Perovskite-based electrochemiluminescence analysis of H2O2

Ziyi Jia , Hui Zhang, Yuxin Chen, Yuan Fang, Junnan Zhang and Shanwen Hu*
Department of Health Inspection and Quarantine, School of Public Health, Fujian Medical University, Fuzhou, Fujian 350122, P. R. China. E-mail: shanwenhu@fjmu.edu.cn

Received 18th May 2024 , Accepted 12th June 2024

First published on 19th June 2024


Abstract

The detection of hydrogen peroxide (H2O2) represents an extensive requirement across various domains, including food, environmental, and medical fields. This study introduces a highly sensitive technique for the quantification of H2O2, integrating the electrochemiluminescence properties of perovskite with bio-catalyzed precipitation. A water-soluble perovskite-based electrochemiluminescence (ECL) biosensing interface was constructed, wherein H2O2 catalyzes a precipitation reaction that leads to the formation of an insoluble precipitate on the electrode surface. This occurrence effectively quenches the electrochemiluminescence signal of the perovskite, thus facilitating the quantitative detection of H2O2. The modified perovskite demonstrated excellent ECL performance, offering a stable signal source, while the bio-catalyzed precipitation reaction significantly amplified the quenching effect, thereby enhancing detection sensitivity. This strategy exhibits excellent stability and sensitivity, presenting a promising method for the detection of hydrogen peroxide, which holds great potential for applications in various fields.


1. Introduction

Hydrogen peroxide (H2O2), a widely employed oxidizing and disinfecting agent, is crucial to the pharmaceutical industry, healthcare, and food manufacturing. Additionally, it serves as a byproduct in various enzymatically catalyzed metabolic processes. Precise measurement of H2O2 levels is vital in diagnosing infections, monitoring wound healing, or evaluating the efficacy of disinfection procedures. Moreover, in the food and pharmaceutical industries, the detection of H2O2 can reveal the efficiency of cleaning and sterilization processes, ensuring product safety and quality. Within environmental analysis, measuring H2O2 aids in water quality assessment, pollution source identification, and monitoring oxidative stress in aquatic ecosystems. Notably, as a by-product of human cellular metabolism, endogenous H2O2 is a natural cellular process product,1 which can become harmful if its levels exceed a certain threshold. Excessive levels of H2O2 can lead to oxidative stress, causing cellular damage and potentially contributing to the development of various diseases. Therefore, accurate monitoring of H2O2 is significant for diverse biological processes.

Approaches for detecting hydrogen peroxide primarily include colorimetry, fluorescence, and ECL. Anteneh F. Baye and colleagues have employed nitrogenous environments for the carbothermal reduction of ZnFe double hydroxide and glucose-derived carbon to fabricate Fe3O4–Fe0/Fe3C nanozymes, constructing a colorimetric sensor for H2O2 that offers high selectivity and ultra-sensitivity.2 Di Cheng and associates reported on CuCo2O4's excellent electrocatalytic activity towards H2O2, establishing an electrochemical sensor based on CuCo2O4 that demonstrated a linear range of detection from 10.0 µM to 8.9 mM and a sensitivity of 94.1 µA mM−1 cm−2.3 Among these methods, colorimetry, with its low cost, convenience, and independence from large-scale instrumentation, is suited for the onsite instantaneous detection of hydrogen peroxide, despite its limited detection range and dissatisfactory detection limits. While fluorescence offers a lower detection limit, its results are susceptible to interference from intrinsic fluorescence backgrounds. ECL (electrochemiluminescence), combining advantageous properties such as good controllability, high electrochemical sensitivity, rapid response, and high specificity, alongside easy operation and ultra-high sensitivity, has been widely applied for H2O2 detection. In ECL detection, the luminescent material and the recognition reaction constitute two crucial factors determining sensitivity. Wang and coworkers' use of a graphene–cadmium sulfide (G–CdS) nanocomposite material reacting with H2O2 produced a strong and stable ECL, thereby enhancing the ECL intensity by approx. 4.3 times and significantly lowering the initiation potential, with excellent reproducibility and long-term stability.4 Wang et al. have developed an ECL biosensor for high-sensitivity detection of H2O2 in living cells. The biosensor is based on CdZnSeS quantum dots (QDs) and ordered mesoporous carbon (OMC). The study demonstrates the potential and reliability of the biosensor in diagnostic applications and physiological research.5 The team of Han, Jee Noon investigated the ECL of luminescent phenols on mesoporous platinum electrodes under neutral conditions. In the presence of hydrogen peroxide, the ECL reaction of luminescent phenols is more intense and exhibits a linear relationship with good reproducibility.6 Liu have developed a non-enzymatic ECL sensor based on single-stranded DNA (ssDNA)/g-C3N4 nanosheets. The sensor exhibits high peroxidase-like activity, which catalyses the ECL reaction of luminescent phenol–H2O2 on a glass carbon electrode after modification. This amplifies the luminescence signal and reduces the detection limit.7

Enhancing the sensitivity of H2O2 detection is paramount for early detection and accurate quantification, and numerous efforts have been devoted to this issue. One approach is to construct an enhanced signal related to the target molecules, such as using more efficient electrode materials in electrochemical sensors. Perovskite compounds have garnered increasing attention for their unique ECL performance. Specifically, halide perovskites, particularly all-inorganic cesium lead halide nanocrystals (CsPbX3, X = Cl, Br, I), have attracted widespread interest due to their adjustable band gaps, narrow emission bandwidths, excellent thermal stability, superior carrier transport properties, and high photoluminescence (PL) quantum yields. These exceptional physical and optical properties of perovskite nanomaterials, especially their ultrahigh PLQYs, render them potential novel ECL luminophores. The pioneering study utilizing perovskites for electrochemical detection of hydrogen peroxide was published by Shimizu in 1996, demonstrating that electrodes based on carbon-loaded La0.6Ca0.4Ni0.7Fe0.3O3 could produce the highest performance potentiometric H2O2 sensors.8 Research by He, J employed 800 nm PMMA microspheres to synthesize 3DOM-SmCoO3 perovskite oxide electrodes for H2O2 detection, not only overcame various drawbacks associated with enzyme-based and precious metal-based sensors but also achieved lower costs and higher speeds.9 Wang, XT, and colleagues utilized a sol–gel method to prepare two differently structured layered perovskite nanocrystals, modified on bare glassy carbon electrodes (GCE), assembling a highly efficient enzyme-free electrochemical sensor for H2O2, achieving trace detection through its high selectivity and stability.10 Wang, HY's team fabricated a non-enzymatic electrochemical sensor for H2O2 detection based on nickel-based rare-earth perovskite LaNiTiO3 nanoparticles, which exhibited good catalytic performance and detection advantages for H2O2 in a 0.1 mol L−1 NaOH solution.11 Different types of perovskite are developed to improve the electrochemistry properties for better analysis performance. Cao et al. have developed a self-enhanced LCPB-OAm ECL system with high-quality CsPbBr3 perovskite nanocrystals as novel ECL emitters, for the bioanalyses of hydrogen peroxide based on inhibition.12 Results of Wang et al. shed the light on the electrochemically switchable H2O2 ECL sensing application of the high-throughput microwave-synthesized IPQDs.13 Efforts have also been devoted to explore the electrochemical process between perovskite and H2O2. Chen studied the electrochemical redox and charge transfer natures of RbxCs1−xPbBr3 NCs in an ECL method, indicating the Rb0.2Cs0.8PbBr3 NCs have a higher ECL intensity than the CsPbBr3 NCs. The tunable ECL properties of the RbxCs1−xPbBr3 NCs are helpful to design novel ECL emitters in H2O2 ECL analysis.14 Qiu et al. unveiled the interfacial electrochemiluminescence behavior of lead halide perovskite nanocrystals and H2O2.15 Huang's team explored hydrogen peroxide involved anodic charge transfer and electrochemiluminescence of all-inorganic halide perovskite CsPbBr3 nanocrystals in an aqueous medium.16

Although there are already numerous methods available for constructing highly sensitive ECL systems using perovskite, there are still need to improve the sensitivity of detecting hydrogen peroxide. In this study, we comprehensively considered both the ECL system and the recognition process of target molecule. On the one hand, we synthesized CTAB-modified three-dimensional perovskites via a hot injection method, enhancing the water solubility of perovskites for more stable electrode modification, and on the other hand, we utilized hydrogen peroxide to induce a precipitating reaction with a high quenching efficiency, to construct a quantitative detection strategy for hydrogen peroxide.

2. Experiment

2.1 Perovskite synthesis

The synthesis of CTAB-PNC was carried out using a classic hot-injection method with slight modifications to the original protocol.17,18 Precursor synthesis: 0.22 g of cesium carbonate (Cs2CO3) was added to a single-neck flask containing 10 mL of octadecene (ODE) and 0.65 mL of oleic acid (OAc). The mixture was heated to 120 °C under a nitrogen atmosphere for 20 minutes, then the temperature was raised to 150 °C and maintained for another 20 minutes. After stopping the reaction, the precursor was kept sealed to minimize air contact and heated to complete dissolution before use. Synthesis of CsPbBr3 perovskite: 0.104 g of lead bromide (PbBr2) was added to a three-neck flask containing 10 mL of octadecene, 1 mL of oleic acid, and 1 mL of oleylamine mixture. The mixture was heated to 120 °C under vacuum and held for 1 hour, then the temperature was raised to 170 °C. After adding 0.8 mL of the precursor, the mixture was kept for 5 seconds then quickly cooled in an ice bath. The crude solution was centrifuged at 8000 rpm for 10 minutes to collect the precipitate, which was then dispersed in n-hexane.

2.2 GO–HRP composite synthesis

To maintain the activity of horseradish peroxidase, the pH of graphene was first adjusted to 7.4. 11 mg of EDC (1-(3-dimethylaminopropyl)-3-ethylcarbodiimide) and 15 mg of NHS (N-hydroxysuccinimide) were mixed with 1 mg mL−1 graphene solution (pH = 7.4) and stirred for 60 minutes. Then, 30 µL of 10 mg mL−1 HRP solution was added dropwise to the stirred graphene solution and stirred at a speed of 150 rpm for 12 hours. After the reaction was completed, it was stored at 4 °C for later use.

2.3 ECL sensing

2.3.1 Preparation of electrodes. Glassy carbon electrodes were polished with 0.3 µm and 0.05 µm alumina slurries, respectively. The polished glassy carbon electrodes were then ultrasonically cleansed in 50% nitric acid solution, anhydrous ethanol, and pure water, then dried with nitrogen gas. A solution of 5 mM potassium ferricyanide in 0.1 M KCl was used as the electrolyte to scan the cyclic voltammogram (scan range from −0.2 V to 0.6 V at a scan rate of 50 mV s−1). The electrochemical impedance was scanned in the range from 0.1 Hz to 1000 kHz with an AC amplitude of 0.005. 2.5 µL of the perovskite dispersion was dropped onto the polished glassy carbon electrode and dried at room temperature to get the perovskite-modified electrode. A three-electrode system was used with the glassy carbon electrode as the working electrode, a platinum wire electrode as the counter electrode, and an Ag/AgCl electrode as the reference electrode.
2.3.2 Construction of ECL biosensing interface. 2.5 µL of the perovskite solution was dropped onto the electrode and air-dried overnight at room temperature before activating with NHS. The prepared GO–HRP composite was then dropped onto the activated electrode and reacted overnight at 4 °C. After the reaction was complete, the electrode was kept aside for further use. The perovskite electrode with HRP was immersed in 10× PBS solution containing 0.2 M potassium persulfate as the electrolyte to measure electrochemical luminescence, scanning the electrode potential from 0 to −2.45 V, while the photomultiplier tube (PMT) voltage remained at 380 V and the scan rate was 0.1 V s−1, producing an initial signal. A mixture of 1 mM 4-CN (4-chloro-1-naphthol) and different concentrations of H2O2 was prepared. 10 µL of the mixture was applied to the HRP-carrying perovskite electrode for a certain period to measure the electrochemical luminescence signal.

3. Results and discussions

3.1 Principle of the ECL analysis of H2O2

The principle of ECL detection in this study is depicted in Fig. 1. Initially, three-dimensional perovskite modified with cetyltrimethylammonium bromide (CTAB–CsPbBr3) was synthesized using a hot-injection method. The pristine CsPbBr3 suffers from poor photo-stability and water-stability due to surface defects. The introduction of CTAB passivates the surface defects of CsPbBr3 by forming a stable self-assembled monolayer on the perovskite surface,19,20 significantly enhancing its photo-stability and water-stability. The synthesized perovskite quantum dots are first drop-cast onto a glassy carbon electrode, which is crucial for generating the ECL signal. To immobilize horseradish peroxidase onto the electrode, a covalent complex of horseradish peroxidase and carboxylated graphene oxide (GO–HRP) was prepared in advance via the carbodiimide reaction. GO–HRP was then immobilized onto the activated electrode modified with CTAB–CsPbBr3. In the presence of hydrogen peroxide and 4-CN, the HRP on the electrode catalyzes the oxidation of 4-CN by hydrogen peroxide, generating insoluble and insulating benzo-4-chlorohexadienone products on the electrode surface. This results in a shading effect that influences the ECL intensity of the immobilized CTAB–CsPbBr3 on the electrode surface, causing a decrease in the ECL signal negatively related with H2O2 concentrations.
image file: d4ra03652b-f1.tif
Fig. 1 Illustration of ECL detection for hydrogen peroxide.

3.2 Characterization of the CTAB–CsPbBr3

Appropriate surface modification can improve the physicochemical properties of nanomaterials, thus enhancing the materials' testing performance.21,22 TAs shown in Fig. 2A, the crystal structure of CTAB–CsPbBr3 under the transmission electron microscope (TEM) exhibits regular rectangular shapes, with a size of approximately 15 nm. To verify the passivation effect of CTAB on the surface of CsPbBr3, Fourier Transform Infrared Spectroscopy was employed to characterize the surface ligands of the synthesized CTAB-modified perovskite.20 Fig. 2B shows absorption bands at 1540 cm−1 and 1400 cm−1 for the CTAB–CsPbBr3 sample, owing to the asymmetric and symmetric stretching vibrations of the carboxylate groups, indicating the complexation of oleate anions on the surface of CsPbBr3 NCs.23 The spectrum of CTAB-modified CsPbBr3 NCs exhibited strong signals at 2850 cm−1 and 2920 cm−1, attributed to the symmetric and asymmetric stretching vibrations of CH2 and CH3, respectively, elucidating the interaction of oleic acid and oleylamine ligands with the perovskite crystals. Absorption peaks at 911 cm−1 and 962 cm−1 corresponding to the stretching vibrations of C–N+ are also visible, signifying typical absorption bands of CTAB. Notably, the spectrum of CTAB–CsPbBr3 features a distinct absorption peak at 1635 cm−1, corresponding to the asymmetric deformation of amino groups.
image file: d4ra03652b-f2.tif
Fig. 2 (A) The TEM image of CTAB–CsPbBr3; (B) the FTIR spectrometer image of CTAB (blue line) and CTAB–CsPbBr3 (green line).

To investigate the effect of CTAB on CsPbBr3, fluorescence lifetimes of both CTAB–CsPbBr3 and CsPbBr3 were tested and compared. The time-resolved PL decay curves in Fig. 3 were fitted with a tri-exponential decay function using a nonlinear fitting algorithm, which yielded good agreement with the experimental data (R2 all above 0.995). The average fluorescence lifetime was calculated using the following formula:

image file: d4ra03652b-t1.tif


image file: d4ra03652b-f3.tif
Fig. 3 (A) Time-resolved PL decay of CsPbBr3; (B) time-resolved PL decay of CTAB–CsPbBr3.

As shown in Fig. 3, the PL decay rate of the CTAB-modified perovskite was slower compared to CsPbBr3, extending the average fluorescence lifetime from 2.03 ns to 4.82 ns. This result demonstrates that CTAB passivation enhanced the stability of CsPbBr3. The three lifetimes, τ1, τ2, and τ3, obtained from the fitting curve using a tri-exponential decay function, represent the intrinsic exciton relaxation, the interaction between excitons and phonons, and the interaction between excitons and defects.24 In Table S1, the τ1, τ2, and τ3 of CsPbBr3 were 2.03 ns, 8.96 ns, and 71.38 ns, respectively; whereas for CTAB–CsPbBr3, τ1, τ2, and τ3 were 4.80 ns, 19.87 ns, and 122.87 ns, respectively. Surface defects on the CsPbBr3 nanocrystal materials not only shorten the PL decay lifetime but also decrease the luminous efficiency of the nanocrystals.25 The surface modification of CTAB–CsPbBr3 exhibited an extended fluorescence lifetime, reducing non-radiative recombination and thereby enhancing the luminous efficiency of CsPbBr3.

3.3 Characterization of the electrode

The characterization was performed using differential pulse voltammetry (DPV) to measure the electrochemical signals of the perovskite-modified electrode with and without the presence of potassium persulfate. The DPV curves in Fig. S2 show the responses of the perovskite-modified electrode in potassium persulfate solution and in PBS solution without potassium persulfate. As demonstrated, a distinct cathodic peak appears for the perovskite-modified electrode in the presence of potassium persulfate solution; in contrast, no significant peak is observed in the absence. The DPV results confirm that the addition of the co-reactant potassium persulfate alters the electrochemical process of the perovskite-modified layer, which promotes electron transfer of modified electrode, thereby generating a significant peak.

Recording electrochemical impedance spectroscopy (EIS) graphs of various modified electrodes is an effective method to study electron transfer efficiency. To demonstrate the occurrence of precipitation reactions on the electrode surfaces, EIS measurements were conducted for differently modified electrodes in a potassium ferricyanide solution containing 0.1 M KCl. Within the impedance spectra, the high-frequency region correlates with rapid charge transfer processes in the electrochemical system, whereas the low-frequency region pertains to slower charge transfer processes, with the semicircular shape at low frequencies representing interface phenomena that restrict charge transfer.26 As shown in Fig. 4A, the impedance spectrum exhibits an increasing semicircle with each successive layer of modification on the electrode surface, indicating the successful assembly of horseradish peroxidase (HRP) and the formation of insoluble precipitate benzo-4-chlorohexadienone.27 The increase in resistance is attributed to the poor conductivity of the enzyme and the inhibition of electron transfer by the gradually increasing resistive effects of the insoluble precipitate layer.28,29


image file: d4ra03652b-f4.tif
Fig. 4 Electrical characterization of modified electrodes. (A) EIS spectra and (B) CV curves of various modified electrodes in 0.1 M KCl solution containing ferricyanide (a–d: bare GCE, CTAB–CsPbBr3/GCE, GO–HRP/CTAB–CsPbBr3/GCE, H2O2/4-CN/CTAB–CsPbBr3/GCE). (C) CV curves and (D) ECL spectra of CTAB–CsPbBr3/GCE (curve a), GO–HRP/CTAB–CsPbBr3/GCE (curve b), and GO–HRP/CTAB–CsPbBr3/GCE (curve c) after incubation with 4-CN and H2O2 in 0.2 M potassium persulfate solution.

