D. Käsera,
T. Van Acker
b,
J. Koch
a,
B. Hattendorf
a and
D. Günther
*a
aLaboratory of Inorganic Chemistry, Department of Chemistry and Applied Biosciences, ETH Zürich, Vladimir-Prelog-Weg 1, 8093 Zurich, Switzerland. E-mail: guenther@inorg.chem.ethz.ch
bAtomic and Mass Spectrometry Research Unit (A&MS), Department of Chemistry, Ghent University, Krijgslaan 281-S12, 9000 Ghent, Belgium
First published on 20th June 2025
The composition of laser-generated aerosols produced during single spot ablation was analyzed via inductively coupled plasma time-of-flight mass spectrometry (LA-ICP-TOFMS) and compared with that of the same material deposited on the sample surface. Differences between the signals recorded from single spot ablation and the subsequent deposit analysis enabled studying and exploring non-stoichiometric processes at the ablation site. These measurements allowed for the determination of the transport efficiencies of the elements present in NIST SRM 610, which were element-specific and always higher than 85%. A mass balance of 238U+/232Th+ signal intensity ratios allowed the differentiation of laser-induced and ICP-induced elemental fractionation using only LA-ICP-TOFMS data, without the need for the off-line analysis of the collected material. All phenomena reported herein were studied using a range of parameters for single spot ablation including different laser fluences (7.4–25 J cm−2) and number of pulses per spot (100–600 pulses). The method presented here provides direct access to important parameters that influence the accuracy of quantitative analysis using LA-ICPMS, such as the amount and composition of the deposited material around the crater and information about the aforementioned transport efficiencies of individual elements. In addition, the proposed analysis and calculation procedure allowed to distinguish between laser- and ICP-induced elemental fractionations for different LA and ICPMS operating conditions. Knowledge about these parameters will help understand the ablation behavior of different matrices and may therefore improve non-matrix matched quantification using LA-ICPMS.
While fractionation effects were shown to be reduced by improving the laser ablation systems used, non-stoichiometric detection and analysis still remained a challenge, with fractionation indices deviating from 1 for various elements.4,5 These studies indicated that the fractionation index describes processes affecting the detection of element intensity ratios and heavily depends on operating conditions. For example, a comparison of fractionation indices for different crater aspect ratios (i.e. depth-to-diameter ratio) and by varying the ablation rates of individual laser ablation systems showed highly variable fractionation effects.9
Another indicator that has been proposed for instrument optimization is the 238U+/232Th+ signal intensity ratio, which must be close to one for the NIST SRM 610 glass standard.10 This is because both elements are present in a very similar concentration in the reference material and have almost equal ionization energies and only a small mass difference. Sensitivity ratios higher than one can be attributed to incomplete aerosol vaporization, which also affects elemental fractionation in LA-ICPMS.11 Incomplete vaporization, atomization and ionization of laser-generated particles are considered less severe for ablation in helium with 193 nm LA in comparison with 266 nm in argon.12 Using the latter, it has been shown that 150 nm is a critical size, above which the laser-generated aerosol particles of NIST SRM 610 show incomplete ionization.13 The previously observed decrease in fractionation effects while using a nanosecond 193 nm laser was then related to smaller particles and narrower particle size distribution of the aerosol generated.14 A better understanding of the conditions affecting these fractionation effects has helped establish the use of shorter wavelength lasers15 and minimize ICP-induced fractionation by adjusting the 238U+/232Th+ signal intensity ratio to their concentration ratio.10
Minimizing elemental fractionation is key to applications such as U–Pb geochronology, where elemental fractionation limits the precision and accuracy of mineral dating.15 For these studies, specific attention must be paid to the sensitivity ratios of U/Th and Pb/U.16 In this context, the aspect ratio of the crater affects the signal of uranium, whereas, lead undergoes constant material loss, independent of the crater size.17 Material losses on the sample surface also lead to non-stoichiometric composition with variations in the Pb/U and Pb/Th ratios of the deposits surrounding the ablation craters.16 This is in general agreement with the preliminary analyses of the surface condensate reported by Eggins et al.,7 demonstrating an enrichment in volatile elements compared to more refractory ones.
ICP-related fractionation effects have, however, never been clearly distinguished from laser-induced fractionation. It has been shown that the aerosol produced by the laser ablation of brass differs in composition from the bulk material.18 Together with the fact that the composition of the deposited material also differs from the bulk composition of the analyzed samples,7,16 this implies that elemental fractionation occurs even during the sampling process. In order to distinguish ICP-related and laser-ablation-dependent elemental fractionation, a better understanding of fractionation that occurs during laser ablation sampling is necessary. By analyzing the composition of the aerosol generated during single spot ablation and the deposited material under identical ICPMS operating conditions, the degree of fractionation at the ablation site can be evaluated.
This study thus employs single spot ablation of the NIST SRM 610 glass standard, followed by re-ablation of the entire deposit around the crater, based on a low-fluence sampling approach (≤1 J cm−2). Signals monitored during single spot ablation and the ablation of the deposits were recorded under identical ICPMS operating conditions. This approach enabled mass balance calculations to determine individual transport efficiencies of all monitored isotopes. By using ICP-TOFMS, the signal intensity ratios of the isotopes of interest could be monitored quasi-simultaneously, allowing a comprehensive description of the composition and spatial distribution of the deposits. These experiments were carried out under various ablation conditions and by varying the laser fluence, number of pulses and crater diameter to investigate their impact on the amount and composition of the individual materials transported and deposited during laser ablation.
Using this experimental setup, we were able to show that laser-induced fractionation reproduces the pattern of fractionation indices published by Fryer in 1995.4 The Laser-Induced Fractionation Indices (LIFI) introduced here indicate that even though the fractionation effects can be reduced significantly, they are associated with laser–matter interactions. The results allow us to distinguish laser-induced and ICP-induced fractionations. Both sources of fractionation can be isolated based on the relative element abundance determined in the deposits and the initial spot.
Laser ablation | ICP-TOFMS | |||
---|---|---|---|---|
Spot ablation | Imaging | |||
Repetition rate | 10 Hz | 5 Hz | Make-up gas flow rate (Ar) | 0.7–0.8 L min−1 |
Spot size | 24–44 μm | 32 μm | Carrier gas flow rate (He) | 1.3–1.4 L min−1 |
Step size | — | 10 μm | Sampling depth | 3.2–4.5 mm |
Fluence | 7.4–25 J cm−2 | ≤1 J cm−2 | Plasma power | 1550 W |
Image size | Auxiliary gas flow rate | 0.8 L min−1 | ||
Raster size/μm | 1000 × 1000 (44 μm craters) | Plasma gas flow rate | 16 L min−1 | |
500 × 500 (24 μm craters) |
The mass fractions and isotope abundance values used in the data evaluation of NIST SRM 610 were obtained from the GeoReM Database.21,22
The area considered for signal integration from the deposits was adjusted to a region of interest (ROI) where the deposition was visible using light microscopy. The signals recorded for each line scan were corrected for co-ablation using the average value obtained from the same line upstream of the carrier gas flow in the ROI to avoid potential contributions from deposits caused by gas dynamics. Despite the use of a lower fluence in the imaging experiments, co-ablation occurred especially for the more volatile elements in NIST SRM 610 (e.g. Pb). In particular, in experiments with a low amount of deposited material (low pulse number and fluence), the correction required for co-ablation can thus be substantial and may affect the results.
N(i)A + N(i)D = N(i)tot | (1) |
This assumption also implies negligible transport losses in comparison with the deposit losses, which was already suggested when helium was initially introduced as ablation environment.7 Transport losses in the tubing between the LA cell and the ICP unit were also considered insignificant because aerosol throughput is independent of tubing length.25 In addition, no significant differences in transport efficiency have been reported between different cell designs. Therefore, we did not consider further losses in the ablation cell and its fittings to the tubing.26 Nevertheless, the ablation cell used in this study was optimized to minimize aerosol dispersion using high velocity but still laminar gas flows,20 making substantial losses downstream of the ablation site highly unlikely. Similarly, the material losses on the crater walls are not considered as these particles are observed only on the upper part of craters with extreme depth-to-diameter ratios.9
Another assumption is that the material removed from the crater represents the exact composition of the sample, i.e., stoichiometric, which allows the expansion of eqn (1) to isotope number ratios, as presented in eqn (2), for isotopes x and y in relation to their concentration (c) ratio.
![]() | (2) |
The respective number of atoms, however, cannot be detected directly because the element-specific ion yields27 and mass-dependent transmission28 manifest in the ICPMS mass spectra. The number of ions (I) detected is thus always lower than the number of atoms delivered to the ICP. This is accounted for by introducing the isotope-specific factor P, as shown in eqn (3):
N(i) × P(i) = I(i) | (3) |
Assuming that factor P(i) is equal for both single spot ablation and deposits analysis (under the same ICPMS conditions), these equations allow the estimation of the transport efficiency (TE(i)) for each isotope from the measured signals:
![]() | (4) |
Notably, the P factors cancel out since they are assumed identical for the single spot and deposit ablation experiments, and the temporal drift was corrected by bracketing. As the only losses considered here are for the deposits next to the ablation site, the relative portion of the material deposited for a given isotope can also be expressed as “deposit fraction” (DF(i)), which is defined similarly to the TE.
![]() | (5) |
The laser-induced fractionation index (LIFI) then represents the relative degree of preferential deposition of the elements and can be defined as the ratio of DFs for any combination of isotopes x relative to a reference isotope y:
![]() | (6) |
As shown by the equations above, element-specific sensitivity cannot be derived directly, but the relative factors can be calculated using eqn (1)–(3) for isotopes x and y, as shown in eqn (7).
![]() | (7) |
This expression resembles the absolute sensitivity ratio for the isotopes as it includes the entire ablated material and not just the transported fraction.
![]() | ||
Fig. 2 232Th+ raw signal intensity maps of 600-pulses craters of 44 μm diameter ablated using different fluences. |
Based on these maps, the signal intensities corrected for co-ablation were integrated across the respective ROIs to determine the deposited material fraction. As illustrated in Fig. 3, the 232Th+ and 238U+ signal intensities increased for both single spot and deposit ablation with increasing laser fluences. However, the deposit signals at fluences higher than 15 J cm−2 increased more steeply, while the slope of the signal from single spot ablation continuously decreased with increasing fluences.
Based on these data, the transport efficiencies of Pb, Th and U were calculated. As displayed in Fig. 4, the transport efficiencies of all elements decreased with increasing fluence and their losses increased with their volatility. The TEs calculated from these experiments were in the order Th > U > Pb, with values between 99% and 95% for Th, 99% and 93% for U and 95% to 90% for Pb.
As seen in Fig. 3, the change in signal intensity with increasing fluence was not equal for all investigated isotopes. While Th exhibited a relatively stronger increase than U during single spot ablation, its ion signals from the deposits were always lower. The corresponding intensity ratios 238U+/232Th+ and 208Pb+/238U+ are listed in Table 2. The variability was smaller for the single spot ablation experiments, but a decreasing trend of 238U+/232Th+ signal intensity ratio was observed at higher fluence, while only a minute increase was observed for 208Pb+/238U+. The ion signals from deposit ablation, however, indicated a substantial decrease in both ratios. The deposit signals revealed a trend towards the expected concentration ratios of the isotopes (≈1.01 for 238U+/232Th+ and ≈ 0.49 for 208Pb+/238U+),21,22 along with the depletion of the more volatile element with increasing fluence, as shown in Table 2. As for the single spot ablation experiments, on the other hand, there was no unique trend with respect to element volatility. While U/Th decreased continuously and reached a value below the concentration ratio of the isotopes, the Pb/U ratio remained almost stable (or rather appeared to increase) with fluence and was always higher than the concentration ratio in NIST SRM 610.
