Issue 11, 2018

Evaluation of a cooling/heating-assisted microextraction instrument using a needle trap device packed with aminosilica/graphene oxide nanocomposites, covalently attached to cotton

Abstract

A low-cost and reliable cooling/heating-assisted microextraction (CHaME) instrument was designed and fabricated for use in different modes of microextraction methods. The CHaME setup is able to cool down the sorbent and simultaneously heat the sample in a wide temperature range. Consequently, it can create a large thermal gap between the sorbent and the sample matrix, to promote the release of analytes from the sample tissue and enhance their effective trapping on the microextraction phase. The primary versions of the instrument have previously been evaluated, coupled with different modes of solid- and liquid-phase microextraction strategies. Compared with conventional microextraction systems, it is able to extract volatile organic compounds from complicated solid matrices more effectively, rapidly and without any need for a sample preparation step. In this research, the final and compact version of the CHaME instrument was fabricated and employed in a cooling/heating-assisted needle trap device (CHaME-NTD) for microextraction of polycyclic aromatic hydrocarbons (PAHs) in contaminated soil samples, prior to GC-FID determination. An aminosilica/graphene oxide nanocomposite was synthesized, covalently attached to cotton (Am-Si/GO/Cot), packed inside a needle, and applied as an effective sorbent for trapping of the analytes. The influence of experimental parameters on the extraction efficiency of the TC-NTD-GC-FID strategy was evaluated and optimized. Under the optimal conditions, linear dynamic ranges (LDRs), limits of detection (LODs), and relative standard deviations (RSDs) for the PAHs were 0.001–2.0 μg g−1, 5–38 pg g−1, and 6.2–9.8% (n = 6), respectively. The CHaME-NTD-GC-FID procedure was compared with the traditional NTD-GC-FID method. Additionally, the Am-Si/GO/Cot nanocomposite sorbent was compared with the most frequently used commercial sorbents. The results demonstrated the remarkable performance of the CHaME-NTD procedure and the Am-Si/GO/Cot composite sorbent. The developed setup was also used for the extraction and determination of PAHs in contaminated soil samples, through the CHaME-NTD-GC-FID procedure.

Graphical abstract: Evaluation of a cooling/heating-assisted microextraction instrument using a needle trap device packed with aminosilica/graphene oxide nanocomposites, covalently attached to cotton

Supplementary files

Article information

Article type
Paper
Submitted
11 Jan 2018
Accepted
14 Apr 2018
First published
18 Apr 2018

Analyst, 2018,143, 2632-2640

Evaluation of a cooling/heating-assisted microextraction instrument using a needle trap device packed with aminosilica/graphene oxide nanocomposites, covalently attached to cotton

A. R. Ghiasvand, N. Heidari, S. Abdolhosseini, A. Hamdi and P. R. Haddad, Analyst, 2018, 143, 2632 DOI: 10.1039/C8AN00063H

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.