Issue 12, 2024

A mesothelin microsensor based on an embedded thionine electronic medium within an imprinted polymer on an acupuncture needle electrode

Abstract

An electrochemical microsensor for mesothelin (MSLN) based on an acupuncture needle (AN) was constructed in this work. To prepare the microsensor, MSLN was self-assembled on 4-mercaptophenylboronic acid (4-MPBA) by an interaction force between the external cis-diol and phenylboronic acid. This was followed by the gradual electropolymerization of thionine (TH) and eriochrome black T (EBT) around the anchored protein. The thickness of the surface imprinted layers influenced the sensing performance and needed to be smaller than the height of the anchored protein. The polymerized EBT was not electrically active, but the polymerized TH provided a significant electrochemical signal. Therefore, electron transfer smoothly proceeded through the eluted nanocavities. The imprinted nanocavities were highly selective toward MSLN, and the rebinding of insulating proteins reduced the electrochemical signal of the embedded pTH. The functionalized interface was characterized by SEM and electrochemical methods, and the preparation conditions were studied. After optimization, the sensor showed a linear response in the range of 0.1 to 1000 ng mL−1 with a detection limit of 10 pg mL−1, indicating good performance compared with other reported methods. This microsensor also showed high sensitivity and stability, which can be attributed to the fine complementation of the imprinted organic nanocavities. The sensitivity of this sensor was related to the nanocavities used for electron transport around the AuNPs. In the future, microsensors that can directly provide electrochemical signals are expected to play important roles especially on AN matrices.

Graphical abstract: A mesothelin microsensor based on an embedded thionine electronic medium within an imprinted polymer on an acupuncture needle electrode

Supplementary files

Article information

Article type
Paper
Submitted
29 Mar 2024
Accepted
20 Apr 2024
First published
22 Apr 2024

Analyst, 2024,149, 3309-3316

A mesothelin microsensor based on an embedded thionine electronic medium within an imprinted polymer on an acupuncture needle electrode

Y. Zhang, X. Kong, H. Guo, J. Wang and Z. Yin, Analyst, 2024, 149, 3309 DOI: 10.1039/D4AN00490F

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.