instrument,5 which in most cases is not exactly correct. This drawback is solved in the third and the most state-of-the-art method for IIF correction – the optimized regression model (ORM).10–13 It utilizes the significant temporal correlated drift of the IIF of both elements when using MC-ICP-MS and turns this adverse phenomenon into an advantage. The only shortcoming of this correction model is that it is time-consuming. The combination of IS or ORM models with SSB provides even better results of IIF correction, and was successfully used by our research group.14,15
Still in both combined correction strategies (IS-SSB and ORM-SSB) the application of the isotopic pair of known isotopic ratios is required. For some analytes it is not difficult to find an appropriate neighboring element (Tl for Pb;7 Br, Rb or Sr for Se,14 etc.). However, there are some opposite elements in the periodic table, e.g., Mg or Ba, that do not have any options of calibrators with a close atomic mass. Consequently, for several elements there is no certified isotopic reference standard with established absolute isotope ratios approved by the IUPAC Commission on Isotopic Abundances and Atomic Weights (CIAAW).16 So no calibrated absolute isotope ratio measurements can be carried out for them on MC-ICP-MS using a second element as a calibrator. This is one of the key problems that limit the further development of analytical chemistry of isotope ratios. Although some efforts have been made recently to overcome this limitation, e.g., by the application of the triple-isotope calibration approach,17,18 this problem is still far from being solved.
In this work a completely new (according to our knowledge) strategy for accurate measurements of isotope ratios without referring to isotopic standards will be tested. Ultimately this will be achieved through the use of gravimetric mixtures of monoisotopic elements as IIF calibrators in MC-ICP-MS. The possible mixtures of monoisotopic elements that can be used as calibrators in isotope ratio measurements are summarized in Table 1. A possible big advantage of using monoisotopic elements for calibration is their much lower uncertainty on the atomic masses, which would certainly have an impact on the uncertainty of a measured isotope ratio after data processing, especially if absolute isotope ratios are being assessed. We used mixtures of monoisotopic elements with a composition determined by calibration to verify if it is possible to use such mixtures as calibrators in the ORM method and the Internal Standard method.
Studied elements | Fe | Sr or Zr | Dy or Er | Hg, Tl or Pb |
Proposed calibrators | Mn and Co | Y and Nb | Tb, Ho or Tm | Au and Bi |
Calibrators' atomic mass estimates19 | Mn 54.938043 ± 0.000002 | Y 88.905838 ± 0.000002 | Tb 158.925354 ± 0.000007 | Au 196.966570 ± 0.000004 |
Co 58.933194 ± 0.000003 | Nb 92.90637 ± 0.00001 | Ho 164.930329 ± 0.000005 | Bi 208.98040 ± 0.00001 | |
Tm 168.934219 ± 0.000005 | ||||
Possible spectral interferences20 | For Mn: 40Ar14N1H+, 39K16O+, 37Cl18O+, 40Ar15N+, 38Ar17O+, 36Ar18O1H+, 38Ar16O1H+, 37Cl17O1H+, 23Na32S+, 36Ar19F+ | For Y: 51V38Ar+, 53Cr36Ar+, 49Ti40Ar+, 73Ge16O+; 72Ge17O+, 71Ga18O+ | For Tb: 143Nd16O+ | For Au: 181Ta16O+ |
For Co: 43Ca16O+, 42Ca16O1H+, 24Mg35Cl+, 36Ar23Na+, 40Ar18O1H+, 40Ar19F+ | For Nb: 55Mn38Ar+, 57Fe36Ar+, 53Cr40Ar+, 77Se16O+, 75As18O+ | For Ho: 149Sm16O+ | For Bi: 193Ir16O+ | |
For Tm: 153Eu16O+ | ||||
Faraday cup configuration | Not studied yet | H8 – 93Nb; H6 – 92Zr; H4 – 91Zr; H2 – 90Zr; Ax – 89Y; L2– 88Sr; L4 – 87Sr; L5 – 86Sr | H8 – 170Er; H7 – 169Tm; H6 – 168Er; H5 – 167Er; H4 – 166Er; H3 – 165Ho; H2 – 164Er; L3 – 159Tb | H6 – 209Bi; H5 – 208Pb; H4 – 207Pb; H3 – 206Pb; H2 – 205Tl, H1 – 204Pb; Ax – 203Tl; L5 – 197Au |
Intensity (dry plasma), V/10 ppb | — | Y – 5.0 | Tb – 6.7 | Au – 3.8 |
Nb – 5.0 | Ho – 6.7 | Bi – 19 | ||
Tm – 6.7 |
Internal Standard (IS) and Optimized Regression Model (ORM) calibration protocols were both used in parallel to find out whether any of these calibration methods allow the use of a gravimetric mixture of two monoisotopic elements as an internal calibrator (ORM). These calibration methods are known for their robustness and ability to correct for matrix effects,5,21,22 but the ORM method in particular may be attractive for this purpose because it is free from the assumptions burdening Russell's law, especially the assumption of uniformity of IIF for the calibrator and analyte.
The mixtures of monoisotopic elements were usually prepared by gravimetric weighing of corresponding single element standards (0.5 g each) in one vessel (usually a 1.5 mL tube, Eppendorf, Hamburg, Germany).
RF power | 1300 W |
Coolant flow (Ar) | 13 L min−1 |
Auxiliary flow (Ar) | 1.1 L min−1 |
Nebulizer gas (Ar) | 36.0 psi |
Interference cones | Nickel |
Resolution mode | Low (∼300) |
Number of measurement blocks | 1 |
Number of measurements in a block | 20 |
Magnet delay time, s | 3 |
Integration time, s | |
– Zero measurement | 30 |
– Sample measurement | 10 |
(1) |
(2) |
In all IS experiments, the entire measurement cycle consisted of 20 replicates with a constant integration time (t = 10 s per replicate) and blank correction by ESA (Electrostatic analyzer) deflection. All measurements for monoisotopic elements and the most abundant isotope of polyisotopic elements were taken at a signal intensity between 40 and 50 V. The blank signals for different isotopes were in the range of several millivolts, so the signal-to-blank ratio for studied elements was up to 5000.
The relationship of the natural logarithms of the measured isotope ratios at different RF power is linear and can be expressed by the equation:
ln(ri/j) = a + b × ln(rk/l) | (3) |
The actual isotope ratio R is proportional to the measured ratio r, and further equations can be derived from the above equation to calculate isotope ratios free from instrumental fractionation (mass discrimination effect) and independent of the assumption that different isotope pairs undergo identical isotopic fractionation, as assumed in the traditional IS model:
Ri/j = ea × Rk/lb | (4) |
In fact, the advantage of ORM over the Internal Standard method is the lack of assumption of atomic mass-dependent fractionation.5,25,26
The ORM measurements were carried out using the Time Resolved Analysis mode (TRA) and real zero blank correction.
The 93Nb/89Y ratio was initially determined by measurement using the Internal Standard method with the 88Sr/86Sr (NIST 987) pair as a calibrator. As a next step the 87Sr/86Sr isotope ratio was measured using the 93Nb/89Y mixture as an internal calibrator. The average result for the 87Sr/86Sr ratio obtained using the 93Nb/89Y mixture for IS correction was equal to 0.70966 (2SD 0.00179). The difference between this result and the literature value (0.71025 (ref. 27)) is considerable (δ = −0.83‰). However, the most important observation was that the stability of subsequent measurements of the same solution under optimized conditions was poor and a constant drift of the results was observed (Fig. 1).
A similar phenomenon was observed when zirconium was measured instead of strontium using the same monoisotope mixture (Fig. 2). The average result (n = 8) for the 92Zr/90Zr ratio (0.33340; 2SD 0.0010) lies within the uncertainty range of the previously published result (0.33305 ± 0.00052 (ref. 28)) and the extended uncertainty range from the corresponding certified values of ZIRC-1 (0.33305 ± 0.00104). But the more results obtained, the further they deviate from these uncertainty ranges.