Fluence | 238U+/232Th+ | 208Pb+/238U+ | ||
---|---|---|---|---|
Single spot ablation | Deposit removal | Single spot ablation | Deposit removal | |
7.4 J cm−2 | 1.06 | 1.22 | 0.67 | 3.06 |
11.5 J cm−2 | 1.02 | 1.28 | 0.67 | 2.20 |
15 J cm−2 | 0.99 | 1.17 | 0.67 | 1.61 |
20 J cm−2 | 0.98 | 1.11 | 0.68 | 1.10 |
25 J cm−2 | 0.96 | 1.12 | 0.68 | 0.94 |
The signal intensity ratios (Table 3) in the 600-pulses experiment were generally comparable to the values obtained with 7.4 J cm−2 in the previous experiment (Table 2). However, the data generated with the same conditions (600 pulses, 7.4 J cm−2) differed between the experiments, highly likely because of the different operating conditions of the instrument, as the experiments were carried out on different days. During single spot ablation, the 238U+/232Th+ signal intensity ratios were lower than the ones obtained in the fluence variation experiment and dropped to values below the concentration ratio. The 208Pb+/238U+ signal intensity ratios, on the other hand, were higher than those observed in the fluence variation experiment and showed a steeper increase with the number of pulses. The intensity ratio, in this case, was also always higher than the concentration ratio. Moreover, the intensity ratios of 238U+/232Th+ in the deposits were higher than the concentration ratio and dropped below this value as the number of pulses increased. The 208Pb+/238U+ intensity ratio was always substantially higher than the concentration ratio of 0.49 and showed a decreasing trend when the number of pulses was increased beyond 200. At a lower number of laser pulses, we assume that the signals were affected to a greater extent by co-ablation correction and were, therefore, not considered.
Crater depth | 238U+/232Th+ | 208Pb+/238U+ | ||
---|---|---|---|---|
Single spot ablation | Deposit removal | Single spot ablation | Deposit removal | |
100 pulses | 1.03 | 2.46 | 0.69 | 4.04 |
200 pulses | 0.98 | 1.42 | 0.70 | 4.57 |
300 pulses | 0.94 | 1.13 | 0.71 | 4.69 |
400 pulses | 0.91 | 1.05 | 0.73 | 4.24 |
500 pulses | 0.90 | 0.94 | 0.75 | 4.12 |
600 pulses | 0.89 | 0.95 | 0.77 | 4.16 |
To investigate the effect of the depth-to-diameter ratio of the crater in more detail, the same experiment was repeated for craters with a diameter of 24 μm (Table S1†). The intensity ratios obtained from single spot ablation were lower and less variable for 238U+/232Th+, while the 208Pb+/238U+ intensity ratios were more than 10% higher and increased notably with increasing crater depth. Notably, the signal intensity ratios were similar to those obtained for a crater diameter of 44 μm with twice the number of laser pulses, which would hence have a similar depth-to-diameter ratio. The ratios obtained from deposit analyses exhibited a greater scattering, which can be attributed to the comparably low amount of the deposited material. The enrichment of the more volatile element in the deposits, however, appeared to be more pronounced in this case.
The 88Sr+/43Ca+ and 139La+/140Ce+ intensity ratios were also evaluated in detail (Table S2†), which indicated that Sr and Ca experienced very similar fractionation trends, only little variation between different crater aspect ratios and even between the single spot ablation and deposit analyses. The ratios, however, differed substantially from their concentration ratio because of the mass discrimination in ICP-TOFMS. The measured 139La+/140Ce+ signal intensity ratios, alternatively, demonstrated the relative enrichment of Ce in both deposits and single spot ablation in comparison with the corresponding concentration ratio. The enrichment decreased in both cases as the aspect ratio of the crater increased and was more pronounced in the deposits, in general. These observations are consistent with the additional +4 oxidation state of Ce and its higher volatility compared to La.29,30
The three previous experiments indicate common trends in the measured intensity ratios of 238U+/232Th+ and 208Pb+/238U+ when the aspect ratio of the laser-generated crater increases, either using higher pulse energies for the same number of pulses or by applying more laser pulses of the same fluence. For 208Pb+/238U+, this is in good agreement with all previous reports on elemental fractionation, that is, the intensity ratio obtained from single spot ablation generally increases with increasing crater aspect ratios.16 The corresponding decrease observed for 238U+/232Th+ also agrees with previous observations and can be attributed to a change in the particle size distribution, that is, a greater fraction of smaller particles are formed from deeper craters, as suggested by Kuhn et al.14
Notably, the measurements were carried out after optimizing the He carrier gas and Ar make-up gas flows in combination with the sampling depth to ensure a 238U+/232Th+ sensitivity ratio near 1 for single pulse ablation, i.e. always ablating the pristine surface of a NIST SRM 610 glass standard, which led to a substantial fraction of large particles in the aerosol. This was done to ensure that particles of all sizes could be efficiently vaporized in the ICP. As the crater aspect ratio increased during the experiments, the particle size distribution during single spot ablation changed to smaller average sizes,14 and the earlier vaporization of 238U+ vs. 232Th+ led to a stronger loss by diffusion within the ICP. At the same time, however, the deposits formed from craters with a higher aspect ratio were found to be increasingly depleted in U relative to Th even though they were all ablated from the sample surface directly, regardless of the crater aspect ratio. Yet, all 238U+/232Th+ intensity ratios determined for deposits from craters with the lowest aspect ratio were substantially higher than the concentration ratio. This implies an enrichment of Th vs. U in the aerosol transported to the ICP from the ablation site during single spot ablation. Based on this reasoning, the measured 208Pb+/238U+ intensity ratios imply that an even stronger enrichment of the more volatile element Pb occurred in the deposited material. While the intensity ratios recorded during single spot ablation were higher by 30–50% than the isotope concentration ratios in NIST SRM 610, as listed in Table 3, the deposits exhibited 8.5 to 9.5-fold higher values at low laser fluences, which decreased to approximately 2-fold at the highest fluence used.
![]() | ||
Fig. 6 Laser-induced fractionation indices normalized to 88Sr for craters generated with different fluences and 600 pulses (top) and with 7.4 J cm−2 and an increasing number of pulses (bottom). |
At first glance, it can be surprising that more volatile elements were enriched in the deposits and that their LIFI decreased with increasing crater aspect ratios. This is because their condensation is expected to occur at a later stage than that for the refractory elements, and they would thus get more diluted in the expanding plasma plume. However, Eggins et al.7 have already proposed that vapor deposition is responsible for the enrichment of more volatile elements in the vicinity of the ablation crater, as observed here.
![]() | ||
Fig. 7 Comparison of the laser-induced fractionation indices normalized to 88Sr after optimization for 238U+/232Th+ signal intensity ratios near 1 (depth and fluence) or 1.75. |
Evaluation of laser-induced fractionation is, therefore, only possible when the ICP-induced fractionation is minimized and not masking the laser-induced fractionation. In our experiments, based on the composition measured during single spot and deposit ablation, we found that laser- and ICP-induced elemental fractionation have opposite effects. The 238U+/232Th+ signal intensity ratio of 1.75 instead of near 1 was caused by the ICP because the aerosol from single spot analysis is depleted in uranium, as indicated by the deposit analysis. By definition, the LIFI is independent of ICPMS parameters. However, the distinct variation in their values may be linked to the optimization of ICP and related processes therein. Therefore, further investigations are needed to better understand ICP-induced effects on laser aerosols.
A mathematical approach to assess inter-elemental variations caused by the laser ablation process is proposed using a mass balance approach for samples with known concentrations. This method can describe the laser-induced fractionation and is independent of isotope-specific sensitivities of the ICPMS instrument used. The LIFI of the deposits obtained using this method, in fact, closely resemble the element-specific pattern reported by Fryer based on arbitrarily selected periods of the transient signal alone.4 However, the numeric values obtained by Fryer and the LIFI determined in our study cannot be compared because of the different experimental approaches and instrumentation.
Higher energy densities led to increased material deposition, whose chemical composition was closer to that of the bulk sample, which indicates that the extent of deposition is not directly related to the magnitude of laser-induced fractionation.
Our findings confirm that the aerosol is not only affected by laser-induced fractionation but also fractionates within the ICP. While the effects related to the ablation process highlight the intrinsic nature of the fractionation process linked to laser-matter interaction, ICP-induced fractionation can be minimized by adjusting the ICPMS operating conditions for a 238U+/232Th+ sensitivity ratio of one. The lack of such an optimization target can lead to significant fractionation in the ICP. However, the effects of ICP-induced fractionation are currently not well comprehended and detailed studies must be carried out.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: https://doi.org/10.1039/d5ja00173k |
This journal is © The Royal Society of Chemistry 2025 |