Abstract

Continuous flow polymerase chain reaction (CFPCR) devices are compact reactors suitable for microfabrication and the rapid amplification of target DNAs. For a given reactor design, the amplification time can be reduced simply by increasing the flow velocity through the isothermal zones of the device; for flow velocities near the design value, the PCR cocktail reaches thermal equilibrium at each zone quickly, so that near ideal temperature profiles can be obtained. However, at high flow velocities there are penalties of an increased pressure drop and a reduced residence time in each temperature zone for the DNA/reagent mixture, that potentially affect amplification efficiency. This study was carried out to evaluate the thermal and biochemical effects of high flow velocities in a spiral, 20 cycle CFPCR device. Finite element analysis (FEA) was used to determine the steady-state temperature distribution along the micro-channel and the temperature of the DNA/reagent mixture in each temperature zone as a function of linear velocity. The critical transition was between the denaturation (95 °C) and renaturation (55 °C–68 °C) zones; above 6 mm s−1 the fluid in a passively-cooled channel could not be reduced to the desired temperature and the duration of the temperature transition between zones increased with increased velocity. The amplification performance of the CFPCR as a function of linear velocity was assessed using 500 and 997 base pair (bp) fragments from λ-DNA. Amplifications at velocities ranging from 1 mm s−1 to 20 mm s−1 were investigated. The 500 bp fragment could be observed in a total reaction time of 1.7 min (5.2 s cycle−1) and the 997 bp fragment could be detected in 3.2 min (9.7 s cycle−1). The longer amplification time required for detection of the 997 bp fragment was due to the device being operated at its enzyme kinetic limit (i.e., Taq polymerase deoxynucleotide incorporation rate).

Article information

Article type
Paper
Submitted
07 May 2004
Accepted
13 Aug 2004
First published
19 Oct 2004

Lab Chip, 2004,4, 638-645

Rapid PCR in a continuous flow device

M. Hashimoto, P. Chen, M. W. Mitchell, D. E. Nikitopoulos, S. A. Soper and M. C. Murphy, Lab Chip, 2004, 4, 638 DOI: 10.1039/B406860B

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements