Daisuke Babaab,
Asep Sugih Nugrahaab,
Muhammad Iqbalab,
Jiang Boa,
Cuiling Lia,
Abdulmohsen Ali Alshehric,
Jungmok Youd,
Victor Malgras*a,
Yusuke Yamachi*de and
Toru Asahi*ab
aInternational Center for Materials Nanoarchitectonics (WPI-MANA), National Institute for Materials Science (NIMS), 1-1 Namiki, Tsukuba, Ibaraki 305-0044, Japan. E-mail: MALGRAS.Victor@nims.go.jp; tasahi@waseda.jp
bFaculty of Science and Engineering, Waseda University, 3-4-1 Okubo, Shinjuku, Tokyo 169-8555, Japan
cDepartment of Chemistry, King Abdulaziz University, P. O. Box. 80203, Jeddah 21589, Saudi Arabia
dDepartment of Plant & Environmental New Resources, Kyung Hee University, 1732 Deogyeong-daero, Giheunggu, Yongin-si, Gyeonggi-do 446-701, South Korea
eSchool of Chemical Engineering, Australian Institute for Bioengineering and Nanotechnology (AIBN), The University of Queensland, Brisbane, QLD 4072, Australia. E-mail: y.yamauchi@uq.eudu.au
First published on 14th March 2018
Mesoporous metal films can detect biomarkers with high sensitivity. Further coating the mesoporous metal with polymers enhances sensing selectivity by favoring specific biomarkers against other interferents. In the present study, we report the fabrication of a Nafion®-coated mesoporous Pd film to filtrate interferents present in sweat during non-invasive biosensing. By using a Nafion®-coated mesoporous Pd film, lactic acid, a metabolite present in sweat, can be successfully detected with high sensitivity.
These reported devices use enzymes to detect biomarkers from sweat without having to filtrate any interferents. Despite the presence of interfering species in the medium, enzymes can form enzyme–substrate complexes with specific biomarkers, resulting in high selectivity. However, these sensors have some disadvantages in terms of cost and stability (requiring suitable pH and temperature to prevent degradation), which make their practical use less attractive. On the other hand, non-enzymatic sensors, such as mesoporous metal films and nanoparticles, have recently been reported as a promising alternative.3 Compared with enzymatic sensors, their selectivity is still relatively low because they cannot form complexes with biomarkers. Under the application of an optimal voltage, however, the influence of interferents can be reduced. In addition, high sensitivity can be achieved because of the large surface area reacting with the biomarkers.
The typical biomarkers in sweat are lactic acid, glucose, interleukin and other proteins.4 Among these biomarkers, lactic acid is one of the metabolites whose concentration in sweat is different before and after exercise.5 Furthermore, the relationship between the lactic acid content and the metabolic activity is well-known.6 Therefore, if a biosensor could detect and monitor the exact content of lactic acid in sweat, it could be used for health management purposes.
In this study, we aim to detect biomarkers from sweat after filtrating out the interferents with a Nafion®-coated mesoporous Pd film. Nafion® is a polymer which is typically used for ion-exchange.7 Mesoporous Pd films can be fabricated in nonionic solution consisting of P123 surfactant and PdCl2 metal precursor. The resulting large surface area is suitable for applications in biosensor. Bare mesoporous Pd films have already a strong potential for biosensing, but further coating the film with a Nafion® layer with mesoporous roughness is critical to remove interferents present in sweat.
Although Nafion® has been utilized as a part of electrode materials for biosensing,8 the hybrid structure consisting of a mesoporous metal combined with a Nafion® layer has never been reported. Composites made of mesoporous materials and polymers have previously been reported for drug delivery and photochemical applications,9 but not for electrochemical biosensing. Integrating filtration and detection in one functional system using mesoporous metal films is a novel approach. Filtration generally needs complicated process, but Nafion® coating is simple and can be used in a “disposable device” technology. In this study, after filtration by Nafion®, lactic acid is monitored by chronoamperometry under voltage which is tuned to minimize the influence of interferents such as glucose, urea, ammonia and ethanol present in sweat. To benchmark the sensitivity of our Nafion®-coated mesoporous Pd film, bare mesoporous Pd, Nafion®-coated nonporous Pd and bare nonporous Pd films are also compared.
The electrochemical response of the mesoporous Pd films, with and without Nafion® coating were measured in NaCl solution (0.4%) to mimic the typical composition of human sweat. The CV measured in NaCl solution is shown in Fig. S5† and the area under the CV curve for the Nafion®-coated film was smaller than that for the uncoated film. This can be explained as the anionic Nafion® layer has abundant sulphonate groups as well as anionic ions such as chloride, which cannot easily penetrate the membrane. For the bare mesoporous Pd film, the peak observed at around 0.25 V is due to the oxidation of Pd. For reference, Pd in KCl solution is oxidized into K2PdCl4 with an oxidation peak observed at around 0.25 V.11 Furthermore, the smooth peaks spanning from −0.2 V to −0.6 V are also reported in the reference and can be ascribed the reduction of Pd salt into Pd.11 These peaks cannot be observed when the mesoporous film is coated with Nafion® which prevents Pd from redox reaction. Previous works reported that protein adhesion is hindered by the presence of Nafion® coating.12 Only small molecules such as lactate and glucose can be transported through this polymer,13 which is why good selectivity is expected for lactic acid sensing.
The electrochemical surface area (ECSA) of the catalyst films were investigated by carrying out CV in 0.5 M H2SO4 with a scan rate of 50 mV s−1 between −0.2 and 1.2 V (vs. Ag/AgCl) (Fig. 2a). The ECSA is estimated by calculating the charge associated with PdO reduction between the potential of 0.35 and 0.70 V (vs. Ag/AgCl) from the negative potential scan of the CV. By assuming that the conversion factor for an oxide monolayer reduction is 420 μC cm−2 on a smooth Pd surface, the mass-normalized ECSA for mesoporous Pd films is calculated to be 37.9 m2 g−1, which is 10 times higher than that of the nonporous Pd films (3.56 m2 g−1). As shown in Fig. 2b, CVs are measured in NaCl solution containing lactic acid (10 mM) by using mesoporous and nonporous Pd films as the working electrode. Anodic current and cathodic peaks can be observed, corresponding respectively to the oxidation of lactic acid and reduction of pyruvic acid. The electro-oxidation of lactic acid into pyruvic acid is likely to occur at the surface of the Pt electrode, as previously reported.14 Because of a high ECSA, the mesoporous film displays larger current peaks and superior catalytic activity towards the oxido-reduction reaction of lactic acid and pyruvic acid compared to that of the nonporous film. Both the dependence of the ECSA and anodic peak current density by lactic acid oxidation on the deposition time are shown in Fig. S6.† Since the size of lactic acid molecules is much bigger than that of H+ ion, the anodic peak current density due to oxidation of lactic acid becomes saturated when the film thickness is larger than 600 nm. Therefore, we used the mesoporous Pd film prepared for 600 s for further electrochemical study.
From Fig. 2b, it can be observed that the oxidation of lactic acid and reduction of pyruvic acid on mesoporous Pd film take place between −0.5 V and 0.5 V (vs. Ag/AgCl). Therefore, the chronoamperometry is measured under the same voltage window to minimize the influence of glucose, urea, ammonia or ethanol. These interferents, present in human sweat,15 and cannot be removed by the Nafion® coating because they are too small to be filtrated. The current response due to lactic acid and other interferents are compared in Fig. 3a, thus confirming that the contribution from the interferents becomes negligible under an applied potential of 0.4 V (vs. Ag/AgCl).
Therefore, chronoamperometry was measured at each addition of 1 mM of lactic acid and the results are shown in Fig. 3b. To evaluate the sensitivity of the Nafion®-coated mesoporous Pd film to lactic acid, bare mesoporous Pd, Nafion®-coated nonporous Pd and bare nonporous Pd films were also tested. The limit of detection (LOD) is calculated to be 0.34 mM (for the Nafion®-coated mesoporous Pd film), 0.48 mM (for the bare mesoporous Pd film), 0.75 mM (for the Nafion®-coated nonporous Pd film), and 1.5 mM (for the bare nonporous Pd film). The LOD using Nafion®-coated mesoporous Pd film (0.34 mM) is superior to what was previously reported (0.5 mM).16 The relationship between the concentration of lactic acid and the current is shown in Fig. 3c. The mesoporous Pd film shows a higher sensitivity than the nonporous Pd film, due to a higher ECSA. Interestingly, the Nafion®-coated mesoporous Pd film shows a higher sensitivity than the bare mesoporous Pd film. To discuss this result, adsorption of lactic acid on both electrodes in NaCl solution should be considered. In 0.4% NaCl solution, the pH is almost 7.0 and lactic acid releases H+ ion because the pKa is 3.86. Therefore, negatively charged lactic acid ions cannot approach the Nafion® surface due to electrostatic repulsion. It would then be expected that the bare mesoporous Pd film shows higher sensitivity. This is due to chloride ions being critical interferents during the sensing of lactic acid. When Nafion® is present, the chloride ions are removed. We tested the repeatability of this sensor using the relationship between the amperometric response and the lactic acid concentration. Even after addition of 15 mM, the linearity is maintained showcasing high repeatability (Fig. S7†).