Cyclic voltammetry (CV) is an important method for validating electrochemiluminescence sensors. Experiments were performed in a potassium ferricyanide solution with 0.1 M KCl, and the CV test results are presented as indicated in Fig. 4B. Compared to the bare electrode (curve a), the redox peak current of the perovskite-modified electrode significantly decreased, confirming the successful modification with ECL material. Subsequently, the current response decreased after immobilizing GO–HRP on the electrode surface, due to the insulating properties of HRP affecting electron transfer at the electrode surface. Finally, in the presence of both hydrogen peroxide and 4-CN, HRP on the electrode surface catalyzed the oxidation of 4-CN by hydrogen peroxide to produce a precipitate, resulting in a continuous decrease in peak current response. Both electrochemical impedance and cyclic voltammetry outcomes confirm the occurrence of the precipitation reaction on the electrode surface.

To assess the ECL performance of the sensor and verify the impact of precipitation reaction on the perovskite signal, experiments were conducted using differently modified electrodes in a 0.2 M potassium persulfate solution. Initially, CV curves were collected (Fig. 4C), where the electrode solely modified with CTAB–CsPbBr3 exhibited a strong cathodic peak in the solution containing the co-reactant due to the facilitated charge transfer process. Subsequently, upon immobilizing GO–HRP on the electrode, a layer was formed obstructing the electrode surface, which restricted the charge transfer rate, resulting in a slight decrease in the CV curve peak. Finally, when the GO–HRP/CTAB–CsPbBr3/GCE was reacted with H2O2 in the presence of 4-CN, a significant decrease in the CV curve peak was observed, indicating the oxidation of 4-CN at the electrode surface forming an insoluble precipitate, thereby inhibiting electron transfer.30

ECL curves of different electrodes in PBS solution containing potassium persulfate were collected. As shown in Fig. 4D, the ECL signal of CTAB–CsPbBr3/GCE was remarkably significant, demonstrating the strong ECL property of perovskite in the presence of the co-reactant. With the immobilization of GO–HRP, the ECL signal slightly decreased due to the increased resistance at the electrode surface, which slowed down the electron transfer. When GO–HRP/CTAB–CsPbBr3/GCE reacted with H2O2 in the presence of 4-CN, the formation of an insoluble precipitate inhibited charge transfer, leading to a rapid decrease in ECL signal.31 These results collectively prove that the precipitation reaction on the electrode was successfully completed as expected, ultimately leading to a notable negative correlation signal.

3.4 Condition optimization

Fig. S3 respectively illustrates the effects of scan rate and scan potential range on the electrochemiluminescence intensity of the perovskite-modified electrode. As shown in Fig. S3A, at very low scan rates, due to the slower rate of charge transfer at the electrode surface, fewer potassium persulfate molecules participate in the reaction, leading to a lower number of electrons and holes injected into CsPbBr3, thus resulting in low and unstable electrochemiluminescence efficiency. With the increase in scan rate, more potassium persulfate molecules are involved in the reaction, enhancing the efficiency of the charge injection process, thereby yielding stable and higher electrochemiluminescence signals. However, when the scan rate exceeds 0.1 V s−1, the rate of the electrochemical reaction may surpass the recombination rate of electrons and holes on the perovskite surface, leading to ineffective recombination of electrons and holes within the crystal and consequently diminishing the electrochemiluminescence efficiency and signal strength.

As depicted in Fig. S3B, when the scan potential range is too narrow, the speed of the electrochemical reaction is low, thus limiting the quantity of electrons and holes injected into CsPbBr3, resulting in a weak ECL signal due to the limited formation of excitons in CsPbBr3 NC*. With an increase in the scan potential range, the rate of charge transfer accelerates, enhancing the amount of electrons and holes injected into CsPbBr3, and subsequently increasing the formation of CsPbBr3 NC* products, leading to a gradual increase in the electrochemiluminescence signal. On the other hand, an excessively wide scan potential range may lead to unstable intermediates or side reactions, adversely affecting the electrochemiluminescence signal. Following optimization of the scan rate and scan voltage, the optimal scan rate was determined to be 0.1 V s−1 with a scan voltage range from 0 to −2.45 V.

In order to achieve optimal analytical performance, the study conducted an optimization of the concentrations of HRP and 4-CN, as well as the reaction time involved. HRP plays a pivotal role in the precipitation reaction that occurs on the electrode surface. During the synthesis of GO–HRP, products synthesized with various concentrations of HRP were used. As demonstrated in Fig. S3C, the GO–HRP synthesized using an ultimate concentration of 300 µg mL−1 of HRP, when immobilized on the electrode, exhibited the most effective catalysis of the H2O2-mediated oxidation of 4-CN, resulting in the greatest decrease in ECL signal and the highest quenching efficiency.

With the GO–HRP synthesized using the optimum HRP concentration immobilized on the electrode, the concentration of 4-CN was varied. Fig. S3D indicates that at a 4-CN concentration of 1 mM, the most significant decrease in ECL signal was observed, suggesting the most pronounced effect of the precipitation reaction on the ECL signal.

Finally, at room temperature, H2O2 and 4-CN were added to the electrolyte solution, and the variations in the ECL signal of the GO–HRP/CTAB–CsPbBr3/GCE were monitored over a reaction time of 0–40 minutes. As indicated by Fig. 5 A, the ECL signal showed a pronounced decreasing trend over the time period from 0 to 17 minutes; after 17 minutes, the ECL signal gradually stabilized.


image file: d4ra03652b-f5.tif
Fig. 5 (A) Optimization of reaction time for detecting H2O2 by electrochemical luminescence method; (B) standard curves obtained by the detection of different concentrations of H2O2; (C) ECL-time response curve of H2O2 sensor after 17 consecutive scanning cycles; (D) ECL sensor specificity and selectivity (where H2O2 concentration was 0.35 µM; the concentration of other interfering substances was 3.5 µM.).

3.5 ECL detection performance

Under the optimized conditions elucidated above, the sensing performance of the ECL sensor was evaluated. Initially, various concentrations of H2O2 were detected. As shown in Fig. 5B, an increase in the concentration of H2O2 led to an increase in the precipitate reaction products on the electrode, which in turn enhanced the quenching effect on the perovskite ECL, resulting in a more pronounced decrease in the ECL signal. Within the range of 0.05 µM to 0.5 µM, a favorable linear relationship between ΔECL intensity and H2O2 concentration was observed. By calculating the difference in ECL signal intensity before and after the reaction with H2O2, a linear regression equation was constructed for ΔIECL against H2O2 concentration, with the equation being ΔIECL = 0.24213C + 0.11211 (R2 = 0.997). The detection limit (LOD) was determined to be 50 nM.

Additionally, assessments of stability and selectivity are indispensable. To evaluate the stability, the prepared electrode was continuously scanned for 17 cycles in a potassium persulfate solution. No significant variation was noted in the ECL sensor signal over 17 consecutive cycles, with the relative standard deviation (RSD) calculated at 1.64%, indicating good stability, as illustrated in Fig. 5C. And the long-time stability was demonstrated in Fig. S4. For selectivity, potential interferents such as dopamine, ascorbic acid, and glucose were used. The concentration of H2O2 was set at 0.35 µM, whereas the concentrations of the other interfering substances were established at 3.5 µM each. As depicted in Fig. 5D, the ECL responses to the interfering substances showed no significant deviation from the ECL signal of the blank sample. However, upon the addition of H2O2 at a concentration tenfold lower than that of the interfering substances, the sensor still demonstrated a strong response, signifying outstanding selectivity for H2O2.

3.6 Sample validation of the approach

To further evaluate the capability of the ECL sensor in complex samples, the method was applied to the detection of H2O2 in milk samples under optimal conditions using the standard addition method. As shown in Table 1, the recovery rates ranged between 98% and 106%, with RSD values below 4%. These results indicate a potential application in food analysis.
Table 1 Sample detection results of H2O2
Sample Spiked (µM) Detected (µM) Recovery (%) RSD (n = 3, %)
1 0.07 0.069 98.5 1.7
2 0.16 0.170 106 3.9
3 0.36 0.369 103 2.6


4. Conclusions

In conclusion, perovskite-based ECL approach for the detection of hydrogen peroxide was proposed. A water-soluble perovskite was used to construct a sensitive ECL interface on electrode, and the H2O2-catalyzed precipitation reaction, resulting in the formation of an insoluble precipitate on the electrode surface and quenching the ECL signal, thereby enabling precise detection, offering a novel approach for the monitoring of hydrogen peroxide, which holds significant potential for applications in food safety, environmental monitoring, and other fields.

Data availability

The data that support the findings of this study are available on request from the corresponding author.

Author contributions

Ziyi Jia and Hui Zhang: proposing and conducting experiments, writing original draft. Yuxin Chen: investigation. Junnan Zhang and Yuan Fang: samples analysis and investigation. Shan-Wen Hu: writing-review & editing, supervision, project administration. All authors discussed the results and reviewed the manuscript.

Conflicts of interest

The authors declare no competing financial interest.

Acknowledgements

This work was supported by the Undergraduate Innovation and Entrepreneurship Training Program of Fujian Medical University (S202310392001) and Undergraduate Innovation and Entrepreneurship Training Program, School of Public Health, Fujian Medical University (xy202210014).

Notes and references

  1. L. P. Zong, L. Y. Ruan, J. J. Li, R. S. Marks, J. S. Wang, S. Cosnier, X. J. Zhang and D. Shan, Fe-MOGs-based enzyme mimetic and its mediated electrochemiluminescence for in situ detection of H2O2 released from Hela cells, Biosens. Bioelectron., 2021, 184, 113216 CrossRef CAS.
  2. A. F. Baye, H. T. Nguyen and H. Kim, Fe0/Fe3C-assisted Fe3O4 redox sites as robust peroxidase mimics for colorimetric detection of H2O2, Sens. Actuators, 2023, 377, 133097 CrossRef CAS.
  3. D. Cheng, T. Wang, G. X. Zhang, H. M. Wu and H. Mei, A novel nonenzymatic electrochemical sensor based on double-shelled CuCo2O4 hollow microspheres for glucose and H2O2, J. Alloys Compd., 2020, 819, 153014 CrossRef CAS.
  4. K. Wang, Q. A. Liu, X. Y. Wu, Q. M. Guan and H. N. Li, Graphene enhanced electrochemiluminescence of CdS nanocrystal for H2O2 sensing, Talanta, 2010, 82, 372–376 CrossRef CAS.
  5. L. Wang, X. H. Shi, Y. F. Zhang, A. A. Liu, S. L. Liu, Z. G. Wang and D. W. Pang, CdZnSeS quantum dots condensed with ordered mesoporous carbon for high-sensitive electrochemiluminescence detection of hydrogen peroxide in live cells, Electrochim. Acta, 2020, 362, 137107 CrossRef CAS.
  6. J. N. Han, J. Jang, B. K. Kim, H. N. Choi and W. Y. Lee, Detection of hydrogen peroxide with luminol electrogenerated chemiluminescence at mesoporous platinum electrode in neutral aqueous solution, J. Electroanal. Chem., 2011, 660, 101–107 CrossRef CAS.
  7. Z. J. Liu, L. Wang, P. F. Liu, K. R. Zhao, S. Y. Ye and G. X. Liang, Rapid, ultrasensitive and non-enzyme electrochemiluminescence detection of hydrogen peroxide in food based on the ssDNA/g-C3N4 nanosheets hybrid, Food Chem., 2021, 357, 129753 CrossRef CAS.
  8. Y. Shimizu, H. Komatsu, S. Michishita, N. Miura and N. Yamazo, Sensing characteristics of hydrogen peroxide sensor using carbon-based electrode loaded with perovskite-type oxide, Sens. Actuators, B, 1996, 34, 493–498 CrossRef CAS.
  9. J. He, W. Zhou, J. Sunarso, X. M. Xu, Y. J. Zhong, Z. P. Shao, X. J. Chen and H. Zhu, 3D ordered macroporous SmCoO3 perovskite for highly active and selective hydrogen peroxide detection, Electrochim. Acta, 2018, 260, 372–383 CrossRef CAS.
  10. X. T. Wang, B. Liu, D. R. Kong, Z. Y. Zhang, X. F. Zhang, Z. P. Deng, L. H. Huo and S. Gao, T- and T -type layered perovskite Ln2CuO4 nanocrystals for enhanced sensing detection of hydrogen peroxide, J. Alloys Compd., 2022, 911, 165037 CrossRef CAS.
  11. H. Y. Wang, X. L. Zhu, M. L. Xin and Y. H. Xu, A Nonenzymatic Sensor for H2O2 Detection Based on Rare-earth Perovskite LaNiTiO3 Containing Ni, Chinese J. Anal. Chem., 2014, 42, 847–852 CAS.
  12. Y. Cao, W. L. Zhu, L. L. Li, Z. Y. Zhang, Z. X. Chen, Y. H. Lin and J. J. Zhu, Size-selected and surface-passivated CsPbBrsub3 perovskite nanocrystals for self-enhanced electrochemiluminescence in aqueous media, Nanoscale, 2020, 12, 7321–7329 RSC.
  13. Y. F. Wang, T. Y. Chen, C. X. Huang, Y. R. Wang, J. Wu and B. Sun, Electrochemically switchable electrochemiluminescent sensor constructed based on inorganic perovskite quantum dots synthesized with microwave irradiation, J. Electroanal. Chem., 2020, 867, 114181 CrossRef CAS.
  14. L. Chen, Q. Kang, Z. Li, B. Zhang, G. Z. Zou and D. Z. Shen, Tunable electrochemiluminescence properties of CsPbBr3 perovskite nanocrystals using mixed-monovalent cations, New J. Chem., 2020, 44, 3323–3329 RSC.
  15. L. H. Qiu, L. H. Lin, Y. P. Huang, Z. W. Lai, F. M. Li, S. Y. Wang, F. Y. Lin, J. F. Li, Y. R. Wang and X. Chen, Unveiling the interfacial electrochemiluminescence behavior of lead halide perovskite nanocrystals, Nanoscale Adv., 2019, 1, 3957–3962 RSC.
  16. Y. Huang, X. Y. Long, D. Z. Shen, G. Z. Zou, B. Zhang and H. S. Wang, Hydrogen Peroxide Involved Anodic Charge Transfer and Electrochemiluminescence of All-Inorganic Halide Perovskite CsPbBr3 Nanocrystals in an Aqueous Medium, Inorg. Chem., 2017, 56, 10135–10138 CrossRef CAS.
