Robust determination of low Ti contents in zircon using LA–ICP–MS/MS with NH3 reaction gas†
Abstract
The Ti-in-zircon thermometer is widely used in geoscience for constraining the temperatures of geological settings. Precise and accurate determination of Ti contents (∼1–10 μg g−1) in zircon is a prerequisite for the application of this thermometer; however, traditional laser ablation–single quadrupole–inductively coupled plasma–mass spectrometry (LA–SQ–ICP–MS) analysis of Ti contents in zircon is challenging due to the low abundance of the interference-free isotope of 49Ti (isotope abundance: 5.51%), resulting in poor precision (>30%, 2RSD), in particular for the high spatial resolution (laser spot size <20 μm) analysis. We have developed a robust technique for determining low Ti contents in zircon using LA–ICP–MS/MS with NH3 as the reaction gas. Using an iCAP TQ ICP–MS/MS instrument (Thermo Scientific, USA), high-purity NH3 was a more effective reaction gas than the commonly used 1 : 9 NH3 : He mixture gas, and a three-fold improvement in sensitivity was achieved using an N2 flow rate of 5.0 mL min−1. The reaction products of Ti, Ca, and Zr with NH3 were identified over a mass range of 40 to 160 amu. The reaction product 48Ti(14N1H)(14N1H3)4 (a mass shift of +83, expressed as (48+83)Ti) was measured to separate 48Ti+ from 48Ca+ and 96Zr2+. Interfering ions 48Ca+ and 96Zr2+ have low reaction rates (∼0.0046% and ∼0.07%) at a mass shift of +83, and corrections are not required for glass reference materials and zircon samples. Five zircon reference materials with Ti contents of 2.5–5.0 μg g−1 were analysed, and the analytical precision was better than 10%. Compared with the single quadrupole (SQ) mode, our method is more robust for determining low Ti contents in zircon at high spatial resolutions (laser spot size <20 μm), making it useful for the analysis of complex zircon grains, e.g., metamorphic zircon.