Paulina
Janicka†
ab,
Sylwia
Baluta
*c,
Juliusz
Winiarski
d,
Kinga
Halicka-Stępień
c,
Aleksandra
Pogorzelska
a,
Joanna
Cabaj
c,
Katarzyna
Pala
e and
Barbara
Bażanów
a
aDepartment of Pathology, University of Environmental and Life Sciences in Wrocław, Norwida 31, 50-375 Wrocław, Poland
bFood4Future Technologies Sp. z o. o., ul. Tarasa Szewczenki 24, 51-351 Wrocław, Poland
cInstitute of Advanced Materials, Wrocław University of Science and Technology, Wybrzeże Wyspiańskiego 27, 50-370 Wrocław, Poland. E-mail: sylwia.baluta@pwr.edu.pl
dGroup of Surface Technology, Department of Advanced Material Technologies, Wroclaw University of Science and Technology, Wybrzeże Wyspiańskiego 27, Wroclaw 50-370, Poland
eWater Science and Technology Institute- H2O SciTech, ul. Tarasa Szewczenki 24, 51-351 Wrocław, Poland
First published on 16th February 2024
Norovirus (NoV) infection is one of the most common non-bacterial causes of gastroenteritis among the population worldwide. From the point of view of medical diagnostics, it is important to develop a system that would sensitively and selectively detect norovirus from a patient's sample in order to control and limit its spread. In this paper, we present a stable and sensitive NoV (mouse model) detection matrix in infected food samples. The bio-platform was made of a modified gold electrode with a self-assembled L-cysteine monolayer, covered with gold nanoparticles, a linker and an antibody specific to the VP1 surface protein of the virus. Binding of the VP1 protein to the antibody caused a decrease in the current strength confirmed by electrochemical techniques – cyclic voltammetry (CV) and differential pulse voltammetry. The reduction of the current was proportional to the concentration of NoV sample. The biosensors showed high sensitivity and linearity in a range from 1 × 10−9 to 1 × 10−18 TCID50, with the detection limit of 1 × 10−18 TCID50. CV showed a diffusion-controlled process. In addition, each modification step was confirmed by scanning electron microscopy, electrochemical impedance spectroscopy, and CV. The described immunosensor showed excellent recovery values, good linearity and long-term stability, crucial parameters for biosensor construction.
Paulina Janicka Paulina Janicka, MSc, holds a Biology degree with a focus on laboratory techniques from the University of Life Sciences in Wrocław. Since 2019, she's pursued a PhD at the Faculty of Veterinary Medicine, concentrating on norovirus biology and occurrence. Her doctoral work centers on creating a swift norovirus detection method for various samples. Presently, she's employed at Food for Future Technologies (formerly Fish Farm Solutions), contributing to the development of rapid pathogen detection tests for salmon and shrimp. Paulina Janicka has co-authored a scientific publication and holds a patent. |
Sylwia Baluta Sylwia Baluta, PhD, graduated in Biotechnology and Medicinal Chemistry in 2014, followed by a PhD from Wrocław University of Science and Technology in 2021. Her research focuses on developing biosensors for neurotransmitters, hormones, and viruses through electrochemical and optical methods using nanoparticles like graphene quantum dots (GQDs) and semi-conductive polymers. She has co-authored 15 publications, holds 6 patents, and contributed to patent declarations. Sylwia Baluta honed her skills through research internships with Food for Future Technologies (formerly Fish Farm Solutions) and at the University of Sassari, Italy, working with Prof. Maria Pilo. |
Juliusz Winiarski Juliusz Winiarski, PhD, DSc, is a Associate Professor in Department of Advanced Material Technologies at Wroclaw University of Science and Technology (WUST), Wroclaw, Poland. Prof. Winiarski received his MSc (2008) and PhD (2012) from WUST. After habilitation (2020) he is currently serving as a head of the Group of Surface Technology. His recent research are the modern electroplating, industrial corrosion problems, coatings and electrochemistry. |
Kinga Halicka-Stępień Kinga Halicka-Stępień, MSc, is a graduate of Pharmaceutical Biotechnology at Wrocław University of Science and Technology. Since 2019, she has been a PhD Student at the Faculty of Chemistry of Wrocław University of Science and Technology, where her research concerns the application of nanomaterials in enzymatic detection systems for the electrochemical determination of neurotransmitters. |
Aleksandra Pogorzelska Aleksandra Pogorzelska, MSc, is a graduate in Microbiology at the University of Wrocław, Wrocław, Poland. As of 2019, she is an employee of the Faculty of Veterinary Medicine, Wrocław University of Life Sciences, Wrocław, Poland. She conducts research in classical and molecular virology with a special focus on the family Adenoviridae. |
Joanna Cabaj Joanna Cabaj, PhD, DSc, earned her PhD in organic chemistry in 2005 and habilitated in biotechnology in 2012 at Wrocław University of Science and Technology, Poland. She currently serves as a professor at the Faculty of Chemistry there. Her research focuses on developing semiconducting materials for molecular electronics and sensors, with a specialization in thin Langmuir–Blodgett and Langmuir-Schaefer films for gas sensors and molecular biosensors. Joanna Cabaj has co-authored over 70 publications, numerous book chapters, and holds more than 40 patents. Her work combines organic chemistry and sensor technology to advance materials for various applications. |
Katarzyna Pala Katarzyna Pala, PhD, is a Biotechnology graduate from Wrocław University of Science and Technology (2008) who completed her PhD at the University of Wrocław in 2014 and engaged in an intellectual property venture at Jagiellonian University in Krakow. She underwent an extensive internship at the University of Wollongong. Currently, she serves as the Co-Founder and CEO of Food for Future Technologies (formerly Fish Farm Solutions). Additionally, she holds the position of Chair of the Board of Directors at the Water Science and Technology Institute in Wrocław. Food for Future Technologies, a Polish-Norwegian startup, specializes in sustainable food production, adhering to circular economy principles. Dr Katarzyna Pala has authored scientific publications and patents. |
Barbara Bażanów Barbara Bażanów, PhD, MVD, is an associate professor in the Department of Microbiology at Wroclaw University of Environmental and Life Sciences, specializing in topics such as abortogenic viruses in horses, cytotoxicity of chemical compounds and plant extracts, epidemiological studies in various animal species, viral infections' impact on oxidative stress markers, graphene oxide and nanoceria toxicity, adenovirus 36-induced obesity, and noroviruses. She's actively involved in various research projects and grants, holds memberships in the Polish Society of Veterinary Sciences, and the Expert Group on Veterinary Medicinal Products. With over 40 scientific publications, she's contributed to successive editions of the Polish Pharmacopoeia. |
Electrochemical immunosensors demonstrate unparalleled advantages over traditional techniques, prominently characterized by their superior properties. First and foremost, these innovative sensors allow for efficient miniaturization, enhancing portability, a crucial aspect that facilitates their broader applicability.10 Their inherent cost-effectiveness, when compared to conventional counterparts, positions electrochemical biosensors as a practical and economically viable solution for diagnostic applications. Additionally, the heightened sensitivity of electrochemical immunosensors deserves explicit attention. This distinctive feature allows for the accurate detection of target analytes even at low concentrations, a critical capability, especially in scenarios like norovirus infections where the viral load is relatively minimal. Furthermore, the compatibility of electrochemical immunosensors with various matrices and their versatility in accommodating diverse biomolecules contribute to their significance. These sensors can be tailored to specific applications, offering a flexible platform for the detection of a wide array of analytes.11,12
The future of medical diagnostics, particularly in the detection of viral diseases, may relate to innovative biosensors. One of the promising elements used in the bioreceptor layer is antibodies, due to their high specificity towards the analyte to be detected.13 The basic factor in the fabrication of an electrochemical immunosensor is a proper preparation of a chemically modified electrode by the immobilization of a biologically active compound, an antibody, on its surface. One of the commonly used strategies is the covalent binding based on strong, covalent interactions between the surface of the modified electrode and the functional groups of the antibody, as it creates more stable and better oriented immunosensing platforms.14–16 Since bare electrodes do not possess proper sites for covalent bonding, it is crucial to successfully modify their surface to prepare it for protein anchoring. For this purpose, some reagents can be applied, such as glutaraldehyde, carbodiimide succinimide ester, or N-hydroxysuccinimide.17 Another often used strategy is the application of the self-assembled monolayer (SAM) method to obtain functionalized thin film rich in groups desired for the antibody immobilization.18,19 The result is the formation of highly organized SAMs due to semi-covalent binding between a sulfide group (e.g. thiols) and the noble metal surfaces (e.g. Au). Thiol functionalized antibodies possess high affinity to Au surfaces, and therefore they can be easily immobilized.20
In the advancement of electrochemical immunosensing strategies, enhancing signal amplification, sensitivity, and stability emerges as pivotal for optimizing immunosensor performance. Signal amplification within electrochemical immunosensors can be effectively realized through the incorporation of e.g. nanomaterials.21
To additionally enhance important working parameters of an electrochemical immunosensor, semi-conductive properties need to be improved. Commonly, it is achieved by the application of semi-conductive materials to the electrode surface. Nowadays, considerable focus has been devoted to nanomaterials, which could be applied in biosensors due to their exquisite sensitivity in chemical and biological sensing.22 A wide range of nanomaterials could be applied in the biosensor systems, however gold nanoparticles (AuNPs) have received great interest due to their many beneficial characteristics, such as high surface-to-volume ratio or high surface energy, which allow for a stable immobilization of biomolecules with simultaneous retention of their bioactivity.23–25 Additionally, they permit fast and direct electron transfer between a broad range of electrochemically detected analytes and electrode surface.25
In this study, we present an electrochemical biosensor consisting of a self-assembled monolayer on a gold electrode, modified with gold nanoparticles, a linker, and an antibody specific for the VP1 protein of norovirus based on a murine model. Scanning electron microscopy (SEM), electrochemical impedance spectroscopy (EIS), as well as cyclic voltammetry (CV) were applied for the visualization of each step of electrode preparation. The novelty of presented research lies in the preparation of a new, very stable matrix for the NoV detection, which could be used as a biorecognition part in a biosensorics device. Differential pulse voltammetry was used for the NoV detection and characterization of the immunosensor working parameters, such as linearity, sensitivity, selectivity, and accuracy. Described immunosensor is distinguished by good stability obtained by properly performed modification of the electrode surface, which gave very promising results in possible application as a future point-of-care device.
MNV-F CCGCAGGAACGCTCAGCAG |
MNV-R GGYTGAATGGACGGCCTG |
MNV-TPd FAM-ATGAGTGATGGCGCA |
The qPCR reaction conditions were as follows: 10 min at 45 °C and 2 min at 95 °C, followed by 45 cycles of 5 s at 95 °C and 20 s at 55 °C.
Fig. 1 (A) Schematic preparation of the immunosensor for the NoV determination; (B) proposition of the chemical groups involved in the process of creating a stable matrix. |
The electrode was rinsed with 0.1 M PBS (pH 7.0) and 10 mM pH 8.0 Tris–HCl buffers for 10 minutes each and with deionized water after each step of the modification process to wash unbound molecules from the electrode surface and remove impurities or any cross-linker byproducts.
Each step of the modification was checked using CV and EIS analysis. SEM was adopted for the visualization of the surface of the gold electrode after modification. Prepared electrode was stored at 4 °C when not in use.
For the characterization of the immunosensor's working parameters, the differential pulse voltammetry technique was used. DPV for a range of dilutions (10−1–10−10) of the NoV (1 × 10−8 TCID50 – tissue culture infective dose) was performed in the potential range of 0–0.6 V vs. Ag|AgCl and with the amplitude of 50 mV. To test the immunosensor's ability to work in open-air conditions and at room temperature, all electrochemical measurements were performed under such conditions. Samples of the NoV in the concentration range of 1 × 10−9 to 1 × 10−18 TCID50 were prepared by serially diluting norovirus samples in 0.1 M PBS buffer at pH 7.0 containing 0.1 M KCl (supporting electrolyte).
The stability test of the immunosensor was conducted 2 weeks after the electrode preparation using the CV analysis with AuE/SAM/AuNPs/EDC/NHS/Ab/BSA in the presence of 5 mM ferro/ferricyanide redox couple in 0.1 M KNO3 for 60 cycles in a potential range of −0.4–0.6 V vs. Ag|AgCl reference electrode with the scan rate of 50 mV s−1.
SEM was used to monitor the changes in surface morphology after each modification step. The first step led to the formation of SAMs with a fairly diverse morphology, i.e. the surface was dotted with spheroidal and randomly distributed clusters (few micrometers in size) embedded in the L-cysteine layer (Fig. 2a). For comparison, the AuNPs deposited on the raw gold substrate formed a generally uniform layer with scattered very fine agglomerates and relatively large and flat clusters with a diameter of several micrometers (Fig. 2b). Combining these two steps together resulted in a denser, than in Fig. 2a, surface coverage with several micrometer nodules with a flocculent and (probably) porous structure (Fig. 2c). Addition of EDC/NHS layer caused that the surface after SAM/AuNPs modification was covered with a quite differentiated layer (Fig. 2d). Namely, the nodules were covered (the smaller ones) or partially embedded (the larger ones) in a fairly smooth and continuous EDC/NHS layer. Outside the EDC/NHS clusters, the layer was discontinuous, of varying thickness, and granular in structure (Fig. 2d). Immobilization of the antibody (Fig. 2e) resulted in fine granularity in places where the EDC/NHS layer was the thinnest. In addition, few, flat clusters with sizes exceeding 100 micrometers were visible – probably formed after dropping the antibody onto the surface.
Fig. 2 SEM images of surface morphology of the electrode modified with: SAM (a), AuNPs (b), SAM/AuNPs (c), SAM/AuNPs/EDC/NHS (d), SAM/AuNPs/EDC/NHS/antibody (e). |
Impedance spectra recorded for all samples with a modified surface are characterized by a similar shape that slightly differs from that of unmodified Au electrode (Fig. 3a–c). In the Nyquist plot (Fig. 3a), one time constant can be distinguished in the medium frequency range. A further increase in impedance, starting from a few Hz towards mHz, can be interpreted in two ways. On the one hand, it may be a fragment of the second semicircle. On the other hand, this fragment is almost perfectly inclined (for the first three samples only) at an angle of 45° and may indicate the existence of diffusion constraints. Regardless of the interpretation, the impedance characteristics are similar (in shape) to those in the literature.29–31 While the presence of a second time constant is hard to see on the Nyquist plot (Fig. 3a), the second maximum of the phase angle could be found on the Bode plot at about 0.1 Hz (Fig. 3c). It is especially visible for the electrodes after modification. Apart from that, all samples reach rather high impedances (Fig. 3b).
Fig. 3 Nyquist (a) and Bode (b and c) representations of the impedance spectra recorded for unmodified and modified gold electrodes at EOC after 30 min of exposure in a solution of 0.1 M KNO3 with 5 mM [Fe(CN)6]3−/4−. Solid fitting lines refer to Model 2 from Fig. 4b. |
The use of the simplest model with one time constant containing an open Warburg element connected in series after the resistor was initially considered (Fig. 4a). In this circuit the physical sense of the elements could be as follows: R1 – electrolytic solution resistance, CPE1 and R2 – the capacitance and resistance of the charge transfer reaction, Wo – Warburg diffusion element, where: R reflects diffusion impedance, T reflects time of diffusion of the particle through layer thickness and P is the Wo exponent.
Fig. 4 Electric equivalent circuits used for fitting the experimental impedance spectra recorded for both modified (a) and unmodified (b) gold electrodes. |
Quite good fit (chi square ca. 4.3 × 10−5–2.5 × 10−4) and low residual errors were obtained. Fitting results, shown only for exploratory and illustrative purposes, are summarized in Table 1, without delving into the physical meaning of the elements.
Sample | R 1 (Ω cm2) | CPE1–T (Ω−1 cm−2 sP) | CPE1–P | R 2 (kΩ cm2) | W o–R | W o–T | W o–P |
---|---|---|---|---|---|---|---|
a Values in brackets are the % error. | |||||||
AuE | 269 (0.7) | 4.90 × 10−7 (0.8) | 0.93 (0.1) | 44.77 (1.5) | 2.79 × 106 (4.5) | 164 (9.7) | 0.49 (0.3) |
AuE + SAM | 254 (0.5) | 7.49 × 10−7 (3) | 0.94 (0.4) | 7.44 (4) | 1.16 × 106 (4.5) | 118 (9.6) | 0.51 (0.3) |
AuE + SAM + AuNPs | 238 (0.5) | 7.89 × 10−7 (3.8) | 0.94 (0.5) | 8.01 (10) | 1.95 × 106 (4.3) | 124 (9.5) | 0.49 (0.3) |
AuE + SAM + AuNPs + Ab | 226 (0.7) | 6.10 × 10−7 (1) | 0.90 (0.2) | 304.9 (3.6) | 3.42 × 106 (32) | 268 (81) | 0.49 (2.3) |
Fitting was repeated, but the circuit was modified to the one consisting of two time constants connected in parallel (Fig. 4b). Using this physical model, the best fit (chi square ca. 3–9 × 10−5) and low residual errors were obtained in this work. Fitting results are summarized in Table 2, and the fitting lines visible on Nyquist and Bode plots in Fig. 3a–c are related directly to these results. In this circuit, the physical sense of the elements was adopted, according to 29–31, as follows: R1 – electrolytic solution resistance, CPE1 and R2 – the frequency response of the charge transfer process, while CPE2 and R3 – the frequency response of the diffusion process. Normally, if P = 0.8–0.9, then CPE2 would describe some real capacitance. However, here P ∼ 0.5 so this constant phase element behaves rather like a Warburg element. This assumption is supported by the previous fitting using the circuit from Fig. 4a.
Sample | R 1 (Ω cm2) | CPE1–T (Ω−1 cm−2 sP) | CPE1–P | R 2 (kΩ cm2) | CPE2–T (Ω−1 cm−2 sP) | CPE2–P | R 3 (MΩ cm2) |
---|---|---|---|---|---|---|---|
a Values in brackets are the % error. | |||||||
AuE | 269 (0.3) | 4.94 × 10−7 (0.9) | 0.93 (0.1) | 45.95 (1.9) | 4.35 × 10−6 (0.3) | 0.50 (0.6) | 27.5 (22) |
AuE + SAM | 252 (0.8) | 8.14 × 10−7 (1.8) | 0.93 (0.2) | 8.98 (2.3) | 9.49 × 10−6 (0.7) | 0.53 (0.4) | 2.91 (6) |
AuE + SAM + AuNPs | 237 (0.2) | 8.46 × 10−7 (1.5) | 0.94 (0.2) | 11.24 (3.7) | 5.32 × 10−6 (0.2) | 0.51 (0.3) | 5.04 (4.6) |
AuE + SAM + AuNPs + Ab | 224 (0.2) | 6.29 × 10−7 (0.5) | 0.90 (0.1) | 377.2 (1.5) | 3.89 × 10−6 (1.9) | 0.60 (1.9) | 2.49 (5.5) |
Table 2 shows that the deposition of SAM layer reduced the resistance of R2 five times, below 10 kΩ cm2, but more importantly, reduced the resistance R3 by almost one order of magnitude. If one associates R2 with electron transfer and R3 with diffusion constraints, then their large decrease would confirm that one has achieved the goal of this sub-step of modification. The next step, the deposition of AuNPs on the surface, caused only a slight increase in both resistances (Table 2). It can therefore be confirmed that surface functionalization with gold nanoparticles also facilitates electron transfer between detected analytes and Au electrode surface, in accordance with the literature.25 A significant change was caused by the immobilization of the antibody. Although R2 has increased more than thirty times, R3 remained at a low level, in the order of 2.5 MΩ cm2 (Table 2). This means that although the electron transfer resistance has increased (because the antibody inhibited electron transfer), the diffusion constraints have decreased slightly. In addition, the CPE1–P parameter slightly decreased, which would suggest greater heterogeneity (distribution of reaction rates) within the surface. All these findings would confirm that the antibodies have been successfully grafted onto the surface of the modified electrode.
Using this physical model, a CV was also used to characterize the gold electrode after each assembly step. As can be observed in Fig. 5, the cyclic voltammogram of ferri/ferrocyanide redox couple demonstrated reversible behavior on a bare Au electrode, with a peak-to-peak separation of c.a. 30 mV (Fig. 5, gold curve). When AuE was modified by the immobilization of SAM and AuNPs onto its surface, the peak current increased noticeably (Fig. 5, green curve) because of an exquisite electron transfer performance of AuNPs.32 As expected, the immobilization of the biorecognition element for the NoV detection (the antibody) onto the electrode surface caused a current signal decrease, with peak-to-peak separation c.a. 60 mV (Fig. 5, black curve). The covalent assembly of the antibody onto the electrode could form a closely packed film which passivates the electrode, hindering the transfer of electrons between the redox probe and the electrode. This phenomenon was observed in numerous other studies.30,33,34
Results described above corroborate that all molecules were successfully used to modify the AuE surface, and the immunosensor could be applied for the VP1 determination.
Furthermore, the scan rate and the correlation between the current of anodic/cathodic peak were also investigated at every step during the immunosensor fabrication. CV has been performed in the same potential range at different scan rates (10–300 mV s−1). As can be observed in Fig. 6a, current signals obtained from both cathodic and anodic peaks (Ipc, Ipa, respectively) increase along with the scan rates. In combination with the results obtained from the relationship of Ipc and Ipa values vs. square root of scan rates (v1/2), presented in Fig. 6b, which shows good linearity (R2 = 0.995 for Ipa and R2 = 0.997 for Ipc), it may suggest that the electrochemical process of the interface for the immunosensor was a diffusion-controlled process.35 Because of that, results obtained in the next step are only from the tested VP1 protein of NoV.
As can be seen in Fig. 7a, obtained DPV peaks were inversely proportional to the exposed concentrations of NoV, due to the fact that binding of the VP1 protein of NoV to the antibody causes an increase in resistance and therefore a decrease in current values.30 The corresponding calibration curve, shown in Fig. 7b, represents good linearity in the investigated range (1 × 10−9–1 × 10−18 TCID50) with R2 = 0.992. The detection limit (LOD) was the lowest concentration of virus detected by presented method – 1 × 10−18 TCID50.
Fig. 7 (a) Representative DPV-scans for different NoV dilutions in the potential range of 0–0.6 V vs. Ag/AgCl; (b) linear relationship between the current and NoV dilution. |
Regeneration of the immunosensor was done by soaking the platform in 200 mM KSCN solution36 for 4 min and then washing it with doubly distilled water and PBS solution for 30 seconds.
Some examples of electrochemical immunosensors and biosensors for NoV detection can be found in recent literature. Hong et al. reported biosensor based on concanavalin-A (Con-A) immobilized onto a gold electrode to detect human NoV, applying an electrochemical transducer.31 The developed biosensor gave a response of NoV detection after approx. 1 h and presented good parameters of sensitivity and selectivity. Accuracy test provided in lettuce gave a very good response with a detection limit of 60 genomic copies per mL. Another electrochemical approach presented by Wang et al. was based on the utilization of a modified aptamer in the biorecognition part for murine norovirus detection (as an alternative to human NoV).37 Gold electrode was modified by a thiolated aptamer AG3 and specific binding to murine NoV. The researchers applied the cyclic voltammetry method to characterize the aptamer binding and square wave voltammetry (SWV) was employed to check the NoV capture. Such immunosensor gave a sensitive and fast response and was simple to prepare. Chand and Neethirajan described a microfluidic chip consisting of screen-printed carbon electrodes and polydimethylsiloxane for the electrochemical detection of human NoV.38 Silica magnetic beads were used to fill the chip for concentrating the virus samples and also for the filtration of clinical samples. The biorecognition part was prepared as follows: carbon electrode was modified with gold nanoparticles and thiolated streptavidin, onto which the human NoV-specific aptamers labelled with ferrocene (as redox probes) and biotin were immobilized. Scientists conducted their research using differential pulse voltammetry and obtained the detection limit equal to 100 pM.
The reproducibility of the Au/SAM/AuNPs/EDC/NHS/Ab/BSA immunosensor for the NoV determination was tested by the DPV method by checking the current of the same concentration of NoV (1 × 10−8 TCID50) with four, modified in the same way, gold electrodes. The average current was calculated as 1554 μA. The results exhibited a 5.12% relative standard deviation (RSD), showing very good reproducibility.
Good, long-term stability is one of the issues biosensors are facing. The stability was tested after storing the immunosensor at 4 °C (humid conditions) for 2 weeks after all experiments. The stability test was carried out using the CV method (Fig. 9). The average electrical current decreased in comparison with the original signal by approx. 4%, which shows that the constructed immunosensor is characterized by very high stability, possibly resulting from a very strong and well-prepared modified electrode for the Ab immobilization.
Western blot is used in research to separate and identify proteins. In scientific studies of human as well as mouse norovirus, the western blot technique is frequently used.
The created sensor detected the VP1 protein of mouse norovirus, and to confirm the presence of this protein in the sample, a number of dilutions were tested by western blot technique. In the experiment, 3 well visible bands and one faintly visible band were observed (Fig. 11). All striations are located at 70 kDa, which corresponds to the size of the VP1 protein.
Footnote |
† PhD Student in the 4th Edition of the Implementation Doctorate Programme—Ministry of Education and Science. |
This journal is © The Royal Society of Chemistry 2024 |