Calcium isotope determination in urine samples via the monitoring of 44CaF and 40CaF molecules by high-resolution continuum source graphite furnace molecular absorption spectrometry

Abstract

This study investigates the possibility of obtaining Ca isotopic information by monitoring the absorption spectrum of a diatomic molecule generated in a graphite furnace using a high-resolution continuum source spectrometer (HR CS GFMAS). The method proposed enables calcium isotopic analysis using CaF as the target species. The difference in wavelength between 40CaF (maximum at 628.534 nm) and 44CaF (maximum at 628.191 nm) absorption signals was found to be 337.8 pm, evidencing that both species can be determined simultaneously. The method developed relies on the determination of 40Ca and 44Ca concentrations based on calibration functions specific to each isotope and makes use of 12 μg of F as a molecule forming element without further chemical modifiers. The sensitivity of the method depends on the pH. It increases at pH values close to 2, and decreases until reaching a plateau at pH 6. The characteristic mass found for 40CaF was 0.11 μg for solutions with pH 2 and 0.18 μg for alkaline solutions. The respective limits of detection were 1 and 2 mg L−1. The method was applied to the analysis of three urine samples, spiked with different 40Ca and 44Ca amounts. Due to chlorine interference, before the analysis of urine, calcium must be separated from the samples by precipitation with ammonium oxalate. Recovery tests provided results between 95 and 109%, highlighting the potential of the method for 40Ca and 44Ca determination in tracer experiments.

Graphical abstract: Calcium isotope determination in urine samples via the monitoring of 44CaF and 40CaF molecules by high-resolution continuum source graphite furnace molecular absorption spectrometry

Article information

Article type
Paper
Submitted
06 Jul 2019
Accepted
06 Sep 2019
First published
06 Sep 2019

J. Anal. At. Spectrom., 2019,34, 2280-2287

Calcium isotope determination in urine samples via the monitoring of 44CaF and 40CaF molecules by high-resolution continuum source graphite furnace molecular absorption spectrometry

M. B. T. Zanatta, F. V. Nakadi, M. Resano and M. A. M. S. da Veiga, J. Anal. At. Spectrom., 2019, 34, 2280 DOI: 10.1039/C9JA00233B

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements