Determination of tantalum from tantalum oxide nanoparticle X-ray/CT contrast agents in rat tissues and bodily fluids by ICP-OES

Abstract

Accurate and precise means for quantifying Ta in tissues, bodily fluids, and bone is critical in understanding anticipated safety-profiles for tantalum oxide (TaO) nanoparticle-based X-ray/CT diagnostic imaging agents and has prompted the development of three digestion methods which are the focus of this work. Spike recovery studies were employed to evaluate bias, precision, and sample matrix effects in the quantification of Ta in (1) liver, blood and femur by microwave-assisted digestion, (2) urine by open-beaker digestion, and (3) carcass, liver and feces by dry-ash digestion. All analyses were performed with inductively coupled plasma optical emission spectrometry (ICP-OES). Spike recoveries were 98.5–102.3% for all biological matrices except femur (91.6%); however, a modified version of the original microwave digestion procedure improved the recovery of Ta in femur to 103.8%. Precision of spike recovery reported as one standard deviation ranged from 0.1 to 3.8% for within-run and from 0.5 to 3.3% for overall recovery depending on the tissue type and digestion method. Limit of detection (LOD) was 0.006 to 6 μg Ta per g and limit of quantification (LOQ) was 0.02 to 20 μg Ta per g depending on the method. The presented methods were applied to the determination of Ta in liver, kidney, spleen and carcass from an in vivo TaO nanoparticle retention study, and the results for percent injected dose (% ID) of Ta retained are given.

Graphical abstract: Determination of tantalum from tantalum oxide nanoparticle X-ray/CT contrast agents in rat tissues and bodily fluids by ICP-OES

Supplementary files

Article information

Article type
Technical Note
Submitted
07 Nov 2015
Accepted
19 Apr 2016
First published
22 Apr 2016

J. Anal. At. Spectrom., 2016,31, 1311-1317

Author version available

Determination of tantalum from tantalum oxide nanoparticle X-ray/CT contrast agents in rat tissues and bodily fluids by ICP-OES

J. M. Crowder, N. Bates, J. Roberts, A. S. Torres and P. J. Bonitatibus, J. Anal. At. Spectrom., 2016, 31, 1311 DOI: 10.1039/C5JA00446B

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements