Sepsis diagnosis requires development of methods to identify rare pathogen cells in small samples of human blood. Magnetic beads functionalized with pathogen-binding ligands have been used to rapidly isolate microbes from blood; however, it is commonly difficult to optically detect the captured species because the excess numbers of beads required for pathogen binding physically interfere with light transmission after they have been concentrated. Here we describe a microdevice that uses microfluidics combined with optimized magnetic field concentrators and magnetic beads coated with a generic blood opsonin to efficiently capture unknown blood pathogens and spread them into a thin layer suitable for automated optical detection. Using this device, we have been able to detect fungal pathogens in less than three hours after sample collection compared to days with current technology, and with an extremely high sensitivity (<1 cell mL−1 of human blood).
To solve the excess bead problem, we modified the microdevice design to better balance the force of the applied magnetic field gradient on the beads with fluidic drag forces so that the collected beads could be spread thinly enough to visualize the pathogens using fluorescence microscopy (Fig. 2b). We fabricated a polysulfone holder to position a magnetic flux concentrator (MFC) directly beneath a 3/4′′ neodymium cube magnet (K&J Magnetics, grade N42 magnet, magnetization 875352 A m−1) and placed it directly above the ceiling of a microfluidic channel with the same dimensions as the channel described above (Fig. 2a and b). The ceiling was made thin (300 μm) to minimize dissipation of the generated magnetic field. The shape of the MFC (5 mm wide × 1.55 mm thick × 25 mm long) was optimized in COMSOL using finite element modelling, and it was CNC-machined from a single bar of EFI alloy 79 (Ed Fagan, Inc.). The front end of the MFC was tapered to a wedge to decrease the initial magnetic field gradient spike produced by the magnet, and the remaining area of the MFC was covered by rows of parallel grooves (400 μm deep and 400 μm wide, with 800 μm center-to-center spacing) (Fig. 2c). The trailing edge of the MFC has a sharp, untapered edge to ensure capture of the magnetic beads with a strong magnetic field gradient spike. The MFC was designed to be wider than the channel width to produce an even magnetic field gradient over the entire channel width to reduce clumping on the sides of the channel. We tested a range of periodic groove sizes in the MFC and found that use of 800 μm wide grooves provided only a minimal advantage because there were relatively few places where gradient spikes formed along the length of the device and the beads still formed relatively dense piles. With the 200 μm spacing, we found that the bead capture efficiency was poor and finite element model simulations showed that the beneficial gradient spikes were smaller with the shallower grooves and dropped to low levels before penetrating the height of the channel. Thus, the 400 μm spacing was the best compromise to create an even bead distribution, while still ensuring bead capture in the channel.
We also modified the internal surface of the middle 25 mm section of the microfluidic channel by micromolding a repeating array of closed packed, circular wells (50 μm diameter × 20 μm deep, spaced 100 μm apart with each subsequent row 50 μm offset from the other) (Fig. 2d). We tested a range of well depths (10 through 80 μm deep) and found that bead capture was the most reliable with the 20 μm deep wells. The MFC created places where the magnetic field lines diverged outwards periodically, resulting in production of high magnetic field gradient zones that promoted efficient bead capture and alignment within the circular microwells (Fig. 2d).
The magnetic force acting on the superparamagnetic beads is proportional to the local magnetic field gradient, so we produced a finite element model (COMSOL) of the magnetic field (∇B) in the various device configurations (magnet with or without MFC) (Fig. 3a). This analysis revealed that when the magnet is directly apposed to the channel without a MFC, it produces a large field gradient spike near its upstream and downstream edges. In contrast, the simulation predicted that the presence of a MFC with a tapered leading edge and a grooved washboard structure along its length will produce multiple smaller gradient spikes distributed along the entire length of the magnet, which should create a more uniform bead distribution (Fig. 3a).