  17. Y. X. Chen, X. Y. Zhong, Q. L. Yang, H. P. Chen, N. Hao and S. W. Hu, A perovskite-based electrochemiluminescence aptasensor for tetracycline screening, Luminescence, 2024, 39, e4717 CrossRef CAS.
  18. G. Nedelcu, L. Protesescu, S. Yakunin, M. I. Bodnarchuk, M. J. Grotevent and M. V. Kovalenko, Fast Anion-Exchange in Highly Luminescent Nanocrystals of Cesium Lead Halide Perovskites (CsPbX3, X = Cl, Br, I), Nano Lett., 2015, 15, 5635–5640 CrossRef CAS.
  19. S. H. Chi, S. Yang, Y. S. Sun, Z. Y. Pang, X. X. Sun, L. Fan, F. Y. Wang, X. Y. Liu, M. B. Wei, J. H. Yang, N. N. Yang and L. L. Yang, Synthesis and Improved Photoluminescent Properties and Stability of Bromine‐Rich CsPbX3 Nanocrystals Via using CTAB as Additive, Cryst. Res. Technol., 2022, 57, 2200051 CrossRef CAS.
  20. L. Yang, Y. Liu, P. F. Tan, Y. Lu, Q. G. Ding and J. Pan, Anchoring TiO2@CsPbBr3 on g-C3N4 nanosheet for enhanced photocatalytic degradation activity of tetracycline hydrochloride, Diamond Relat. Mater., 2023, 133, 109727 CrossRef CAS.
  21. S. Ntakirutimana, W. Tan and Y. Wang, Enhanced surface activity of activated carbon by surfactants synergism, RSC Adv., 2019, 9, 26519–26531 RSC.
  22. W. Tan, T. Gao and Y. Wang, Influence of Surface Potential on the Capacitive Performance of the TiO2 Thin-Film Electrode with Different Crystalline Forms, Langmuir, 2020, 36, 3836–3842 CrossRef CAS.
  23. Y. T. Cai, P. J. Zhang, W. H. Bai, L. Lu, L. Wang, X. Chen and R. J. Xie, Synthesizing Bright CsPbBr3 Perovskite Nanocrystals with High Purification Yields and Their Composites with In Situ-Polymerized Styrene for Light-Emitting Diode Applications, ACS Sustain. Chem. Eng., 2022, 10, 7385–7393 CrossRef CAS.
  24. Y. F. Zhao, R. R. Yang, W. Wan, X. P. Jing, T. Wen and S. Ye, Stabilizing CsPbBr3 quantum dots with conjugated aromatic ligands and their regulated optical behaviors, Chem. Eng. J., 2020, 389, 124453 CrossRef CAS.
  25. Z. T. Wang, X. Yuan, L. Han, J. M. Li, X. Z. Zhang, J. Hua, J. Wang, J. L. Zhao and H. B. Li, A-Site FA+ Doping-Enhanced Photoluminescence Efficiency and Photostability of Mn-Doped Perovskite Nanocrystals, J. Phys. Chem. C, 2022, 126, 3582–3590 CrossRef CAS.
  26. L. Hou, Y. L. Cui, M. D. Xu, Z. Q. Gao, J. X. Huang and D. P. Tang, Graphene oxide-labeled sandwich-type impedimetric immunoassay with sensitive enhancement based on enzymatic 4-chloro-1-naphthol oxidation, Biosens. Bioelectron., 2013, 47, 149–156 CrossRef CAS.
  27. W. W. Zhao, P. P. Yu, J. J. Xu and H. Y. Chen, Ultrasensitive photoelectrochemical biosensing based on biocatalytic deposition, Electrochem. Commun., 2011, 13, 495–497 CrossRef CAS.
  28. B. R. Adhikari, M. Govindhan and A. C. Chen, Carbon Nanomaterials Based Electrochemical Sensors/Biosensors for the Sensitive Detection of Pharmaceutical and Biological Compounds, Sensors, 2015, 15, 22490–22508 CrossRef CAS.
  29. L. Zhang, Y. B. Ying, Y. B. Li and Y. C. Fu, Integration and synergy in protein-nanomaterial hybrids for biosensing: Strategies and in-field detection applications, Biosens. Bioelectron., 2020, 154, 112036 CrossRef CAS.
  30. N. Hao, X. Zhang, Z. Zhou, J. Qian, Q. Liu, S. B. Chen, Y. Zhang and K. Wang, Three-dimensional nitrogen-doped graphene porous hydrogel fabricated biosensing platform with enhanced photoelectrochemical performance, Sens. Actuators, B, 2017, 250, 476–483 CrossRef CAS.
  31. X. Wang, X. C. Liao, B. J. Zhang, S. Y. Chen, M. M. Zhang, L. S. Mei, L. Zhang, X. W. Qiao and C. L. Hong, Fabrication of a novel electrochemical immunosensor for the sensitive detection of carcinoembryonic antigen using a double signal attenuation strategy, Anal. Chim. Acta, 2022, 1232, 340455 CrossRef CAS.

Footnotes

Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d4ra03652b
Ziyi Jia and Hui Zhang contributed equally to this work.

This journal is © The Royal Society of Chemistry 2024