The size of these deviations during the application of a monoisotopic mixture is easier to visualize when compared with the results of intra-elemental calibration of strontium. The 87Sr/86Sr ratios were measured simultaneously from the same solution using different pairs of isotopes as calibrators: 93Nb/89Y and 88Sr/86Sr (Fig. 3). It is clear that the 93Nb/89Y pair cannot act as an internal calibrator as a significant drift is observed for the corresponding curve in Fig. 3 in comparison with the calibration using 88Sr/86Sr. Besides, 87Sr/86Sr vs. 93Nb/89Y results are very sensitive to changes in measurement conditions (e.g., gas flows, plasma power).
The same phenomenon was observed when zirconium isotope ratios were measured with an IIF correction using strontium isotopes and the 93Nb/89Y pair (Fig. 4). Beside a significant drift, the 92Zr/90Zr ratio measured versus 93Nb/89Y was also sensitive to changes in measurement conditions.
It can be clearly observed that the measured 93Nb/89Y isotope ratio does not remain stable under the selected working conditions. This may be due to the difference in chemical properties of these two elements. The most important reason for the observations is the fact that different chemical elements are subjected to different drifts in the instrument, although they have similar atomic weight values. These slight, systematic changes in sensitivity are different for each element. As long as a pair of isotopes of the same element is measured, this drift has a minimal effect on the measured isotope ratio because both isotopes of the same element are affected in a very similar way. On the basis of all performed experiments we observed the phenomenon that when the calibrator is a pair of isotopes of two different elements, those small changes in sensitivity to individual elements (drift of the instrument) result in a continuous and significant change in the measured ratio between these two elements.
It can be seen that for isotopes of the same element the instrument behaves as a simultaneous one (signal changes are strongly correlated), while for two different elements (Nb and Y) signal changes are not correlated. Probably a different mathematical description is needed to deal with this type of drift.
As a next step we attempted to correct the observed drift and performed REE measurements by combining the Internal Standard and Standard Sample Bracketing methods (IS SSB). The solution of gadolinium standard GADS-1 (50 μg kg−1 Gd) was spiked with an appropriately prepared 165Ho/159Tb mixture (and served as a bracketing standard). Nine repeated measurements (brackets) of Gd isotope ratios were taken using 165Ho/159Tb as an internal calibrator. Four ratios of Gd isotopes were studied. The results and respective SD values are presented in Table 3. The results indicate that the use of the IS-SSB combination in the case of matrix samples allowed results with much better trueness and precision to be obtained compared to the results obtained with SSB only. The use of this combination made it possible to significantly reduce the impact of the drift of monoisotopic elements on the measurement result of the analyte isotopic ratio in model matrix samples.
δ | δ 160Gd/158Gd | δ 160Gd/157Gd | δ 160Gd/156Gd | δ 160Gd/155Gd |
---|---|---|---|---|
IS SSB (Tb/Ho corrected) delta values | ||||
Average (n = 9) | 0.099 | 0.148 | 0.201 | 0.199 |
SD | 0.022 | 0.041 | 0.060 | 0.080 |
SSB delta values | ||||
Average (n = 9) | −0.166 | −0.250 | −0.312 | −0.480 |
SD | 0.280 | 0.413 | 0.567 | 0.687 |
Based on the results from Table 3, we summarize that the use of an internal calibrator in the form of a Tb/Ho mixture allowed for a significant improvement in both the precision and trueness of the results obtained for matrix samples in comparison to results not corrected by any internal calibrator. We emphasize that this observation is true when conducting measurements in the IS SSB regime. The results obtained in this way do not differ significantly in terms of precision and trueness from the results obtained for samples without a matrix (average delta +0.10 ± 0.12, n = 9). At the same time, the results obtained for matrix samples using the SSB method without correction with an internal calibrator show a significant systematic error, and the results of subsequent repeated measurements differ significantly, resulting in very low precision. We explain this observation by the occurrence of serious matrix effects, the impact of which the SSB method is unable to correct.
To further explore the capabilities of the ORM method a solution containing Gd (NRC GADS-1), Er, Ho and Tb was prepared. As the 167Er/166Er ratio was unknown, first ORM with 160Gd/158Gd (NRC GADS-1 isotopic standard) as a calibrator was used to estimate the 167Er/166Er ratio value. At the same time, we collected data enabling the calculation of the Er isotope ratio using Russell's law. The obtained results are presented in Table 4 and ESI 1.†
ORM 167Er/166Er vs. 160Gd/158Gd | IS 167Er/166Er vs. 160Gd/158Gd | |
---|---|---|
Mean value | 0.68236 | 0.68287 |
SD | 0.00034 | 0.00007 |
N | 34 | 58 |
It should be noted that the results for the IS and ORM methods differ significantly. Since examining whether the Er–Gd system meets the assumptions made for Russell's law is beyond the scope of this work (and we do not discuss it here), in the rest of the work we adopted the result obtained by the ORM method. We also emphasize that the aim of this work is not to determine the true values for the studied isotopic systems, but to perform a critical evaluation of the applicability of the two discussed calibration methods.
Later in the experiment, we measured the 165Ho/159Tb ratio using the ORM method with 160Gd/158Gd (NRC GADS-1) as a calibrator. Examples of data processing according to the ORM method are presented in ESI 2.† In the following measurement session (different day, different conditions) we used the same solution, but this time we used the ORM method with 165Ho/159Tb as a calibrator to measure the 167Er/166Er isotope ratio. As the reference value of the 165Ho/159Tb ratio, we used the value determined by the ORM method in the measurement session preceding these measurements.
Based on the results summarized in Table 5, it can be found that the average value (0.68223) lies within a standard deviation of the value presented in Table 3. However, it should be stressed that the results obtained from individual sessions differ significantly and do not meet the conditions of high reproducibility imposed on isotopic measurements. Additionally, it was observed that the precision of determining the 165Ho/159Tb ratio (here we mean a single point in the logarithmic coordinates) is one order of magnitude worse than that of the gadolinium or erbium isotope ratio and further decreases as the plasma power value moves away from optimal plasma power value. Ultimately, the reproducibility of the ORM results obtained for the system with the 165Ho/159Tb mixture as the calibrator is worse than that of the ORM results obtained for the system with Gd as the calibrator and for the Gd/Er system with correction based on Russell's law. Despite many attempts, at this stage of the experiment, we were not able to find such compromise measurement conditions to ensure equally high precision of measurement of both the Er isotope pair and the 165Ho/159Tb pair. We emphasize, however, that the aim of our work was to investigate the possibility of using the Internal Standard method and the optimized regression model for measurements with a mixture of two monoisotopic elements as a calibrator. It is necessary to carry out selected parameters of validation of the ORM method used that way and to check in detail which of the factors is directly responsible for the worse reproducibility and general difficulties in obtaining ORM curves of satisfactory quality (e.g., with a high coefficient of determination).
ORM 167Er/166Er vs. 165Ho/159Tb as a calibrator | ||||
---|---|---|---|---|
Session number | 1 | 2 | 3 | 4 |
167Er/166Er mean | 0.68224 | 0.68257 | 0.68235 | 0.68174 |
SD | 0.00035 | 0.00024 | 0.00028 | 0.00518 |
N | 7 | 3 | 8 | 7 |
δ vs. 0.68236 | −0.18 | 0.30 | −0.02 | −0.92 |
Average value with SD: 0.68223 ± 0.00351 |
During the preparation of an in-house isotope standard, as done in this research, all possible factors contributing to the total measurement uncertainty must be considered.29 The primary source of uncertainty in preparing monoisotope mixtures likely originates from the weighing of the respective solutions of each element. Therefore, only relatively large amounts of solutions (>0.5 g) were weighed on scales with high precision (up to 0.01 mg).
Despite this, the significant discrepancy between the theoretical and experimental isotope ratios in the in-house standard that we observed could not be solely attributed to weighing uncertainty. In fact, even after accounting for differences in the first ionization potential between two monoisotopic elements, the results improved but were still not accurate enough. This suggests that factors such as second and consecutive ionization energies, mass discrimination effect, and other physicochemical properties of the studied atoms may need to be considered for a more accurate estimation of the true isotope ratio in the prepared mixtures.
Sufficient effects were not achieved, and in our opinion, the most important reasons for the observations are the fact that different chemical elements undergo instrumental drift differently. Consequently, the measured ratio of two isotopes of different elements changes significantly during the measurement session. This leads to results of unsatisfactory precision and potentially subject to systematic error (depending on at what point in the measurement session the measurement is performed).
Our main conclusion is that maybe only a possible revised mathematical description will allow the use of a monoisotopic element pair as an internal calibrator due to the significant drift of results and the inability to maintain adequate repeatability of the results. An important observation is also the fact that small changes in measurement parameters, such as plasma power or nebulizer gas flow rate, cause very large changes in the obtained result, and therefore do not make it possible to maintain the results independent of changes in measurement conditions, which is one of the main functions of the internal calibrator. However, the use of a pair of monoisotopic elements as internal calibrators, combined with the SSB method (IS SSB), makes it possible to reduce the impact of the discussed drift on the results of relative measurements. Additionally, we observe the fact that the internal calibrator in the form of a pair of monoisotopic elements makes it possible to correct the influence of matrix effects in the case of relative measurements carried out in the IS SSB regime. This observation can be summarized as follows: SSB is so sensitive to matrix influences that it is better to correct them even with an imperfect internal calibrator like a monoisotopic mixture than to apply no correction at all. So in isotopic systems where it is not possible to use a “classic” calibrator, monoisotope calibrators make it possible to improve precision, trueness and robustness to matrix influences.
ORM measurements revealed that changes in plasma power induce larger changes in the measured ratios of monoisotopic elements (165Ho/159Tb) compared to a pair of isotopes of the same element. The obtained ORM dependencies were linear, but their coefficients a and b were radically different from the theoretical ones (e.g., slope around 0.077 instead of 0.34, or 0.067 instead of 0.162). Additionally, the precision of measuring the 165Ho/159Tb ratio was an order of magnitude lower than that of Gd or Er, which results in the location of each regression point on the X axis being more random and, consequently, the obtained R coefficient was worse and the ORM result more aleatory. The lower precision and consequently lower reproducibility of measuring the ratios of monoisotopic elements compared to the precision of measuring Gd or Er may be explained by the drift during the integration time of each ORM point for each plasma power (200 seconds). At this stage of the experiment, we were unable to ensure equally high precision in measuring both the analyte (Gd or Er) and the calibrator (165Ho/159Tb). Secondly, the exact ratio between monoisotopic elements remains unknown, as there are described above problems with measuring it. Such a mixture can be prepared gravimetrically, but in this case there are also some limitations which are beyond the scope of this work. For these reasons, we are currently unable to decide with certainty whether the use of ORM with a 165Ho/159Tb mixture as a calibrator can give true and precise isotopic results and what is the potential of this calibration procedure to correct matrix influences. In the near future, we plan to continue our research on this topic.
Based on the described research, we currently believe that Russell's law does not provide a correct description for systems with a mixture of two monoisotopic elements, probably due to the assumption of equal IIF for the analyte and standard (which, in our opinion, does not occur in the case of two monoisotopic elements). The introduction of a suitable correction could be enabled by a new mathematical description of such a system, e.g., by applying the multivariate meta-regression approach30 or the GIMiCK algorithm.31 At this stage, ORM looks encouraging and shows the potential to be usable even with a mixture of monoisotopic elements as calibrators. The current challenge is to ensure a sufficiently high precision of calibrator ratio measurement and, consequently, better regression linearity and higher precision of the obtained results.