Issue 9, 2024

Potential of external (in air) particle induced gamma-ray emission method for the preparation of isotopic composition of boron in-house reference standard in boron carbide matrix for quality control work

Abstract

The quality control (QC) of an analytical method for any analytical data measurement depends directly on the matrix matching standard and certified reference material (CRM). In many cases, obtaining a matrix match standard and an RM or CRM as a control sample in sufficient amounts for QC analysis is difficult. It is also not advised to use a CRM as a standard for routine analysis owing to its high cost and availability in lower quantities. In view of this, in-house reference standards/materials for isotopic compositions of boron (IC, 10B/11B atom ratio) in natural and enriched boron carbide matrix were prepared in the present work. These in-house reference standards/materials will be helpful for method validation, quality control and quality assurance in the IC determination of boron carbide samples using particle induced gamma-ray emission (PIGE) method. A total of five different isotopic compositions, namely natural (19.8 atom% of 10B) and 30, 40, 50 and 67 atom% of 10B-enriched B4C, were prepared in sufficient quantities (0.8 kg each). Homogeneity and stability tests were performed for the prepared materials with respect to their isotopic composition by analysing direct powder samples using the external PIGE method, and they were found sufficiently homogeneous and stable for the property value for use as in-house reference standards. Assigned values of isotopic compositions were determined through external PIGE using both powder and pellet samples. The obtained values of the isotopic compositions are 0.247(1), 0.427(2), 0.662(3), 0.991(4), and 2.028(7) for the natural and 30, 40, 50 and 67 atom% of 10B-enriched boron carbide, respectively.

Graphical abstract: Potential of external (in air) particle induced gamma-ray emission method for the preparation of isotopic composition of boron in-house reference standard in boron carbide matrix for quality control work

Article information

Article type
Paper
Submitted
25 Apr 2024
Accepted
24 Jun 2024
First published
08 Jul 2024

J. Anal. At. Spectrom., 2024,39, 2278-2289

Potential of external (in air) particle induced gamma-ray emission method for the preparation of isotopic composition of boron in-house reference standard in boron carbide matrix for quality control work

S. W. Raja, R. Acharya, A. D. Sonawane, T. S. R. C. Murthy and S. Majumdar, J. Anal. At. Spectrom., 2024, 39, 2278 DOI: 10.1039/D4JA00156G

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.