To validate the simulation, we used a magnetometer (Bell 5080 gaussmeter with STD58-0404 transverse probe) to scan over the surface of the magnet with and without the MFC in place. These results revealed a close correlation between experimental measurements and predictions from the simulation (Fig. 3b). Furthermore, when we flowed magnetic beads through a microfluidic channel positioned in the device, a dense pile of beads formed at the upstream edge of the magnet when only the magnet was used, whereas there was much more uniform spread of the captured beads when the MFC was in place (Fig. 3c).
These results confirm that modulating the magnetic gradient can greatly improve the bead distribution in the capture chamber. More specifically, altering the magnetic field gradient distributions with the MFC allowed us to spread the beads within microwells distributed along the channel much more effectively than was possible with the magnet alone using the same channel dimensions (Fig. 2d and 3c). By experimenting with different channel sizes, we determined that a channel 280 μm high by 3 mm wide produced optimal spreading of the magnetic beads and tagged pathogens in a thin layer, while allowing us to operate at a flow rate of 10 ml h−1 with minimal variation across the width of the channel.
Fluorescent images obtained from this automatic scanning were stitched back together and analyzed using an automated image capture and analysis protocol (Zeiss Axio 2 microscope with a Zeiss EC-Plan NEOFLUAR 10× objective, numerical aperture of 0.3, depth of field 8 μm, Metamorph for image acquisition and stitching to form composite image, ImageJ for thresholding and image registration, MATLAB for cell quantification). Pathogen counts obtained by manually scanning the entire device on the microscope were compared with results obtained in the same device using the automated image capture and analysis protocol. After setting thresholding values appropriately, it was possible to correctly quantify the captured pathogens within ±10% of the manual count, and to consistently detect when no fungi were present in the device. Red blood cells were lysed during washing process with 1% Triton X-100 to minimize interference because they tended to settle in the bottom of the device, making it difficult to image the magnetically tagged pathogens on the ceiling of the chamber.
Using the improved detector design and pathogen counting methods, we were able to detect significantly (p < 0.05) more C. albicans fungi in saline with the MFC compared to the magnet alone (Fig. 4b). With the magnet directly adjacent to the PDMS microdevice, the beads formed a pile under the upstream edge that buried a large fraction of the bead-bound fungal cells (Fig. 3c) so that we could only detect 47% of the 850 fungal cells that were initially seeded into the 10 mL saline fluid sample. Adding the MFC between the channel and the magnet boosted the detection sensitivity significantly (p < 0.05), resulting in detection of 68% of the pathogens (Fig. 4b). Inclusion of the MFC also significantly (p < 0.05) improved the detection of fungi captured from human blood samples: using pre-bound C. albicans, we could detect up to 67% of the cells added to a 5 mL blood sample (diluted 1:1 with saline to bring total volume to 10 mL), which was nearly equivalent to our detection rate in saline, whereas only 43% of the spiked pathogens could be detected in blood using the magnet alone (Fig. 4b). The spread of the beads in blood was similar to that seen in saline, with piling of beads at the leading edge only being observed in the absence of the MFC and the captured pathogens were localized almost exclusively within the micropatterned wells in the presence of the MFC.
After optical identification of the pathogens, the magnet and MFC can be removed from the device, which allows the magnetically tagged pathogens to drop from the ceiling and to be flushed out of the device with a 1 ml saline wash. The captured beads and pathogens were then spread on blood agar plates and cultured for 36 hours in a 30 °C incubator; colonies of the original C. albicans can then be used antibiotic susceptibility testing.
Our early device prototypes captured more than 99% of bound pathogens from flowing blood, but the unmodulated magnetic field caused the magnetically tagged pathogens and excess beads to form a dense pile in which only 15% of the captured cells were detectable. To overcome this limitation, we balanced the magnetic pull on the beads and the downstream fluidic drag by inserting a microengineered ferromagnetic flux concentrator (MFC) between the magnet and the upper surface of the channel, and by adjusting the channel geometry and surface topography. This caused the captured beads to spread out evenly into a thin layer through which we could easily visualize the fluorescently labelled pathogens. Finite element models of the magnetic gradient allowed us to determine the MFC shape that would create a high, relatively uniform gradient along the entire length of the channel so that the force on the beads was more uniform to make them spread out in the microwells in the ceiling of the device rather than piling up at the upstream edge of the magnet. The channel dimensions were adjusted so that the beads would only stop in one of the micropatterned wells where they were sheltered from the fluidic shear that tries to push them downstream. Moreover, if one microwell fills, subsequent beads and magnetically tagged pathogens are pushed downstream and collected in the next open well, rather than forming a dense pile. The main problem with the magnet alone was that there was no way to prevent the beads from piling up regardless of the micropatterned wells in the region of the gradient spike. Inserting the MFC between the magnet and the microchannel allowed us to decrease this initial spike and to boost the gradient across the rest of the device so that the bead spread was far more uniform and suitable for optical detection
The ability to capture and concentrate the beads and bound pathogens from a 10 mL fluid sample, and to spread them into a thin layer inside our microdevice, made it possible to detect most of the bound pathogens when stained with fluorescent dyes. Using this approach, we could detect more than two-thirds of the pathogen cells spiked into a 10 mL volume of saline using the magnet with the MFC, whereas most of the captured pathogens were buried by beads (and hence, undetectable) with the magnet alone. The pathogens were labelled with three fluorophores, calcofluor as a general fungal stain, GFP and a Cy3 anti-C. albicans antibody. The calcofluor and GFP stains were used for manually counting the captured pathogens and immunofluorescent Cy3 stain (Fig. 4a) was used for our automated detection protocol.
Importantly, inclusion of the MFC in the device significantly increased detection of pathogens in both saline and blood, when compared with magnet alone. By focusing on pathogens pre-bound with beads in blood, we were able to probe only the pathogen detection sensitivity, rather than having to assess the cumulative effects of bead–pathogen binding in blood and detection simultaneously. Using this approach, we obtained similar detection sensitivities in blood and saline using the MFC. But even with this improved sensitivity, about a third of the captured pathogens could not be visualized in the device. There are two possible explanations for this: 1) some of the bead-bound pathogens might be surrounded by multiple beads, which still obscure their fluorescent signature, or 2) some of the bead piles are still deep enough to bury the pathogens since the wells are 20 μm deep. Shallower wells were not sufficient to arrest the beads and bound pathogens, which resulted in a decrease in bead capture, causing them to be lost downstream in the waste flow. The 50 μm diameter of the wells was determine by the limit of our photolithographic transparency mask; however, mirror chrome masks which can produce smaller and shallower wells could be used in the future. This would give the wells similar dimensions to a single pathogen, which should lower the chance of any pile being deep enough to completely obscure pathogens.
An advantage of this method over rapid molecular diagnostic techniques is that pathogen viability is not significantly affected by the magnetic capture and concentration process, so viable pathogens can be recovered after imaging for conventional antibiotic susceptibility testing. This could allow clinicians to rapidly diagnose sepsis and remove the need for blood culture before performing sub-culturing for antibiotic susceptibility testing, cutting approximately 24 hours off the total processing time.
In summary, we have shown that our technology can detect the presence of dilute pathogens in blood using a simple optical readout that can be detected by optical microscopy. Using this approach, we were able to detect C. albicans fungal cells in human blood at concentrations down to 1 fungal cell mL−1 in less than three hours. We believe that this platform could form the basis of a rapid diagnostic device because the beads are coated with a generic opsonin and thus, the identity of the pathogen does not need to be known for capture to be carried out. Moreover, it has the ability to concentrate rare pathogens from a large fluid volume (e.g., 20–30 mL clinical blood sample) for rapid optical detection. While our results were obtained with fungal cells, it should be possible to create a version of the device to reliably detect bacteria as well by modifying flow rate, magnetic field and channel geometry. The ability to detect rare pathogens directly within human blood within a few hours after collection would represent a major advance relative to current pathogen detection methods that often require a day or more to complete.