Haojun Yuan‡
a,
Baihui Liang‡bc,
Ping Yanga,
Zhiwei Yanga,
Xinyi Caoa,
Yangbo Wua,
Jie Zoua,
Qinghui Jin*a and
Wanlei Gao*a
aCollege of Information Science and Engineering, Ningbo University, Ningbo 315211, Zhejiang, China. E-mail: gaowanlei@nbu.edu.cn; jinqinghui@nbu.edu.cn
bHealthy & Intelligent Kitchen Engineering Research Center of Zhejiang Province, Ningbo 315336, Zhejiang, China
cNingbo Fotile Kitchen Ware Company, Ningbo 315336, Zhejiang, China
First published on 19th January 2024
The electrochemiluminescence (ECL) behavior of a tri(2,2′-bipyridyl)ruthenium(II) (Ru(bpy)32+)/tripropylamine (TPrA) system was investigated in sensor chips with two kinds of integrated two-electrode systems, which included screen-printed electrodes (SPE) and physical vapor deposition (PVD) electrodes. Firstly, under excitation with an optimal transient potential (TP) within 100 ms, the ECL assay could be carried out on the microchips using an Au & Au electrode system, emitting strong and stable light signal. Secondly, on the PVD chip, the ECL intensity initiated by optimal TP was eight times stronger than the peak light signal emitted by the linear sweep voltammetry model. Finally, the logarithmic ECL intensities exhibited a linear increase with the logarithmic concentrations of Ru(bpy)32+ in both the SPE and PVD chips without any reference electrode (RE). Typically, the integration of an interdigital two-electrode system in the microchip significantly enhanced the ECL sensitivity of Ru(bpy)32+ because the large relative area between the working electrode (WE) and counter electrode (CE) achieved a highly efficient mass transfer. This improvement enabled the establishment of a reliable linear relationship across a wide concentration range, spanning from 1 pM to 1 μM (R2 = 0.998). Therefore, the exceptional ECL response of the Ru(bpy)32+/TPrA system on microfluidic chips using a two-electrode system and the TP excitation model has been demonstrated. This suggests that ECL chips without a RE have broad potential for the rapid and sensitive detection of multiple targets.
Recent developments in materials have further expanded the application of ECL technology.8–10 The original ECL luminophore, tris(2,2′-bipyridine) ruthenium(II) dichloride, also known as Ru(bpy)32+, only produced fluorescence in aprotic electrolytes.11 However, the introduction of co-reactants, such as amines, alcohols, and amino acids, has enabled ECL to occur in aqueous solutions, leading to a breakthrough in the scale of detection.12 Additionally, the luminol and H2O2 ECL system, a classic ECL system, exhibits excellent ECL activity.13 Although Ru(bpy)32+ and luminol, along with their co-reactants, have limited applications due to difficulties in labeling analytes, the use of nano-materials such as CdSe quantum dots (QDs) has shown promise in revealing ECL activity, and can be easily modified with biomolecules.12,14
In addition to the development of new ECL systems, extensive research has been conducted on the electrode material, surface status and structure to improve ECL signal strength.15–19 For example, Neužil's group presented a nanostructured gold amalgam microelectrode array with a high surface-to-volume ratio, which enhanced the electrical current for ECL.11 In ECL experiments, large-sized, Ag/AgCl electrode filled with saturated KCl solution and platinum (Pt) electrodes were used as reference and counter electrodes, respectively. With the continuous development of microfluidic technology and microelectro-mechanical systems (MEMS) technology, many micro-electrode systems have been integrated into microfluidic chips for the sensitive detection of biomolecules.20–22 Satienperakul's group developed a simple 3D-printed platform with a three-electrode system, which was used to detect sibutramine in dietary supplements, demonstrating the broad potential for micro-electrode systems in ECL-based biosensor applications.23 Although current microchips utilize a three-electrode system to achieve ECL reaction, three-electrode system increases the complexity of multi-detection applications on a single microchip.
Bipolar electrode (BPE) based chip, requires only two driving electrodes to exert potential over many BPE arrays, presenting an excellent option for realizing multiplex ECL assays.24 Compared to traditional three-electrode system, BPE-ECL chips significantly simplify detecting unit and operation. For the two poles of BPE in the same electrolyte solution, the target will react on the driving electrode, bringing a strong background signal and reducing the sensitivity of detection. Xu's group designed an Indium Tim Oxide (ITO) single-electrode electrochemical system (SEES) for high-throughput ECL tests, which was free from the ECL background problem in BPE system.25 ECL assays in each cell on one SEES chip were initiated by the same pair wires, therefore light signals for each detection cell only can be measured simultaneously by charge coupled device (CCD), not photomultiplier tube (PMT).
If a two-electrode system without a RE is able to replace the three-electrode system and ensure a stable electrochemical potential for light emission, it may be readily to allow for a two-electrode system array (at least four units) in one ECL microchip, making them suitable for the simultaneous detection of multiple parameters. Currently, there is a lack of detailed research on the impact of integrated electrode structures and constituents on the efficiency of ECL assay. Gold is a preferred metal for sensor fabrication due to its simplicity, stability, and potential for surface chemistry via thiol cross-linkers. To investigate the performance of different microchip configurations, simple sensor chips with integrated electrodes were fabricated from gold using screen printing and physical vapor deposition (PVD) to create a two-electrode system, respectively. And the ECL performance of Ru(bpy)32+/tripropylamine (TPrA) system was studied on different chips. Under the excitation with an optimal transient potential (TP), ECL assay could be carried out on the chips with two-electrode systems. The effect of electrodes' size and buffer's ionic concentration on the ECL behavior of Ru(bpy)32+/TPrA was confirmed with PVD microchip. Furthermore, there was a strong linear correlation between the logarithmic ECL intensity and the concentration of Ru(bpy)32+ in all chips, without including any RE. These findings suggest that a microchip utilizing a two-electrode system for ECL is a promising possibility.
A stock solution of Ru(bpy)32+ (0.1 mM) was prepared by dissolving the appropriate quality of tris(2,2′-bipyridyl)dichloro ruthenium(II) hexahydrate in the 0.01 M PBS buffer (pH 7.0). A series of Ru(bpy)32+ reagent solution ranging from 1 pM to 1 μM were prepared with the appropriate dilution of the Ru(bpy)32+ stock solution in the PBS buffer. TPrA reagent solution for ECL assay was prepared with adding the appropriate amount of TPrA solution in the working solution of Ru(bpy)32+.
Structure no. | Type | Electrode system | Working electrode sizea | The area of working electrode | Distancea | Microchip size |
---|---|---|---|---|---|---|
a R means the radius of circular electrode. W means the width of finger in interdigital electrodes. L means the length of finger in interdigital electrodes. D means the distance between working electrode and counter electrode. | ||||||
#1 | SPE | Three-electrode (Au & Au & Ag/AgCl) (Fig. 1B-#1) | R = 1.5 mm | 7.065 mm2 | D = 750 μm | 12 mm × 45 mm |
#2 | SPE | Three-electrode (Au & Au & Au) (Fig. 1B-#2) | R = 1.5 mm | 7.065 mm2 | D = 750 μm | 12 mm × 45 mm |
#3 | SPE | Two-electrode (Fig. 1B-#3) | R = 1.5 mm | 7.065 mm2 | D = 750 μm | 12 mm × 45 mm |
#4 | PVD | Two-electrode (Fig. 1B-#4) | R = 900 μm | 2.543 mm2 | D = 200 μm | 13 mm × 13 mm |
#5 | PVD | Two-electrode (Fig. 1B-#5) | R = 800 μm | 2.010 mm2 | D = 200 μm | 13 mm × 13 mm |
#6 | PVD | Two-electrode (Fig. 1B-#6) | R = 650 μm | 1.327 mm2 | D = 200 μm | 13 mm × 13 mm |
#7 | PVD | Interdigital two-electrode (ten fingers) (Fig. 1B-#7) | W = 200 μm | 6 mm2 | D = 150 μm | 6 mm × 8 mm |
L = 3 mm |
Four kinds of microchips with two-electrode system were fabricated through physical vapor deposition (PVD) method. Structure sizes of PVD microchips are listed in Table 1. Among PVD microchips, three of them contain electrode array, which includes four circular WE and one shared CE (Fig. 1B#4, 5 and 6). For PVD microchips #4, #5 and #6, the radius of WE are 900 μm, 800 μm and 650 μm, respectively. The distance between WE and CE in PVD microchips #4, #5 and #6 are all 200 μm. The whole size of PVD microchips #4, #5 and #6 are 13 mm × 13 mm. In addition, a PVD microchip #7 with interdigital electrode system was developed, which contains ten fingers (Fig. 1B#7). The width and length of every finger is 200 μm and 3 mm. The spacing between each finger in interdigital electrode system is 150 μm. The whole size of PVD microchip #7 is 6 mm × 8 mm. Additional five PVD microchips of various sizes, labeled as #8, #9, #10, #11 and #12, were listed in Table S1.†
The fabrication process of PVD microchips is shown in Fig. 1A. Firstly, the BF33 glass substrate was cleaned using a plasma degumming machine (Fig. 1Aa). A 10 nm chromium layer (Cr) as an adhesion layer and a 200 nm Au layer as electrode layer were added to the wafer, respectively, via sputter coating (Fig. 1Ab and c). The wafer was then coated with AZ5214 and micro-structures were patterned through the lithography process (Fig. 1Ad and e). Next, the metal (Au and Cr) layers were removed from the wafer via ion beam etching (IBE) (Fig. 1Af). The electrode structures were finally immersed in acetone, which was agitated using ultrasound, to dissolve the photoresist for a duration of 5 minutes (Fig. 1Ag).
For each microchip, a PDMS well was designed based on the area of the electrode system. The diameters of the well for SPE microchip, PVD microchip and PVD microchip with interdigital electrode system were 5 mm, 1.5 mm and 5 mm, respectively. The fabrication process of PDMS well was as follows: a PDMS mixture (in a 10:1 ratio of base and curing agent) was poured on the top of the silicon mold, degassed in vacuum and cured in an 80 °C oven for 1 h. After curing, the PDMS was cut into pieces as the size of microchip and punched with needle to form a well. The tailored PDMS chip and microchip were aligned together through plasma treatment. At last, the device was incubated on a hot plate at 120 °C for 8 h to strengthen the bonding between the PDMS chip and microchip.
TPrA → TPrA˙+ + e− | (1) |
TPrA˙+ → TPrA + H+ | (2) |
Ru(bpy)32+ → Ru(bpy)33+ + e− | (3) |
Ru(bpy)32+ + TPrA˙ → Ru(bpy)3+ + DPrA/propanal | (4) |
Ru(bpy)33+ + Ru(bpy)3+ → Ru(bpy)32+* + Ru(bpy)32+ | (5) |
Ru(bpy)32+* → Ru(bpy)32+ + hν | (6) |
First the oxidation of TPrA is occurred on the surface of the WE (eqn (1)). At the same time, Ru(bpy)32+ loses an electron to form Ru(bpy)33+ (eqn (3)). Then Ru(bpy)32+ is reduced by TPrA radical to turn into Ru(bpy)3+. Ru(bpy)3+ continued to reduce Ru(bpy)33+ to Ru(bpy)32+ and Ru(bpy)32+* (eqn (5)). When the excited-state molecule returns to its ground state, light of 620 nm is produced (eqn (6)). The ground-state Ru(bpy)32+ can go on to react with TPrA.26 Therefore, if TPrA is excess in the solution, this reaction process can continue to produce light.
The ECL behavior of the Ru(bpy)32+ and TPrA system was studied on the circular Au WE with a radius of 900 μm, which was located in PVD chip #4. During linear sweep voltammetry (LSV) process, the potential reached 2.2 V, the current of Ru(bpy)32+ oxidation reached its peak value and at the same time the ECL signal reached the highest (∼1422 RLU, see curve a and b in Fig. 3B). To evaluate the ECL behavior of Ru(bpy)32+ and TPrA system under the transient potential (TP) excitation, the ECL signal was recorded every 10 ms for 100 ms. The dots in the curve c of Fig. 3B represented the ECL light signal under different TP excitation. In the optimal excitation potential optimization experiment, to eliminate the effect of TPrA's concentration, the mixture of Ru(bpy)32+ and TPrA was replaced after each measurement. The PVD microchip #4 was utilized with a transient potential (TP) ranging from 1.5 V to 2.2 V, resulting in increased light signal intensity as the TP increased. By further increasing the TP value, a continuous increase in the ECL signal was found. When the TP continued to increase (>2.6 V), the ECL signal became reduced. The ECL signal (∼12138 RLU) under the optimal TP (∼2.6 V) excitation was about eight times as large as the peak intensity of ECL signal (∼1422 RLU) during LSV process. Therefore, strong ECL signal from Ru(bpy)32+ and TPrA system in a two-electrodes based microchip can be initiated by TP excitation, but the best TP value is not consistent with the peak potential obtained from the oxidative process by LSV.
To validate the effect of distance between WE and CE on ECL assay, PVD chips with identical WE area (a 900 μm radius of WE) but varying sizes of CE were employed for the ECL test. As shown in Fig. S2,† as the distance decreased from 750 μm to 500 μm, the optimal voltage for the ECL assay dropped from 2.7 V to 2.6 V, resulting in a slight increase in the ECL signal (an improvement of 1.5%, see curve #11, and #12 in Fig. S2†). Then the distance continued to reduce, the optimal voltage remained constant with a minimal improvement in ECL intensity (see curve #4, #10, and #11 in Fig. S2†). Compared with the area of the WE, the distance between the WE and CE in the PVD chip had a relatively low impact on the ECL assay. Herein, the results demonstrated PVD chip with circular integrated electrodes can be applied for ECL detection and has the similar performance with the SPE chip.
For another PVD chip with interdigital electrode system, the intensity of ECL signal become stronger with the increasing of TP. While the potential exceeded 2.2 V, some bubbles appeared on the surface of the electrodes due to the electrolysis of water in the solution (see curve #7 in Fig. 3C). Therefore, we chose 2.2 V as the best TP for PVD chip # 7. The interdigital electrode system in PVD hip #7 has a large relative area between WE and CE, achieving a highly efficient mass transfer.27 Therefore, the light emitted from the ECL system on interdigital electrode was the strongest among all microchips with two-electrode system.
EC behavior of Ru(bpy)32+/TPrA system in different buffer was tested on PVD chip #4. The LSV showed that the ionic concentration of reaction system has little impact on the oxidation peak potential of Ru(bpy)32+ (Fig. 4A). As the increase of PBS concentration, the peak current increased obviously. In addition, the best working voltage under TP model was tested for ECL assay in buffers with different ionic concentration. From Fig. 4B, we can see that the increase of ionic strength in ECL system has improved the intensity of emitted light and reduced the optimal TP value. The concentrations of H2PO4− and HPO42− in PBS buffer increased, resulting in the increase of deprotonation of TPrA+˙ and therefore a high concentration of Ru(bpy)32+* and an improvement of ECL signal.28,29 Therefore, ionic concentration of buffer has significant effect on increasing the catalytic current of Ru(bpy)32+/TPrA system and ECL signal in microchips with two electrodes.
In biochemical analysis, 10 mM PBS (1× PBS) is a commonly used buffer solution for diluting samples. Generally, 0.1 M PBS (10× PBS) is used for long-term storage. And the 10× PBS buffer is diluted to form a 1× PBS buffer for detection. For our further study, the PVD microchip based ECL detection method would be applied for biomarker (protein) detection. Therefore, 10 mM PBS was chosen to be used as ECL buffer for detection.
The ECL behavior of Ru(bpy)32+/TPrA system was measured with four PVD chips through TP activation. Because of the reduction in electrodes' size, the optimal working potentials for PVD chips with circular WE increased obviously compared to that of SPE chip. But the PVD chip with interdigital electrode has the similar optimal working potential to that of SPE chip. Four PVD chips with two electrodes exhibited higher normalized ECL responses compared to the SPE chips when the intensity was normalized by the area of WE (the data of area in Table 1). PVD chips with two-electrode system demonstrated larger optimal TP (2.2–2.9 V) for Ru(bpy)32+/TPrA system compared to the screen-printed electrodes (SPE) chips with two-electrode system (∼2.1 V), leading to more than 39.9% improvement in normalized ECL signal intensity. The ECL response on interdigital electrodes was the strongest among all microchips (Fig. 5A inset).
To evaluate the linearity of Ru(bpy)32+ on microchips, ECL responses were measured for different concentration of Ru(bpy)32+ ranging from 1 pM to 1 μM on three SPE chips and four PVD chips, each using their respective optimal potential settings. As shown in Fig. 5B, the linear of Ru(bpy)32+ for all SPE chips extended from 0.1 nM to 1 μM (see curve #1, #2, and #3 in Fig. 5B). The results revealed that RE was not indispensable for ECL assay in integrated microchips. Under the excitation of the optimal TP, microchips integrated with WE and CE can be applied for highly efficient ECL detection.
Furthermore, reducing the WE's radius from 1.5 mm to 900 μm did not significantly impact the ECL signal intensity observed on PVD chip #4. The linear relationship between logarithmic ECL intensity and logarithmic Ru(bpy)32+ concentration was good in the range of 0.1 nM to 1 μM (R2 = 0.995, see curve #4 in Fig. 5B). As the radius of WE in PVD chip #5 was reduced to 800 μm, the sensitivity decreased to 1 nM. The sensitivity of PVD chip #6 was the lowest because the radius of WE was smallest (650 μm). For PVD chips #4, #5 and #6, there were four WE and a shared CE. Every two-electrode unit in one chip showed the similar ECL response when detecting Ru(bpy)32+. While the PVD chip with interdigital electrode showed a highest ECL sensitivity of Ru(bpy)32+ and a wide linear range between 1 pM to 1 μM (R2 = 0.998). An interdigital WE demonstrated higher catalytic efficiency for the Ru(bpy)32+/TPrA system compared to a circular WE, resulting in improved sensitivity. To detect rare target molecule, microchips with interdigital electrodes may be a best choice for ECL assay. The rectangular shape of the WE and the small distance between the WE and CE make it difficult to selectively immobilize capture probes for ECL assays on the WE while excluding the CE using traditional modification methods. And the structure of interdigital electrode is not readily integrated with other units of interdigital electrode, limiting its application in multi-parameter detection.
For further application in detecting TPrA, PVD chip #4 was used to analyze different concentration of TPrA in 0.01 M PBS containing 10 μM Ru(bpy)32+. As shown in Fig. S3,† the ECL intensity increased with the concentration of TPrA rose. There was a strong linear relationship between the logarithmic TPrA concentration and logarithmic ECL intensity across the range of 10 μM to 10 mM (R2 = 0.998, l g (ECL intensity) = 0.99 ± 0.02 l g (TPrA) +6.47 ± 0.08). Therefore, due to the good consistency and high sensitivity, the PVD microchip with two-electrode array can be applied for multi-parameter detection in further study.
Footnotes |
† Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d3ra07298c |
‡ These authors contributed equally. |