Issue 9, 2018

A narrow open tubular column for high efficiency liquid chromatographic separation

Abstract

We report a great feature of open tubular liquid chromatography when it is run using an extremely narrow (e.g., 2 μm inner diameter) open tubular column: more than 10 million plates per meter can be achieved in less than 10 min and under an elution pressure of ca. 20 bar. The column is coated with octadecylsilane and both isocratic and gradient separations are performed. We reveal a focusing effect that may be used to interpret the efficiency enhancement. We also demonstrate the feasibility of using this technique for separating complex peptide samples. This high-resolution and fast separation technique is promising and can lead to a powerful tool for trace sample analysis.

Graphical abstract: A narrow open tubular column for high efficiency liquid chromatographic separation

Supplementary files

Article information

Article type
Communication
Submitted
21 Dec 2017
Accepted
15 Mar 2018
First published
15 Mar 2018

Analyst, 2018,143, 2008-2011

Author version available

A narrow open tubular column for high efficiency liquid chromatographic separation

H. Chen, Y. Yang, Z. Qiao, P. Xiang, J. Ren, Y. Meng, K. Zhang, J. Juan Lu and S. Liu, Analyst, 2018, 143, 2008 DOI: 10.1039/C7AN02065A

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements