Syed Asad Manzoor Bukhari,
M. Faheem Khan,
Ankur Goswami*,
Ryan McGee and
Thomas Thundat*
Department of Chemical and Materials Engineering, University of Alberta, T6G 2R3, Edmonton, Canada. E-mail: agoswami@ualberta.ca; thundat@ualberta.ca
First published on 24th January 2017
Microchannel cantilevers are an emerging platform for physical characterization of materials at the picogram level. Here we report on detecting multiple thermal transitions in picogram amounts of two well-known polymers, semicrystalline poly(L-lactide) (PLA) and amorphous poly(methylmethacrylate) (PMMA), using this platform. The polymer samples, when loaded inside the cantilever, affect its resonance frequency due to changes in its total mass and stiffness. When taken through a thermal cycle, the resonance response of the cantilever further changes due to multiple thermal transitions of the samples. Continuous monitoring of the resonance frequency provides information about β-transition (Tβ), glass transition (Tg), crystallization (Tc), and melting (Tm) of the confined polymer samples. The measured Tg, Tc, and Tm for PLA were ∼60, 78, and 154 °C, respectively, while the Tg and Tβ for PMMA were 48 and 100 °C, respectively. These results are in an agreement with the data obtained from differential scanning calorimetry (DSC). Because of its high sensitivity, this technique is capable of detecting the weaker β-transitions that cannot be observed with conventional DSC.
The microchannel cantilevers were fabricated using conventional top-down microfabrication techniques, as described in our previous study.21 The dimensions of the cantilevers were 200 μm (length) × 20 μm (width) × 3 μm (microchannel height) as shown in Fig. 1a. The calculated mass of the cantilevers were 14.6 pg (considering ρSi3N4 = 3.44 g cm−3). Prior to loading the polymer samples, each cantilever was cleaned with piranha (in order to remove organic contaminants) and heated on a hot plate at 200 °C for two hours to drive out any residual moisture. The cantilevers were filled with the polymer samples through a melt infiltration technique, where a small amount of the samples was placed on inlets (located on backside on the chip). The chips were heated on a hot plate to 150–170 °C. At the melting temperature, due to strong capillary forces, the polymer is able to flow inside the microchannel.
Fig. 1a shows a schematic of the microchannel cantilever based thermal analysis platform used in this study. In order to apply a controlled thermal cycle, the cantilever was firmly placed on top of a Peltier element, and the temperature was continuously monitored by a Pt-1000 resistance temperature detector (RTD) mounted on top of the chip. Temperature was logged in real time by monitoring the RTD's resistance using a Keithley 197 multimeter interfaced with LabView software. The cantilever was heated from room temperature to 180 °C at a heating rate of 4 °C min−1. In order to enhance the signal to noise ratio (SNR), the cantilever was mechanically excited by an external piezo actuator. The resonance frequency was measured optically using a laser Doppler vibrometer (LDV, Polytec, USA) and was monitored continuously during the experiments. As the polymer samples are viscous, a cantilever cannot be effectively flushed and be used for another sample. Therefore, to avoid cross contamination, two similar cantilevers were used to characterize the two polymer samples. Prior to loading each sample, the empty cantilevers were characterized to determine their resonance frequency and phase. In an effort to enhance the quality factor (and thus the sensitivity) of the cantilevers, all the experiments were carried out at a vacuum level of ∼10−6 mbar. SEM of the top view of the cantilever is shown in Fig. 1b. A cross sectional SEM view of the channel (showing a filled and an empty channel) is shown in Fig. 1c. For reference measurements, the polymers were characterized using DSC (TAQ100) with 5–7 mg of sample, with a heating rate of 10 °C min−1.
(1) |
Filling the cantilever with a polymer changes the stiffness and the mass of the cantilever, causing a shift in its resonance frequency. If the polymer stiffness and mass are indicated as kp and mp, respectively then by assuming a parallel spring-mass model, the effective resonance frequency fc+p(T0) of the polymer-filled cantilever would follow eqn (2).
(2) |
In our case, since , it is evident from eqn (1) and (2) that fc(T0) > fc+p(T0) (ESI†).
As compared to the structural material (silicon nitride) of the cantilever, the polymer exhibits a higher change in its stiffness due to heating. Therefore, as the temperature increases, the change in kp dominates over kc, due to significant changes in the elastic modulus of the polymer. Hence, by monitoring fc+p(T), the transitions of the polymer can be determined. From eqn (3) it is evident that kp can be determined by monitoring the resonance frequency of a cantilever before and after loading the polymer (ESI for detailed derivation†).
(3) |
Cantilevers with similar mechanical properties were used in order to perform the thermal analyses of the polymers. Prior to loading the polymers, resonance frequency and phase response of the empty cantilevers were recorded (shown in Fig. 2a and b). The quality factor of the empty cantilevers was found to be ∼6000. However, after loading the polymers into the microchannel, the resonance frequency of the cantilevers decreased by 14–16 kHz and the quality factor reduced to ∼20% of their original values. These changes ensure that the polymer successfully entered the cantilever. Unlike a cantilever (placed inside a liquid), which experiences a large viscous damping, a microchannel cantilever loaded with a viscous polymer maintains a quality factor well above 1000.
To accurately perform experiments, stability of the device temperature and its effect on the resonance frequency were first quantified by heating the cantilever in a stepwise manner while simultaneously recording the temperature and frequency response. A 4 °C step was applied to each cantilever and their corresponding temperature and frequency responses were recorded as shown in Fig. 3a. Temperature stability was achieved after ∼60 s. However, as shown in Fig. 3a, the resonance frequency of the cantilever is material dependant and show stabilization times of 100, 150, and 270 s for the empty (τc,empty), PLA (τc+PLA), and PMMA (τc+PMMA) filled cantilevers, respectively. Therefore, for subsequent measurements we have used 100, 180 and 300 s as stabilization times for the empty, PLA, and PMMA filled cantilevers respectively. The variation in time constant is directly related to the specific heat capacities of the cantilever material (SiN) and the polymers (PMMA and PLA) stored within. The trend in the observed time constants, τc,empty < τc+PLA < τc+PMMA, follows the samples' specific heat capacity as cp,SiN < cp,SiN+PLA < cp,SiN+PMMA.31,32
Fig. 3 (a) Stabilization in temperature and frequency measured for empty, PLA-filled, and PMMA-filled cantilevers. (b) Thermal characterization of empty, PLA-filled, and PMMA-filled cantilevers. |
The stabilization time for the resonance frequency of the cantilever (filled with PLA or PMMA) is much longer than 60 s. If a measurement is performed by achieving full stabilization (3τ) in the resonance frequency, each measurement would take multiple hours. In order to reduce the measurement time and maintain the detection of small transitions (such as Tβ), the cantilever was heated at the rate of 4 °C min−1. This heating rate and stabilization time enabled rapid measurement with sufficient sensitivity and resolution.
Temperature dependant resonance spectra of the empty and polymer-filled cantilevers were measured and plotted as shown in Fig. 3b. In the case of the empty cantilevers, the resonance frequency initially increases with temperature, but no significant changes occur at higher temperatures. This is in stark contrast to the results from the polymer-filled cantilevers, which show a significant reduction in the resonance frequency at specific temperatures, corresponding to various thermal transitions. It is evident that the polymers undergo thermal transitions with increasing temperature, which reduces the effective stiffness of the filled cantilever, resulting in a reduction in the resonance frequency of the cantilever.
Further, the stiffness of the polymer inside the cantilever was determined using eqn (3) and plotted in Fig. 4a and b. Considering that the total mass (mp + mc) and the stiffness (kc) of the cantilever material (SiN) are constant, the stiffness of the polymer follows the same trend as that of the resonance frequency of the cantilever. For a comparison, Fig. 4a shows the stiffness of the PLA sample (extracted from the results of the cantilever), as well as the measurements carried out with DSC. The overall stiffness of the PLA-filled cantilever initially increases (up to 60 °C), which is most likely due to a release of stress (in the silicon nitride). At ∼60 °C, the stiffness of the cantilever decreases sharply which is the onset of Tg of PLA. Here the resonance frequency of the cantilever decreases by 60 Hz. The results from DSC indicate that the Tg of the sample lies within the same region. As the temperature is increased further, kPLA starts to increase at 78 °C, which indicates the onset of crystallization (Tc) of PLA. This can be explained as due to an increase in the elastic modulus of PLA during crystallization, which in turn increases the stiffness of the polymer. The Tc of the sample is also evident at ∼87 °C according to the data obtained from DSC. Melting point of the sample is determined by increasing the temperature in a continuous fashion. The onset of melting appears at ∼154 °C (compared to 160 °C obtained from DSC) as seen by a sharp decrease in kPLA. Upon melting, the viscosity of the sample decreases which explains the reduction in its stiffness.
Fig. 4 Heat flow (from DSC) and calculated stiffness of (a) PLA and (b) PMMA are plotted as functions of temperature. |
Fig. 4b shows the thermal transitions of PMMA, measured by the cantilever as well as DSC. As the cantilever is heated, the sample goes through a small but recoverable change in its stiffness (kPMMA) at 48 °C. Such a change is not visible in the data obtained using the DSC. This can be attributed to the β-transition of PMMA and can be explained as due to hindrance of the molecular motion of the –COOCH3 group (attached to the main chain C–C bond).33 Though this transition cannot be probed using DSC, it can be measured by DMA and DEA, as reported in literature.34,35 As the PMMA filled cantilever is further heated to ∼100 °C, a sharp reduction in kPMMA occurs, which is identified as the Tg of PMMA. Such a change is also observed in the DSC data at 103 °C. Any small discrepancies between the temperatures reported between these two techniques may be due to the method used to determine the temperature within the microchannel.
Overall, the thermal transitions of these two polymers measured by suspended microchannel cantilever are in excellent agreement with established techniques such as dilatometry, DMA, and DES as reported in literature as shown in Table 1.
Transitions temperature (°C) | PMMA | PLA | ||||||||
---|---|---|---|---|---|---|---|---|---|---|
Dilatometer34 | DSC (measured and reported) | DMA34 | DES36,37 | Cantilever (measured) | Dilatometer38 | DSC (measured and reported) | DMA39 | DES40 | Cantilever (measured) | |
Glass transition | 103 | 103 (ref. 41) | 106 | 100 (1 kHz) | 100 | 40 | 58 | 61 | 66 (1 kHz) | 60 |
Crystallization | Usually does not exist | 85 | 86.5 | 85 (1 Hz) | 80 (1 kHz) | 78 | ||||
β-Transition | Cannot be detected | Cannot be detected | 35–40 (1 Hz) | 20–50 (1 kHz) | 45–50 (120 kHz resonance frequency) | Cannot be detected | Cannot be detected | Not reported | −63 (1 kHz) | Not measured |
Melting | Not reported | 150–170 | Not reported | Not reported | 140 | Not reported | 154 | Not reported | Not reported | 160 |
Footnote |
† Electronic supplementary information (ESI) available. See DOI: 10.1039/c6ra25455a |
This journal is © The Royal Society of Chemistry 2017 |