O. Strohmeierab,
S. Keila,
B. Kanata,
P. Pateld,
M. Niedrigd,
M. Weidmanne,
F. Hufertf,
J. Drexlerg,
R. Zengerleabc and
F. von Stetten*ab
aHSG-IMIT-Institut für Mikro- und Informationstechnik, Georges-Koehler-Allee 103, 79110 Freiburg, Germany. E-mail: felix.von.stetten@hsg-imit.de
bLaboratory for MEMS Applications, IMTEK-Department of Microsystems Engineering, University of Freiburg, Georges-Koehler-Allee 103, 79110 Freiburg, Germany
cBIOSS – Centre for Biological Signalling Studies, University of Freiburg, 79110 Freiburg, Germany
dCentre for Biological Threats and Special Pathogens 1, Robert Koch-Institute, Nordufer 20, 13353 Berlin, Germany
eInstitute of Aquaculture, University of Stirling, Pathfoot Building, FK9 4LA Stirling, UK
fInstitute of Microbiology and Virology, Brandenburg Medical School Theodor Fontane, Campus Senftenberg, Germany
gQIAGEN Lake Constance GmbH, Jacques-Schiesser-Straße 3, 78333 Stockach, Germany
First published on 27th March 2015
We present total nucleic acid extraction from whole blood, Gram-positive Bacillus subtilis, Gram-negative Escherichia coli, and Rift Valley fever RNA virus on a low-cost, centrifugal microfluidic LabDisk cartridge processed in a light-weight (<2 kg) and portable processing device. Compared to earlier work on disk based centrifugal microfluidics, this includes the following advances: combined lysis and nucleic acid purification on one cartridge and handling of sample volumes as large as 200 μL. The presented system has been validated for logarithmic dilutions of aforementioned bacteria and viruses from various sample matrices including blood plasma and culture media and extraction of human DNA from whole blood. Recovered DNA and RNA concentrations in the eluate were characterized by quantitative PCR to: 58.2–98.5%, 45.3–102.1% and 29.5–34.2% versus a manual reference for Bacillus subtilis, Escherichia coli and Rift Valley fever virus, respectively. For extraction of human DNA from whole blood, similar results for on-disk ((10.1 ± 7.6) × 104 DNA copies) and manual reference extraction ((10.2 ± 6.3) × 104 DNA copies) could be achieved. Eluates from on-disk extraction show slightly increased ethanol concentrations of 4.1 ± 0.3% to 5.5 ± 0.2% compared to a manual reference (2.0 ± 0.5% to 3.6 ± 0.6%). The complete process chain for sample preparation is automatically performed within ∼30 minutes, including ∼15 minutes lysis time. It is amenable to concatenation with downstream modules for multiplex nucleic acid amplification as recently demonstrated for panel testing of various pathogens at the point of care.
Centrifugal microfluidics3–6 recently demonstrated its potential to automate complex process chains in nucleic acid analysis including polymerase chain reaction (PCR),7–11 nested PCR,12 digital PCR13 and isothermal recombinase polymerase amplification (RPA).14 Disadvantageously, these systems required already purified nucleic acid samples which were prepared off-chip.
Suitable process chains for upstream target-unspecific nucleic acid extraction from intact, non-lysed cells have not been presented so far. To the best of our knowledge, all presented solutions within centrifugal microfluidics are limited to perform either cell lysis15–17 or purification without chip-integrated lysis.18–21 Mechanical lysis on centrifugal microfluidic cartridges has been demonstrated using collision and friction of glass beads in a rimming flow15 or agitation of beads by an integrated magnetic element.16,17,22 Afterwards, the obtained eluates had to be post treated prior to PCR analysis. In another approach, antibody labelled beads were used for off-chip pathogen specific capturing and concentration of pathogens and subsequent on-chip thermal lysis.23 Purification of RNA and DNA from virus lysates and pre-lysed whole blood, respectively, has been demonstrated using sol–gel,20 glass beads21 or silica membranes as solid phase.18 The most comprehensive approach showed the microfluidic integration of pathogen specific DNA extraction from Hepatitis B DNA viruses and Gram-negative E. coli bacteria from a whole blood sample using anti-body coated magnetic beads. After separation, captured pathogens were thermally lysed by laser irradiation.1 However, the need for target-specific antibody coating does not allow the detection of a certain pathogen from a broad panel of causative agents.
In this work, we demonstrate a target independent, comprehensive and generic approach for total nucleic acid extraction from whole blood, Gram-positive Bacillus subtilis and Gram-negative Escherichia coli bacteria, and the RNA Rift Valley fever virus (RVFV) on a centrifugal microfluidic “LabDisk” platform. The application of magnetic beads as solid phase allows for a concentration of the nucleic acids from sample volume (200 μL) to elution volume (100 μL). DNA recovery and eluate purity of the extracted DNA and RNA were characterised by quantitative real-time PCR, residual ethanol- and spectro-photometric A260/280 measurements, respectively. Compared to our previous work, we report the following advances: (1) microfluidic integration of sample lysis prior to nucleic acid purification; (2) handling of extended sample (200 μL) and reagent volumes (up to 500 μL) comparable to standard bench top extraction protocols; and (3) validation for logarithmic dilutions of different target organisms including Gram-positive and Gram-negative bacteria, RNA viruses and human cells. The developed LabDisk cartridges can be processed in a small and light weight (<2 kg) centrifugal processing device, suitable for application at the point-of-care. After loading of sample and extraction reagents and starting of a predefined frequency protocol, no further manual interaction is required. The presented process chain for total nucleic acid extraction was designed to be combinable with downstream microfluidic modules for multiplex nucleic acid analysis by PCR7–9,12 for future sample-to-answer pathogen panel testing in one unit use cartridge. Here, the pathogen-unspecific extraction principle would allow extraction of DNA and/or RNA from a group of pathogens related to a certain disease.
Fig. 1 Schematic of the microfluidic structure for nucleic acid extraction. Radial position of the transport magnet is r1 = 42 mm and of the collection magnet r2 = 52.5 mm (see also ESI Fig. S1†). Azimuthal distance between both magnets is 42°. Sample liquid and nucleic acid extraction reagents are pipetted to the inlets (a) and the disk is rotated, transferring the liquids radially outwards (b). After a short halt in the frequency protocol and capillary priming of the siphon, the disk is rotated again and the binding buffer is merged with the lysed sample (c). |
(1) |
Fsurface = 61/3·π2/3σliquid·Vcluster1/3 | (2) |
χmag and χliquid are the magnetic volume susceptibilities of the magnetic beads and the surrounding liquid, σ is the surface tension of the surrounding liquid and (grad(B)·B) is the strength and gradient of the magnetic field.
Gram-positive Bacillus subtilis (DSM 4451, Leibniz-Institute DSMZ – German collection of Microorganisms and Cell Cultures, Germany) and Gram-negative Escherichia coli DH5αZ1 (provided by Prof. Gescher, Karlsruhe Institute of Technology, Germany) were cultivated over night (30 °C and 37 °C respectively) in LB media. Afterwards, the B. subtilis and E. coli stocks were cooled down on ice and a log10 dilution series was prepared. For preservation of the bacteria during storage at −80 °C, each dilution was mixed with glycerol (99.5%, Carl Roth, Germany) for a final glycerol concentration of 35%. The mixture was aliquoted into 250 μL subvolumes and stored at −80 °C. Cell count of the undiluted B. subtilis and E. coli glycerol stock solution was determined by plate-counting to 1.07 × 108 colony forming units (CFU) mL−1 and 1.45 × 108 CFU mL−1, respectively.
Inactivated RVFV samples with a stock concentration of 3.3 × 106 genome equivalents mL−1 (Robert Koch Institute, Germany) were diluted logarithmically in DI-water, aliquoted into 20 μL aliquots and stored at −80 °C.
For automated on-disk extraction, the whole blood sample, pre-mixed with protease was added to the sample inlet port. Lysis buffer AL (300 μL), ethanol for DNA binding (450 μL), washing buffers AW 1 and AW 2 (2× 500 μL) and RNAse free water (100 μL) were loaded to the corresponding inlets of the disk and the predefined spin protocol was started. Prestored silica coated magnetic beads served as the solid phase.
For DNA and RNA extraction from the bacterial (B. subtilis and E. coli) and viral (RVFV) targets, extraction reagents from a commercially available magnetic bead based nucleic acid extraction kit (innuPREP MP Basic Kit A, Analytik Jena AG, Germany) were used. For a single extraction experiment for B. subtilis or E. coli, a 200 μL aliquot of the bacterial stock dilution was taken from the freezer, thawed on ice and shaken for homogenisation. Each 200 μL aliquot was then split into two 100 μL portions, one for manual reference and one for automated on-disk extraction. 100 μL portions of bacterial stock dilution were then mixed with 100 μL pure LB media. RVFV samples were composed from 20 μL virus stock spiked in 180 μL blood plasma. Bacterial and viral samples were afterwards mixed with 5 μL lambda phage DNA and 20 μL Proteinase K. Manual reference extractions from these bacterial and viral samples were conducted as described for the genomic DNA extraction from whole blood but using reagents from the innuPREP MP Basic Kit (300 μL lysis buffer RL, 450 μL binding buffer RBS, 2× 500 μL washing buffer HS and LS and 100 μL RNAse free water for elution). For automated on-disk extraction, the sample, and equivalent volumes of lysis, binding, washing and elution buffer were added to the corresponding inlet of the LabDisk. Magnetic beads were already prestored in the disk as dry pellet and the pre-defined spin protocol was started.
Extracted DNA from human whole blood was quantified on a Rotor-Gene 2000 real-time PCR thermocycler (formerly Corbett Research, acquired by QIAGEN, Germany) starting with hot start (5 min/95 °C) followed by 45 cycles denaturation (30 s/94 °C) and annealing/extension (45 s/55 °C). Composition of a single PCR reaction and used primer and probe sequences are depicted in ESI Table S3 and S4.† DNA yield was quantified on a Rotor-Gene 6000 real-time PCR thermocycler (QIAGEN, Germany). Thermocycling protocol started with an initialisation step (2 min/50 °C) followed by hot-start (10 min/95 °C) and 50 cycles of denaturation (15 s/95 °C) and annealing/extension (60 s/60 °C). Thermocycling protocol for amplification of RNA from Rift Valley fever virus started with reverse transcription (30 min/50 °C) followed by hot-start (15 min/95 °C) and 45 cycles of denaturation (15 s/95 °C) and annealing/extension (60 s/60 °C).
Measurements were conducted as follows: 8.6 μL eluate were mixed with 16.4 μL of water and 25 μL of Biochain reagent A. After 20 minutes of incubation, reactions were stopped with 25 μL of Biochain reagent B. Absorption was then determined at 570 nm in a spectrophotometer (Wallac 1420 Multilable counter, Perkin Elmer, Finland) and compared to a standard curve with known ethanol concentrations.
Finally, we conducted extraction experiments with the gold standard method using QIAGEN spin columns according to manufacturer's protocol (ESI section 6†). Spin columns contain an integrated silica membrane as solid phase for DNA binding. The recovery was measured to (28.2 ± 10.7) × 104 extracted DNA copies, which is almost three times higher than the manual reference in the test tube. The higher recovery from the spin column experiments is very likely attributed to the increased available surface compared to a 20 μL silica bead suspension.
For characterization of the on-disk extraction performance, logarithmic dilutions of Gram-positive B. subtilis (1.1 × 106–1.1 × 102; 5.3 × 102 and 5.3 × 101 colony forming units per sample) and Gram-negative E. coli (1.5 × 106–1.5 × 102 CFU per sample) were prepared and subjected to automated extraction. Each experiment was conducted in replicates of three per dilution. In parallel, manual reference extractions with similar reagent volumes were conducted as described in materials and methods. The recovered number of DNA copies from manual reference extraction versus automated on-disk extraction is depicted in Fig. 4 and 5. For the investigated concentrations that were qualified by real-time PCR, recovery of extracted DNA for B. subtilis was between 58.2% and 98.5% with respect to the manual reference (calculated from real-time PCR results as depicted in ESI, section 9†). The three lowest concentrated B. subtilis dilutions, containing 5.3 × 102 CFU, 1.1 × 102 CFU and 5.3 × 101 CFU per sample yielded eluates that did not result in positive PCR amplification in all replicates and therefore were quantified by PROBIT regression analysis. It has to be considered that for PCR quantification, only 1 μL from the 100 μL total elution volume was used and thus, the number of negative PCR reactions per se increases with decreasing CFU concentration due to statistical DNA distribution in the PCR reaction. Recovery of extracted DNA for E. coli was between 45.3% and 102.1% (see ESI†). Here, only the dilution with a concentration of 1.5 × 102 CFU had to be quantified by PROBIT analysis. The deviation in performance between manual reference and on-disk extraction is comparable to whole blood extraction experiments and might also be explained by insufficient resuspension of the bead pellet consequently providing less available surface for DNA binding. This problem can be solved by either increasing the amount of bead suspension in the pellet or by improving the spin protocol for resuspension.
Fig. 4 Yield of manual DNA reference extraction (diamond) versus on-disk extraction (triangle) from B. subtilis dilutions. Pelleted magnetic beads were used for all experiments. |
Fig. 5 Yield of manual DNA reference extraction (diamond) versus on-disk extraction (triangle) from E. coli dilutions. Pelleted magnetic beads were used for all experiments. |
The RNA extraction performance was characterized using dilutions of RVFV in blood plasma samples (6.6 × 104–6.6 × 102 genome equivalents per sample). For each dilution, three independent extraction experiments were conducted on the LabDisk. In parallel, three manual reference extractions were performed. Yield of extracted RNA was afterwards determined by real-time RT-PCR for the highest two Rift Valley fever virus concentrations (6.6 × 103 and 6.6 × 104 genome equivalents). For determination of the RNA extraction yield from the lowest concentration, 6.6 × 102 genome equivalents, PROBIT regression analysis was used as explained in ESI† (Fig. 6). LabDisk based extraction yielded between 29.5% and 34.2% RNA copies compared to manual reference in the tube for those samples that were quantified by real-time RT-PCR. Compared to the extraction of DNA from bacterial pathogens, the ratio of on-disk extraction is significantly lower what might be explained by the higher sensitivity of RNA to degradation.
Fig. 6 Yield of manual RNA reference extraction (diamond) versus on-disk extraction (triangle) from Rift Valley fever virus dilutions. Pelleted magnetic beads were used for all experiments. |
On-disk [% w/v] | Manual ref. [% w/v] | |
---|---|---|
Whole blood | 4.1 ± 0.3 | 2.0 ± 0.5 |
B. subtilis | 5.5 ± 0.2 | 3.6 ± 0.6 |
E. coli | 5.3 ± 0.6 | 2.8 ± 0.7 |
RVFV | 4.7 ± 0.7 | 3.0 ± 1.6 |
Measured concentrations for manual reference extraction reveal to be lower as ethanol residuals can evaporate during a five minute air drying step after the second washing step which was not implemented in the on-disk extraction protocol. However, as only 1 μL eluate from the total volume of 100 μL is subjected to quantification in a 10 μL PCR reaction, the ethanol content is diluted well below the critical level for inhibition of 2.5%.28,29
Furthermore, the absorbance ratio at 260 nm to 280 nm (A260/280) in the eluates from whole blood extraction was measured (Nanovue plus spectrophotometer; GE lifescience, USA). Nucleic acids are regarded as pure if the A 260/280 ratio is between ∼1.7 and ∼1.9 while lower values indicate contaminations with proteins or phenols,31,32 further PCR inhibitors. Measured A 260/280 values were 2.05 ± 0.25 (n = 4) for automated on-disk extraction with bead pellet, 1.77 ± 0.18 (n = 2) for on-disk extraction with liquid beads, 1.88 ± 0.22 (n = 2) for reference extraction in tube, and 1.93 ± 0.09 (n = 5) for spin column extraction. All measured values are within the expected range. Measurements were not conducted for eluates from the bacterial and viral extraction as the required nucleic acid concentration was below the measurement range of the device.
We thoroughly characterized the system by extracting human DNA from whole blood and genomic DNA from dilutions of Gram-positive and Gram-negative bacteria in stock culture, B. subtilis and E. coli, respectively. We demonstrated that on-disk extraction and manual reference extraction performed similar when a suspension of magnetic beads is used for extraction yielding (10.1 ± 7.6) × 104 DNA copies extracted on-disk vs. (10.2 ± 6.3) × 104 DNA copies extracted by manual reference. On-disk extraction with prestored, pelleted magnetic beads yielded about 30% compared to the set-up using a suspension of magnetic beads. Therefrom, we hypothesize that available surface on the beads might be reduced after drying. On-disk extraction for B. subtilis yielded 58.2–98.5% versus manual reference. Recovery for E. coli was 45.3–102.1% compared to reference. Furthermore, we extracted RNA from dilutions of RVFV samples spiked in blood plasma with a yield of 29.5% and 34.2% compared to the reference. All on-disk extractions showed increased ethanol concentrations between 4.1 ± 0.3% and 5.5 ± 0.2% compared to manual reference extractions (2.0 ± 0.5% to 3.6 ± 0.6%).
The presented implementation of a microfluidic process chain for pathogen unspecific, total DNA and RNA extraction from various sample matrices might be concatenated with downstream microfluidic modules for PCR or RPA based multiplex amplification for full sample-to-answer panel testing of various pathogens at the point-of-care in the near future. Additionally, the LabDisk cartridge can be equipped with liquid reagent prestorage in miniature stick packs as recently presented by van Oordt et al.33
Footnote |
† Electronic supplementary information (ESI) available. See DOI: 10.1039/c5ra03399c |