Molar mass measurement of a 28Si-enriched silicon crystal with high precision secondary ion mass spectrometry (SIMS)

Abstract

Several highly-enriched 28Si crystals were produced to enable a better determination of the Avogadro constant through removing the uncertainty associated with the abundance determination of the minor isotopes 29Si and 30Si. Previous determinations have used inductively coupled plasma mass spectrometry (ICP-MS), which typically requires rather large amounts of material and complex chemical pre-treatment. A secondary ion mass spectrometry (SIMS) protocol for ultra-high precision and accuracy Si isotope measurements of a 28Si-enriched crystal (AVO28) was developed in this study using a reverse-geometry ion microprobe SHRIMP RG. An ≈8 nA primary beam intensity was used to sputter the 28Si-enriched crystal. Secondary ions were measured on a combination of Faraday cup, for 28Si+, and electron multiplier for the low abundance 29Si+ and 30Si+ signals respectively. The secondary ion intensities and isotope ratios were determined following baseline correction, gain correction, tailing correction and instrumental mass fractionation correction. An internal precision of ±0.000016 (2 sm) and external precision (reproducibility) of ±0.000043 (2 s) were achieved for 29Si/28Si of spot-to-spot analyses on natural silicon (WASO 04). The Si isotope ratios of AVO28 are n(29Si)/n(28Si) = 0.00004107(63) mol mol−1 and n(30Si)/n(28Si) = 0.00000107(4) mol mol−1, 2 s, respectively. As a result, an average molar mass of the AVO28 material was determined to be 27.97696972 g mol−1 with an associated relative standard uncertainty of 1.3 × 10−8 (k = 1). This newly developed SIMS method avoids the complex chemical pre-treatment and the molar mass was in good agreement (∼10−8) with solution multi-collection-ICP-MS results.

Graphical abstract: Molar mass measurement of a 28Si-enriched silicon crystal with high precision secondary ion mass spectrometry (SIMS)

Article information

Article type
Paper
Submitted
04 Jul 2022
Accepted
21 Sep 2022
First published
22 Sep 2022

J. Anal. At. Spectrom., 2022,37, 2546-2555

Molar mass measurement of a 28Si-enriched silicon crystal with high precision secondary ion mass spectrometry (SIMS)

Y. Gao, T. Ren, I. S. Williams, T. R. Ireland, T. Long, O. Rienitz, A. Pramann, S. Wang, P. Song and J. Wang, J. Anal. At. Spectrom., 2022, 37, 2546 DOI: 10.1039/D2JA00229A

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements