Volume 205, 2017

The effect of STM parameters on tip-enhanced Raman spectra

Abstract

In this work, we evaluate the dependence of tip-enhanced Raman (TER) spectra of a monolayer of thiophenol at a Au(111) electrode on the scanning tunneling microscope’s tunneling current set-point and bias voltage parameters. We find an increase of the TER intensity upon set-point increase or bias decrease as expected from a gap-distance reduction. The relations obtained follow a theoretical model considering a simple gap-distance change when tuning the mentioned parameters. We find that the value of the bias voltage affects the TER intensity to a larger extent than the current set-point. Therefore it is advisable to work in a low-bias regime when aiming for ultrasensitive TER measurements.

Associated articles

Supplementary files

Article information

Article type
Paper
Submitted
08 May 2017
Accepted
12 May 2017
First published
17 May 2017
This article is Open Access
Creative Commons BY license

Faraday Discuss., 2017,205, 233-243

The effect of STM parameters on tip-enhanced Raman spectra

N. M. Sabanés, A. Elizabeth, J. H. K. Pfisterer and K. F. Domke, Faraday Discuss., 2017, 205, 233 DOI: 10.1039/C7FD00164A

This article is licensed under a Creative Commons Attribution 3.0 Unported Licence. You can use material from this article in other publications without requesting further permissions from the RSC, provided that the correct acknowledgement is given.

Read more about how to correctly acknowledge RSC content.